← Product Code [CAC](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/CAC) · K221483

# FlowSaver Blood Return System (K221483)

_Inari Medical, Inc. · CAC · Feb 17, 2023 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94therapeutic-devices/CAC/K221483

## Device Facts

- **Applicant:** Inari Medical, Inc.
- **Product Code:** [CAC](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/CAC.md)
- **Decision Date:** Feb 17, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5830
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

The FlowSaver Blood Return System is used with Triever Catheters for autologous blood transfusion.

## Device Story

The FlowSaver Blood Return System is an accessory used during embolectomy procedures to facilitate autologous blood transfusion. It processes aspirated blood from the FlowTriever Retrieval/Aspiration System using a dual-layer 40 µ/200 µ filtration mechanism. This filtration minimizes intraprocedural blood loss by cleaning the aspirated blood for return to the patient. The device is operated by clinicians in a procedural setting. The system's output is filtered blood, which is re-infused into the patient to reduce the need for donor blood products. The current modification removes the requirement for a secondary filtration procedure previously performed after the initial FlowSaver filtration.

## Clinical Evidence

No new clinical trials were conducted for this submission. Safety and effectiveness were supported by post-market clinical data from three studies: PEERLESS (NCT05111613), FLAME (NCT04795167), and FLASH (NCT03761173). These studies demonstrated the filtration efficiency and safety of the FlowSaver system when used without the previously required secondary filtration procedure.

## Technological Characteristics

The device utilizes a dual-layer filtration system (40 µ/200 µ) to process aspirated blood. It is designed for use with Triever Catheters. The device is a mechanical accessory; it does not utilize software, electronics, or external energy sources. It is a sterile, single-use device.

## Regulatory Identification

An autotransfusion apparatus is a device used to collect and reinfuse the blood lost by a patient due to surgery or trauma.

## Predicate Devices

- FlowSaver Blood Return System ([K210176](/device/K210176.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with an abstract design. To the right of the seal, there is the FDA logo in blue, followed by the words "U.S. FOOD & DRUG" in a larger font and "ADMINISTRATION" in a smaller font. The logo is simple and professional, and it is easily recognizable.

February 17, 2023

Inari Medical, Inc. Kaitlyn Weinkauf Sr. Regulatory Affairs Specialist 6001 Oak Canyon, Suite 100 Irvine, California 92618

Re: K221483

Trade/Device Name: FlowSaver Blood Return System Regulation Number: 21 CFR 868.5830 Regulation Name: Autotransfusion Apparatus Regulatory Class: Class II Product Code: CAC Dated: December 16, 2022 Received: December 20, 2022

Dear Kaitlyn Weinkauf:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Nicole M. Gillette -S

Nicole Gillette Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known)

#### K221483

Device Name FlowSaver Blood Return System

Indications for Use (Describe)

The FlowSaver Blood Return System is used with Triever Catheters for autologous blood transfusion.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(K) SUMMARY

| Date prepared                                                                  | February 16, 2023                                                                                                                                                                                                                                          |
|--------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name                                                                           | Inari Medical, Inc.<br>6001 Oak Canyon, Suite 100<br>Irvine, CA 92618<br>949.600.8433                                                                                                                                                                      |
| Contact person                                                                 | Kaitlyn Weinkauf<br>Sr. Regulatory Affairs Specialist                                                                                                                                                                                                      |
| Trade name                                                                     | FlowSaver Blood Return System                                                                                                                                                                                                                              |
| Common name                                                                    | Blood filter                                                                                                                                                                                                                                               |
| Regulation name                                                                | Autotransfusion apparatus                                                                                                                                                                                                                                  |
| Classification number                                                          | 21 CFR 868.5830                                                                                                                                                                                                                                            |
| Product code                                                                   | CAC                                                                                                                                                                                                                                                        |
| Regulatory class                                                               | II                                                                                                                                                                                                                                                         |
| Predicate device                                                               | FlowSaver Blood Return System (K210176)                                                                                                                                                                                                                    |
| Description                                                                    | The FlowSaver Blood Return System accessory allows for autologous<br>injection of aspirated blood from the FlowTriever Retrieval/Aspiration<br>System embolectomy procedure by dual layer 40 µ/200 µ filtration to<br>minimize intraprocedural blood loss. |
| Indications for Use                                                            | The FlowSaver Blood Return System is used with Triever Catheters for<br>autologous blood transfusion.                                                                                                                                                      |
| Device modification                                                            | The purpose of this submission is a labeling change to remove the secondary<br>filtration procedure that is performed after the initial filtration with the<br>FlowSaver Blood Return System.                                                              |
| Comparison of<br>Technological<br>Characteristics with<br>the Predicate Device | There is no change to the principles of operation, indications for use,<br>intended use, fundamental scientific technology, or technological<br>characteristics between the proposed FlowSaver Blood Return System and<br>the predicate device.            |

{4}------------------------------------------------

| Summary of<br>substantial<br>equivalence | The proposed FlowSaver Blood Return System accessory and the predicate device have the same indications for use, intended use, principles of operation, and fundamental scientific technology.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                          | <b>Non-Clinical Testing</b><br>The following verification test demonstrated compliance with product requirements:<br>• Filtration Efficiency<br>The following testing was leveraged from the predicate device (K210176):<br>• Visual Inspection<br>• Dimensional Inspection<br>• Engagement & Disengagement Force Testing<br>• Flow Rate Testing<br>• Media Integrity testing<br>• Leakage Testing<br>• Vacuum Testing<br>• Clot Burden Filtration Validation<br>• Simulated Use and Tensile Testing<br>• Simulated Use and Torque Testing<br>• Burst Testing<br>• Hematocrit Testing<br>• Mechanical Hemolysis Testing<br>Test results demonstrated that all acceptance criteria were met; therefore, the device conforms to established product specifications.<br><b>Clinical Testing</b><br>Post-market clinical data was used to demonstrate the safety and effectiveness of the FlowSaver Blood Return System's use without a secondary filtration procedure. Post-market clinical data from PEERLESS (NCT05111613), FLAME (NCT04795167), and FLASH (NCT03761173) studies demonstrated the FlowSaver's filtration efficiency and safety without use of the second filter in support of substantial equivalence.<br><b>Conclusion</b><br>The proposed labeling change to the FlowSaver Blood Return System does not change its intended use nor does it change the principles of operation. The verification testing results demonstrate that the proposed FlowSaver Blood Return System is substantially equivalent to the predicate device. |

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94therapeutic-devices/CAC/K221483](https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94therapeutic-devices/CAC/K221483)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
