K101586 · Sorin Group Italia S.R.L. · CAC · Oct 5, 2010 · Anesthesiology
Device Facts
Record ID
K101586
Device Name
XTRA AUTOTRANSFUSION SYSTEM
Applicant
Sorin Group Italia S.R.L.
Product Code
CAC · Anesthesiology
Decision Date
Oct 5, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5830
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The XTRA Autotransfusion System is indicated for intraoperative recovery of blood, washing of blood collected in the postoperative period, and preoperative sequestration (with indirect patient connection). Typical clinical applications of autotransfusion include the following surgical specialties: Cardiovascular, Orthopedics, Thoracic, Transplant Surgery, Emergency (Trauma), Neurosurgery, Obstetrics and gynecology, and Urology.
Device Story
XTRA Autotransfusion System is an automated device for blood recovery and processing. Hardware includes centrifuge, blood pump, automatic clamps, sensors, and user interface (display/keyboard). Disposables feature simplified installation design. System processes blood via centrifugation and washing; output is processed blood for reinfusion. Used in clinical settings (OR, etc.) by trained personnel. Device automates blood management to reduce need for allogeneic transfusion, potentially benefiting patients by minimizing transfusion-related risks.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993-1 (hemolysis, hemocompatibility, cytotoxicity, reactivity, sensitization, toxicity, mutagenicity, sterility, pyrogenicity). Electrical and electromagnetic compatibility testing conducted. Performance testing confirmed comparability to the predicate device.
Technological Characteristics
Automated autotransfusion system comprising hardware (centrifuge, pump, sensors, UI) and disposable sets. Materials tested per ISO 10993-1. Connectivity includes user interface panel. Sterilization and specific software architecture details not provided.
Indications for Use
Indicated for intraoperative blood recovery, postoperative blood washing, and preoperative sequestration in patients undergoing cardiovascular, orthopedic, thoracic, transplant, emergency/trauma, neurosurgical, obstetric/gynecologic, and urologic procedures.
Regulatory Classification
Identification
An autotransfusion apparatus is a device used to collect and reinfuse the blood lost by a patient due to surgery or trauma.
{0}------------------------------------------------
| | 510(k) SUMMARY | K101586 |
|----------------------|------------------------------------------------------------------------------------|------------|
| SUBMITTER: | Sorin Group Italia S.r.I.<br>86, Via Statale 12 Nord<br>41037 Mirandola (MO) Italy | |
| CONTACT PERSON: | Luigi Vecchi<br>Phone: 011 39 0535 29811<br>Fax: 011 39 0535 25229 | |
| DATE PREPARED: | June 3, 2010 | OCT 5 2010 |
| DEVICETRADE NAME: | XTRA | |
| COMMON NAME: | Autotransfusion System | |
| CLASSIFICATION NAME: | Apparatus, Autotransfusion | |
| PREDICATE DEVICE: | Dideco ELECTA<br>K020647 | |
# DEVICE DESCRIPTION:
XTRA Autotransfusion System consists of hardware and disposables. It is the next generation of the Sorin autotransfusion device family. The main elements of the hardware include the centrifuge, blood pump, automatic clamps, control and monitoring sensors, and an user interface (display panel and keyboard).
The main modifications to the disposables are the elimination of the bowl snap on bayonet fitting system to fix the device into the centrifuge, the elimination of the silcone bayon inting bystem to and the newly designed system that allows the set up of the disposable set to simplify the installation of disposable.
# INDICATION FOR USE:
INDIOATION CON OOL recovery of blood, washing of the processed blood and preoperative sequestration (with indirect patient connection). Typical clinical applications of autotransfusion include the following surgical specialties: Cardiovascular, Orthopedics, Thoracic, Transplant following "Sarglour" opeolation" "Sansargery, Obstetrics and Gynecology, and Urology.
# TECHNOLOGICAL CHARACTERISTICS:
Sorin makes the claim of substantial equivalence to cited predicates based on intended Oom makes the olain of echnological characteristics, and operational characteristics.
{1}------------------------------------------------
| XTRA Autotransfusion System | Traditional 510(k) ~ |
|-----------------------------|----------------------|
| Sorin Group Italia S.r.l. | June 3, 2010 |
Sorin Group Italia Srl believes that the XTRA Autotransfusion Svstem is substantially equivalent to the Dideco ELECTA and to other currently marketed automated autotransfusion devices, that any differences are minor, and raise no new issues of safety and effectiveness.
# BIOCOMPATIBILITY TEST RESULTS:
A complete battery of tests was carried out in accordance with the requirements of ISO 10993-1 and the FDA May 1, 1995 Memorandum on the use of the ISO 10993 standard for biocompatibility testing on the raw materials. Testing was performed on the Disposable set of XTRA Autotransfusion System (accelerated aging). The device was tested for Hemolysis, Hemocompatibility, Cytotoxicity, Intracutaneous Reactivity, Sensitization, Acute Systemic Toxicity and Mutagenicity, Sterility, Pyrogenicity, ETO residuals and package integrity testing were also conducted. The results of the testing met established specifications.
## IN VITRO TEST RESULTS:
Testing supplied in the 510(k) premarket notification for the XTRA Autotransfusion System includes electrical testing, electromagnetic compatibility testing, and performance testing that demonstrate compliance with performance specifications. The results of the study showed the device characteristics between XTRA Autotransfusion System and Dideco ELECTA were comparable.
## CONCLUSIONS:
The results of in vitro studies demonstrate that the XTRA Autotransfusion System is substantially equivalent to the predicate device in terms of safety and effectiveness.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three wing-like shapes. The bird is positioned within a circular border. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circular border.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Sorin Group Italia S.r.l. c/o Mr. Barry Sall 195 West Street · Waltham, MA 02451
OCT. 5 2010
Re: K101586
Trade/Device Name: XTRA Autotransfusion System Regulation Number: 21 CFR 868.5830 Regulation Name: Autotransfusion apparatus Regulatory Class: II Product Code: CAC Dated: September 3, 2010 Received: September 7, 2010
Dear Mr. Sall:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for abounding to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I va mayy attrerey misions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability adulteration: Trease noter u, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I least of davisou that I Drivision that your device complies with other requirements of the Act that I DX has made a devel regulations administered by other Federal agencies. You must or any Fedelal statutes and regulations and admited to: registration and listing (21
{3}------------------------------------------------
Page 2 - Mr. Barry Sall
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part . 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
suma R. Vc. Aner
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
XTRA Autotransfusion System Sorin Group Italia S.r.I.
Traditional 510(k) June 3, 2010
Device Name: XTRA Autotransfusion System
(K101586)
Indications for Use:
The XTRA Autotransfusion System is indicated for intraoperative recovery of blood, washing of blood collected in the postoperative period, and preoperative sequestration (with indirect patient connection). Typical clinical applications of autotransfusion include the following surgical specialties:
- Cardiovascular
- Orthopedics
- Thoracic
- Transplant Surgery
- Emergency (Trauma)
- Neurosurgery
- Obstetrics and gynecology
- Urology
X Prescription Use _ (Part 21 CFR 801 Subpart D) C)
Over-the-Counter Use AND/OR (21 CFR 807 Subpart
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Dian R. Vilmer
(Division Sign-Off) (Division Sign of Cardiovascular Devices
540(k) Number Kvolts & C
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.