K243484 · Piccolo Medical, Inc. · LJS · May 29, 2025 · General Hospital
Device Facts
Record ID
K243484
Device Name
PM2+ System and SmartPICC Stylet
Applicant
Piccolo Medical, Inc.
Product Code
LJS · General Hospital
Decision Date
May 29, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5970
Device Class
Class 2
Indications for Use
The PM2+ System and SmartPICC Stylet is indicated for guidance and positioning of commercially available central venous catheters. The Stylet provides stiffness for use in placement of the catheter, intravascular capability for ECG detection and recording for catheter guiding and positioning, and intravascular ionic dilution for supplemental catheter navigation. The SmartPICC Stylet, when used with the PM2+ System, provides real-time catheter tip location information by using the patient's physiological (cardiac electrical activity and blood flow) information. When relying on the patient's ECG signal, the PM2+ System and SmartPICC Stylet is indicated for use as an alternative method to fluoroscopy or chest x-ray for central venous catheter tip placement confirmation in adult patients. The SmartPICC Stylet is indicated for use with PICC lines having a minimum lumen diameter of 0.021". Limiting but not contraindicated situations for this technique are in patients where alterations of cardiac rhythm change the presentation of the P-wave as in atrial fibrillation, atrial flutter, severe tachycardia and pacemaker driven rhythm. In such patients, who are easily identifiable prior to central venous catheter insertion the use of an additional method is required to confirm catheter tip location.
Device Story
Bedside navigation and tip location system for central venous catheters/PICCs; utilizes intravascular ECG (ivECG) and optional ionic dilution. Stylet provides mechanical stiffness for catheter insertion; transmits physiological signals (cardiac electrical activity and blood flow) to PM2+ system. System processes, displays, and stores ECG waveforms and ionic dilution data for real-time tip location. Used by clinicians during PICC installation; stylet removed post-confirmation. Output enables clinicians to confirm catheter tip position without fluoroscopy or chest x-ray, reducing radiation exposure and improving procedural efficiency.
Clinical Evidence
No human clinical data provided. Substantial equivalence based on non-clinical performance testing: tensile strength, insertion force, corrosion resistance, catheter compatibility, leak testing, infusion rate, ionic dilution performance, algorithm verification, electrical hardware verification, sterile barrier packaging, biocompatibility, electrical safety, EMC, software V&V, and human factors validation.
Technological Characteristics
Intravascular ECG and ionic dilution sensing; electrical connectivity to PM2+ system. Materials include biocompatible components. System provides real-time waveform processing and display. Software-based algorithm for signal analysis. Sterilized components. Compatible with PICC lines (min 0.021" lumen).
Indications for Use
Indicated for guidance and positioning of central venous catheters in adult patients. Used as an alternative to fluoroscopy or chest x-ray for tip placement confirmation. Compatible with PICC lines (min lumen 0.021"). Contraindicated/limited in patients with atrial fibrillation, atrial flutter, severe tachycardia, or pacemaker-driven rhythms where P-wave morphology is altered; requires secondary confirmation method in these cases.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
K200037 — Piccolo Medical SmartPICC System · Piccolo Medical, Inc. · Jan 22, 2021
K160925 — VPS Rhythm Device with TipTracker Technology · Arrow International, Inc. (A Subsidiary of Teleflex, Inc.) · Dec 22, 2016
K112744 — SAPIENS TIP CONFIRMATION SYSTEM · C.R. Bard, Inc. · Oct 20, 2011
K252792 — PM2 System with ECGuide Connector · Piccolo Medical, Inc. · Nov 17, 2025
K220363 — VPS Rhythm DLX Device with TipTracker Technology · Arrow International, LLC (A subsidiary of Teleflex, Inc.) · Aug 22, 2022
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 29, 2025
Piccolo Medical, Inc.
Augustus Shanahan
CEO
218 Mississippi St
San Francisco, California 94107
Re: K243484
Trade/Device Name: PM2+ System and SmartPICC Stylet
Regulation Number: 21 CFR 880.5970
Regulation Name: Percutaneous, Implanted, Long-Term Intravascular Catheter
Regulatory Class: Class II
Product Code: LJS, DRX, DSA
Dated: April 30, 2025
Received: April 30, 2025
Dear Augustus Shanahan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243484 - Augustus Shanahan
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K243484 - Augustus Shanahan
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Aneesh S. Deoras -S
Aneesh Deoras
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K243484
Device Name
PM2+ System and SmartPICC Stylet
Indications for Use (Describe)
The PM2+ System and SmartPICC Stylet is indicated for guidance and positioning of commercially available central venous catheters. The Stylet provides stiffness for use in placement of the catheter, intravascular capability for ECG detection and recording for catheter guiding and positioning, and intravascular ionic dilution for supplemental catheter navigation. The SmartPICC Stylet, when used with the PM2+ System, provides real-time catheter tip location information by using the patient's physiological (cardiac electrical activity and blood flow) information. When relying on the patient's ECG signal, the PM2+ System and SmartPICC Stylet is indicated for use as an alternative method to fluoroscopy or chest x-ray for central venous catheter tip placement confirmation in adult patients. The SmartPICC Stylet is indicated for use with PICC lines having a minimum lumen diameter of 0.021".
Limiting but not contraindicated situations for this technique are in patients where alterations of cardiac rhythm change the presentation of the P-wave as in atrial fibrillation, atrial flutter, severe tachycardia and pacemaker driven rhythm. In such patients, who are easily identifiable prior to central venous catheter insertion the use of an additional method is required to confirm catheter tip location.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
| 510(k) #: K243484 | 510(k) Summary | Prepared on: 2025-04-29 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | Piccolo Medical, Inc. | |
| Applicant Address | 218 Mississippi St San Francisco CA 94107 United States | |
| Applicant Contact Telephone | 6178424168 | |
| Applicant Contact | Mr. Augustus Shanahan | |
| Applicant Contact Email | ashanahan@piccolomedical.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | PM2+ System and SmartPICC Stylet | |
| Common Name | Tip Location System | |
| Classification Name | Percutaneous, implanted, long-term intravascular catheter | |
| Regulation Number | 21 CFR 880.5870 | |
| Product Code(s) | LJS, DRX, DSA | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K200037 | Piccolo Medical SmartPICC System | LJS |
| K240486 | PM2 System and ECGuide Connector | LJS |
| K201286 | 0.9% Sodium Chloride I.V. Flush Syringe | NGT |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| The PM2+ System and SmartPICC Stylet is a device used by clinicians for bedside navigation and tip location confirmation of central venous catheters including PICCs. The PM2+ System and SmartPICC Stylet provide real-time catheter tip location information by using intravascular ECG (ivECG). The SmartPICC Stylet provides intravascular ECG for catheter tip location confirmation as well as an optional, ionic dilution navigation feature. The SmartPICC Stylet is used only during PICC installation, and removed once the PICC tip position is confirmed to be at the target location. The SmartPICC Stylet is electrically connected to the PM2+ system which processes, displays, and stores the ECG waveforms as well as the ionic dilution data. The PM2+ System is also compatible with the commercially available ECGuide Connector. | | |
| Intended Use/Indications for Use | | 21 CFR 807.92(a)(5) |
| The PM2+ System and SmartPICC Stylet is indicated for guidance and positioning of commercially available central venous catheters. The Stylet provides stiffness for use in placement of the catheter, intravascular capability for ECG detection and recording for catheter guiding and positioning, and intravascular ionic dilution for supplemental catheter navigation. The SmartPICC Stylet, when used with the PM2+ System, provides real-time catheter tip location information by using the patient's physiological (cardiac electrical activity and blood flow) information. When relying on the patient's ECG signal, the PM2+ System and SmartPICC Stylet is indicated for use as an alternative method to fluoroscopy or chest x-ray for central venous catheter tip placement confirmation in adult patients. The SmartPICC Stylet is indicated for use with PICC lines having a minimum lumen diameter of 0.021". | | |
{5}
Limiting but not contraindicated situations for this technique are in patients where alterations of cardiac rhythm change the presentation of the P-wave as in atrial fibrillation, atrial flutter, severe tachycardia and pacemaker driven rhythm. In such patients, who are easily identifiable prior to central venous catheter insertion the use of an additional method is required to confirm catheter tip location.
| Indications for Use Comparison | 21 CFR 807.92(a)(5) |
| --- | --- |
| The indications for use for the predicate device and the subject device are the same. | |
| Technological Comparison | 21 CFR 807.92(a)(6) |
| The subject device is identical to the predicate K200037 SmartPICC System in its use of intravascular ECG for central venous catheter tip location confirmation as well as its optional ionic dilution navigation feature. In addition, the interaction with the PICC is identical between the two devices. There are minor differences between the packaging configuration, accessories, software system and user interface; however these minor differences do not raise different questions of safety and effectiveness.
The subject device is also equivalent to the additional predicate device (PM2 System with ECGuide Connector). There is a minor change in software between the subject PM2+ and predicate PM2 System and both are used with the same ECGuide Connector disposable component. | |
| Non-Clinical and/or Clinical Tests Summary & Conclusions | 21 CFR 807.92(b) |
| The determination of substantial equivalence was based on an assessment of non-clinical performance data. Performance testing performed on the PM2+ System and SmartPICC Stylet verified all device performance requirements including tensile strength, insertion force, corrosion resistance, catheter compatibility, leak testing, infusion rate testing, ionic dilution performance testing, algorithm verification, electrical hardware verification, sterile barrier packaging testing, biocompatibility, electrical safety, EMC, software verification and validation, design validation and human factors validation. All tests were successfully completed. The passing results supported the determination of substantial equivalence with the predicate device.
No human clinical data was provided to support substantial equivalence.
The information included in this premarket notification supports the substantial equivalence of the subject PM2+ System and SmartPICC Stylet to the stated predicate device. The listed reference device also provides substantial equivalence support for the electrical hardware utilized in the PM2+ System. The subject device has the same intended use, similar indications for use and incorporates the same fundamental technology as the legally marketed predicate device and the reference devices to which it was compared. The results of the testing included in this premarket notification support a determination of substantial equivalence. | |
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.