K113428 · Medrad, Inc. · QEZ · Dec 2, 2011 · Cardiovascular
Device Facts
Record ID
K113428
Device Name
ANGIOJET ULTRA DVX THROMBECTOMY SET
Applicant
Medrad, Inc.
Product Code
QEZ · Cardiovascular
Decision Date
Dec 2, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The AngioJet Ultra DVX Thrombectomy Set is intended for use with the AngioJet Ultra Console to break apart and remove thrombus from : upper and lower extremity peripheral arteries ≥ 3.0mm in diameter, . upper extremity peripheral veins ≥ 3.0mm in diameter, ileofemoral and lower extremity veins ≥ 3.0mm in diameter, . A-V access conduits > 3.0mm in diameter and o for use with the AngioJet Ultra Power Pulse Kit for the control and selective infusion of e physician specified fluids, including thrombolytic agents, into the peripheral vascular system.
Device Story
Sterile, single-use disposable thrombectomy catheter and pump unit; used with AngioJet Ultra Console. Operates via mechanical thrombectomy to break apart and remove thrombus from peripheral vasculature; supports selective infusion of fluids/thrombolytics via Power Pulse Kit. Used by physicians in clinical settings. Distal section design modified for improved guide wire robustness. Output is physical thrombus removal and fluid delivery; clinical decision-making informed by real-time procedural use. Benefits include minimally invasive thrombus clearance.
Clinical Evidence
Bench testing only. Testing included catheter operational characteristics, leak testing, guide wire compatibility, tracking, extended use, hemolysis ratio, distal emboli, catheter tip temperature, and mechanical integrity (tensile strength, compression/buckling, torque).
Technological Characteristics
Sterile, single-use disposable catheter and pump unit. Mechanical thrombectomy principle. Designed for use with AngioJet Ultra Console. Includes modifications to distal section for guide wire robustness. Mechanical integrity testing performed for tensile strength, compression, buckling, and torque.
Indications for Use
Indicated for thrombus removal in upper/lower extremity peripheral arteries, upper extremity peripheral veins, ileofemoral/lower extremity veins, and A-V access conduits (all ≥ 3.0mm diameter); also indicated for selective infusion of physician-specified fluids (including thrombolytics) into the peripheral vascular system when used with the AngioJet Ultra Power Pulse Kit.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" above the word "ADMINISTRATION".
February 8, 2021
MEDRAD, Inc. Mike Burnside Manager 9055 Evergreen Blvd NW Minneapolis, Minnesota 55433-8003
Re: K113428
Trade/Device Name: AngioJet Ultra DVX Thrombectomy Set Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: QEZ, KRA
Dear Mike Burnside:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated December 02, 2011. Specifically, FDA is updating this SE Letter as an administrative correction because FDA has created a new product code to better categorize your device technology.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Gregory O'Connell, OHT2: Office of Cardiovascular Devices, (301) 796-6075. Gregory. Oconnell(@FDA.HHS.gov.
Sincerely,
Gregory W. Digitally signed by
Gregory W. Gregory W. O'connell -S ○'connell -S Date: 2021.02.08
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three wavy lines representing the feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
DEC - 2 2011
MEDRAD. Inc. c/o Mike Burnside Manager, Regulatory Affairs 9055 Evergreen Boulevard NW Minneapolis, MN 55433-8003
Re: K113428
Trade/Device Name: Angiojet Ultra DVX Thrombectomy Set Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: DXE, KRA Dated: November 18, 2011 Received: November 21, 2011
Dear Mr. Burnside:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. However, we remind you that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Page 2 - Mr. Burnside
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
W. H. Hillen
Bram Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
{3}------------------------------------------------
### Indications for Use
2113428 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: AngioJet® Ultra DVX® Thrombectomy Set
### Indications for Use:
The AngioJet Ultra DVX Thrombectomy Set is intended for use with the AngioJet Ultra Console to break apart and remove thrombus from :
- upper and lower extremity peripheral arteries ≥ 3.0mm in diameter, .
- . upper extremity peripheral veins ≥ 3.0mm in diameter,
- ileofemoral and lower extremity veins ≥ 3.0mm in diameter, .
- A-V access conduits > 3.0mm in diameter and o
- for use with the AngioJet Ultra Power Pulse Kit for the control and selective infusion of e physician specified fluids, including thrombolytic agents, into the peripheral vascular system.
Prescription Use X (Part 21 CFR 801 Subpart D)
ﺮ
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Office of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular Devices
| 510(k) Number | K113428 |
|---------------|---------|
|---------------|---------|
{4}------------------------------------------------
# ·K113428
DEC - 2 2011
:
ﭩﺮ
## Section 5 – 510(k) Summary
| Submitter: | MEDRAD, Inc.<br>9055 Evergreen Boulevard NW<br>Minneapolis, MN 55433-8003 USA |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Mike Burnside<br>Manager, Regulatory Affairs<br>Phone: (763) 717.1077<br>Fax: (763) 780.2227<br>Email: XXXX |
| Date Prepared: | November 18, 2011 |
| Trade Name: | AngioJet® Ultra DVX® Thrombectomy Set |
| Classification: | 870.5150 and 870.1210 |
| Product Code: | DXE and KRA |
| Predicate Device(s): | The subject device is equivalent to the following devices:<br>• K091593 AngioJet Ultra DVX Thrombectomy Set<br>• K101406 AngioJet Solent Proxi Thrombectomy Set |
| Device Description: | AngioJet Ultra DVX Thrombectomy Set is a sterile, single use,<br>disposable set that includes a Thrombectomy Catheter and Pump in<br>one combined unit. The AngioJet Ultra DVX Thrombectomy Set is<br>used with the AngioJet Ultra Console. |
| Intended Use: | The AngioJet Ultra DVX Thrombectomy Set is intended for use with<br>the AngioJet Ultra Console to break apart and remove thrombus from:<br>• upper and lower extremity peripheral arteries ≥ 3.0mm in<br>diameter,<br>• upper extremity peripheral veins ≥ 3.0mm in diameter,<br>• ileofemoral and lower extremity veins ≥ 3.0mm in diameter,<br>• A-V access conduits ≥ 3.0mm in diameter and<br>• for use with the AngioJet Ultra Power Pulse Kit for the control<br>and selective infusion of physician specified fluids, including<br>thrombolytic agents, into the peripheral vascular system. |
| Comparison to<br>predicate: | Design changes were made to the distal section of the AngioJet Ultra<br>DVX device to make the device more robust to guide wire use. |
| Performance Data: | Bench testing was performed to support a determination of substantial<br>equivalence to the predicate device. Results from the testing provide<br>assurance that the proposed device conforms to the requirements for<br>its intended use. This included the following testing:<br>• Catheter operational characteristics<br>• Leak testing |
{5}------------------------------------------------
- Guide wire compatibility tests ●
- Tracking .
- Extended use .
- Hemolysis ratio .
- Distal emboli 0
- Catheter tip temperature o
- Mechanical integrity (tensile strengths, compression / . buckling, torque testing)
### MEDRAD considers the AngioJet Ultra DVX Thrombectomy Set to Conclusion: be substantially equivalent to the predicate devices listed above. This conclusion is based upon the devices' similarities in functional design, materials, indications for use, and principles of operation.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.