K162777 · Merit Medical Systems, Inc. · QEY · Oct 31, 2016 · Cardiovascular
Device Facts
Record ID
K162777
Device Name
Squirt Fluid Delivery System
Applicant
Merit Medical Systems, Inc.
Product Code
QEY · Cardiovascular
Decision Date
Oct 31, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Squirt Fluid Delivery System is intended for the controlled administration of thrombolytic agents into the peripheral vasculature.
Device Story
Hand-held instrument for consistent, forceful, pulsed injections during thrombolysis procedures; attaches to syringe and catheter to infuse thrombolytic agents into peripheral vasculature. Device facilitates controlled delivery of medication to aid in clot dissolution. Used by clinicians in clinical settings. Benefits include optimized thrombolysis through controlled, pulsed administration.
Clinical Evidence
Bench testing only. Performance testing conducted per ISO 8536-4, ISO 8536-10, and ISO 10993 series. Tests included rotational torque, hydrostatic pressure, vacuum leak, particulate, leakage, and comprehensive biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemocompatibility). Results demonstrated compliance with all predetermined acceptance criteria.
Technological Characteristics
Hand-held fluid delivery system. Materials evaluated per ISO 10993 and ASTM F756-13. Sterilization via ethylene oxide (ISO 11135). Mechanical components include a rotator assembly with an O-ring. No electronic or software components.
Indications for Use
Indicated for the controlled administration of thrombolytic agents into the peripheral vasculature.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
K981417 — Squirt Fluid Delivery System · Merit Medical Systems, Inc. · Jul 17, 1998
K111182 — ANGIOJET SOLENT OMNI THROMBECTOMY SET · Medrad, Inc. · May 24, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines extending from the profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 31, 2016
Merit Medical Systems, Inc. Ms. Angela Brady Senior Regulatory Affairs Specialist 1600 West Merit Parkway South Jordan, UT 84095
Re: K162777
Trade/Device Name: Squirt Fluid Delivery System Regulation Number: 21 CFR 870.1210 Regulation Name: Continuous Flush Catheter Regulatory Class: Class II Product Code: KRA Dated: September 30, 2016 Received: October 4, 2016
Dear Ms. Brady:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Brian D. Pullin -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K162777
Device Name Squirt Fluid Delivery System
Indications for Use (Describe)
The Squirt Fluid Delivery System is intended for the controlled administration of thrombolytic agents into the peripheral vasculature.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
| General<br>Provisions | Submitter Name:<br>Address:<br>Telephone Number:<br>Fax Number:<br>Contact Person:<br>Date of Preparation:<br>Registration Number: | Merit Medical Systems, Inc.<br>1600 West Merit Parkway<br>South Jordan, UT 84095<br>(801) 316-4818<br>(801) 316-4878<br>Angela Brady<br>September 30, 2016<br>1721504 |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject Device | Trade Name:<br>Common/Usual Name:<br>Classification Name:<br>Regulatory Class:<br>Product Code:<br>21 CFR §:<br>Review Panel: | Squirt® Fluid Delivery System<br>Fluid Delivery System<br>Catheter, Continuous Flush<br>II<br>KRA<br>870.1210<br>74 Cardiovascular |
| Predicate<br>Device | Trade Name:<br>Classification Name:<br>Premarket Notification:<br>Manufacturer: | Squirt® Fluid Delivery System<br>Catheter, Continuous Flush<br>K981417<br>Merit Medical Systems, Inc. |
| The predicate has not been subject to a design-related recall. | | |
| Device<br>Description | The Squirt® Fluid Delivery System is a hand held instrument to provide<br>consistent, forceful, pulsed injections for optimal thrombolysis<br>procedures. It's designed to be attached to a syringe and a catheter, to<br>infuse controlled administration of thrombolytic agents into the<br>peripheral vasculature. | |
| Indications for<br>Use | The Squirt® Fluid Delivery System is intended for the controlled<br>administration of thrombolytic agents into the peripheral vasculature.<br>There is no change in the Indications for Use Statement from the<br>predicate to the subject device. | |
| Comparison to<br>Predicate<br>Device | The technological characteristics of the subject Squirt® Fluid Delivery<br>System are identical to the predicate device. Both devices use the<br>same components and materials, with the exception of the O-Ring<br>within the rotator assembly that is bonded on the end of the Squirt®<br>Fluid Delivery System which has undergone a chemical formulation<br>change in the subject device. The indications for use, principle of<br>operation, and technological characteristics of the subject device are<br>identical to the predicate device.<br>Both devices have the same mode of operation and indication for use. | |
| Performance<br>Data | No performance standards have been established under Section 514 of<br>the Food, Drug and Cosmetic Act for these devices. Performance<br>testing of the subject Squirt® Fluid Delivery System was conducted<br>based on the risk analysis and based on the requirements of the<br>following international standards:<br>● ISO 8536-4:2010, Infusion equipment for medical use - Part 4:<br>Infusion sets for single use, gravity feed<br>● ISO 8536-10:2015, Infusion equipment for medical use - Part 10:<br>Accessories for fluid lines for single use with pressure infusion<br>equipment<br>● ISO 11135:2014, Sterilization of health care products – Ethylene<br>oxide - Requirements for development, validation and routine<br>control of a sterilization process for medical devices<br>● ISO 10993-1:2009, Biological Evaluation of Medical Devices Part<br>1: Evaluation and Testing within a risk management process, and<br>FDA guidance Required Biocompatibility Training and Toxicology<br>Profiles for Evaluation of Medical Devices, May 1, 1995<br>● ISO 10993-4:2002 (Amd.1:2006), Biological evaluation of medical<br>devices – Part 4: Selection of tests for interaction with blood<br>● ISO 10993-5:2009, Biological evaluation of medical devices - Part<br>5: Tests for in vitro cytotoxicity<br>● ISO 10993-7:2008, Biological evaluation of medical devices – Part<br>7: Ethylene oxide sterilization residuals<br>● ISO 10993-10:2010, Biological evaluation of medical devices -<br>Part 10: Tests for irritation and skin sensitization<br>● ISO 10993-11:2006, Biological evaluation of medical devices -<br>Part 11: Tests for systemic toxicity<br>● ASTM F756-13, Standard Practice for Assessment of Hemolytic<br>Properties of Materials<br>● United States Pharmacopeia 38, National Formulary 33, 2015<br><151> Pyrogen Test<br>● AMMI TIR 28:2009, Product adoption and process equivalency for<br>ethylene oxide sterilization | |
| • Merit Rotational Torque Test<br>• Merit Hydrostatic Pressure Test<br>• Merit Vacuum Leak Test<br>• ISO 8536-4 Chemical Requirements<br>• ISO 8536-10 Particulate<br>• ISO 8536-10 Leakage | Biocompatibility<br>• Cytotoxicity<br>• Sensitization<br>• Irritation<br>• Acute Systemic Toxicity<br>• Pyrogenicity<br>• Hemocompatibility<br>• Chemical Characterization | |
| Safety &<br>Performance<br>Tests cont. | | |
| The results of the testing demonstrated that the subject Squirt® Fluid<br>Delivery System met the predetermined acceptance criteria and thus is<br>substantially equivalent to the predicate device. | | |
| Summary of<br>Substantial<br>Equivalence | Based on the indications for use, design, and performance testing, the<br>subject Squirt® Fluid Delivery System meets the requirements that are<br>considered essential for its intended use and is substantially equivalent<br>to the predicate device, the Squirt® Fluid Delivery System, K981417<br>manufactured by Merit Medical Systems, Inc. | |
{4}------------------------------------------------
{5}------------------------------------------------
### Performance Testing
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.