← Product Code [QEY](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/QEY) · K162771

# EkoSonic Endovascular System with Control Unit 4.0 (K162771)

_Btg International, Inc. · QEY · Feb 24, 2017 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/QEY/K162771

## Device Facts

- **Applicant:** Btg International, Inc.
- **Product Code:** [QEY](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/QEY.md)
- **Decision Date:** Feb 24, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5150
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The EkoSonic® Endovascular System is indicated for the: - Ultrasound facilitated, controlled and selective infusion of physician-specified fluids, including thrombolytics, into the vasculature for the treatment of pulmonary embolism. - Infusion of solutions into the pulmonary arteries. - Controlled and selective infusion of physician-specified fluids, into the peripheral vasculature.

## Device Story

System comprises disposable infusion catheter with removable ultrasound core and Control Unit 4.0 (CU 4.0). Catheter features multiple side holes for fluid delivery; ultrasound core contains up to 30 elements for energy delivery. Thermal sensors in treatment zone monitor temperature. CU 4.0 generates/controls radiofrequency energy delivery to ultrasound core; monitors/controls treatment zone temperature to limit surface temperature to 43°C. Used in clinical settings by physicians to facilitate thrombolytic agent delivery into vascular blood clots. CU 4.0 provides power and user interface for operator control. System includes internal lithium-ion backup battery. Output power limited by software to ~50W. Benefits include enhanced thrombolytic delivery via ultrasound facilitation.

## Clinical Evidence

Bench testing only. No clinical data provided. Verification testing included hardware (system tip, battery, electrical safety, EMC/EMI, RF board, environmental) and software (self-test, UI, firmware, event logging, system connect, RF channel, graphing, therapy support, workflow, startup, communication, source code inspection) performance tests. All tests passed.

## Technological Characteristics

Materials: Aluminum chassis/front panel (powder-coated or clear anodized). Energy: Radiofrequency energy for ultrasound transducers. Temperature control: Feedback loop limits surface temperature to 43°C. Power: Internal lithium-ion backup battery; software-limited output (~50W). Connectivity: Standalone control unit. Software: Firmware-based control system.

## Regulatory Identification

An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.

## Predicate Devices

- EKOS EkoSonic Endovascular System with the EkoSonic Control Unit (PT-3B) ([K140151](/device/K140151.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three abstract human figures connected by flowing lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

## February 24, 2017

BTG International, Inc. Curtis Jensen Senior Regulatory Affairs Associate 11911 N Creek Pkwy S Bothell, Washington 98011

Re: K162771

Trade/Device Name: EkoSonic Endovascular System with Control Unit 4.0 Regulation Number: 21 CFR 870.1210 Regulation Name: Continuous Flush Catheter Regulatory Class: Class II Product Code: KRA Dated: January 18, 2017 Received: January 23, 2017

Dear Curtis Jensen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K162771

### Device Name

EkoSonic Endovascular System with Control Unit 4.0

Indications for Use (Describe)

The EkoSonic® Endovascular System is indicated for the:

- · Ultrasound facilitated, controlled and selective infusion of physician-specified fluids, including thrombolytics, into the vasculature for the treatment of pulmonary embolism.
- · Infusion of solutions into the pulmonary arteries.
- · Controlled and selective infusion of physician-specified fluids, into the peripheral vasculature.

| Type of Use (Select one or both, as applicable)                                                                   |                                                                                                                  |
|-------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Section 5: 510(k) Summary

#### Submitter l.

EKOS Corporation, Inc. 11911 North Creek Parkway S Bothell, WA 98011

Phone: 425-415-3122 Fax: 425-415-3105

Contact Person: Curtis Jensen Date Prepared: September 30, 2016

#### ll. Device

EkoSonic® Endovascular System with Control Unit 4.0 Proprietary Name: Common or Usual Name: Continuous Flush Catheter Catheter, Continuous Flush (21 CFR §870.1210) Classification Name: Product Code: KRA

Continuous Flush Catheters have been classified by the FDA Cardiovascular Panel as Class II.

#### ���. Predicate Devices

The EkoSonic Endovascular System with Control Unit 4.0 (CU 4.0) is substantially equivalent to another legally marketed device. This predicate device is the EKOS EkoSonic Endovascular System with the EkoSonic Control Unit (PT-3B) (K140151).

This predicate device has been subject to one design-related Class II recall. Recall Z-1752-2008 was determined to be Software Design related. The device design used as a predicate device for this submission represents the most current software and hardware design.

No reference devices were used in this submission.

#### IV. Device Description

The EkoSonic Endovascular System consists of an EkoSonic Device and the Control Unit 4.0 (CU 4.0). The EkoSonic Device consists of a disposable infusion catheter with removable ultrasound core. The infusion catheter contains multiple side holes distributed over the length of the treatment zone. The ultrasound core contains up to

{4}------------------------------------------------

30 ultrasound elements, evenly spaced over the treatment zone. Thermal sensors in the treatment zone monitor catheter temperature. The CU 4.0 generates and controls the delivery of radiofrequency energy to the ultrasound core while monitoring and controlling the temperature of the treatment zone.

It is the modifications to the control unit that are the subject of this submission.

#### V. Intended Use/Indications for use

The EkoSonic® Endovascular System is indicated for the:

- · Ultrasound facilitated, controlled and selective infusion of physician-specified fluids, including thrombolytics, into the vasculature for the treatment of pulmonary embolism.
- · Infusion of solutions into the pulmonary arteries.
- Controlled and selective infusion of physician-specified fluids, including thrombolytics, into the peripheral vasculature.

|                           | Predicate Devices                                                                                                                                                                                                                                                                                                                                                                                                                                                | Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name               | EkoSonic Endovascular System with PT3-B                                                                                                                                                                                                                                                                                                                                                                                                                          | EkoSonic Endovascular System with CU 4.0                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| 510(k) Number(s)          | K140151 - Amendment 002                                                                                                                                                                                                                                                                                                                                                                                                                                          | To be assigned                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Product Code              | KRA                                                                                                                                                                                                                                                                                                                                                                                                                                                              | KRA<br>(This is equivalent to the predicate<br>device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Indications for Use       | The EkoSonic® Endovascular System is<br>indicated for the:<br>• Ultrasound facilitated, controlled and<br>selective infusion of physician-<br>specified fluids, including<br>thrombolytics, into the vasculature for<br>the treatment of pulmonary embolism.<br>• Infusion of solutions into the<br>pulmonary arteries.<br>• Controlled and selective infusion of<br>physician-specified fluids, including<br>thrombolytics, into the peripheral<br>vasculature. | The EkoSonic® Endovascular System is<br>indicated for the:<br>• Ultrasound facilitated, controlled and<br>selective infusion of physician-<br>specified fluids, including<br>thrombolytics, into the vasculature for<br>the treatment of pulmonary embolism.<br>• Infusion of solutions into the<br>pulmonary arteries.<br>• Controlled and selective infusion of<br>physician-specified fluids, including<br>thrombolytics, into the peripheral<br>vasculature.<br>(These are equivalent to the predicate<br>device) |
| Principle of<br>Operation | The EkoSonic Endovascular System<br>employs ultrasound to facilitate the delivery<br>of thrombolytic agents into vascular blood<br>clots.                                                                                                                                                                                                                                                                                                                        | The EkoSonic Endovascular System<br>employs ultrasound to facilitate the delivery<br>of thrombolytic agents into vascular blood<br>clots.<br>(This is equivalent to the predicate<br>device)                                                                                                                                                                                                                                                                                                                          |

#### VI. Comparison of Technological Characteristics with the Predicate Devices

{5}------------------------------------------------

|                                        | Predicate Devices                                                                                                                               | Proposed Device                                                                                                                                                                           |
|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Energy Source                          | The EkoSonic Control Unit (PT-3B) provides<br>the power and user interface for operator<br>control.                                             | The upgraded EkoSonic Control Unit (CU<br>4.0) provides the power and user interface<br>for operator control.                                                                             |
|                                        |                                                                                                                                                 | (This is equivalent to the predicate<br>device)                                                                                                                                           |
| Battery Backup                         | No internal battery backup                                                                                                                      | Internal, lithium-ion backup battery                                                                                                                                                      |
| Maximum Output<br>Power Limit          | Power limited to ~50 W Pulses                                                                                                                   | Power is available for ~100W Pulses. The<br>power output is limited by software to ~50W,<br>which is identical to the predicate device.                                                   |
| Maximum EkoSonic<br>Device Temperature | Temperature monitoring, feedback and<br>control system limits the surface<br>temperature of the IDDC to 43°C during<br>operation.               | Temperature monitoring, feedback and<br>control system limits the surface temperature<br>of the IDDC to 43°C during operation.<br>(This is equivalent to the predicate<br>device)         |
| Accessories                            | The predicate control unit is designed to<br>provide electrical energy to drive the<br>ultrasound transducers in EKOS Corporation<br>catheters. | The Control Unit 4.0 is designed to provide<br>electrical energy to drive the ultrasound<br>transducers in EKOS Corporation catheters.<br>(This is equivalent to the predicate<br>device) |
| Materials                              | Chassis and Front Panel: Die Cast and<br>sheet Aluminum, Painted                                                                                | Chassis - Sheet aluminum, powder coated<br>or clear anodized<br>Front Panel - Black Powder-coated<br>Aluminum                                                                             |

The device modifications described in the submission do not affect the intended use, indications for use or the technological characteristics for the EkoSonic System.

#### VII. Performance Data

Testing has confirmed that the EkoSonic Endovascular System with CU 4.0 functions as intended and is substantially equivalent to the predicate device.

| Hardware Testing Performed         |      |
|------------------------------------|------|
| System Tip Test                    | Pass |
| Secondary Cell Battery Test        | Pass |
| Reliability Prediction             | Pass |
| Battery Life Prediction            | Pass |
| Electrical Safety                  | Pass |
| Hardware Inspection                | Pass |
| EMC/EMI                            | Pass |
| RF Board Tests                     | Pass |
| System Integration                 | Pass |
| CIC Hardware Tests                 | Pass |
| Shipping Package Tests             | Pass |
| Design Verification Hardware Tests | Pass |
| UI Board Hardware Tests            | Pass |

{6}------------------------------------------------

| Hardware Testing Performed    |      |
|-------------------------------|------|
| Environmental Tests           | Pass |
| Power Management Module Tests | Pass |

| Software Testing Performed |      |
|----------------------------|------|
| Periodic Self-Test         | Pass |
| User Interface             | Pass |
| Firmware Upgrade           | Pass |
| Event Logging              | Pass |
| System Connect Validation  | Pass |
| Single Channel RF          | Pass |
| Graphing                   | Pass |
| Device Compatibility       | Pass |
| Dual Channel RF            | Pass |
| Therapy Support            | Pass |
| Workflow                   | Pass |
| System Startup             | Pass |
| Processor Communication    | Pass |
| Constant Parameter         | Pass |
| Source Code Inspection     | Pass |
| AC and Battery Power       | Pass |

Performance standards have not been promulgated for Continuous Flush Catheters.

#### VIII. Conclusions

The EkoSonic Endovascular System with Control Unit 4.0 is substantially equivalent to the predicate device. The modifications in the control unit do not affect the intended use or the technological characteristics for the system and do not raise different questions of safety or effectiveness.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/QEY/K162771](https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/QEY/K162771)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
