← Product Code [QEY](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/QEY) · K111705

# EKOSONIC ENDOVASCULAR SYSTEM (K111705)

_Ekos Corporation · QEY · Aug 3, 2011 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/QEY/K111705

## Device Facts

- **Applicant:** Ekos Corporation
- **Product Code:** [QEY](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/QEY.md)
- **Decision Date:** Aug 3, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5150
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The EkoSonic Endovascular System is intended for the infusion of solutions into the pulmonary arteries.

## Device Story

The EkoSonic Endovascular System is an infusion device designed to deliver fluids into the pulmonary arteries. It utilizes a multi-sidehole catheter containing a MicroSonic device, which features linearly distributed ultrasound transducers. The system is operated by physicians in a clinical setting to deliver physician-specified agents or fluids. The ultrasound energy is intended to facilitate the delivery of these agents. The device is used for peripheral vasculature applications. Clinical benefit is derived from the targeted infusion of therapeutic agents.

## Clinical Evidence

No clinical data provided. Evidence consists of preclinical bench and animal studies demonstrating that the device meets design specifications and is safe and effective for its intended use.

## Technological Characteristics

Infusion device consisting of a multi-sidehole catheter and a MicroSonic device containing linearly distributed ultrasound transducers. Operates via ultrasound-assisted fluid delivery. Class II device.

## Regulatory Identification

An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.

## Predicate Devices

- EKOS EkoSonic Endovascular System ([K081467](/device/K081467.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services-USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a recognizable symbol of the agency responsible for regulating food and drugs in the United States.

November 22, 2021

EKOS Corporation Jocelyn Kersten VP. Quality Assurance, Clinical Regulatory Affairs 11911 North Creek Parkway South Bothell, Washington 98011

Re: K111705

Trade/Device Name: EkoSonic Endovascular System Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: QEY, KRA

Dear Jocelyn Kersten:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated August 3, 2011. Specifically. FDA is updating this SE Letter as an administrative correction because FDA has created a new product code to better categorize your device technology.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Gregory O'Connell, OHT2: Office of Cardiovascular Devices, (301) 796-6075, Gregory.Oconnell(@FDA.HHS.gov.

Sincerely,

Image /page/0/Picture/9 description: The image shows a digital signature. The signature is from Gregory W. O'connell -S. The date of the signature is 2021.11.22, and the time is 13:39:32 -05'00'.

Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and tail feathers. The eagle is enclosed in a circle of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MI) 20993-0002

EKOS Corporation c/o Jocclyn Kersten 11911 North Creek Parkway South Bothell, WA 9801 1

AUG - 3 2011

Re: K111705

Trade/Device Name: EkoSonic Endovascular System Regulation Number: 21 CFR 870.1210 Regulation Name: Continuous Flush Catheter Regulatory Class: Class II (two) Product Code: KRA Dated: June 17, 2011 Received: June 17, 2011

Dear Ms. Kersten:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Please be advised that I'DA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

/
Bram D. Zuckerm

- Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## 1.10 Indications for Use Statement

EKOS Corporation 11911 North Creek Parkway South Bothell, WA 98011

## Indications for Use Statement

510(k) Number (if known):

Device Name: EkoSonic Endovascular System

Indications for Use:

The EkoSonic Endovascular System is intended for the infusion of solutions into the pulmonary arteries.

| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------------------------------|
|--------------------------------------------------------|

| Prescription Use (Per 21 CFR 801.109) | X |
|---------------------------------------|---|
| OR                                    |   |
| Over-The-Counter Use                  |   |

| Division Sign-Off                  |
|------------------------------------|
| Division of Cardiovascular Devices |

| 510(k) Number | K111705 |
|---------------|---------|
|---------------|---------|

{4}------------------------------------------------

K111705

AUG - 3 2011

## 510(k) Summary 8.0

-

",

| Applicant Name:     | EKOS Corporation                                                                                                                                                                                                                                                                                                                                                                                                         |  |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Address:            | 11911 North Creek Parkway South<br>Bothell, WA 98011                                                                                                                                                                                                                                                                                                                                                                     |  |
| Contact Person:     | Jocelyn Kersten<br>Vice President, Quality Assurance, Regulatory and Clinical Affairs                                                                                                                                                                                                                                                                                                                                    |  |
| Telephone:          | (425) 415-3132                                                                                                                                                                                                                                                                                                                                                                                                           |  |
| Fax:                | (425) 415-3102                                                                                                                                                                                                                                                                                                                                                                                                           |  |
| Device:             | EkoSonic Endovascular System                                                                                                                                                                                                                                                                                                                                                                                             |  |
| Classification:     | CFR 870.1210 - Continuous Flush Catheter                                                                                                                                                                                                                                                                                                                                                                                 |  |
| Panel:              | Cardiovascular                                                                                                                                                                                                                                                                                                                                                                                                           |  |
| Product Code:       | KRA                                                                                                                                                                                                                                                                                                                                                                                                                      |  |
| Intended Use:       | The EkoSonic Endovascular System is intended for the infusion of<br>solutions into the pulmonary arteries.                                                                                                                                                                                                                                                                                                               |  |
| Device Description: | The EkoSonic Endovascular System is an infusion Device designed to<br>deliver fluids via a multi sidehole catheter. The fluid is dispersed via<br>multiple ultrasound transducers distributed linearly along the length of an<br>MicroSonic Device which is placed into the center lumen of the catheter.<br>This device is intended to deliver physician-specified agents or fluids<br>into the peripheral vasculature. |  |
| Predicate Basis:    | The EkoSonic Endovascular System is substantially equivalent to<br>another legally marketed device. These devices include<br>EKOS EkoSonic Endovascular System (EKOS Corporation,<br>K081467)                                                                                                                                                                                                                            |  |
| Performance:        | EKOS has conducted preclinical bench and animal studies with the<br>EkoSonic Endovascular System. These studies demonstrate that the<br>performance of the EkoSonic Endovascular System meets its design<br>specifications and is safe and effective for its intended use.                                                                                                                                               |  |

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/QEY/K111705](https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/QEY/K111705)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
