← Product Code [QEW](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/QEW) · K223929

# Laguna Clot Retriever System and Malibu Aspiration Catheter System (Laguna Thrombectomy System) (K223929)

_Innova Vascular, Inc. · QEW · Jun 1, 2023 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/QEW/K223929

## Device Facts

- **Applicant:** Innova Vascular, Inc.
- **Product Code:** [QEW](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/QEW.md)
- **Decision Date:** Jun 1, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5150
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Laguna Thrombectomy System is indicated for: - The non-surgical removal of emboli and thrombi from blood vessels. - Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel. The Laguna Thrombectomy System is intended for use in the peripheral vasculature. The Malibu Aspiration Catheter™ System is indicated for: - The non-surgical removal of emboli and thrombi from blood vessels. - Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel. The Malibu Aspiration Catheter™ System is intended for use in the peripheral vasculature.

## Device Story

Laguna Thrombectomy System comprises Laguna Clot Retriever and Malibu Aspiration Catheter. Used in peripheral vasculature for non-surgical thrombus/embolus removal and fluid management. Clot retriever features self-expanding laser-cut Nitinol structure with flexible pusher; delivered via sheath. Aspiration catheter features single-lumen design with metallic reinforcement, PTFE liner, and hydrophilic coating; includes two-way flow control valves for aspiration/infusion. Operated by physicians in clinical settings under fluoroscopic guidance using radiopaque markers. System enables mechanical engagement and retrieval of clots into aspiration catheter via vacuum syringe. Benefits include minimally invasive clot removal, restoring blood flow in peripheral vessels.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing (dimensional, mechanical, integrity, and usability) and GLP-compliant porcine animal studies evaluating radiopacity and safety at acute and chronic time points.

## Technological Characteristics

Materials: Nitinol (retriever), stainless steel reinforcement, PTFE liner, polymeric jacket. Design: Self-expanding laser-cut closed-cell retriever; single-lumen aspiration catheter with hydrophilic coating. Energy: Mechanical. Connectivity: None. Sterilization: Ethylene Oxide. Standards: ISO 10993-1 (biocompatibility), ISO 10993-7 (EO residuals), AAMI/ANSI ST72 (endotoxins).

## Regulatory Identification

An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.

## Predicate Devices

- FlowTriever Retrieval/Aspiration System ([K211013](/device/K211013.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 1, 2023

Innova Vascular, Inc. Ryan Kenney Director, Regulatory Affairs 15375 Barranca Parkway, B-101 Irvine, California 92618

Re: K223929

Trade/Device Name: Laguna Clot Retriever™ System and Malibu Aspiration Catheter™ System (Laguna Thrombectomy System) Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: QEW, KRA Dated: May 2, 2023 Received: May 4, 2023

Dear Ryan Kenney:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Eleni
For
Whatley -S
Digitally signed by
Eleni Whatley -S
Date: 2023.06.01
20:54:07 -04'00'

Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

### 510(k) Number (if known) K223929

#### Device Name

Laguna Clot Retriever™ System and Malibu Aspiration Catheter™ System (Laguna Thrombectomy System )

Indications for Use (Describe)

The Laguna Thrombectomy System is indicated for:

- · The non-surgical removal of emboli and thrombi from blood vessels.
- · Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.

The Laguna Thrombectomy System is intended for use in the peripheral vasculature.

The Malibu Aspiration Catheter™ System is indicated for:

- · The non-surgical removal of emboli and thrombi from blood vessels.
- · Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.

The Malibu Aspiration Catheter™ System is intended for use in the peripheral vasculature.

Type of Use (Select one or both, as applicable)

| <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------------------|
| <span style="text-decoration: overline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)  |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K223929

### 510(k) Summary

| Date Prepared:                | June 1, 2023                                                                                         |
|-------------------------------|------------------------------------------------------------------------------------------------------|
| Device Classification Name:   | Peripheral Mechanical Thrombectomy with Aspiration                                                   |
| Device Name:                  | Laguna Clot Retriever™ System and Malibu Aspiration Catheter™ System<br>(Laguna Thrombectomy System) |
| Applicant:                    | Innova Vascular, Inc.<br>15375 Barranca Pkwy Ste B101<br>Irvine, CA 92618                            |
| Applicant Contact:            | Ryan Kenney<br>Director, Regulatory Affairs                                                          |
| Regulation Number:            | 870.5150                                                                                             |
| Product Code(s):              | QEW, KRA                                                                                             |
| Regulation Medical Specialty: | Cardiovascular                                                                                       |
| Review Panel:                 | Cardiovascular                                                                                       |
| Predicate Device:             | FlowTriever Retrieval/Aspiration System (K211013)                                                    |

### Device Description:

The Laguna Clot Retriever™ System is a single use, sterile, non-pyrogenic medical device designed for use in the peripheral vasculature. The Laguna Clot Retriever™ System is intended be used in conjunction with the Malibu Aspiration Catheter™ System. The combined use of the Laguna Clot Retriever™ System and Malibu Aspiration Catheter™ System is referred to as the Laguna Thrombectomy System.

The Laguna Clot Retriever™ is used to engage and retrieve emboli and thrombi into the Malibu Aspiration Catheter™. The Laguna Clot Retriever™ is designed with a flexible pusher attached to a distal self-expanding, laser cut, closed cell design Nitinol structure. The self-expanding portion is delivered inside a delivery sheath. The flexible pusher is designed with stainless steel reinforcement and a polymer. The retriever head is designed with and without radiopaque body markers. The retriever head designed with radiopaque body markers is intended to visualize the expansion and collapse of the Laguna Clot Retriever™.

The Malibu Aspiration Catheter™ System is a single use, sterile, non-pyrogenic medical device designed for use in the peripheral vasculature. The Malibu Aspiration Catheter™ System is intended be used independently and in conjunction with the Laguna Clot Retriever™ System.

The Malibu Aspiration Catheter™ is intended for retrieving thrombi or emboli and/or infusing fluids into the peripheral vasculature. The Malibu Aspiration Catheter™ is designed with a single lumen and comprised of tubes for aspiration and infusion connected proximally to its hub. Two-way flow control valves are located at the end of aspiration and infusion tubes, respectively. The aspiration tube can be connected to a locking vacuum syringe via its two-way valve. Similarly, the infusion tube can be connected to a syringe to infuse fluids through the catheter. The Malibu Aspiration Catheter™ is designed with a radiopaque marker band at the distal tip for fluoroscopic visualization. In addition, the Malibu Aspiration Catheter™ is also available with a second marker band located proximally to the distal marker band. The Malibu Aspiration Catheter™ is supplied with a dilator and a 60cc locking

{4}------------------------------------------------

## innova

syringe. The Malibu Aspiration Catheter™ incorporates a metallic reinforcement made of stainless steel, an inner liner made of polytetrafluoroethylene, and polymeric jacket material having gradient softness. The distal segment of the Malibu Aspiration Catheter™ is coated with a hydrophilic coating to optimize lubricity and reduce friction.

Indications for Use Statement:

The Laguna Thrombectomy System is indicated for:

- . The non-surgical removal of emboli and thrombi from blood vessels.
- Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.

The Laguna Thrombectomy System is intended for use in the peripheral vasculature.

The Malibu Aspiration Catheter™ System is indicated for:

- . The non-surgical removal of emboli and thrombi from blood vessels.
- . Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.

The Malibu Aspiration Catheter™ System is intended for use in the peripheral vasculature.

{5}------------------------------------------------

# เทกปังค

Comparison of Technological Characteristics:

|                                                                           | FlowTriever Retrieval/Aspiration System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Laguna Clot Retriever™ System/Malibu Aspiration Catheter™ System<br>(Laguna Thrombectomy System)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|---------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                                           | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| 510(k) Number:                                                            | K211013                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | K223929                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Product Code(s):                                                          | QEW, KRA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same as K211013                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Indications for Use<br>Statement:                                         | The FlowTriever Retrieval/Aspiration System is indicated for:<br>The non-surgical removal of emboli and thrombi from blood vessels. Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.          The FlowTriever Retrieval/Aspiration System is intended for use in the peripheral vasculature and for the treatment of pulmonary embolism.<br><br>Triever Catheters (Triever 16, Triever 20, Triever 20 Curve, and Triever 24) are also intended for use in treating clot in transit in the right atrium but not in conjunction with FlowTriever Catheters. | The Laguna Thrombectomy System is indicated for:<br>The non-surgical removal of emboli and thrombi from blood vessels. Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.          The Laguna Thrombectomy System is intended for use in the peripheral vasculature.<br><br>The Malibu Aspiration Catheter™ System is indicated for:<br>The non-surgical removal of emboli and thrombi from blood vessels. Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.          The Malibu Aspiration Catheter™ is intended for use in the peripheral vasculature |
|                                                                           | FlowTriever Catheter/Laguna Clot Retriever™ System – Dimensional                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Size<br>(Vessel Diameter Range)                                           | S (6 – 10 mm)<br>M (11 – 14 mm)<br>L (15 – 18 mm)<br>XL (19 – 25 mm)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 10 mm (≥ 6 mm)<br>14 mm (≥ 10 mm)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Outer Diameter                                                            | 4 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 10 mm: 3.1 mm<br>14 mm: 4 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                                                           | FlowTriever Retrieval/Aspiration System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Laguna Clot Retriever™ System/Malibu Aspiration Catheter™ System<br>(Laguna Thrombectomy System)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                                                           | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Usable Length                                                             | 115 cm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 10 mm: 127 cm<br>14 mm: 119 cm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Guidewire Compatibility                                                   | 0.035"                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                                                                           | FlowTriever Catheter/Laguna Clot Retriever™ System - Material                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Self-Expanding Nitinol<br>Retriever                                       | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Polymeric Delivery<br>Sheath                                              | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                                                                           | FlowTriever Catheter/Laguna Clot Retriever™ System - Supplied                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Non-Pyrogenic                                                             | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Sterile                                                                   | Ethylene Oxide                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Single Use                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                                                                           | Triever Aspiration Catheter/Malibu Aspiration Catheter™ System - Dimensional                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Outer Diameter                                                            | 16 Fr                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 12 Fr                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                                           | 20 Fr                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 16 Fr                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                                           | 24 Fr                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 20 Fr                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 24 Fr                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Usable length                                                             | 16 Fr: 107 cm<br>20 Fr: 90 cm<br>24 Fr: 90 cm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 12 Fr: 115 cm<br>16 Fr: 107 cm<br>20 Fr: 100 cm<br>24 Fr: 90 cm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Guidewire Compatibility                                                   | 0.035"                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                                                                           | Triever Aspiration Catheter/Malibu Aspiration Catheter™ System - Material                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Polymeric Jacket with<br>Multiple Durometers                              | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Metallic Reinforcement                                                    | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Liner                                                                     | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                                                                           | Triever Aspiration Catheter/Malibu Aspiration Catheter™ System - Packaged With                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Dilator                                                                   | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Syringe                                                                   | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                                                                           | FlowTriever Retrieval/Aspiration System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Laguna Clot Retriever™ System/Malibu Aspiration Catheter™ System<br>(Laguna Thrombectomy System)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                                                           | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Triever Aspiration Catheter/Malibu Aspiration Catheter™ System - Supplied |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Non-Pyrogenic                                                             | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Sterile                                                                   | Ethylene Oxide                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Single Use                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |

{6}------------------------------------------------

# เททปองค

K223929

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "innova" in a stylized font. The letters are brown, and the "o" is replaced with a brown teardrop shape with an orange circle inside. The teardrop shape is slightly taller than the other letters.

{8}------------------------------------------------

### INNOVA

### Sterilization:

The Laguna Clot Retriever™ System and the Malibu Aspiration Catheter™ System are ethylene oxide sterilized. The Laguna Clot Retriever™ System and the Malibu Aspiration Catheter™ System met the requirements for ethylene oxide and ethylene chlorohydrin residuals per ISO 10993-7:2008/Amd1:2019. In addition, the Laguna Clot Retriever™ System and the Malibu Aspiration Catheter™ System met the requirements for bacterial endotoxin per AAMI/ANSI ST72:2019.

### Shelf-Life:

The Laguna Clot Retriever™ System and the Malibu Aspiration Catheter™ System have a shelf-life of one (1) year.

### Biocompatibility:

Biocompatibility was conducted for the Laguna Clot Retriever™ System and the Malibu Aspiration Catheter™ System to meet the requirements of ISO 10993-1:2018 and Guidance for Industry and Food and Drug Administration Staff, Use of International Standard ISO 10993-1. The following biological endpoints were evaluated:

- Cytotoxicity
- Sensitization
- Intracutaneous Reactivity ●
- Acute Systemic Toxicity ●
- Material-Mediated Pyrogenicity
- Hemocompatibility
	- o Hemolysis
	- o Complement Activation
	- o Thrombogenicity

### Performance Data - Bench:

The Laguna Clot Retriever™ System was evaluated for the following:

- Dimensional Verification ●
- Visual Inspection
- Tensile Strength ●
- Turns to Failure ●
- Coating Integrity
- Corrosion Resistance
- Kink to Failure ●
- Liquid Leakage ●
- Air Leakage into Hub Assembly
- Radial Force
- Unsheathing and Resheathing Force ●
- Unsheathing and Resheathing Integrity
- Luer Fittings .

The Malibu Aspiration Catheter™ System was evaluated for the following:

- Dimensional Verification
- Visual Inspection
- Tensile Strength
- Torque Transmission ●
- Turns to Failure
- Coating Integrity
- . Corrosion Resistance
- Kink to Failure
- Liquid Leakage ●
- Air Leakage into Hub Assembly ●
- Lumen Collapse
- Luer Fittings ●

{9}------------------------------------------------

### innova

The Laguna Thrombectomy System was evaluated for the following:

- . Particulate Evaluation
- Usability
- . Simulated Use
### Performance Data – Animal:

The radiopacity and safety of the Laguna Thrombectomy System was evaluated at acute and chronic time points in a porcine model and was conducted in accordance with 21 CFR Part 58 for Good Laboratory Practice (GLP) for Non-Clinical Laboratory Studies.

### Performance Data – Clinical:

A determination of substantial equivalence is based upon successful completion of performance data – bench and animal.

### Conclusion:

The Laguna Thrombectomy System and Malibu Aspiration Catheter™ System are substantially equivalent to the FlowTriever Retrieval/Aspiration System.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/QEW/K223929](https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/QEW/K223929)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
