Indigo® Aspiration System – Lightning Bolt Aspiration Tubing
K223186 · Penumbra, Inc. · QEW · Feb 13, 2023 · Cardiovascular
Device Facts
Record ID
K223186
Device Name
Indigo® Aspiration System – Lightning Bolt Aspiration Tubing
Applicant
Penumbra, Inc.
Product Code
QEW · Cardiovascular
Decision Date
Feb 13, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
INDIGO Aspiration Catheters and Separators: As part of the INDIGO Aspiration System, the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems, and for the treatment of pulmonary embolism. INDIGO Aspiration Tubing: As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Aspiration Pump. Penumbra Aspiration Pump: The Penumbra Aspiration Pump is indicated as a vacuum source for the Penumbra Aspiration Systems.
Device Story
Lightning Bolt Aspiration Tubing acts as a conduit between Penumbra Aspiration Pump and Indigo Aspiration Catheter; facilitates vacuum transfer for thrombus removal. Enables intermittent, continuous, or modulated aspiration; modulated mode alternates connection between catheter and sterile saline bag at ambient pressure. Used in clinical settings by physicians; connects to existing Indigo System components. Output is vacuum-mediated aspiration of thrombus from occluded vessels. Benefits include effective thrombus removal from peripheral vasculature and pulmonary arteries. Device is single-use and disposable.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench-top testing including dimensional/visual inspection, pressurization, simulated use (thrombus removal), valve testing, and tensile testing. Biocompatibility testing (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemocompatibility) confirmed safety. Electrical safety and EMC testing performed per IEC 60601-1 and IEC 60601-1-2.
Technological Characteristics
Biocompatible materials; single-use, sterile (EO). Dimensions: ID 0.125 in, distal OD 0.225 in, paratubing OD 0.230 in, saline tubing OD 0.225 in, length 120 in. Connectivity: connects catheter to vacuum pump. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, AIM 7351731, and ISO 10079-1.
Indications for Use
Indicated for removal of fresh, soft emboli and thrombi from peripheral arterial and venous systems and treatment of pulmonary embolism in patients requiring mechanical aspiration.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
Predicate Devices
Indigo Aspiration System – Lightning Aspiration Tubing (K210323)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.
February 13, 2023
Penumbra, Inc. Deanna Kimlinger Senior Regulatory Specialist One Penumbra Place Alameda, California 95132
Re: K223186
Trade/Device Name: Indigo® Aspiration System - Lightning Bolt Aspiration Tubing Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: QEW Dated: January 13, 2023 Received: January 17, 2023
# Dear Deanna Kimlinger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Gregory W. Digitally signed by
Gregory W. Gregory W. O'connell -S O'connell -S Date: 2023.02.13
12-05'00'
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223186
Device Name
Indigo® Aspiration System - Lightning Bolt Aspiration Tubing
Indications for Use (Describe)
INDIGO Aspiration Catheters and Separators:
As part of the INDIGO Aspiration System, the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems, and for the treatment of pulmonary embolism.
INDIGO Aspiration Tubing:
As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Aspiration Pump.
Penumbra Aspiration Pump:
The Penumbra Aspiration Pump is indicated as a vacuum source for the Penumbra Aspiration Systems.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### ਹ 510(k) Summary
(as required by 21 CFR 807.92)
Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, Penumbra, Inc. is providing the summary of Substantial Equivalence for the Indigo "Aspiration System - Lightning Bolt Aspiration Tubing.
### 1.1 Sponsor/Applicant Name and Address
Penumbra, Inc. One Penumbra Place Alameda, CA 94502 USA
### 1.2 Sponsor Contact Information
Deanna Kimlinger Senior Regulatory Affairs Specialist Phone: (925) 212-9088 FAX: (510) 217-6414 Email: dkimlinger@penumbrainc.com
### 1.3 Date of Preparation of 510(k) Summary
February 08, 2023
### 1.4 Device Trade or Proprietary Name
Indigo Aspiration System - Lightning Bolt Aspiration Tubing
### 1.5 Device Classification
| Regulatory Class: | II |
|-----------------------|-----------------------|
| Classification Panel: | Cardiovascular |
| Classification Name: | Catheter, Embolectomy |
| Regulation Number: | 21 CFR §870.5150 |
| Product Code: | QEW |
### 1.6 Predicate and Reference Devices – Lightning Bolt Aspiration Tubing
| 510(k) Number | Name of Device | |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|
| Lightning Bolt Aspiration Tubing | | |
| Predicate | K210323 Indigo Aspiration System – Lightning Aspiration Tubing | |
| Catheter 7 Aspiration Catheter | | |
| Reference | K210083 Indigo Aspiration System Catheter 7, Indigo Aspiration System Separator 7,<br>Lightning 7 | |
| System Name | Indigo® Aspiration System | |
| | Lightning Aspiration Tubing<br>[Predicate] | Lightning Bolt Aspiration Tubing¹<br>[Subject] |
| Classification | Class II, QEW | SAME |
| 510(k) no. | K210323 | K223186 |
| Indication | INDIGO Aspiration Catheters and<br>Separators:<br>As part of the INDIGO Aspiration<br>System, the INDIGO Aspiration<br>Catheters and Separators are indicated<br>for the removal of fresh, soft emboli and<br>thrombi from vessels of the peripheral<br>arterial and venous systems, and for the<br>treatment of pulmonary embolism.<br><br>INDIGO Aspiration Tubing:<br>As part of the INDIGO Aspiration<br>System, the INDIGO Sterile Aspiration<br>Tubing is indicated to connect the<br>INDIGO Aspiration Catheters to the<br>Penumbra Aspiration Pump.<br><br>Penumbra Aspiration Pump:<br>The Penumbra Aspiration Pump is<br>indicated as a vacuum source for the<br>Penumbra Aspiration Systems. | SAME AS PREDICATE K210323 |
| Aspiration Tubing | Lightning Aspiration Tubing<br>[Predicate] | Lightning Bolt Aspiration Tubing<br>[Subject] |
| 510(k) No. | K210323 | K223186 |
| Materials | | |
| Materials | Biocompatible, commonly utilized for<br>interventional devices | SAME |
| Dimensions | | |
| Tubing Inner Diameter (ID) | 0.125 in MIN | SAME |
| Distal Tubing Outer Diameter<br>(OD) | 0.220 in MIN | 0.225 ± 0.007 in |
| Paratubing Outer Diameter<br>(OD) | 0.223 in MIN | 0.230 + 0.009/- 0.007 in |
| Saline Tubing Outer Diameter<br>(OD) | N/A - Not present | 0.225 ± 0.007 in |
| Overall Length | 100 + 7 in | 120 + 7 in |
| System Name | Indigo® Aspiration System | |
| | Lightning Aspiration Tubing<br>[Predicate] | Lightning Bolt Aspiration Tubing1<br>[Subject] |
| Indigo System Attributes | | |
| Packaging Materials | Commonly used materials for<br>interventional devices | SAME |
| Aspiration Source | Penumbra Aspiration Pump | SAME |
| Sterilization | EO | SAME |
| Shelf-Life | 36 Months | 12 Months |
| Use | Single use, disposable | SAME |
{4}------------------------------------------------
### 1.7 Predicate Comparison
{5}------------------------------------------------
4 The Indigo System Aspiration Catheters, Separators, and Penumbra Aspiration Pumps are unchanged and remain identical to those currently cleared in Section 1.6.
#### 1.8 Device Description
The INDIGO® Aspiration System is comprised of the several devices:
- INDIGO Aspiration Catheter
- Penumbra Aspiration Pump
- · INDIGO Aspiration Pump Canister
- INDIGO Aspiration Tubing
- INDIGO Separator
The INDIGO Aspiration System is designed to remove thrombus from the vasculature using mechanical aspiration. The INDIGO Aspiration Catheter targets aspiration from the pump directly to the thrombus. The INDIGO Separator may be used to clear the lumen of the INDIGO Aspiration Catheter should it become blocked with thrombus. The INDIGO Aspiration Catheter is introduced through a guide catheter or vascular sheath into the peripheral vasculature and guided over a guidewire to the site of the primary occlusion. The INDIGO Aspiration Catheter is used with the Penumbra Aspiration Pump to aspirate thrombus from an occluded vessel. As needed, an INDIGO Separator may be deployed from the INDIGO Aspiration Catheter to assist with thrombus removal. The INDIGO Separator is advanced and retracted through the INDIGO Aspiration Catheter at the proximal margin of the primary occlusion to facilitate clearing of the thrombus from the INDIGO Aspiration Catheter tip. The INDIGO Aspiration Catheter may be provided with a steam shaping mandrel, rotating hemostasis valve, and introducer. The INDIGO Separator may be provided with an introducer and torque devices are visible under fluoroscopy. For the aspiration source, the INDIGO Aspiration Catheter is used in conjunction with the Penumbra Aspiration Pump, which is connected using the INDIGO Aspiration Tubing and the INDIGO Aspiration Pump Canister.
{6}------------------------------------------------
## Lightning Bolt Aspiration Tubing
The Lightning Bolt Aspiration Tubing (INDIGO Aspiration Tubing) is designed to serve as a conduit to assist in thrombus removal, facilitating transfer of vacuum between the Penumbra Aspiration Pump and the INDIGO Aspiration Catheter while providing intermittent, continuous, or modulated aspiration. Modulated aspiration is provided when the Lightning Bolt Aspiration Tubing alternates between connecting the INDIGO Aspiration Catheter to the Penumbra Aspiration Pump and a sterile saline intravenous (IV) bag at ambient pressure.
### 1.9 Indications for Use
## INDIGO Aspiration Catheters and Separators
As part of the INDIGO Aspiration System, the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems, and for the treatment of pulmonary embolism.
## INDIGO Aspiration Tubing
As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Aspiration Pump.
## Penumbra Aspiration Pump
The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.
## 1.10 Summary of Non-Clinical Data/ Performance Data
As required under Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, a summary of any information regarding substantial equivalence of the devices follows. Included in this section are summary descriptions of the testing which substantiates the performance of the subject Lightning Bolt Aspiration Tubing.
## 1.10.1 Biocompatibility
Biocompatibility testing was conducted on the subject Lightning Bolt Aspiration Tubing. The following tests were performed:
- o Cytotoxicity
- Sensitization o
- o Irritation
- O Systemic Toxicity
{7}------------------------------------------------
- Acute Systemic Injection
- I Material Mediated Pyrogenicity
- Hemocompatibility o
- . Indirect contact
The results from the testing performed showed the Lightning Bolt Aspiration Tubing to be biocompatible.
### 1.10.2 Design Verification (Bench-Top) Testing
Non-clinical laboratory testing was performed on the subject Lightning Bolt Aspiration Tubing device to determine substantial equivalence. The following tests were performed:
- O Dimensional/Visual Inspection
- O Pressurization Testing
- Indigo Aspiration System Compatibility/Simulated Use Testing O
- . Thrombus Removal Testing
- O Valve testing
- Tensile Testing O
- Post Destructive Testing Dimensional Inspection o
The in vitro bench-top tests demonstrated that the subject Lightning Bolt Aspiration Tubing met all acceptance criteria and performed similarly to the predicate devices. Performance data demonstrate that the Lightning Bolt Aspiration Tubing device functions as intended and has a safety and effectiveness profile that is similar to the predicate devices.
### 1.10.3 Performance Data - Clinical
No clinical study was conducted as bench and previously performed animal testing was determined sufficient for verification and validation purposes.
### 1.10.4 Electrical Safety/EMC Testing
Electrical Safety and EMC testing were conducted on the subject Lightning Bolt Aspiration Tubing. The subject device complies with the requirements of IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, AIM 7351731, and ISO 10079-1.
### 1.10.5 Software
Software verification and validation testing and documentation for the subject Lightning Bolt Aspiration Tubing was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (issued May 11, 2005).
{8}------------------------------------------------
# 1.11 Summary of Substantial Equivalence
The subject Lightning Bolt Aspiration Tubing is substantially equivalent to the predicate devices, provided in Section 1.6 with regards to intended use, operating principle, design concept, materials, sterilization processes and packaging processes.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.