← Product Code [QEW](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/QEW) · K211492

# Pounce Thrombectomy System (K211492)

_Surmodics,Inc. · QEW · Jul 30, 2021 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/QEW/K211492

## Device Facts

- **Applicant:** Surmodics,Inc.
- **Product Code:** [QEW](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/QEW.md)
- **Decision Date:** Jul 30, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5150
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Pounce™ Thrombectomy System is intended for the non-surgical removal of thrombi and emboli from the peripheral arterial vasculature.

## Device Story

Percutaneous catheter system for mechanical thrombus removal in peripheral arterial vasculature; comprises 5Fr basket delivery catheter, basket wire assembly with two distal self-expanding baskets, and trumpet assembly for collection/retrieval; includes basket loading tool. Operated by physician in clinical setting; introduced via minimum 7Fr guide sheath. Baskets capture thrombus; trumpet assembly facilitates retrieval. Radiopaque components allow visualization under fluoroscopy. System enables non-surgical intervention, potentially reducing need for invasive surgery and improving patient outcomes in arterial occlusion cases.

## Clinical Evidence

No clinical data required. Substantial equivalence supported by non-clinical bench testing (basket cage radial force, simulated use) and animal testing previously submitted under K192814. All pre-defined acceptance criteria met.

## Technological Characteristics

Percutaneous catheter system; 5Fr delivery catheter; self-expanding nitinol-based basket wire assembly (two baskets in series); trumpet assembly for retrieval. Radiopaque markers for visualization. Compatible with 7Fr guide sheaths. Mechanical operation; no energy source required. Non-implantable.

## Regulatory Identification

An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.

## Predicate Devices

- Pounce™ Thrombectomy System ([K192814](/device/K192814.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.

July 30, 2021

Surmodics, Inc. Holly Ramirez Senior Principal Regulatory Affairs Specialist 9924 W 74th Street Eden Prairie, Minnesota 55344

# Re: K211492

Trade/Device Name: Pounce™ Thrombectomy System Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: QEW Dated: June 9, 2021 Received: June 10, 2021

# Dear Holly Ramirez:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K211492

Device Name Pounce™ Thrombectomy System

Indications for Use (Describe)

The Pounce™ Thrombectomy System is intended for the non-surgical removal of thrombi and emboli from the peripheral arterial vasculature.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Surmodics. The logo consists of a blue and green geometric shape on the left, followed by the word "SURMODICS" in a sans-serif font. The geometric shape is made up of two overlapping trapezoids, one blue and one green, creating a sense of depth and movement. The word "SURMODICS" is in a dark gray color.

#### Pounce™ Thrombectomy System 510(k) Summary

#### SUBMITTER [807.92(a)(1)]

| Submitter's Name: | Surmodics, Inc.                       |
|-------------------|---------------------------------------|
| Address:          | 9924 W 74th St, Eden Prairie MN 55344 |
| Contact Person:   | Holly Ramirez                         |
| Telephone:        | 952-500-7545                          |
| Email:            | hramirez@surmodics.com                |
| Date Prepared:    | May 12, 2021                          |

| DEVICE [807.92(a)(2)]                                   |                             |
|---------------------------------------------------------|-----------------------------|
| Trade Name:                                             | Pounce™ Thrombectomy System |
| Common Name:                                            | Thrombectomy catheter       |
| Regulation Name:                                        | Embolectomy catheter        |
| Regulation Number:                                      | 21 CFR 870.5150             |
| Classification:                                         | Class II                    |
| Product Code:                                           | QEW                         |
| Review Panel:                                           | Cardiovascular              |
| PREDICATE [807.92(a)(3)]                                |                             |
| Predicate Device: Pounce™ Thrombectomy System (K192814) |                             |

#### DEVICE DESCRIPTION [807.92(a)(4)]

The Pounce™ Thrombectomy System is a percutaneous catheter system designed to facilitate mechanical thrombus removal in the peripheral arterial vasculature. The system is comprised of three separate components: a 5Fr Basket Delivery Catheter, a Basket Wire Assembly, and a Trumpet Assembly. The system also includes a Basket Loading Tool accessory for loading the Basket Wire Assembly into the Basket Delivery Catheter. The system contains the necessary radiopaque components to conduct the procedure and the system should be introduced through a minimum 7Fr guide sheath.

The Basket Delivery Catheter is a flexible 5Fr catheter designed to deliver the Basket Wire Assembly to the location of the thrombus.

The Basket Wire Assembly is comprised of two distal self-expanding baskets mounted in series on a wire for capturing thrombus.

The Trumpet Assembly is used for thrombus collection and retrieval. The Trumpet Assembly is made of an

{4}------------------------------------------------

inner trumpet catheter and an outer trumpet delivery catheters work together to allow deployment and retraction of the trumpet feature.

# INDICATION FOR USE [807.92(a)(5)]

The Pounce™ Thrombectomy System is intended for the non-surgical removal of thrombi and emboli from the peripheral arterial vasculature.

# PREDICATE DEVICE COMPARISON [807.92(a)(6)]

The subject and predicate device are both intended for non-surgical removal of thrombi and emboli from the peripheral arterial vasculature, and they have the same technological characteristics. The only difference is a revised contraindication for the smaller native vessel size from 4.5 mm.

## NONCLINICAL PERFORMANCE TESTING SUMMARY [807.92(b)]

Based on the outcome of design control and risk assessment activities the following non-clinical bench testing was conducted to evaluate the performance of the Pounce™ Thrombectomy System:

- . Basket Cage Radial Force
- . Thrombectomy System Simulated Use

Test results demonstrated that all pre-defined acceptance criteria were met. Therefore, conforms to established product specifications and intended use and no new questions of safety and effectiveness were raised.

## Animal Testing:

Animal testing previously submitted as part of K192814 was also utilized to demonstrate substantial equivalence.

## Clinical Data:

No clinical data was required to demonstrate substantial equivalence.

## Conclusions:

Based on the risk assessment and subsequent design verifications it is concluded that revising the contraindication did not impact the safety or effectiveness of the Pounce Thrombectomy System and it is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/QEW/K211492](https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/QEW/K211492)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
