Browse hierarchy: [Cardiovascular (CV)](/submissions/CV) → [Subpart F — Cardiovascular Therapeutic Devices](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices) → [21 CFR 870.5210](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/870.5210) → PMK — Cpr Aid Device Without Feedback

# PMK · Cpr Aid Device Without Feedback

_Cardiovascular · 21 CFR 870.5210 · Class 1_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/PMK

## Overview

- **Product Code:** PMK
- **Device Name:** Cpr Aid Device Without Feedback
- **Regulation:** [21 CFR 870.5210](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/870.5210)
- **Device Class:** 1
- **Review Panel:** [Cardiovascular](/submissions/CV)

## Identification

CPR Aid device with feedback is a device that provides real-time feedback to the rescuer regarding the quality of CPR being delivered to the victim, and provides either audio and/or visual information to encourage the rescuer to continue the consistent application of effective manual CPR in accordance with current accepted CPR guidelines (to include, but not be limited to, parameters such as compression rate, compression depth, ventilation, recoil, instruction for one or multiple rescuers, etc.). These devices may also perform a coaching function to aid rescuers in the sequence of steps necessary to perform effective CPR on a victim.

## Classification Rationale

Class II (special controls). The special controls for this device are:

## Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(i) Nonclinical performance testing under simulated physiological or use conditions must demonstrate the accuracy and reliability of the feedback to the user on specific compression rate, depth and/or respiration over the intended duration, and environment of use.
(ii) Labeling must include the clinical training, if needed, for the safe use of this device and information on the patient population for which the device has been demonstrated to be effective (including patient size and/or age limitations,
*e.g.,* adult, pediatric and/or infant).(iii) For devices that incorporate electrical components, appropriate analysis and testing must demonstrate that the device is electrically safe and electromagnetically compatible in its intended use environment.
(iv) For devices containing software, software verification, validation, and hazard analysis must be performed.
(v) Components of the device that come into human contact must be demonstrated to be biocompatible.
(vi) Human factors testing and analysis must validate that the device design and labeling are sufficient for effective use by the intended user.
(3)
*Premarket notification.* The CPR Aid with feedback device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter if it does not contain software (*e.g.,* is mechanical or electro-mechanical) and is in compliance with the special controls under paragraph (b)(2) of this section, subject to the limitations of exemptions in § 870.9.

## Recent Cleared Devices (2 of 2)

| Record | Device Name | Applicant | Decision Date | Decision |
| --- | --- | --- | --- | --- |
| [K944667](https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/PMK/K944667.md) | LIFE PULSE TESUSCITATION TIMER | Green Field Medical Sourcing, Inc. | Oct 4, 1994 | SESE |
| [K932867](https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/PMK/K932867.md) | LIFE PULSE RESUSCITATION TIMER | Green Field Medical Sourcing, Inc. | Jul 25, 1994 | SESE |

## Top Applicants

- Green Field Medical Sourcing, Inc. — 2 clearances

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/PMK](https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/PMK)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
