K180244 · Advanced Cooling Therapy, Inc. D/B/A Attune Medical · PLA · Feb 28, 2018 · Cardiovascular
Device Facts
Record ID
K180244
Device Name
EnsoETM
Applicant
Advanced Cooling Therapy, Inc. D/B/A Attune Medical
Product Code
PLA · Cardiovascular
Decision Date
Feb 28, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.5910
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Model # ECD03-A: The EnsoETM is a thermal regulating device, intended to: · connect to a Gaymar Medi-Therm III Conductive HyperHypothermia System or Stryker Altrix Precision Temperature Management System to control patient temperature, · allow enteral administration of fluids, · and provide gastric decompression and suctioning. Model # ECD04-A: The EnsoETM is a thermal regulating device, intended to: · connect to a Cincinnati Sub-Zero Blanketrol III Hyper-Hypothermia System to control patient temperature. · allow enteral administration of fluids, · and provide gastric decompression and suctioning.
Device Story
EnsoETM is a multi-lumen silicone tube inserted into the esophagus like a standard orogastric tube. Two lumens connect to an external heat exchanger (Gaymar Medi-Therm III, Stryker Altrix, or Cincinnati Sub-Zero Blanketrol III) to circulate coolant/warmant for patient temperature modulation. A third central lumen connects to wall suction for gastric decompression or to an enteral system for fluid administration. Used by clinicians in hospital settings; provides simultaneous thermal regulation and gastric access. Benefits include non-invasive temperature control while maintaining standard gastric functionality. Single-use, disposable, non-implantable; duration of use up to 72 hours.
Clinical Evidence
No clinical data; bench testing only. Substantial equivalence is based on design identity to predicate devices and performance history of reference devices.
Technological Characteristics
Multi-lumen medical-grade silicone tube. Connects to external hyper/hypothermia systems via Clik-Tite or Colder PLC series fittings. Gastric lumen features male ENFit connector. Single-use, non-implantable. Duration of use: 72 hours.
Indications for Use
Indicated for patients requiring esophageal thermal regulation, gastric decompression, or enteral fluid administration. Used in clinical settings for temperature management via connection to external heat exchange systems.
Regulatory Classification
Identification
An esophageal thermal regulation device is a prescription device used to apply a specified temperature to the endoluminal surface of the esophagus via an external controller. This device may incorporate a mechanism for gastric decompression and suctioning. The device is used to regulate patient temperature.
Special Controls
In combination with the general controls of the FD&C Act, the Esophageal Cooling Device is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) The patient contacting materials must be demonstrated to be biocompatible.
(2) Non-clinical performance evaluation must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Mechanical integrity testing.
(ii) Testing to determine temperature change rate(s).
(iii) Testing to demonstrate compatibility with the indicated external controller.
(iv) Shelf life testing.
(3) Animal testing must demonstrate that the device does not cause esophageal injury and that body temperature remains within appropriate boundaries under anticipated conditions of use.
(4) Labeling must include the following:
(i) Detailed insertion instructions.
(ii) Warning against attaching the device to unintended connections, such as external controllers for which the device is not indicated, or pressurized air outlets instead of vacuum outlets for those devices, including gastric suction.
(iii) The operating parameters, name, and model number of the indicated external controller.
(iv) The intended duration of use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 28, 2018
Advanced Cooling Therapy, Inc. d/b/a Attune Medical Erik Kulstad Chief Medical Officer 3440 S. Dearborn St. #215-South Chicago, Illinois 60616
Re: K180244
Trade/Device Name: EnsoETM Regulation Number: 21 CFR 870.5910 Regulation Name: Esophageal Thermal Regulation Device Regulatory Class: Class II Product Code: PLA Dated: January 26, 2018 Received: January 29, 2018
Dear Erik Kulstad:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Nicole G. Ibrahim -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180244
Device Name EnsoETM
Indications for Use (Describe)
Model # ECD03-A:
The EnsoETM is a thermal regulating device, intended to:
· connect to a Gaymar Medi-Therm III Conductive HyperHypothermia System or Stryker Altrix Precision Temperature Management System to control patient temperature,
· allow enteral administration of fluids,
· and provide gastric decompression and suctioning.
Model # ECD04-A:
The EnsoETM is a thermal regulating device, intended to:
· connect to a Cincinnati Sub-Zero Blanketrol III Hyper-Hypothermia System to control patient temperature.
· allow enteral administration of fluids,
· and provide gastric decompression and suctioning.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Company: | Advanced Cooling Therapy, Inc. d/b/a Attune Medical |
|------------------------------|---------------------------------------------------------|
| Address: | 3440 S. Dearborn St.<br>#215-South<br>Chicago, IL 60616 |
| Phone: | +1-708-651-0736 |
| Contact Person: | Erik Kulstad<br>Chief Medical Officer |
| Date Prepared: | January 26, 2018 |
| Device Name & Classification | |
| Trade Name: | EnsoETM |
| Model Number(s): | ECD03-A, ECD04-A |
| Classification Name: | Esophageal Thermal Regulation Device (21 CFR 870.5910) |
| Product Code: | PLA |
| Class: | II |
| Predicate Devices | |
# SECTION 5: 510(k) Summary of Safety and Effectiveness
### Submitter / 510(k) Holder
Attune Medical EnsoETM (model #s: ECD03-A & ECD04-A); K172029
# Reference Devices
Attune Medical EnsoETM (model #s: ECD01-A & ECD02-A); K172493
# Device Description
The EnsoETM is a multi-lumen silicone tube that is placed in the esophagus in a similar manner to a standard orogastric tube to control a patient's temperature, while simultaneously maintaining access to the stomach to allow gastric decompression, drainage, and enteral administration of fluids, thereby maintaining the functionality of the standard orogastric tube. Modulation and control of patient temperature is achieved by connecting the device to an external heat exchanger. Two lumens connect to the external heat exchanger. A third central lumen connects to wall suction to allow standard gastric decompression or an enteral administration system to allow enteral administration of fluids. The device is made of standard medical-grade silicone. It is a single-use, disposable, non-implantable device with a duration of use of 72 hours or less.
# Indications for Use (Subject Device 1; Model #: ECD03-A)
The EnsoETM is a thermal regulating device, intended to:
- . connect to a Gaymar Medi-Therm III Conductive Hyper/Hypothermia System or Stryker Altrix Precision Temperature Management System to control patient temperature,
- allow enteral administration of fluids, and
{4}------------------------------------------------
- provide gastric decompression and suctioning.
#### Indications for Use (Subject Device 2; Model #: ECD02-A)
The EnsoETM is a thermal regulating device, intended to:
- . connect to a Cincinnati Sub-Zero Blanketrol III or Blanketrol III Hyper-Hypothermia System to control patient temperature,
- . allow enteral administration of fluids, and
- . provide gastric decompression and suctioning.
### Technological Characteristics
The EnsoETM product family currently consists of four models:
- the ECD03-A (Predicate Device 1 cleared under K172029),
- . the ECD04-A (Predicate Device 2 cleared under K172029),
- . the ECD01-A (Reference Device 1 cleared under K172493), and
- . the ECD02-A (Reference Device 2 cleared under K172493).
The purpose of this Special 510(k) submission is to increase the duration of use for the ECD03-A and ECD04-A models from 36 hours to 72 hours. The subject devices are identical to the predicate devices except for the change to the duration of use.
The subject devices have the same duration of use as the reference devices and are identical to the reference devices except for the gastric lumen connector and the inclusion of enteral administration of fluids in the indication statement.
Table 1 details the differences between the subject, predicate, and reference devices. Device characteristics not described in Table 2 are identical for all devices.
| Model # | Coolant<br>Lumen<br>Fittings | Gastric Lumen<br>Fittings | Intended heat exchanger | Enteral<br>administration<br>of fluids? | Duration of<br>use |
|------------------------------------|------------------------------------|---------------------------|-----------------------------------------------------------|-----------------------------------------|--------------------|
| ECD03-A<br>(Predicate<br>Device 1) | Clik-Tite<br>connectors | Male ENFit<br>connector | Stryker/Gaymar Medi-<br>Therm III or Stryker Altrix | Yes | 36 hours |
| ECD04-A<br>(Predicate<br>Device 2) | Colder PLC<br>series<br>connectors | Male ENFit<br>connector | Cincinnati Sub-Zero<br>Blanketrol II or Blanketrol<br>III | Yes | 36 hours |
| ECD01-A<br>(Reference<br>Device 1) | Clik-Tite<br>connectors | 5° silicone<br>taper | Stryker/Gaymar Medi-<br>Therm III or Stryker Altrix | No | 72 hours |
| ECD02-A<br>(Reference<br>Device 2) | Colder PLC<br>series<br>connectors | 5° silicone<br>taper | Cincinnati Sub-Zero<br>Blanketrol II or Blanketrol<br>III | No | 72 hours |
Table 1: Differences Between EnsoETM Models
{5}------------------------------------------------
| Model # | Coolant<br>Lumen<br>Fittings | Gastric Lumen<br>Fittings | Intended heat exchanger | Enteral<br>administration<br>of fluids? | Duration of<br>use |
|----------------------------------|------------------------------------|---------------------------|-----------------------------------------------------------|-----------------------------------------|--------------------|
| ECD03-A<br>(Subject<br>Device 1) | Clik-Tite<br>connectors | Male ENFit<br>connector | Stryker/Gaymar Medi-<br>Therm III or Stryker Altrix | Yes | 72 hours |
| ECD04-A<br>(Subject<br>Device 2) | Colder PLC<br>series<br>connectors | Male ENFit<br>connector | Cincinnati Sub-Zero<br>Blanketrol II or Blanketrol<br>III | Yes | 72 hours |
### Performance Testing
No additional performance testing was conducted.
#### Conclusion
Existing data demonstrate Subject Device 1 and Subject Device 2 are substantially equivalent to Predicate Device 1 and Predicate Device 2 respectively.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.