← Product Code [MMX](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/MMX) · K222083

# LimFlow V-Ceiver (K222083)

_LimFlow, Inc. · MMX · Aug 9, 2022 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/MMX/K222083

## Device Facts

- **Applicant:** LimFlow, Inc.
- **Product Code:** [MMX](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/MMX.md)
- **Decision Date:** Aug 9, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5150
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

The LimFlow V-Ceiver™ is intended for use in the cardiovascular system to manipulate and retrieve guidewires specified in the IFU.

## Device Story

The LimFlow V-Ceiver is an intravascular catheter featuring a distal nitinol basket (snare) used to manipulate and retrieve guidewires within the cardiovascular system. The device includes a coaxial sheathing system and radiopaque markers for visualization. It is compatible with 4 Fr introducer sheaths and has a 100 cm working length. Operated by clinicians in a clinical setting, the device is inserted into the vasculature to grasp guidewires. The snare is opened, closed, and retracted to facilitate wire retrieval. The device benefits patients by enabling precise guidewire management during cardiovascular procedures.

## Clinical Evidence

Bench testing included tensile strength, leak testing, simulated use, kink testing, basket opening/closing, retraction force, radiopacity, corrosion, and packaging validation. Biocompatibility was confirmed per ISO 10993 standards (cytotoxicity, sensitization, irritation, acute systemic toxicity, pyrogenicity, hemolysis, PTT, complement activation, thromboresistance). A pre-clinical acute swine study evaluated performance, confirming the device's ability to snare various guidewires and demonstrating acceptable thrombogenicity.

## Technological Characteristics

Intravascular catheter with distal nitinol basket; 6 mm basket diameter; 100 cm working length; compatible with 4 Fr introducer sheaths. Features coaxial sheathing system and radiopaque markers. Materials evaluated for biocompatibility per ISO 10993. Mechanical testing per ISO 10555-1. Sterilization method not specified.

## Regulatory Identification

An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.

## Predicate Devices

- CloverSnare™ (FourSnare™) ([K112185](/device/K112185.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 9, 2022

LimFlow Inc. % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114

Re: K222083

Trade/Device Name: LimFlow V-Ceiver Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: MMX Dated: July 14, 2022 Received: July 15, 2022

Dear Prithul Bom:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K222083

Device Name LimFlow V-Ceiver™

Indications for Use (Describe)

The LimFlow V-Ceiver™ is intended for use in the cardiovascular system to manipulate and retrieve guidewires specified in the IFU.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### 510(k) Summary K222083 Date Prepared: 15 June 2022 Company LimFlow Inc. 3031 Tisch Wav 110 Plaza West San Jose. CA 95127 Contact: Zachary Woodson VP of Regulatory Affairs & Quality Phone: +1 707-328-6522 Email: zwoodson@limflow.com Device Trade Name: LimFlow V-Ceiver Device Common Name: Embolectomy Catheter Device Classification: Device, Embolectomy Catheter Product Code: MMX 21 CFR 870.5150 Class II Review Panel: Cardiovascular Devices K112185, CloverSnare™1 Vascular Retrieval Snare, Predicate Device: originally manufactured by Cook Incorporated Description of the Device: The LimFlow V-Ceiver consists of an intravascular catheter which incorporates a nitinol basket (the "Snare") at its distal tip. The nitinol basket has an outer diameter of 6 mm. The catheter has a built-in coaxial sheathing system and incorporates radiopaque markers at the distal tip on either end of the basket for added visibility. The catheter if compatible with 4 Fr introducer sheaths and has a working length of 100 cm. Intended Use / The LimFlow V-Ceiver is intended for use in the Indication for Use: cardiovascular system to manipulate and retrieve guidewires specified in the IFU. Technological A comparison of the technological characteristics of Characteristics: the subject device and the predicate device shows the LimFlow V-Ceiver to be substantially equivalent to the current marketed predicate device.

<sup>1</sup> CloverSnare™ is the commercial name of the device that was approved under the name "FourSnare™" per 510(K) reference K112185

{4}------------------------------------------------

Equivalence is based on the product performance, design, and intended use. The LimFlow V-Ceiver and the predicate device have similar materials of construction, dimensional specifications, designs, and sterilization process.

## Performance Tests (Non-Clinical):

No performance standards have been established under section 514, performance standards, of the Food, Drug and Cosmetic Act for these devices. A series of testing was conducted in accordance with protocols based on requirements outlined in guidance and industry standards and the below were shown to meet the acceptance criteria that were determined to demonstrate substantial equivalence.

The following tests were performed under the specified testing parameters to support the LimFlow V-Ceiver device substantial equivalence.

Performance Testing, including:

- Tensile Strength In conformance with the . applicable sections of ISO 10555-1, joint strengths were evaluated and met the predetermined acceptance criteria.
- Leak Testing In conformance with the applicable . sections of ISO 10555-1, leak testing was performed and met the predetermined acceptance criteria.
- Simulated Use Testing demonstrated the device ● can grasp guidewires during simulated clinical use, meeting the predetermined acceptance criteria.
- . Kink Testing – Testing demonstrated the catheter would not kink during clinical use, meeting the predetermined acceptance criteria.
- . Opening and Closing Basket Width - Testing demonstrated the basket could be opened and closed, meeting the predetermined acceptance criteria.
- Retraction Force Testing demonstrated that the ● snare can be retracted with a clinically acceptable amount of force, meeting the predetermined acceptance criteria.
- Radiopacity Testing confirmed acceptable ● radiopacity during clinical use.
- Corrosion Testing confirmed acceptable ● corrosion performance.

{5}------------------------------------------------

| Packaging Validation - Testing demonstrated the |
|-------------------------------------------------|
| packaging system can maintain package integrity |
| and met the predetermined acceptance criteria.  |

- Biocompatibility In conformance with the . applicable sections of ISO 10993-1, the following testing was performed confirming biocompatibility of the device:
	- Cytotoxicity (ISO 10993-5) o
	- Sensitization (ISO 10993-10) O
	- Irritation (ISO 10993-10) o
	- Acute Systemic Toxicity (ISO 10993-11) O
	- Material Mediated Pyrogenicity (ISO 10993-11) O
	- O Hemolysis (ISO 10993-4)
	- PTT (ISO 10993-4) o
	- Complement Activation (ISO 10993-4) O
	- Thromboresistance (ISO 10993-4) O

A pre-clinical study was performed in an acute swine model to evaluate the performance of the LimFlow V-Ceiver. Snaring of various guidewires was performed to demonstrate the device is appropriate with the stated indications. Furthermore, thrombogenicity was evaluated and confirmed to be acceptable.

Based on the Indication for Use, design, safety, and Substantial Equivalence: performance testing, the LimFlow V-Ceiver device met the requirements for its intended use and is substantially equivalent to the predicate device.

- Conclusion: The result of all testing demonstrates that the LimFlow V-Ceiver device is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/MMX/K222083](https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/MMX/K222083)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
