K022201 · Radius Medical Technologies, Inc. · MMX · Feb 3, 2003 · Cardiovascular
Device Facts
Record ID
K022201
Device Name
RADIUS MICRO SNARE
Applicant
Radius Medical Technologies, Inc.
Product Code
MMX · Cardiovascular
Decision Date
Feb 3, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5150
Device Class
Class 2
Indications for Use
The Radius Micro Snare is intended for use to retrieve and/or manipulate objects in the distal peripheral vessels of the cardiovascular system and hollow viscus. Manipulation procedures include retrieval and/or repositioning of intravascular foreign objects such as coils, balloons, catheters and/or guidewires within the peripheral and cardiovascular system.
Device Story
Radius Micro Snare is a catheter-based retrieval device for use in distal peripheral vessels and hollow viscus. It consists of an outer sheath tube and a core wire with a snare loop attached to the distal end. The device is operated by a physician who advances or retracts the outer sheath via a molded ABS handle to deploy or capture objects. The core wire is a .008" diameter stainless steel wire. The device is used to retrieve or reposition foreign objects such as coils, balloons, catheters, and guidewires. It is provided sterile for single use.
Clinical Evidence
Bench testing only. Performance testing included tensile strength, torque strength, tip flexibility, and biocompatibility.
Technological Characteristics
Materials: stainless steel core, polyimide tube, PTFE coating, ABS handle. Dimensions: .008" core diameter, 190 cm length, 2-7 mm loop sizes. Delivery: catheter/outer sheath system. Sterilization: ETO (SAL 10-6).
Indications for Use
Indicated for retrieval and/or manipulation of intravascular foreign objects (coils, balloons, catheters, guidewires) in distal peripheral vessels of the cardiovascular system and hollow viscus.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
K071457 — MODIFICATION TO RADIUS SNARE · Radius Medical Technologies, Inc. · Sep 25, 2007
K021441 — RADIUS SNARE · Radius Medical Technologies, Inc. · Jun 14, 2002
K163307 — Expro Elite Snare, Sympro Elite Snare · Vascular Solutions, Inc. · Mar 8, 2017
K014109 — IN-TIME RETRIEVAL DEVICE · Boston Scientific, Target · Mar 12, 2002
K033188 — TEXAN FOREIGN BODY RETRIEVAL DEVICE, MODEL TX30060050 · Idev Technologies, Inc. · Nov 19, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
## FEB 0 3 2003
## 510(K) SUMMARY
Pursuant to §513(i)(3)(A) of the Food, Drug, and Cosmetic Act, Radius Medical Technologies, Inc. is required to submit with this Premarket Notification either an "...adequate summary of any information respecting safety and effectiveness or state that such information will be made available upon request of any person." Radius Medical Technologies, Inc. chooses to submit a summary of information respecting safety and effectiveness.
- 1. Sponsor Name
| Submitter's Name: | Radius Medical Technologies, Inc |
|----------------------|------------------------------------|
| Address: | 63 Great Road<br>Maynard, MA 01754 |
| Contact Person: | Maureen A. Finlayson |
| Date of Preparation: | June 28, 2002 |
- 2. Device Name
| Device Generic Name: | Snare |
|----------------------|-------------------------------------|
| Device Trade Name: | Radius Micro Snare |
| Classification Name: | Wire, Guide, Cardiovascular (74DQX) |
- Identification of Predicate or Legally Marketed Device 3.
The Radius Micro Snare is substantially equivalent to the Microvena Amplatz Goose Neck Micro Snare (K970688), the Radius Snare (K021441) and the Radius PTCA Guidewire (K970466).
- 4. Device Description
The Radius Micro Snare is composed of two primary parts: an outer sheath tube and a core wire with a snare attached to the distal end. The outer sheath acts as a catheter through which the core wire, with snare, slides.
The outer sheath is a stainless steel tube joined to a polyimide tube. The stainless steel and polyimide tubes are covered with PTFE.
The stainless steel core is a solid .008" diameter stainless steel core that is a smaller version of the solid core wire in the FDA cleared Radius PTCA Guidewire
{1}------------------------------------------------
and the FDA cleared Radius Snare. A stainless steel loop is soldered to the distal end of the core. The Radius Micro Snare has loop sizes, which range from 2 - 7 millimeters. The over-all length of the device is 190 centimeters. A molded ABS operating handle is included on the proximal end of the Micro Snare to aid in advancing and retracting the outer sheath over the snare loop.
The Radius Micro Snare will be packaged in a Mylar/Tyvek pouch and ETO sterilized to SAL 10-6.
- న్. Intended Use
The Radius Micro Snare is intended for use to retrieve and/or manipulate objects in the distal peripheral vessels of the cardiovascular system and hollow viscus. Manipulation procedures include retrieval and/or repositioning of intravascular foreign objects such as coils, balloons, catheters and/or guidewires within the peripheral and cardiovascular system.
- 6. Comparison of Technological Characteristics
The Radius Micro Snare is substantially equivalent in material, design and function to the following predicate devices: the Microvena Amplatz Goose Neck Micro Snare (K970688), the Radius Snare (K021441), and the Radius PTCA Guidewire (K970466).
The Radius Micro Snare is similar in materials, function and design to predicate Radius PTCA Guidewire (K970466) and Radius Snare (K021441). The Radius Micro Snare uses the Radius PTCA Guidewire as its base and adds a snare onto the distal tip of the core guidewire.
The design and function of the Radius Micro Snare is the same as the Microvena Amplatz Goose Neck Micro Snare (K970668). Both are delivered via a cathether/outer sheath tube and both have a snare at the distal end. The intended use of both devices is the same: to retrieve and/or manipulate objects in the distal peripheral vessels of the cardiovascular system and hollow viscus.
- 7. Performance Testing
The following in vitro performance tests were performed on the Radius Micro Snare:
Tensile Strength Torque Strength Tip Flexibility Biocompatability
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized image of a human figure with three faces in profile, arranged in a row. The figure is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" written around it.
Public Health Service
FEB 0 3 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Maureen A. Finlayson President Radius Medical Technologies, Inc. 63 Great Road Maynard, MA 01754
Re: K022201
> Trade/Device Name: Radius Micro Snare Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter. Regulatory Class: Class II Product Code: MMX Dated: October 14, 2002 Received: November 6, 2002
Dear Ms. Finlayson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Ms. Maureen A. Finlayson
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4586. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address
http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Kaitlin Tea
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
KO2201 510(k) Number (if known):
Device Name: Radius Micro Snare
Indications For Use:
The Radius Micro Snare is intended for use to retrieve and/or manipulate objects in the distal peripheral vessels of the cardiovascular system and hollow viscus. Manipulation procedures include retrieval and/or repositioning of intravascular foreign objects such as coils, balloons, catheters and/or guidewires within the peripheral and cardiovascular system.
(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use
Norke Tlu
(Division Sign-Off)
Division of Cardiovascular Devices
**510(k) Number** K022201
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches