LIFEPAK 1000 DEFIBRILLATOR

K122600 · Physio-Control, Inc. · MKJ · Oct 11, 2012 · Cardiovascular

Device Facts

Record IDK122600
Device NameLIFEPAK 1000 DEFIBRILLATOR
ApplicantPhysio-Control, Inc.
Product CodeMKJ · Cardiovascular
Decision DateOct 11, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5310
Device ClassClass 3
AttributesTherapeutic, Pediatric

Indications for Use

The LIFEPAK 1000 defibrillator is intended for use by personnel who are authorized by a physician/medical director and are trained in CPR and the use of the LIFEPAK 1000 defibrillator. The manual and ECG modes of the LIFEPAK 1000 defibrillator are intended for use by health care providers trained in ECG rhythm recognition.

Device Story

Semi-automatic external defibrillator; provides manual mode and ECG monitoring. Inputs: ECG signals via defibrillation pads. Operation: AED mode (automated rhythm analysis/shock delivery) or manual mode (operator-controlled rhythm interpretation/shock). Output: Defibrillation energy or ECG rhythm/heart rate display. Used by trained personnel/healthcare providers in clinical or emergency settings. Device aids in terminating fatal arrhythmias; provides real-time monitoring to guide clinical intervention. Subject device adds rechargeable lithium-ion battery option to predicate design.

Clinical Evidence

No clinical data. Substantial equivalence demonstrated via bench testing confirming performance with the new battery option.

Technological Characteristics

External defibrillator; semi-automatic with manual override. Power: Lithium-manganese (non-rechargeable) or lithium-ion (rechargeable). Waveform: Defibrillation waveform (unchanged). Connectivity: ECG display. Software: ECG analysis system (unchanged).

Indications for Use

Indicated for defibrillation of patients in cardiopulmonary arrest (unresponsive, apneic, pulseless) with ventricular fibrillation or symptomatic ventricular tachycardia. AED mode for adults/children >8 years or >25kg; use Infant/Child Reduced Energy Electrodes for <8 years or <25kg. ECG monitoring indicated for conscious/unconscious patients of all ages for rhythm and heart rate monitoring.

Regulatory Classification

Identification

An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows a logo for Physio Control. The logo is in a black box with white text. The word "PHYSIO" is on the top line, and the word "CONTROL" is on the bottom line. 122600 Physio-Control Inc. LIFEPAK 1000 defibrillator 510(k) Premarket Notification ## SECTION E -- 510(K) SUMMARY OCT 11 2012 ### Submitter: Physio-Control, Inc. 11811 Willows Road Northeast P.O. Box 97006 Redmond, Washington 98073-9706 ### Contact Person: Michelle Ackermann Principal Regulatory Affairs Specialist Phone: 425-867-4744, Fax: (425) 867-4154 ### Date Summary Prepared: August 24th, 2012 ## Device Trade Naíne: LIFEPAK®1000 defibrillator #### Device Common Name: External defibrillator ### Device Classification: | Classification Name | Class | |----------------------------------------------------------------------------------------|-------| | Low Energy DC-Defibrillator (Including Paddles), (21CFR<br>870.5300, Product Code LDD) | II | | Automatic External Defibrillators (Non-Wearable) (21CFR<br>870.5310, Product Code MKJ) | III | {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows a black square with the words "PHYSIO CONTROL" stacked on top of each other in white, sans-serif font. The word "PHYSIO" is on the top line, and the word "CONTROL" is on the bottom line. The letters are all capitalized and evenly spaced within the square. ## SECTION E - 510(K) SUMMARY ### Predicate Devices: The features and functions of the proposed LIFEPAK 1000 defibrillator are substantially equivalent to the previously cleared LIFEPAK 1000 defibrillator. The 510(k) number for the predicate LIFEPAK 1000 defibrillator is K042404 (cleared on 12/22/05). #### Description: The LIFPAK 1000 defibrillator is a semi-automatic defibrillator with optional manual mode and ECG display. It is intended for use by personnel who are authorized by a physician/medical director and are trained in CPR and the use of the LIFEPAK 1000 defibrillator. There are 3 modes of operation for the LIFEPAK 1000 defibrillator: AED mode (automated external defibrillation), manual mode (operator ECG interpretation, operator control of charge and shock functions), and ECG mode (ECG display allows for rhythm and heart rate monitoring). The currently marketed LIFEPAK 1000 defibrillator is powered by a non-rechargeable lithium-manganese battery. The proposed LIFEPAK 1000 defibrillator has the same features and functions as the predicate LIFEPAK 1000 defibrillator with an additional battery option - a rechargeable lithium-ion battery. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows a logo for Physio Control. The logo is in a black box with white text. The text is stacked vertically, with "PHYSIO" on top and "CONTROL" on the bottom. # SECTION E - 510(K) SUMMARY ### Intended Use: The LIFEPAK 1000 defibrillator is intended for use by personnel who are authorized by a physician/medical director and are trained in CPR and the use of the LIFEPAK 1000 defibrillator. The manual and ECG modes of the LIFEPAK 1000 defibrillator are intended for use by health care providers trained in ECG rhythm recognition. Indications for Use: ### Defibrillation Defibrillation is a recognized means of terminating certain potentially fatal arrhythmias, such as ventricular fibrillation and symptomatic ventricular tachycardia. The defibrillator is to be used in AED mode only on patients who are in cardiopulmonary arrest. The patient must be unresponsive, not breathing normally, and showing no signs of circulation. The defibrillator may be used with standard defibrillation pads only on adults and children who are 8 years old or more or who weigh more than 25 kg (55lbs). The defibrillator may be used on children who are less than 8 years old or weigh less than 25kg (55lbs) with Infant/Child Reduced Energy Defibrillation Electrodes. #### ECG Monitoring ECG monitoring is for use on conscious and unconscious patients of all ages for the purpose of ECG rhythm recognition and heart rate monitoring {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows a logo for Physio-Control. The logo is in a black square. The text "PHYSIO" is on the top line, and the text "CONTROL" is on the bottom line. # SECTION E - 510(K) SUMMARY ### Technological characteristics of the proposed and predicate device: The only difference between the predicate LIFEPAK 1000 defibrillator and the proposed LIFEPAK 1000 defibrillator is a new battery option. The defibrillation waveform, ECG analysis system, and monitoring and defibrillation features of the proposed LIFEPAK 1000 defibrillator are unchanged from the previously cleared LIFEPAK 1000 defibrillator. ### Conclusion of Testing The information in this 510(k) notification demonstrates that the LIFEPAK 1000 defibrillator is substantially equivalent to the predicate LIFEPAK 1000 device with respect to safety, effectiveness, and performance. {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with its wings spread, rendered in blue lines. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 MAR 2 0 2013 Ms. Michelle Ackermann Principal Regulatory Affairs Specialist Physio-Control, Inc. 11811 Willows Road Northeast P.O. Box 97006 Redmond, WA 98073-9706 Re: K122600 Trade/Device Name: LIFEPAK 1000 defibrillator Regulation Number: 21 CFR 870.5310 Regulation Name: Automatic External Defibrillator (Non-Wearable) Regulatory Class: Class III Product Code: MKJ Dated: August 24, 2012 Received: August 27, 2012 Dear Ms. Ackermann: This letter corrects our substantially equivalent letter of October 11, 2012. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be {5}------------------------------------------------ Page 2 - Ms. Michelle Ackermann found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. M.g. Hilleben Ja Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for Physio Control. The logo is in black and white, with the words "PHYSIO" stacked on top of the word "CONTROL". The text is in a simple, sans-serif font. The logo is simple and recognizable. ## SECTION D - STATEMENT OF INDICATIONS FOR USE 510(k) Number (if known): not yet assigned K ) ≥2600 Device Name: LIFEPAK® 1000 defibrillator Indications For Use: ### Defibrillation Defibrillation is a recognized means of terminating certain potentially fatal arrhythmias, such as ventricular fibrillation and symptomatic ventricular tachycardia. . The defibrillator is to be used in AED mode only on patients who are in cardiopulmonary arrest. The patient must be unresponsive, not breathing normally, and showing no signs of circulation. The defibrillator may be used with standard defibrillation pads only on adults and children who are 8 years old or more or who weigh more than 25 kg (55lbs). The defibrillator may be used on children who are less than 8 years old or weigh less than 25kg (55lbs) with Infant/Child Reduced Energy Defibrillation Electrodes. ### ECG Monitoring ECG monitoring is for use on conscious and unconscious patients of all ages for the purpose of ECG rhythm recognition and heart rate monitoring Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use _ (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) | | Concurrence of CDRH, Office of Device Evaluation (ODE) | | |---------------|-----------------------------------------------------------|-----| | Page 1 of 1 | (Division Sign-Off)<br>Division of Cardiovascular Devices | | | 510(k) Number | K122600 | D-1 |
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...