K110978 · ZOLL Medical Corporation · MKJ · May 13, 2011 · Cardiovascular
Device Facts
Record ID
K110978
Device Name
ONESTEP PEDIATRIC MFE'S
Applicant
ZOLL Medical Corporation
Product Code
MKJ · Cardiovascular
Decision Date
May 13, 2011
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.5310
Device Class
Class 3
Attributes
Therapeutic, Pediatric
Indications for Use
Defibrillation . Cardioversion Noninvasive Pacing ◆ ECG Monitoring . For use with ZOLL® Defibrillators: ZOLL Defibrillators, such as; . R Series M Series . E Series . By Trained Personnel only, Including: Physicians Nurses ● Paramedics . Emergency Medical Technicians Cardiovascular Laboratory Technicians The OneStep Pediatric Electrodes are indicated for use on a patient less than 8 years of age or weighs less than 55 Lbs. (25Kg).
Device Story
OneStep Pediatric Multi-function Electrode is a single-use, disposable accessory for ZOLL defibrillators (R, M, and E Series). Device consists of conductive hydrogel, metal conductor, and adhesive perimeter. Used in hospital environments by trained clinicians (physicians, nurses, paramedics, EMTs) for ECG monitoring, defibrillation, cardioversion, and noninvasive pacing. Electrodes couple to patient skin to deliver electrical therapy or acquire ECG signals. Output is processed by the host defibrillator to assist clinical decision-making during cardiac rescue or treatment. Benefits include providing therapy and monitoring for pediatric patients.
Clinical Evidence
Bench testing only. Validation project (VPN 0731) confirmed device integration with ZOLL defibrillators and performance consistency with predicate devices per ANSI/AAMI DF80 and EN 60601-2-4 standards.
Technological Characteristics
Single-use disposable electrode; conductive hydrogel; metal conductor; adhesive perimeter. Dimensions/conductive area meet ANSI/AAMI DF80 and EN 60601-2-4 standards. Connects to ZOLL R, M, and E Series defibrillators via standard cable interface. No active software or electronic components within the electrode itself.
Indications for Use
Indicated for pediatric patients < 8 years old or < 55 lbs (25 kg) requiring defibrillation, cardioversion, noninvasive pacing, or ECG monitoring. For use by trained personnel (physicians, nurses, paramedics, EMTs, cardiovascular lab technicians) with compatible ZOLL defibrillators.
Regulatory Classification
Identification
An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.
{0}------------------------------------------------
K110978
Image /page/0/Picture/1 description: The image shows the logo for BIO-DETEK INCORPORATED. The words "BIO-DETEK" are in large, bold, sans-serif font. Below that, the word "INCORPORATED" is in a smaller, sans-serif font. The logo is black and white.
MAY 1 3 2011
# 510(k) Summary: Rev. 2
Submitter's Name and Address: Bio-Detek, Inc. A Division of ZOLL®Medical Corporation 525 Narragansett Park Drive Pawtucket. RI 02861 Tel. (401) 729-1400
Contact Person: Robert Morse Director, Regulatory Affairs Tel. (866) 639-0060 x. 224
Date Summary Prepared: March 29, 2011
Device Name: OneStep™ Pediatric Multi-function Electrode
Classification Name: Electrode, Electrocardiograph, Multi-Function; Accessory to an External Defibrillator
#### Substantial Equivalence:
The OneStep Pediatric Electrode System is substantially equivalent to ZOLL ready-padz OneStep Pediatric Electrodes that were FDA cleared on 510(k), K06559. In addition, the OneStep Pediatric Electrode is substantially equivalent to the originally cleared ZOLL pedipadz solid gel identified on 510(k), K931787 and pedi-padz II cleared on K033474. The intent is to show that the OneStep Pediatric Electrodes will assist in providing therapy for defibrillation. cardioversion, Noninvasive pacing and ECG monitoring comparable to the previously cleared devices, for use with ZOLL Defibrillators. All of the cleared devices are in the Regulatory Class III category.
#### Description of Device:
The OneStep pediatricElectrode System is intended for use with ZOLL Defibrillators, such as, R Series M Series, and E Series for ECG monitoring, defibrillation, noninvasive pacing and cardioversion of pediatric patients in either the hospital environment.
525 Narragansell Park Drive, Pawtucket, Rhode Island 02861-4323, Tel. (401) 729-1400 or (800) 729-1408
{1}------------------------------------------------
The system is comprised of a single use, disposable electrode made of conductive hydrogel with a metal conductor and adhesive perimeter suitable for coupling to patient skin during rescue and/or treatment.
#### Indications for Use:
Intended Use:
- o Defibrillation
- o Cardioversion
- o Noninvasive Pacing
- o ECG Monitoring
For use with ZOLL® Defibrillators:
- ZOLL Defibrillators, such as; o
- R Series o
- M Series o
- E Series 0
By Trained Personnel only, Including:
- Physicians 0
- Nurses 0
- Paramedics �
- Emergency Medical Technicians o
- Cardiovascular Laboratory Technicians o
The OneStep Pediatric Electrodes are indicated for use on a patient less than 8 years of age or weighs less than 55 Lbs. (25Kg).
## Comparison of Technological Characteristics:
The intended use of the OneStep Pediatric Multi-Function Electrodes as described in the indications for use, and labeling, has not changed as a result of this submission. For the ZOLL R Series, M Series and E Series, the connection to the defibrillator cable is identical to the predicate devices. All four products use the same hydrogel, adhesive material and have a conductive area similar in size that meet/exceed National and International Standards ANSI/AAMI DF80 and EN 60601-2-4 respectively.
## Nonclinical Testing:
The OneStep Pediatric Electrode System has been subjected to extensive performance testing to ensure the device meets all of its functional requirements and performance specifications as defined in applicable National/International recognized standards.
Validation Project Number (VPN) 0731 addresses design validation to ensure the safety and effectiveness considerations have been successful in integrating the OneStep electrode with previously stated ZOLL defibrillators and performs as well and/or better than the legally marketed predicate devices. Applicable standards are defined in Form FDA 3514, VPN 0731 and Section 2, Device Description, within this premarket submission.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird with three curved lines representing its wings or body. The bird is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Zoll Medical Corporation c/o Mr. Robert Morse Director, Regulatory Affairs Bio-Detek, Inc. 525 Narragansett Park Drive Pawtucket, RI 02861
MAY 1 3 2011
Re: K110978
Trade/Device Name: OneStep Pediatric Multi-function Electrode Regulation Number: 21 CFR 870.5310 Regulation Name: Automated External Defibrillator Regulatory Class: Class III Product Code: MKJ, DRO, DQA, LDD, LIX Dated: April 6, 2011 Received: April 7, 2011
Dear Mr. Morse:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Robert Morse
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
fu-
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
K110978 510(k) number (if known): __
Device Name: OneStep™ Pediatric Multi-Function Electrode
Intended Use:
- Defibrillation �
- . Cardioversion
- Noninvasive Pacing ◆
- ECG Monitoring .
For use with ZOLL® Defibrillators:
- ZOLL Defibrillators, such as; .
- R Series �
- M Series .
- E Series .
By Trained Personnel only, Including:
- Physicians �
- Nurses ●
- Paramedics .
- Emergency Medical Technicians �
- Cardiovascular Laboratory Technicians
The OneStep Pediatric Electrodes are indicated for use on a patient less than 8 years of age or weighs less than 55 Lbs. (25Kg).
(Please do not write below this line - Continue on another page if necessary)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sign-Off Division of Cardiovascular Devices
510(k) Number k//0178
Prescription Use న OR
Over the Counter Use
Rev. 1
OneStep™ Pediatric
(Per 21 CFR 801.109)
Section 4, pg. 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.