ZOLL R SERIES WITH 2010 AHA GUIDELINES SOFTWARE UPDATE
K110361 · Zoll Medical Corporation, World Wide Headquarters · MKJ · Mar 8, 2011 · Cardiovascular
Device Facts
Record ID
K110361
Device Name
ZOLL R SERIES WITH 2010 AHA GUIDELINES SOFTWARE UPDATE
Applicant
Zoll Medical Corporation, World Wide Headquarters
Product Code
MKJ · Cardiovascular
Decision Date
Mar 8, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.5310
Device Class
Class 3
Attributes
Therapeutic
Indications for Use
The ZOLL R Series® External Defibrillator is intended for the defibrillation. Noninyasive Transcutaneous Pacing, multi-parameter monitoring of patient vital signs, including ECG Monitoring, Pulse Oximetry, Non-Invasive Blood Pressure measurement, End Tidal CO2, CPR performance and data printing and recording for resting patients in critical care and transport. The ZOLL R Series is intended for use by qualified medical personnel who are trained and authorized to respond to medical emergencies, to facilitate the ability to monitor and assess the physiological characteristics of the indicated patients in a critical care environment. The design facilitates table top use while still providing a light weight and easy to carry device for transport.
Device Story
The ZOLL R Series is a multi-parameter monitor and defibrillator used by qualified medical personnel in critical care and transport settings. It accepts inputs via ECG leads, hands-free therapy electrodes, paddles, SpO2 sensors, NIBP cuffs, and CO2 sensors. The device provides manual and semi-automatic (AED) defibrillation, synchronized cardioversion, and non-invasive transcutaneous pacing. It features 'Real CPR Help' technology, which uses sensors to provide real-time audio/visual feedback on chest compression rate and depth to guide rescuers. The 2010 AHA Guidelines update adds configuration options for a minimum 2.0-inch compression depth and a 'FULLY RELEASE' prompt to ensure chest recoil. Output is displayed on an integrated screen and documented via a strip recorder. The device assists clinicians in resuscitation and patient monitoring, aiming to improve CPR quality and patient outcomes during cardiac emergencies.
Clinical Evidence
Bench testing only. No clinical data was required or presented to support this software-only update, as the changes involve implementing updated AHA guidelines for CPR performance feedback.
Technological Characteristics
Multi-parameter monitor/defibrillator; DC defibrillator (up to 200J); transcutaneous pacemaker (0-140 mA, 30-180 ppm); ECG, SpO2 (Masimo SET), EtCO2 (Respironics CAPNOSTAT 5), and NIBP monitoring modules. Connectivity via patient cables, electrodes, or paddles. Software-based CPR feedback algorithm. Portable, battery-powered design.
Indications for Use
Indicated for cardiac arrest victims (unconscious, apneic, pulseless) requiring defibrillation or synchronized cardioversion; patients requiring temporary external cardiac pacing for bradycardia, standstill, or tachycardia suppression; and patients requiring monitoring of ECG, SpO2, EtCO2, or NIBP in critical care or transport. CPR monitoring is indicated for adult patients (≥8 years).
Regulatory Classification
Identification
An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.
{0}------------------------------------------------
**ZOLL**
*Advancing Resuscitation. Today.*
MAR -- 8 2011
ZOLL Medical Corpo Worldwide Headquarte 269 Mill Road Chelmsford, MA 01824
U.S.A
# 510(k) Summary:
Submitter's Name and Address:
ZOLL Medical Corporation 269 Mill Road Chelmsford, MA 01824-4105 (978) 421-9655
Contact Person:
Eileen M. Boyle (978) 421-9171
Date Summary Prepared:
January 26, 2011
# Device:
ZOLL R Series® with 2010 AHA Guidelines Software Update
Classification:
Defibrillator, Low-energy - DC : Class II (21 CFR 870.5300)
Automatic External Defibrillators: Class III (21 CFR 870.5310)
Cardiopulmonary Resuscitation Aid: Class III (21 CFR 870.5200)
Cardiac Monitors (including Cardiotachometers and Rate Alarms): Class II (21CFR 870.2300)
External Transcutaneous Cardiac Pacemakers (Non-invasive): Class II (21 CFR 870.5550)
Oximeters: Class II (21 CFR 870.2700)
System, Measurement, Blood-Pressure, Non-Invasive: Class II (21 CFR 870.1130)
Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase: Class II (21 CFR 868.1400)
{1}------------------------------------------------
## Description:
The ZOLL R Series® External Defibrillator is intended for the defibrillation. Noninyasive Transcutaneous Pacing, multi-parameter monitoring of patient vital signs, including ECG Monitoring, Pulse Oximetry, Non-Invasive Blood Pressure measurement, End Tidal CO2, CPR performance and data printing and recording for resting patients in critical care and transport. The ZOLL R Series is intended for use by qualified medical personnel who are trained and authorized to respond to medical emergencies, to facilitate the ability to monitor and assess the physiological characteristics of the indicated patients in a critical care environment. The design facilitates table top use while still providing a light weight and easy to carry device for transport.
previous The recent 2010 American Heart Association Guidelines for CPR changed recommendations from 1.5"-2.0" inches compression depth of at least 2.0". The Guidelines also advise caregivers to allow full recoil of the patient's chest during CPR. As a result, we are proposing modifying the device's software to prompt caregivers to perform CPR in accordance with the new AHA quidelines.
The ZOLL R Series software, which currently supports the 2005 American Heart Association's (AHA) Guidelines for CPR, has been revised to optionally support the new revised 2010 AHA's Guidelines for CPR. The specific changes include:
- Adding, in the form of a configuration option, a minimum CPR target compression depth . of at least 2.0 inches in addition to the 2005 AHA guidelines depth as described in Chapter 6 of the R Series ALS Operator's guide included in Attachment 11-1 and Chapter 7 of of the R Series Plus/ R Series BLS Operator's Guide included in Attachment 11-2. Both depth configurations are necessary since the training to and implementation of the new guidelines occur over a number of years.
- . Adding an optional configuration option to display an informational message to remind the users to release their hands from the chest between compressions in order to allow the chest to recoil. When the R Series is configured to display this new message, the message FULLY RELEASE will be displayed every 45 seconds, for 5 seconds, as long as CPR compressions are detected.
## Intended Use: R Series ALS
#### Defibrillator Function
The R Series product contains a direct current (DC) defibrillator capable of delivering up to 200 joules. It may be used in synchronized mode to perform synchronized cardioversion using the patient's R-wave as a timing reference. The unit uses paddles or disposable, pregelled electrodes for defibrillation.
#### Intended Use - Manual Operation
Use of the R Series products in the manual mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by:
- · Unconsciousness.
- · Absence of breathing.
- · Absence of pulse.
This product should be used only by qualified medical personnel for converting ventricular fibrillation and rapid ventricular tachycardia to sinus rhythm or other cardiac rhythms capable of producing hemodynamically significant heart beats.
In manual mode, the unit can also be used for synchronized cardioversion of certain atrial or ventricular arrhythmias. A qualified physician must decide when synchronized cardioversion is appropriate.
{2}------------------------------------------------
The advisory function should be used to confirm ventricular fibrillation or wide complex ventricular tachycardia (greater than 150 beats per minute) in patients meeting the three conditions indicating lack of circulation (listed above),
## Intended Use - ECG Monitoring
The unit is intended for use when ECG monitoring is indicated to evaluate the patient's heart rate or ECG morphology. In ECG monitoring mode, the unit is intended to be used by personnel who are qualified by training in the use of the R Series defibrillator, basic life and/or advanced life support, or other physician-authorized emergency medical training.
## Intended Use - Real CPR Help
The Real CPR Help function provides visual and audio feedback designed to encourage rescuers to perform chest compressions at the AHA/ERC recommended rate of 100 compressions per minute. Voice and visual prompts encourage a compression depth of 2 inches (5 cm) minimum for adult patients.
Real CPR Help is not intended for use on patients under 8 years of age.
#### Defibrillator Complications
Inappropriate defibrillation or cardioversion of a patient (for example, with no malignant arrhythmia) may precipitate ventricular fibrillation, asystole, or other dangerous arrhythmias.
Defibrillation without proper application of electrodes or paddle electrolyte qel might be ineffective and cause burns, particularly when repeated shocks are necessary. Erythema or hyperemia of the skin under the paddles, or electrodes often occurs; this effect is usually enhanced along the perimeter of the paddles or electrodes. This reddening should diminish substantially within 72 hours.
#### Defibrillator Output Energy
R Series defibrillators can deliver as much as 200 joules into a 50 ohm impedance. The energy delivered through the chest wall, however, is determined by the patient's transthoracic impedance. An adequate amount of electrolyte gel must be applied to the paddles and a force of 10 to 12 kilograms (22 to 26.4 pounds) must be applied to each paddle in order to minimize this impedance. If hands-free therapy electrodes are used, make sure that they are properly applied. (Refer to the instructions on the electrode package).
## External Pacemaker (Optional)
Some R Series products include an optional transcutaneous pacemaker consisting of a pulse generator and ECG-sensing circuitry. Noninvasive transcutaneous pacing (NTP) is an established and proven technique. This therapy is easily and rapidly applied in both emergency and nonemergency situations when temporary cardiac stimulation is indicated.
The output current of the pacemaker is continuously variable from 0 to 140 mA. The rate is continuously variable from 30 to 180 pulses per minute (ppm), by increments of 2.
The pacing output pulse is delivered to the heart via ZOLL hands-free defibrillation/pacing electrodes placed on the patient's back and the precordium.
The characteristics of the output pulse, together with the design and placement of the electrodes, minimize cutaneous nerve stimulation, cardiac stimulation threshold currents, and reduce discomfort due to skeletal muscle contraction.
The unique design of the R Series products allow clear viewing and interpretation of the electrocardiogram on the display without offset or distortion during external pacing.
Proper operation of the device, together with correct electrode placement, is critical to obtaining optimal results. Every operator must be thoroughly familiar with these operating instructions.
{3}------------------------------------------------
## Intended Use — Pacemaker
This product can be used for temporary external cardiac pacing in conscious or unconscious patients as an alternative to endocardial stimulation.
The purposes of pacing include:
· Resuscitation from standstill or bradycardia of any etiology.
Noninvasive pacing has been used for resuscitation from cardiac standstill, reflex vagal standstill, drug-induced standstill (due to procainamide, quinidine, digitalis, b-blockers, verapamil, etc.) and unexpected circulatory arrest (due to anesthesia, surgery, angiography, and other therapeutic or diagnostic procedures). It has also been used for temporary acceleration of bradycardia in Stokes-Adams disease and sick-sinus syndrome. It is safer, more reliable, and more rapidly applied in an emergency than endocardial or other temporary electrodes.
· As a standby when standstill or bradycardia might be expected.
Noninvasive pacing can be useful as a standby when cardiac arrest or symptomatic bradycardia might be expected due to acute myocardial infarction, drug toxicity, anesthesia, or surqery. It is also useful as a temporary treatment in patients awaiting pacemaker implants or the introduction of transvenous therapy. In standby pacing applications, noninvasive pacing might provide an alternative to transvenous therapy that avoids the risks of displacement, infection, hemorhage, embolization, phlebitis, and mechanical or electrical stimulation of ventricular tachycardia or fibrillation associated with endocardial pacing.
· Suppression of tachycardia.
Increased heart rates in response to external pacing often suppress ventricular ectopic activity and might prevent tachycardia.
WARNING! This device must not be connected to internal pacemaker electrodes.
#### Pacemaker Complications
Ventricular fibrillation does not respond to pacing and requires immediate defibrillation. Therefore, the patient's dysrhythmia must be determined immediately, so that you can employ appropriate therapy. If the patient is in ventricular fibrillation is successful but cardiac standstill (asystole) ensues, you should use the pacemaker.
Ventricular or supraventricular tachycardias can be interrupted with pacing, but in an emergency or during circulatory collapse, synchronized cardioversion is faster and more certain.
Pulseless electrical activity (PEA) can occur following prolonged cardiac arrest or in other disease states with myocardial depression. Pacing might then produce ECG responses without effective mechanical contractions, making other effective treatment necessary. Pacing can evoke undesirable repetitive responses, tachycardia, or fibrillation in the presence of generalized hypoxia, myocardial ischemia, cardiac drug toxicity, electrolyte imbalance, or other cardiac diseases.
Pacing by any method tends to inhibit intrinsic rhythmicity. Abrupt cessation of pacing, particularly at rapid rates, can cause ventricular standstill and should be avoided.
Noninvasive temporary pacing can cause discomfort of varying intensity, which occasionally can be severe and preclude its continued use in conscious patients.
Similarly, unavoidable skeletal muscle contraction might be troublesome in very sick patients and might limit continuous use to a few hours. Erythema or hyperemia of the skin under the hands-free therapy electrodes often occurs; this effect is usually enhanced along the perimeter of the electrode. This reddening should lessen substantially within 72 hours.
{4}------------------------------------------------
There have been reports of burns under the anterior electrode when pacing adult patients with severely restricted blood flow to the skin. Prolonged pacing should be avoided in these cases and periodic inspection of the underlying skin is advised.
There are reports of transient inhibition of spontaneous respiration in unconscious patients with previously available units when the anterior electrode was placed too low on the abdomen.
## WARNING! This device must not be connected to internal pacemaker electrodes.
#### Pediatric Pacing
Pacing can be performed on pediatric patients weighing 33 lb. (15 kg) or less using ZOLL pediatric hands-free therapy electrodes. Prolonged pacing (in excess of 30 minutes), particularly in neonates, can cause burns. Periodic inspection of the underlying skin is recommended.
#### Intended Use - SpO2 Monitoring
The R Series pulse oximeter, with the Masimo® SET® technology and the LNCS® series of oximeter sensors, is indicated for the continuous, noninvasive monitoring of arterial oxygen saturation (SpO2) and pulse rate during both no motion and patient motion conditions for adult patients, and no motion conditions for pediatric and neonatal patients in a hospital or prehospital environment.
## Intended Use - EtCO2 Monitoring
The ZOLL R Series EtCO2 option with Respironics Novametrix technology is indicated for the continuous noninyasive monitoring of end tidal carbon dioxide (EtCO2) and respiration rate in patients requiring ventilator support, in-hospital transport, or anesthesia.
This option uses the CAPNOSTAT 5 Mainstream CO2 sensor attached to an airway adapter that connects to an endotracheal tube, mask or disposable mouthpiece.
The R Series EtCO2 option is designed to monitor adult, pediatric, and neonatal patients.
The following substances can influence CO2 measurements made with the CAPNOSTAT 5 CO2 sensor:
- · elevated oxygen levels
• nitrous oxide
- · halogenated agents
The R Series EtCO2 option provides settings for high oxygen and/or nitrous oxide compensation. Halogenated anesthetic agents alter CO2 readings, but the R Series unit will monitor CO2 within specifications when these agents are present at normal clinical levels. The presence of Desflurane in the exhaled breath beyond normal values (5%) may positively bias measured carbon dioxide values by up to an additional 3 mmHg.
The R Series EtCO2 option is intended for use only with the ZOLL/Respironics Novametrix CAPNOSTAT 5 Mainstream CO2 Sensor and mainstream airway adapters.
The R Series EtCO2 option can be used on adult patients (21 years of age and older) and on pediatric patients, as described in the following table:
| Pediatric Subpopulation | Approximate Age Range |
|-------------------------|---------------------------|
| Newborn (neonate) | Birth to 1 month of age |
| Infant | 1 month to 2 years of age |
| Child | 2 to 12 years of age |
| Adolescent | 12-21 years of age |
{5}------------------------------------------------
## Intended Use - NIBP
The ZOLL R Series NIBP option is indicated for the non-invasive measurement of arterial blood pressure for resting patients in critical care and in-hospital transport.
The R Series NIBP option is designed to measure blood pressure for adult patients (21 vears of age and older) and for pediatric patients, as described in the following table:
| Pediatric Subpopulation | Approximate Age Range |
|-------------------------|---------------------------|
| Newborn (neonate) | Birth to 1 month of age |
| Infant | 1 month to 2 years of age |
| Child | 2 to 12 years of age |
| Adolescent | 12-21 years of age |
#### ECG Monitoring
The patient's ECG is monitored by connecting the patient to the unit via a 3- or 5-lead patient cable, hands-free therapy electrodes, or through paddles. Five seconds of ECG is presented on the display along with the following information:
· averaged heart rate, derived by measuring R to R intervals
- · lead selection I, II, III, aVR, aVL, aVF, V (with ECG cable), PADDLES or PADS, P1, P2, P3 (when using OneStep Pacing cable with OneStep Complete electrodes).
P1, P2, and P3 are non-standard ECG leads derived from electrodes within particular OneStep electrodes. While ECG signals acquired from these leads are appropriate for rhythm assessment and determining electrical capture during pacing, they should not be used for ECG morphological evaluation. Attach conventional ECG electrodes for diagnostic purposes.
- · ECG size 0.5, 1, 1.5, 2, 3 cm/mV
- · other operational prompts, messages, and diagnostic codes
Monitoring or diagnostic ECG bandwidth is selectable.
## Recorder Function
The strip recorder is provided to document events. The strip recorder normally operates in the delay mode (6 seconds) to ensure the capture of ECG information immediately preceding critical events. The recorder may be activated manually by pressing the RECORDER button. It is activated automatically whenever a defibrillation SHOCK is delivered, a heart rate alarm occurs, or the rhythm analysis function is activated. The strip recorder may also be configured not to print during these events.
{6}------------------------------------------------
## R Series Plus and BLS Indications for Use:
## Defibrillator Function
The R Series product contains a direct current (DC) defibrillator capable of delivering up to 200 joules. It may be used in synchronized mode to perform synchronized cardioversion using the patient's R-wave as a timing reference. The unit uses paddles or disposable, pregelled electrodes for defibrillation.
## Intended Use - Manual Operation .
Use of the R Series products in the manual mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by:
- · Unconsciousness.
- · Absence of breathing.
- · Absence of pulse.
This product should be used only by qualified medical personnel for converting ventricular fibrillation and rapid ventricular tachycardia to sinus rhythm or other cardiac rhythms capable of producing hemodynamically significant heart beats.
In manual mode, the unit can also be used for synchronized cardioversion of certain atrial or ventricular arrhythmias. A qualified physician must decide when synchronized cardioversion is appropriate.
The advisory function should be used to confirm ventricular fibrillation or wide complex ventricular tachycardia (greater than 150 beats per minute) in patients meeting the three conditions indicating lack of circulation (listed above).
## Intended Use - Semiautomatic Operation (AED)
R Series products are designed for use by emergency care personnel who have completed training and certification requirements applicable to the use of a defibrillator where the device operator controls delivery of shocks to the patient.
They are specifically designed for use in early defibrillation programs where the delivery of a defibrillator shock during resuscitation involving CPR, transportation, and definitive care are incorporated into a medically-approved patient care protocol.
Use of the R Series in the Semiautomatic mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by:
- Unconsciousness.
- · Absence of breathing.
- · Absence of pulse.
Specifications for the ECG rhythm analysis function are provided in the section, "ECG Rhythm Analysis Algorithm Accuracy" on page A-29.
When the patient is less than 8 years of age or weighs less that 55 lbs. (25 Kg), you must use ZOLL pediatric defibrillation electrodes. Do not delay therapy to determine patient's exact age or weight.
#### Intended Use - ECG Monitoring
The unit is intended for use when ECG monitoring is indicated to evaluate the palient's heart rate or ECG morphology. In ECG monitoring mode, the unit is intended to be used by personnel who are qualified by training in the R Series defibrillator, basic life and/or advanced life support, or other physician-authorized emergency medical training.
{7}------------------------------------------------
## Intended Use - Real CPR Help
The Real CPR Help function provides visual and audio feedback designed to encourage rescuers to perform chest compressions at the AHA/ERC recommended rate of 100 compressions per minute. Voice and visual prompts encourage a compression depth of 2 inches (5 cm) minimum for adult patients.
Real CPR Help is not intended for use on patients under 8 years of age.
#### Semiautomatic Operation Contraindications for Use
The rhythm analysis function may not reliably identify ventricular fibrillation in the presence of an implanted pacemaker. Inspection of the electrocardiogram and clinical evidence of cardiopulmonary arrest should be the basis for any treatment of patients with implanted pacemakers.
Do not use the rhythm analysis function during patient movement on a stretcher or in an ambulance or other convevance. A patient must be motionless during ECG analysis. Do not touch the patient during analysis. Cease all movement of the stretcher or vehicle prior to analyzing the ECG. If you are using the R Series in an emergency vehicle, bring the vehicle to a halt before activating the analysis function.
#### Defibrillator Complications
Inappropriate defibrillation or cardioversion of a patient (for example, with no malignant arrhythmia) may precipitate ventricular fibrillation, asystole, or other dangerous arrhythmias.
Defibrillation without proper application of electrodes or paddle electrolyte gel might be ineffective and cause burns, particularly when repeated shocks are necessary. Erythema or hyperemia of the skin under the paddles, or electrodes often occurs; this effect is usually enhanced along the perimeter of the paddles or electrodes. This reddening should diminish substantially within 72 hours.
## Defibrillator Output Energy
R Series defibrillators can deliver as much as 200 joules into a 50 ohm impedance. The energy delivered through the chest wall, however, is determined by the patient's transthoracic impedance. An adequate amount of electrolyte qel must be applied to the paddles and a force of 10 to 12 kilograms (22 to 26.4 pounds) must be applied to each paddle in order to minimize this impedance. If hands-free therapy electrodes are used, make sure that they are properly applied. (Refer to the instructions on the electrode package).
#### External Pacemaker (Optional)
Some R Series products include an optional transcutaneous pacemaker consisting of a pulse generator and ECG-sensing circuitry. Noninvasive transcutaneous pacing (NTP) is an established and proven technique. This therapy is easily and rapidly applied in both emergency and nonemergency situations when temporary cardiac stimulation is indicated.
The output current of the pacemaker is continuously variable from 0 to 140 mA. The rate is continuously variable from 30 to 180 pulses per minute (ppm), by increments of 2.
The pacing output pulse is delivered to the heart via ZOLL hands-free defibrillation/pacing electrodes placed on the patient's back and the precordium.
The characteristics of the output pulse, together with the design and placement of the electrodes, minimize cutaneous nerve stimulation, cardiac stimulation threshold currents, and reduce discomfort due to skeletal muscle contraction.
The unique design of the R Series products allow clear viewing and interpretation of the electrocardiogram on the display without offset or distortion during external pacing.
{8}------------------------------------------------
Proper operation of the device, together with correct electrode placement, is critical to obtaining optimal results. Every operator must be thoroughly familiar with these operating instructions.
## Intended Use - Pacemaker
This product can be used for temporary external cardiac pacing in conscious or unconscious patients as an alternative to endocardial stimulation.
The purposes of pacing include:
· Resuscitation from standstill or bradycardia of any etiology.
Noninvasive pacing has been used for resuscitation from cardiac standstill, reflex vagal standstill, drug-induced standstill (due to procainamide, quinidine, digitalis, b-blockers, verapamil, etc.) and unexpected circulatory arrest (due to anesthesia, surgery, angiography, and other therapeutic or diagnostic procedures). It has also been used for temporary acceleration of bradycardia in Stokes-Adams disease and sick-sinus syndrome. It is safer. more reliable, and more rapidly applied in an emergency than endocardial or other temporary electrodes.
· As a standby when standstill or bradycardia might be expected.
Noninvasive pacing can be useful as a standby when cardiac arrest or symptomatic bradycardia might be expected due to acute myocardial infarction, drug toxicity, anesthesia, or surgery. It is also useful as a temporary treatment in patients awaiting pacemaker implants or the introduction of transvenous therapy. In standby pacing applications, noninvasive pacing might provide an alternative to transvenous therapy that avoids the risks of displacement, infection, hemorrhage, embolization, phlebitis, and mechanical or electrical stimulation of ventricular tachycardia or fibrillation associated with endocardial pacing.
· Suppression of tachycardia.
Increased heart rates in response to external pacing often suppress ventricular ectopic activity and might prevent tachycardia.
WARNING! This device can only be used for external pacing of patients and cannot be used for internal pacing. Do not connect internal pacing lead wires to the ZOLL defibrillator.
#### Pacemaker Complications
Ventricular fibrillation does not respond to pacing and requires immediate defibrillation. Therefore, the patient's dysrhythmia must be determined immediately, so that you can employ appropriate therapy. If the patient is in ventricular fibrillation is successful but cardiac standstill (asystole) ensues, you should use the pacemaker.
Ventricular or supraventricular tachycardias can be interrupted with pacing, but in an emergency or during circulatory collapse, synchronized cardioversion is faster and more certain.
Pulseless electrical activity (PEA) can occur following prolonged cardiac arrest or in other disease states with myocardial depression. Pacing might then produce ECG responses without effective mechanical contractions, making other effective treatment necessary.
Pacing can evoke undesirable responses, tachycardia, or fibrillation in the presence of generalized hypoxia, myocardial ischemia, cardiac drug toxicity, electrolyte imbalance, or other cardiac diseases.
Pacing by any method tends to inhibit intrinsic rhythmicity. Abrupt cessation of pacing, particularly at rapid rates, can cause ventricular standstill and should be avoided.
{9}------------------------------------------------
Noninvasive temporary pacing can cause discomfort of varying intensity, which occasionally can be severe and preclude its continued use in conscious patients.
Similarly, unavoidable skeletal muscle contraction might be troublesome in very sick patients and might limit continuous use to a few hours. Erythema or hyperemia of the skin under the hands-free therapy electrodes often occurs; this effect is usually enhanced along the perimeter of the electrode. This reddening should lessen substantially within 72 hours.
There have been reports of burns under the anterior electrode when pacing adult patients with severely restricted blood flow to the skin. Prolonged pacing should be avoided in these cases and periodic inspection of the underlying skin is advised.
There are reports of transient inhibition of spontaneous respiration in unconscious patients with previously available units when the anterior electrode was placed too low on the abdomen.
WARNING! This device can only be used for external pacing of patients and cannot be used for internal pacing. Do not connect internal pacing lead wires to the ZOLL defibrillator.
#### Pediatric Pacing
Pacing can be performed on pediatric patients weighing 33 lb. (15 kg) or less using ZOLL pediatric hands-free therapy electrodes. Prolonged pacing (in excess of 30 minutes), particularly in neonates, can cause burns. Periodic inspection of the underlying skin is recommended.
#### Intended Use - SpO2 Monitoring
The R Series pulse oximeter, with the Masimo® SET® technology and the LNCS® series of oximeter sensors, is indicated for the continuous, nonitoring of arterial oxygen saturation (SoO2) and pulse rate during both no motion and patient motion conditions for adult patients, and no motion conditions for pediatric and neonatal patients in a hospital or prehospital environment.
### Intended Use - EtCO2 Monitoring
The ZOLL R Series EtCO2 option with Respironics Novametrix technology is indicated for the continuous noninvasive monitoring of end tidal carbon dioxide (EtCO2) and respiration rate in patients requiring ventilator support, in-hospital transport, or anesthesia.
This option uses the CAPNOSTAT 5 Mainstream CO2 sensor attached to an airway adapter that connects to an endotracheal tube, mask or disposable mouthpiece.
The R Series EtCO2 option is designed to monitor adult, pediatric, and neonatal patients.
The following substances can influence CO2 measurements made with the CAPNOSTAT 5 CO2 sensor:
- · elevated oxygen levels
- nitrous oxide
- halogenated aqents
The R Series EtCO2 option provides settings for high oxygen and/or nitrous oxide compensation. Halogenated anesthetic agents alter CO2 readings, but the R Series unit will monitor CO2 within specifications when these agents are present at normal clinical levels. The presence of Desflurane in the exhaled breath beyond normal values (5%) may positively bias measured carbon dioxide values by up to an additional 3 mmHg.
The R Series EtCO2 option is intended for use only with the ZOLL/Respironics Novametrix CAPNOSTAT 5 Mainstream CO2 Sensor and mainstream airway adapters.
{10}------------------------------------------------
The R Series EtCO2 option can be used on adult patients (21 years of age and older) and on pediatric patients, as described in the following table:
| Pediatric Subpopulation | Approximate Age Range |
|-------------------------|---------------------------|
| Newborn (neonate) | Birth to 1 month of age |
| Infant | 1 month to 2 years of age |
| Child | 2 to 12 years of age |
| Adolescent | 12-21 years of age |
#### Intended Use - NIBP
The ZOLL R Series NIBP option is indicated for the non-invasive measurement of arterial blood pressure for resting patients in critical care and in-hospital transport.
The R Series NIBP option is designed to measure blood pressure for adult patients (21 years of age and older) and for pediatric patients, as described in the following table:
| Pediatric Subpopulation | Approximate Age Range |
|-------------------------|---------------------------|
| Newborn (neonate) | Birth to 1 month of age |
| Infant | 1 month to 2 years of age |
| Child | 2 to 12 years of age |
| Adolescent | 12-21 years of age |
#### Substantial Equivalence:
The features and functions of the proposed ZOLL R Series (with 2010 AHA Guidelines software update) are substantially equivalent to the ZOLL R Series (K081574, cleared for use on 9/8/2008 and K081828, cleared for use on 12/9/2008).
#### Comparison of Technological Characteristics
The technological characteristics of the proposed ZOLL R Series (with 2010 AHA Guidelines software update) are substantially equivalent to the currently marketed ZOLL R Series (K081574, cleared for use on 9/8/2008 and K081828, cleared for use on 12/9/2008).
#### Performance Testing:
Extensive performance testing ensures that the ZOLL R Series Defibrillator meets all of its functional requirements and performance specifications.
#### Conclusion
Performance testing of the ZOLL R Series Defibrillator demonstrates that its features and functions are substantially equivalent to that of the indicated commercially distributed predicate devices with regard to performance, safety and effectiveness.
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight. The symbol is composed of several curved lines that suggest the wings and body of the bird.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
MAR - 8 2011
Zoll Medical Corporation c/o Ms. Eileen M. Boyle Regulatory Affairs Specialist 269 Mill Road Chelmsford, MA 01824-4105
Re: K110361
Trade/Device Name: Zoll R Series with 2010 AHA Guidelines Software Update Regulation Number: 21 CFR 870.5310 Regulation Name: Automated External Defibrillator Regulatory Class: Class III (three) Product Code: MKJ Dated: January 26, 2011 Received: February 7, 2011
Dear Ms. Boyle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{12}------------------------------------------------
Page 2 - Ms. Eileen M. Boyle
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
F.l. Gend
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
{13}------------------------------------------------
# SECTION 4 - INDICATIONS FOR USE
510(k) Number (if known):
Device Name: ZOLL R Series
# Defibrillator Function
# Intended Use - Manual Operation
Use of the R Series products in the manual mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by:
- . Unconsciousness
- . Absence of breathing
- . Absence of pulse.
This product should be used only by qualified medical personnel for converting ventricular fibrillation and rapid ventricular tachycardia to sinus rhythm or other cardiac rhythms capable of producing hemodynamically significant heart beats.
In manual mode, the unit can also be used for synchronized cardioversion of certain atrial or ventricular arrhythmias. A qualified physician must decide when synchronized cardioversion is appropriate.
The advisory function should be used to confirm ventricular fibrillation or wide complex ventricular tachycardia (greater than 150 beats per minute) in patients meeting the three conditions indicating lack of circulation (listed above).
# Intended Use - Semiautomatic Operation (AED)
The R Series products are designed for use by emergency care personnel who have completed training and certification requirements applicable to the use of a defibrillator where the device operator controls delivery of shocks to the patient.
They are specifically designed for use in early defibrillation programs where the delivery of a defibrillator shock during resuscitation involing CPR, transportation, and definitive care are incorporated into a medically approved patient care protocol.
Use of the R Series in the Semiautomatic mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by:
- . Unconsciousness
- . Absence of breathing
- Absence of pulse t
Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K\10361
{14}------------------------------------------------
# Indications for Use
Specifications for the ECG rhythm analysis function are provided in the section, "ECG Rhythm Analysis Agorithm Accuracy" on page A-26.
When the patient is less than 8 years of age or weighs less than 55lbs (25Kg), you must use ZOLL pediatric defibrillation electrodes. Do not delay therapy to determine the patient's exact age or weight.
# Intended Use - ECG Monitoring
The unit is intended for use when ECG monitoring is indicated to evaluate the patient's heart rate or ECG morphology. In ECG monitoring mode, the unit is intended to be used by personnel who are qualified by training in the use of the R Series defibrillator, basic life and/or advanced life support, or other physician authorized emergency medical training.
Intended Use - CPR Monitoring
The CPR monitoring function provides visual and audio feedback designed to encourage rescuers to perform chest compressions at the AHA/ERC recommended rate of 100 compressions per minute. Voice and visual prompts encourage a compression depth of 1.5 to 2 inches (3.8 to 5.0 cm) for adult patients.
The CPR monitoring function is not intended for use on patients under 8 years of age.
External Pacemaker (Pacer Version Only)
Intended Use - Pacemaker
This product can be used for temporary external cardiac pacing in conscious or unconscious patients as an alternative to endocardial stimulation.
The purposes of pacing include:
Resuscitataion from standstill or bradycardia of any etiology:
Noninvasive pacing has been used for resuscitation from cardiac standstill, reflex vagal standestill, drug induced standstill (due to procainamide, quinidine, digitalis, b-blockers, verapamil, etc.) and unexpected circulatory arrest (due to anesthesia, surgery, angiography, and other therapeutic or diagnostic procedures). It has also been used for temporary acceleration of bradycardia in Stokes-Adams disease and sick-sinus syndrome. It is safer, more reliable, and more rapidly applied in an emergency than endocardial or other temporry electrodes.
{15}------------------------------------------------
As a standby when standstill or bradycardia might be expected:
Noninvasive pacing may be useful as a standby when cardiac arrest or symptomatic bradycardia might be expected due to acute myocardial infarction, drug toxicity, anesthesia or surgery. It is also useful as a temporary treatament in patients awaiting pacemaker implants or the introduction of transvenous therapy. In standby pacing applications, noninvasive pacing may provide an alternative to transvenous therapy that avoids the risks of displacement, infection, hemorrhage, embolization, perforation, phlebitis and mechanical or electrical stimulation of ventricular tachycardia or fibrillation associated with endocardial pacing.
## Supression of tachycardia:
Increased heart rates in response to external pacing often suppress ventricular ectopic activity and may prevent tachycardia.
## Pediatric Pacing
Pacing can be erformed on pediatric patients weighing 33lb. (15kg) or less using ZOLL pediatric hands-free therapy electrode pads. Prolonged pacing (in excess of 30 minutes), particularly in neonates, could cause burns. Periodic inspection of the underlying skin is recommended.
## Intended Use - SpO2 Monitoring
The R Series pulse oximeter, with the Masimo® SET® technology and the LNOP® series of oximeter sensors, is indicated for the continuous, noninvasive monitoring of arterial oxygen saturation (SpO2) and pulse rate during both no-motion and patient motion conditions for adult patients, and no motion conditions for pediatric and neonatal patients in a hospital or prehospital environment.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.