← Product Code [MKJ](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/MKJ) · K052057

# LIFEPAK 500 AUTOMATED EXTERNAL DEFIBRILLATOR (K052057)

_Medtronic Emergency Response Systems, Inc. · MKJ · Feb 17, 2006 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/MKJ/K052057

## Device Facts

- **Applicant:** Medtronic Emergency Response Systems, Inc.
- **Product Code:** [MKJ](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/MKJ.md)
- **Decision Date:** Feb 17, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5310
- **Device Class:** Class 3
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The LIFEPAK 500 AED is to be used only on patients in cardiopulmonary arrest. The patient must be unconscious, not breathing normally, and showing no signs of circulation (for example, no pulse, and/or no coughing, no movement) before the device is used to analyze the patient's ECG rhythm. With Infant/Child Reduced Energy Defibrillation Electrodes, the specially configured biphasic LIFEPAK 500 AED may be used on children up to 8 years old or 25kg (55 lb).

## Device Story

Portable, battery-operated AED for cardiac arrest treatment; used in hospital and out-of-hospital settings by trained personnel. Device inputs: patient ECG via disposable chest electrodes. Algorithm analyzes ECG to detect shockable rhythms. Output: visual/audible prompts for operator to deliver defibrillation shock. Modified version includes 'cprMax' technology, allowing medical directors/physicians to customize CPR settings to align with local resuscitation protocols. Benefits: provides automated rhythm analysis and defibrillation capability to facilitate rapid intervention in cardiac emergencies.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on design controls and bench testing of the modified software features.

## Technological Characteristics

Portable, battery-operated AED. Biphasic defibrillation waveform. Software-based ECG analysis algorithm. Includes cprMax technology for configurable CPR settings. Connectivity: not specified. Sterilization: not applicable. Software class: embedded firmware.

## Regulatory Identification

An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.

## Predicate Devices

- Medtronic LIFEPAK 500 AED ([K955854](/device/K955854.md))
- Medtronic LIFEPAK 500 AED ([K983393](/device/K983393.md))
- Medtronic LIFEPAK 500 AED ([K012428](/device/K012428.md))
- Philips Medical Systems Heartstream Heartstart FR2+ AED ([K013425](/device/K013425.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Kosz057

## FEB 1 7 2006

## 510(k) SUMMARY

Submitter's Name and Address:

Medtronic Emergency Response Systems 11811 Willows Road Northeast Redmond, WA 98052

Contact Person:

Sherri Pocock Regulatory Affairs Manager

Date Summary Prepared:

December 12, 2005

Device:

Medtronic biphasic LIFEPAK® 500 Automated External Defibrillator (AED)

Classification:

Low Energy DC-Defibrillator: Class II Automatic External Defibrillator: Class III

Substantial Equivalence:

The features and functions of the modified biphasic LIFEPAK 500 AED are I the readies and railer. to the previously cleared Medtronic LIFEPAK 500 AED, Substantially Squarants K955854 (11/4/96), K983393 (05/05/99), K012428 (09/28/01), 310(k) hambers 119690); and the Philips Medical Systems Heartstream Heartstart FR2+ AED, 510(k) number K013425 (01/14/2002).

Description:

The LIFEPAK 500 automated external defibrillator is a small, portable, battery operated device intended for treatment of cardiac arrest. The device uses a opented software algorithm to analyze the patient's electrocardiogram (ECG) to patented solve if a shockable rhythm is present. The LIFEPAK 500 AED will inform determine if it detects a shockable rhythm and the operator can press the shock the operator were energy. The energy is delivered via disposable defibrillation electrodes applied to the chest.

{1}------------------------------------------------

Intended Use:

The LIFEPAK 500 AED intended for use on patients in cardiac arrest. It is The En EFTIN 500 ALD spital and in out-of-hospital environments.

The LIFEPAK 500 AED is intended for use by personnel who are authorized by a The LIFET AN 500 AED to minimum, the following skills and training:

- . CPR training
- CI K training equivalent to that recommended by the American Heart . Association
- Training in the use of the LIFEPAK 500 AED �

Indications For Use:

The LIFEPAK 500 AED is to be used only on patients in cardiopulmonary arrest. The LIFEF AN 500 ALD is to be about exampling normally, and showing no signs of The patient must be unconsolous, no coughing, no movement) before the circulation (for example, no pulse, and of 5 rhythm. With Infant/Child Reduced device is used to analyze the pacially configured biphasic LIFEPAK 500 AED may be used on children up to 8 years old or 25kg (55 lb).

Technological characteristics of new and predicate device:

Defibrillation technology and functionality are unchanged from the previously Declorination technology and Imerely and 500 software has been modified to allow the medical director or physician greater control over CPR settings.

The setup options, collectively known as cprMax technology, are intended to The setup options, occroor physicians to more easily implement resuscitation anow modious in their Emergency Medical Services systems or AED programs.

Summary of Design Controls:

This 510(k) includes a summary of design control activities and a declaration of conformity to design controls.

The information in this 510(k) notification demonstrates that the modified LIFEPAK 500 AED is substantially equivalent to the predicate devices.

K05 2057

{2}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 1 7 2006

Medtronic Emergency Response Systems c/o Ms. Sherri Pocock Regulatory Affairs Manager 11811 Willows Road Northeast Redmond. WA 98052

Re: K052057

: 1892037
Trade Name: LIFEPAK 500 Automatic External Defibrillator Regulation Number: 21 CFR 870.5310 Regulation Name: Automatic External Defibrillator Regulatory Class: Class III Product Code: MKJ Dated: January 9, 2006 Received: January 10, 2006

Dear Ms. Pocock:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate 101 use stated in the enerobars) to stg. 25, 1976, the Medical Device Amendments, or to commerce price to May 20, 1978, in accordance with the provisions of the Federal Food, Drug, devices that have been resulted trequire approval of a premarket approval application (PMA). alla Cosmetic Hot (1101) that to nov sequent to the general controls provisions of the Act. The I ou may, merefore, maniel and include requirements for annual registration, listing of general controls provisions of as tice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is elabilitional controls. Existing major regulations affecting your device can may or subject to back as a regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Sherri Pocock

Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issualled of a babbains with other requirements of the Act
that FDA has made a determination that your device complies with other requirements of that I DA has made a decemination administered by other Federal agencies. You must as a light or any Federal statutes and regulations administered of other its registration and listing (21 let comply with an the Act s requirements, mercains, wantacturing practice requirements as set CFR Part 807), labeling (21 CFR Part 820); and If applicable, the electronic forth in the quanty systems (Sections (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control provisions (Sections of Levice as described in your Section 510(k) This letter will anow you to begin manteing your antial equivalence of your device to a legally premarket notheadon: The PDF miding of basis.if the to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your ac not as a cases note the regulation entitled, contact the Office of Compulance at (210) 216 or art 807.97). You may obtain "Misoranung by televelee to promantee to promantee the Act from the Division of Small other general Informational and Consumer Assistance at its toll-free number (800) 638-2041 or Manufacturers, International and Ochsanter v/swww.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Blumimmofor

Bram D. Zuckerman, MD Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the Medtronic logo. The logo consists of a stylized human figure inside of a circle on the left and the word "Medtronic" in bold, black letters on the right. The human figure appears to be in motion, with arms and legs outstretched.

## SECTION D - STATEMENT OF INDICATIONS FOR USE

510(k) Number (if known):

Device Name: LIFEPAK 500 Automated External Defibrillator

Indications For Use:

The LIFEPAK 500 AED is to be used only on patients in cardiopulmonary arrest. The patient must be unconscious, not breathing normally, and showing ano signs of circulation (for example, no pulse, and/or no coughing, no movement) before the device is used to analyze the patient's ECG rhythm. With Infant/Child Reduced Energy Defibrillation Electrodes, the specially configured biphasic LIFEPAK 500 AED may be used on children up to 8 years old or 25kg (55 lb).

Prescription Use _ (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use_ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Bhumma
Sign-Off

ion of Cardlovascular Devloes ා Number

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/MKJ/K052057](https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/MKJ/K052057)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
