Sprinter Legend Rapid Exchange Balloon Dilatation Catheter

K251970 · Medtronic, Ireland · LOX · Jan 17, 2026 · Cardiovascular

Device Facts

Record IDK251970
Device NameSprinter Legend Rapid Exchange Balloon Dilatation Catheter
ApplicantMedtronic, Ireland
Product CodeLOX · Cardiovascular
Decision DateJan 17, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5100
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Sprinter Legend RX and OTW 1.25mm Balloon Dilatation Catheter is indicated as a pre dilatation catheter for enlarging coronary luminal diameters during PCI procedures. The Sprinter Legend RX 1.5-4.0mm balloon dilatation catheter is indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion. The balloon dilatation catheter (balloon models 2.25 mm - 4.0 mm) is also indicated for post-delivery expansion of balloon expandable stents.

Device Story

Sprinter Legend Rapid Exchange Balloon Dilatation Catheter is a PTCA device used by physicians in clinical settings to treat coronary artery stenosis. The device consists of a proximal luer hub, strain relief, hypotube, and stiffening wire, with a distal end featuring an inflatable balloon, dual-lumen tip, and radiopaque marker bands. A hydrophilic coating on the distal shaft facilitates advancement. The physician tracks the device over a 0.014” guidewire to the target lesion, using radiopaque markers for precise positioning. Once in place, the balloon is inflated to dilate the stenotic vessel or expand a stent. The device improves myocardial perfusion by enlarging luminal diameters. It is a mechanical intervention device; no electronic or software components are involved.

Clinical Evidence

No clinical data was required for this submission. Substantial equivalence was demonstrated through non-clinical bench testing, including biocompatibility (ISO 10993) and performance testing (track assessment, particulate characterization, and interface testing).

Technological Characteristics

PTCA balloon dilatation catheter. Features: hydrophilic-coated distal shaft, dual-lumen tip, radiopaque marker bands, and hypotube/stiffening wire proximal assembly. Materials evaluated per ISO 10993. Mechanical operation via inflation lumen. No energy source, software, or connectivity.

Indications for Use

Indicated for patients undergoing percutaneous coronary intervention (PCI) requiring pre-dilatation of coronary luminal diameters (1.25mm models), balloon dilatation of stenotic coronary arteries or bypass grafts to improve myocardial perfusion (1.5-4.0mm models), or post-delivery expansion of balloon-expandable stents (2.25-4.0mm models).

Regulatory Classification

Identification

Standard PTCA Catheter: A PTCA catheter is a device that operates on the principle of hydraulic pressurization applied through an inflatable balloon attached to the distal end. A PTCA balloon catheter has a single or double lumen shaft. The catheter features a balloon of appropriate compliance for the clinical application, constructed from a polymer. The balloon is designed to uniformly expand to a specified diameter and length at a specific pressure as labeled, with well characterized rates of inflation and deflation and a defined burst pressure. The device generally features a type of radiographic marker to facilitate fluoroscopic visualization of the balloon during use. A PTCA catheter is intended for balloon dilatation of a hemodynamically significant coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. A PTCA catheter may also be intended for the treatment of acute myocardial infarction; treatment of in-stent restenosis (ISR) and/or post-deployment stent expansion. Cutting/scoring PTCA Catheter: A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.

Special Controls

*Classification.* Class II (special controls). The special control for this device is “Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters.” See § 870.1(e) for the availability of this guidance document.(b) *Cutting/scoring PTCA Catheter* —(1)*Identification.* A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.(2) *Classification.* Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 870.3.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION January 17, 2025 Medtronic Ireland Michelle Greaney Senior Regulatory Affairs Specialist Parkmore Business Park West Galway, H91 VY19 Ireland Re: K251970 Trade/Device Name: Sprinter Legend Rapid Exchange Balloon Dilatation Catheter Regulation Number: 21 CFR 870.5100 Regulation Name: Percutaneous Transluminal Coronary Angioplasty (Ptca) Catheter Regulatory Class: Class II Product Code: LOX Dated: June 26, 2025 Received: December 18, 2025 Dear Michelle Greaney: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K251970 - Michelle Greaney Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2} K251970 - Michelle Greaney Page 3 Sincerely, Jenny R. Katsnelson -S Digitally signed by Jenny R. Katsnelson -S Date: 2026.01.17 17:15:13 -05'00' for Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K251970 | | | Device Name Sprinter Legend Rapid Exchange Balloon Dilation Catheter | | | Indications for Use (Describe) The Sprinter Legend RX and OTW 1.25mm Balloon Dilatation Catheter is indicated as a pre dilatation catheter for enlarging coronary luminal diameters during PCI procedures. The Sprinter Legend RX 1.5-4.0mm balloon dilatation catheter is indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion. The balloon dilatation catheter (balloon models 2.25 mm - 4.0 mm) is also indicated for post-delivery expansion of balloon expandable stents. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} Page 1 of 4 K251970 # Traditional 510(k) Summary per 21 CFR 807.92 Submitter: Medtronic Ireland Parkmore Business Park West Galway Ireland Contact Name(s): (1) Michelle Greaney Senior Regulatory Affairs Specialist Phone: +353-87-9978808 Email: michelle.greaney@medtronic.com (2) Laura Guinane Senior Regulatory Affairs Program Manager Phone: +353-86-1051563 Email: laura.m.guinane@medtronic.com Date Prepared: January 14, 2026 Trade Name: Sprinter Legend Rapid Exchange Balloon Dilatation Catheter Common Name: Catheter Device Classification: Class II Classification Name: Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter Classification Panel: Cardiovascular Classification Regulation: 21 CFR 870.5100 Product Code: LOX {5} Page 2 of 4 K251970 ## Predicate Device: | Predicate Device Clearance/Approval # | Predicate Device Name | | --- | --- | | P790017/S096 | Sprinter Legend Rapid Exchange Balloon Dilatation Catheter (1.5 to 4.0mm) | | K103095 | Sprinter Legend Rapid Exchange Balloon Dilatation Catheter (1.25mm) | | *On October 08th 2010, PTCA devices were reclassified to Class II in the U.S. | | ## Device Description: Sprinter Legend Rapid Exchange Balloon Dilatation Catheter is a percutaneous transluminal coronary angioplasty (PTCA) device. The proximal end comprises of a luer hub, strain relief and a hypotube and stiffening wire subassembly (proximal shaft assembly). The distal end comprises of an inflatable balloon, dual lumen tip (guidewire lumen and the inflation lumen) and radiopaque marker bands. The distal shaft includes a hydrophilic coating to aid in device advancement to the target lesion. The guidewire lumen at the distal end enables the use of a 0.014” guidewire to position the device during use. Radiopaque balloon markers enable accurate placement at the target lesion. Exit markers on the proximal shaft indicate the exit of the balloon tip out of the guide catheter for brachial and femoral approaches. ## Indication For Use: | 1.25mm size | The Sprinter Legend RX and OTW 1.25 mm Balloon Dilatation Catheter is indicated as a pre dilatation catheter for enlarging coronary luminal diameters during PCI procedures. | | --- | --- | | 1.5 to 4.0mm sizes | The Sprinter Legend RX 1.5-4.0mm balloon dilatation catheter is indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion. The balloon dilatation catheter (balloon models 2.25mm-4.0mm) is also indicated for post-delivery expansion of balloon expandable stents. | ## Comparison to Predicate Devices: The subject device is substantially equivalent to the predicate device, which is Medtronic’s own legally market device, and have comparable safety and performance. There is no change in sterilization method, sterility assurance level and shelf life. The material difference does not introduce different questions of safety or effectiveness compared to the predicate, as demonstrated by the bench testing. {6} Page 3 of 4 K251970 # Summary of Non-Clinical Data: The following non-clinical tests were performed to evaluate and demonstrate substantial equivalence between the subject device and the predicate device. The results confirmed that the devices are substantially equivalent. The following tests were performed: # Biocompatibility Evaluation (conducted per ISO 10993 standards): - Chemical Characterization - Toxicological Risk Assessment - Cytotoxicity - Acute Systemic Toxicity - Hemocompatibility - Hemolysis - Complement Activation - Partial Thromboplastin Time (PTT) - Platelet/Leucocyte Count - Sensitization - Irritation - Material Mediated Pyrogenicity # Performance Testing included: - Exit Marker Locations - Shaft OD (Outer Diameter) - Coated Hypotube - Product Removal from the Hoop - Product Interface testing - Looper - Y-Adaptors - Guide Catheters - Guidewires - Delivery & Retraction Characterization (Track Assessment) - PTFE Particulate Characterization Testing - PTFE Durability Characterization {7} Page 4 of 4 K251970 All test results met the acceptance criteria, which are consistent with the predicate device, demonstrating that the subject device meets established performance specifications and is suitable for its intended use. ## Summary of Clinical Data: No clinical testing was required for this 510(k) submission. ## Conclusion from Data: The differences between the subject device and predicate device have been evaluated through non-clinical testing, which demonstrates that the subject device is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...