← Product Code [LOX](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/LOX) · K103153

# VOYAGER NC CORONARY DILATATION CATHETER AND NC TREK RX CORONARY DILATATION CATHETER (K103153)

_Abbott Vascular-Cardiac Therapies · LOX · Dec 30, 2010 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/LOX/K103153

## Device Facts

- **Applicant:** Abbott Vascular-Cardiac Therapies
- **Product Code:** [LOX](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/LOX.md)
- **Decision Date:** Dec 30, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5100
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The NC TREK™ RX Coronary Dilatation Catheter is indicated for: a) balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perforusion; b) balloon dilatation of a coronary artery occlusion for the purpose of restoring coronary flow in patients with ST-segement elevation myocardial infarction; c) balloon dilatation of a stent after implantation (balloon models 2.0 mm - 5.0 mm only).

## Device Story

Rapid exchange, co-axial coronary dilatation catheter; features distal balloon and radiopaque markers for positioning. Inner member accommodates guide wire; outer lumen enables balloon inflation/deflation with contrast fluid. Used by physicians in clinical settings (e.g., cath lab) to dilate stenotic coronary arteries or bypass grafts, restore flow in STEMI patients, or expand stents. Device advancement to stenosis facilitated by guide wire; balloon inflation achieves mechanical dilatation. Benefits include improved myocardial perfusion and restored coronary blood flow.

## Clinical Evidence

Bench testing only. Testing included catheter preparation, bond tensile strength, body pressure integrity, kink/flexibility, and torque. Biocompatibility testing included cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, hemolysis, pyrogen, and complement activation. Results met all acceptance criteria.

## Technological Characteristics

Co-axial design with tubular inner/outer members; rapid exchange. Features distal balloon and radiopaque markers. Materials include hypotube proximal shaft. Sterilization method not specified. No software or electronic components.

## Regulatory Identification

Standard PTCA Catheter: A PTCA catheter is a device that operates on the principle of hydraulic pressurization applied through an inflatable balloon attached to the distal end. A PTCA balloon catheter has a single or double lumen shaft. The catheter features a balloon of appropriate compliance for the clinical application, constructed from a polymer. The balloon is designed to uniformly expand to a specified diameter and length at a specific pressure as labeled, with well characterized rates of inflation and deflation and a defined burst pressure. The device generally features a type of radiographic marker to facilitate fluoroscopic visualization of the balloon during use. A PTCA catheter is intended for balloon dilatation of a hemodynamically significant coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. A PTCA catheter may also be intended for the treatment of acute myocardial infarction; treatment of in-stent restenosis (ISR) and/or post-deployment stent expansion. Cutting/scoring PTCA Catheter: A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is “Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters.” See § 870.1(e) for the availability of this guidance document.(b)
*Cutting/scoring PTCA Catheter* —(1)*Identification.* A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.(2)
*Classification.* Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 870.3.

## Predicate Devices

- VOYAGER® NC Coronary Dilatation Catheter

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# KU3153

## DEC 30 2010

l

### 510(k) Summary

The 510(k) Summary is submitted in accordance with 21 CFR §807.92 and the requirements of the Safe Medical Device Act (SMDA) of 1990.

| 1. SUBMITTER'S NAME              | Abbott Vascular Inc.                                                                                  |
|----------------------------------|-------------------------------------------------------------------------------------------------------|
| 2. SUBMITTER'S ADDRESS           | 26531 Ynez Road, Temecula, CA 92591                                                                   |
| 3. TELEPHONE                     | (951) 914-3243                                                                                        |
| 4. FAX                           | (951) 914-0339                                                                                        |
| 5. CONTACT PERSON                | Suzanne Redman                                                                                        |
| 6. DATE PREPARED                 | November 9, 2010                                                                                      |
| 7. DEVICE TRADE NAME             | NC TREK™ RX Coronary Dilatation Catheter                                                              |
| 8. DEVICE COMMON NAME            | • Coronary Dilatation Catheter<br>• Percutaneous Transluminal Coronary<br>Angioplasty (PTCA) Catheter |
| 9. DEVICE CLASSIFICATION<br>NAME | PTCA Catheter (LOX)                                                                                   |
| 10. PREDICATE DEVICE NAME        | VOYAGER® NC Coronary Dilatation Catheter                                                              |

### 11. DEVICE DESCRIPTION

The NC TREK RX Coronary Dilatation Catheters is a rapid exchange co-axial design with a balloon at the distal tip. Table 1 provides a matrix of the balloon diameters and lengths available.

{1}------------------------------------------------

| Balloon<br>Diameter<br>(mm) | Catalog<br>Number | Balloon Length |     |     |     |     |     |
|-----------------------------|-------------------|----------------|-----|-----|-----|-----|-----|
| 1.50                        | 1012444           | -06            | -08 | -12 | -15 | -20 |     |
| 2.00                        | 1012445           | -06            | -08 | -12 | -15 | -20 |     |
| 2.50                        | 1012447           | -06            | -08 | -12 | -15 | -20 | -25 |
| 2.75                        | 1012448           | -06            | -08 | -12 | -15 | -20 |     |
| 3.00                        | 1012449           | -06            | -08 | -12 | -15 | -20 | -25 |
| 3.25                        | 1012450           |                |     | -12 | -15 | -20 |     |
| 3.50                        | 1012451           | -06            | -08 | -12 | -15 | -20 | -25 |
| 3.75                        | 1012452           |                |     | -12 | -15 | -20 |     |
| 4.00                        | 1012453           |                | -08 | -12 | -15 | -20 |     |
| 4.50                        | 1012454           |                | -08 | -12 |     |     |     |
| 5.00                        | 1012455           |                | -08 | -12 |     |     |     |

#### Table 1 NC TREK RX Balloon Sizes

The balloon segment expands to a known diameter and length at a specific inflation pressure and has radiopaque marker(s) under the balloon to aid in positioning the balloon in a stenosis. The coaxial shaft consists of a tubular inner and outer member. The inner member permits the use of a guide wire to facilitate advancement of the catheter to and through the stenosis to be dilated. The outer lumen provides for inflation and deflation of the balloon with contrast fluid. The proximal shaft consists of a hypotube with a hub on the proximal end, a tapered distal section ending distal to the guide wire notch junction, along with brachial and femoral markers.

### 12. INDICATIONS FOR USE

The NC TREK™ RX Coronary Dilatation Catheter is indicated for:

a) balloon dilatation of the stentotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion;

b) balloon dilatation of a coronary artery occlusion of the purpose of restoring coronary flow in patients with ST-segment elevation myocardial infarction;

c) balloon dilatation of a stent after implantation (balloon models 2.0 mm - 5.0 mm only).

#### TECHNOLOGICAL CHARACTERISTICS 13.

Comparisons of the new and predicate devices show that the technological characteristics such as product performance, design and intended use are substantially equivalent to the current marketed predicate devices.

{2}------------------------------------------------

#### 14. PERFORMANCE DATA

The NC TREK RX Coronary Dilatation Catheters were subjected to the following in vitro bench tests according to the requirements of Guidance for Industry and FDA Staff -Class II Special Controls for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters, September 8, 2010:

- Catheter Preparation .
- Catheter Bond Tensile Strength .
- Catheter Body Pressure Integrity .
- . Kink and Flexibility
- . Torque

Biocompatibility testing included cyctotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, hemolysis, pyrogen, and complement activation.

These in vitro bench and biocompatibility tests demonstrated that the NC TREK RX Coronary Dilatation Catheters met all acceptance criteria and performed similarly to the predicate devices. No new safety or effectiveness issues were raised during the testing program and therefore, these devices may be considered substantially equivalent to the predicate devices.

3

{3}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Abbott Vascular, Inc. c/o Ms. Suzanne Redman Regulatory Affairs 26531 Ynez Road Temecula, CA 92591

DEC 30 2010

Re: K103153

Trade/Device Name: NC Trek RX Coronary Dilatation Catheter Regulation Number: 21 CFR 870.5100 Regulation Name: Percutaneous Transluminal Coronary Angioplasty Catheter Regulatory Class: Class II Product Code: LOX Dated: October 13, 2010 Received: October 14, 2010

Dear Ms. Redman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{4}------------------------------------------------

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fdagov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## INDICATIONS FOR USE

#### 510(k) Number (if known): K103153

DEC 30 2010

Device Names: NC TREK™ RX Coronary Dilatation Catheter

Indications for Use:

The NC TREK™ RX Coronary Dilatation Catheter is indicated for: a) balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perforusion;

b) balloon dilatation of a coronary artery occlusion for the purpose of restoring coronary flow in patients with ST-segement elevation myocardial infarction;

c) balloon dilatation of a stent after implantation (balloon models 2.0 mm - 5.0 mm only).

Prescription Use X (Per 21 CFR 801.109)

OR

Over-The-Counter (Optional Format 1-1-96)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

W.M.P.

(Division Sign-Off)
Division of Cardiovascular Devices
K103153

510(k) Number

Page _________________________________________________________________________________________________________________________________________________________________________ ੀ ਸ

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/LOX/K103153](https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/LOX/K103153)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
