700-F SERIES STIMULATION ELECTRODES

K964469 · Cardiotronics Systems, Inc. · LDD · Dec 11, 1996 · Cardiovascular

Device Facts

Record IDK964469
Device Name700-F SERIES STIMULATION ELECTRODES
ApplicantCardiotronics Systems, Inc.
Product CodeLDD · Cardiovascular
Decision DateDec 11, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5300
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The 700-F series electrodes are disposable electrodes for external cardiac stimulation and monitoring. Electrodes of this design are suitable for external pacing, external defibrillation, synchronized cardioversion and ECG monitoring. The expected patient population for the use of these devices includes adults and children. The environment for use includes all areas of the hospital, ambulance, and prehospital (paramedic) situations. The duration for use is intended to be less than 24 hours. Individual catalog numbers are labeled for more specific use based primarily on the interface connector and cable system which defines the host external device, such as an external defibrillator. The devices covered by this premarket notification provide an interface from the skin to a number diagnostic or therapeutic devices and are labeled with one or more of the following intended uses: - External Defibrillation - External Pacing - ECG Monitoring - Synchronized Cardioversion

Device Story

Disposable electrodes for external cardiac stimulation/monitoring; consist of foam backing, metallic substrate, and conductive hydrogel. Applied to patient chest; connected via integral lead wire to host defibrillator/pacemaker/monitor. Facilitates energy delivery for defibrillation/pacing/cardioversion and signal acquisition for ECG monitoring. Used in hospitals, ambulances, and prehospital settings by clinicians/paramedics. Provides interface between skin and therapeutic/diagnostic equipment; enables clinical decision-making based on ECG monitoring and therapeutic intervention. Benefits patient through non-invasive cardiac support.

Clinical Evidence

Bench testing included electrical performance per ANSI/AAMI DF-39 (subsection 3.3.19) and mechanical parameters. Biocompatibility testing (cytotoxicity, skin irritation, sensitization) confirmed material safety. Limited animal study (live pigs, 24-hour duration) demonstrated effective energy delivery, clear ECG signal acquisition, and lack of adverse skin changes.

Technological Characteristics

Foam backing, laminated metallic substrate, conductive hydrogel, adhesive foam ring. Low-impedance, large-surface design. Integral lead wire with interface connectors. Compliant with AAMI DF39 standards. Non-sterile (implied by disposable nature).

Indications for Use

Indicated for external cardiac stimulation (pacing, defibrillation, synchronized cardioversion) and ECG monitoring in adult and pediatric patients in hospital, ambulance, and prehospital settings. Duration of use is less than 24 hours.

Regulatory Classification

Identification

Low-energy DC-defibrillator: A device that delivers into a 50 ohm test load an electrical shock of a maximum of 360 joules of energy used for defibrillating (restoring normal heart rhythm) the atria or ventricles of the heart or to terminate other cardiac arrhythmias. This generic type of device includes low energy defibrillators with a maximum electrical output of less than 360 joules of energy that are used in pediatric defibrillation or in cardiac surgery. The device may either synchronize the shock with the proper phase of the electrocardiogram or may operate asynchronously. The device delivers the electrical shock through paddles placed either directly across the heart or on the surface of the body. High-energy DC-defibrillator: A device that delivers into a 50 ohm test load an electrical shock of greater than 360 joules of energy used for defibrillating the atria or ventricles of the heart or to terminate other cardiac arrhythmias. The device may either synchronize the shock with the proper phase of the electrocardiogram or may operate asynchronously. The device delivers the electrical shock through paddles placed either directly across the heart or on the surface of the body.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} CARDIOTRONICS K964469 # 510(k) Summary Summary of information contained in the 510(k) Premarket Notification Date: November 5, 1996 Submitter: Cardiotronics Systems, Inc. 5966 La Place Court Carlsbad, California 92008 Contact Person Tim J. Way Vice President - Regulatory Affairs Telephone (619) 431-9446 Facsimile (619) 431-1805 Tradename: 700-F Series Disposable Cardiac Stimulation/ Monitoring Electrodes (a number of end products, including those manufactured under private label) Classification: These electrodes are disposable accessories to the following products: | Product | Code | Class | CFR Section | | --- | --- | --- | --- | | DC-Defibrillator, Low Energy (including paddles) | 74LDD II | | 870.530 | | Pacemaker, Cardiac, External Transcutaneous | 74DRO III | | 870.555 | or under specific classification related to ECG monitoring would be considered: Electrode, Electrocardiographic 74DRX II 870.230 The 700-F series electrodes are substantially equivalent to the R2 Medical Systems 600 Series electrode. Description: The 700-F Series electrodes consist of a foam backing material, a laminated metallic substrate which is covered by a conductive adhesive (hydrogel). This hydrogel is then surrounded by an adhesive coated foam ring and covered by a removable release liner. The electrodes have an integral lead wire for attachment to the interface cable of the defibrillator/pacemaker/monitor. 14-1 {1} CARDIOTRONICS S Y S T E M S Upon use, the release liner is removed from the electrode, exposing the hydrogel and adhesive areas. Each electrode of the set is then applied to the patient's chest in the selected set and pressed to ensure adhesion. The electrode is then attached to the host device via the interface cable and treatment is performed in accordance with established protocols. The 700-F Series electrodes are produced in two basic styles; Adult and Child. Leadwire sets having various lengths and interface connectors are added to the basic electrode styles to obtain many combinations of end product design. The product line is further differentiated through trade name and OEM identification. ## Intended Use: The 700-F series electrodes are disposable electrodes for external cardiac stimulation and monitoring. Electrodes of this design are suitable for external pacing, external defibrillation, synchronized cardioversion and ECG monitoring. The expected patient population for the use of these devices includes adults and children. The environment for use includes all areas of the hospital, ambulance, and prehospital (paramedic) situations. Individual catalog numbers are labeled for more specific use based primarily on the interface connector and cable system which defines the host external device, such as an external defibrillator. The devices covered by this premarket notification provide an interface from the skin to a number diagnostic or therapeutic devices and are labeled with one or more of the following intended uses: - External Defibrillation - External Pacing - ECG Monitoring - Synchronized Cardioversion ## Substantial Equivalence: Note: As used herein, the term "substantial equivalence" is only as used in 21 CFR 807.81. The design, materials, and intended uses of the 700-F series electrodes are equivalent to the those of the R2 Medical Systems 600 Series electrodes. As in the 600 Series, the 700-F series are considered to be low impedance, large surface electrodes suitable for external defibrillation, ECG monitoring, and external pacing. Substantial equivalence has been determined primarily on the basis of electrical performance results in conformance to the AAMI standard DF39 (Subsection 3.3.19) and comparisons to materials and physical dimensions of the 600 Series. The 700-F series electrodes differ from the 600-Series electrodes primarily in the conductive adhesive (hydrogel) material. While both hydrogels are of a similar chemical family, the 700-F series utilize a cure-in-place hydrogel manufactured by Cardiotronics while the 600 series contain a commercially available hydrogel suitable for this application. All other primary components and methods of manufacture are equivalent to the 600 series electrodes. 14-2 {2} CARDIOTRONICS S Y S T E M S # Performance Evaluation Bench testing of the 700-F Series electrodes consisted of a battery of both electrical and mechanical tests. These tests include electrical performance parameters as indicated in ANSI/AAMI DF-39 subsection 3.3.19 as well as a number of mechanical performance parameters developed by the manufacturer. These test show the 700-F series electrodes to be suitable for the uses and environment specified. Biocompatibility studies were conducted on the hydrogel conductive adhesive and the adhesive foam ring. The tests on the hydrogel consisted of Cytotoxicity, Primary Skin Irritation, and Delayed Contact Sensitization. Tests on the adhesive foam ring consisted of Primary Skin Insult Patch test. These tests showed all materials which contact the skin to be biocompatible. In addition to the mechanical, electrical, and biocompatibility tests, a limited number of electrodes were tested on live pigs over a twenty four hour period to demonstrate effectiveness at delivering energy, ability to sense clear ECG signal and to ensure that these electrodes do not contribute to any significant skin changes during or after the delivery of energy. The results of the aforementioned battery of tests demonstrate that the 700-F series electrodes are substantially equivalent to the 600 series electrodes and are suitable for the specified intended uses in the environment for use indicated. 237 14-3 {3} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 AUG 26 2011 Cardiotronics Systems, Inc. c/o Mr. Tim J. Way 5966 La Place Court Carlsbad, CA 92008 Re: K964469 700-FR Series Stimulation Electrodes Regulation Number: 21 CFR 870. 5300 Regulation Name: DC-Defibrillator (Including paddles) Regulatory Class: Class II (two) Product Code: LDD Dated: November 5, 1996 Received: December 7, 1996 Dear Mr. Way: This letter corrects our substantially equivalent letter of December 11, 1996. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent [(for the indications for use stated in the enclosure)] to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4} Page 2 - Mr. Tim J. Way Please be advised that FDA’s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act’s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally Part 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4639. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at their toll free number (800) 638-2041 or at (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, ![img-0.jpeg](img-0.jpeg) {5} # Indications for use Statement 510(k) Number K964469 Device Name: 700-F Series Disposable Stimulation Electrodes Indications for Use: The 700-F series electrodes are disposable electrodes for external cardiac stimulation and monitoring. Electrodes of this design are suitable for external pacing, external defibrillation, synchronized cardioversion and ECG monitoring. The expected patient population for the use of these devices includes adults and children. The environment for use includes all areas of the hospital, ambulance, and prehospital (paramedic) situations. The duration for use is intended to be less than 24 hours. Individual catalog numbers are labeled for more specific use based primarily on the interface connector and cable system which defines the host external device, such as an external defibrillator. The devices covered by this premarket notification provide an interface from the skin to a number diagnostic or therapeutic devices and are labeled with one or more of the following intended uses: - External Defibrillation - External Pacing - ECG Monitoring - Synchronized Cardioversion Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription use _____ Over the Counter _____ Richard Phillips (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number K964469 4
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...