K111748 · Joy Long · JOW · Nov 23, 2011 · Cardiovascular
Device Facts
Record ID
K111748
Device Name
RECOVERY+ MODEL PS 511, DVTHERAPY MODEL PS 511A
Applicant
Joy Long
Product Code
JOW · Cardiovascular
Decision Date
Nov 23, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Pulsar Scientific device is not a life-supporting or life-sustaining device. It is a new, prescription use only [21 CFR 801 Subpart D] device for use with restricted mobility and non-ambulatory patients, or other patients as deemed appropriate by the physician. It is not intended for over-the-counter use [21 CFR 801 Subpart C]. The device is intended to function as an intermittent, external pneumatic compression device. The intended therapy of the device is to reduce the risk of the formation of deep venous thrombosis (DVT) by aiding in blood flow back to the heart, to increase blood flow to various body areas, as determined by the physician, to aid in the reduction and control of edema and venous stasis ulcers.
Device Story
Recovery+ (PS511) and DVTherapy (PS511a) are portable, microprocessor-controlled intermittent pneumatic compression and thermal (hot/cold) therapy systems. Device uses inflatable anatomical cuffs/wraps applied to extremities. Pump delivers sequential pneumatic pressure to simulate muscle contraction/venous return and circulates temperature-controlled fluid for thermal therapy. Operated by clinicians or patients under physician prescription in clinical or home settings. System includes sensors for pressure/temperature monitoring, visual/audible alerts, and safety vent valves to prevent overpressure. Thermal unit requires 115 VAC; DVT unit operates via battery or AC. Output provides localized compression and thermal regulation to reduce edema, manage pain, and prevent DVT. Benefits include improved venous/lymphatic flow and accelerated wound healing.
Clinical Evidence
No clinical data. Substantial equivalence established via non-clinical bench testing, including electromagnetic compatibility (UL 60601-1, CSA C22.2 NO 601.1, EN 60601-1), mechanical integrity, environmental/life cycle testing, and verification of pressure delivery, cuff fill time, cycle time, and temperature management.
Technological Characteristics
Microprocessor-controlled pneumatic and thermal system. Materials: medical-grade non-latex, non-woven nylon/polyester cuffs; polyurethane tubing. Energy: 115 VAC or battery. Thermal range: 43ºF to 105ºF. Safety: pressure vent valve, high/low temperature sensors, malfunction indicators. Connectivity: standalone. Sterilization: N/A (single-patient use cuffs).
Indications for Use
Indicated for patients with restricted mobility or non-ambulatory status requiring DVT risk reduction, blood flow enhancement, edema/venous stasis ulcer management, or treatment of pain/swelling from musculoskeletal injury, surgery, or vascular/lymphatic insufficiency. Contraindicated in patients with CHF, active DVT/PE, phlebitis, pulmonary edema, arterial dysregulation, cardiac insufficiency, carcinoma in extremity, acute skin infection, Raynaud's, poor peripheral circulation, cold hypersensitivity, fractures, or sensory deficits in the limb.
Regulatory Classification
Identification
A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.
{1}------------------------------------------------
# 510(k) Summary
# K11748
#### Applicant [21 CFR 807.92 a(1)]:
Pulsar Scientific, LLC 8 Stony Brook Lane Ludlow, MA 01056 (413) 385-0005, ext. 202 - phone (413) 385-0003 - fax
Contact Person:
Joy Long (317) 496-8647 joyslong@gonowmail.com and Fernando Ubidia (413) 385-0005, ext 202 fernando@tandemtech.net
#### Device Name [21 CFR 807.92 a(2)]:
| Regulation No.: | CFR Part 870.5800 |
|------------------------------|---------------------------------------------|
| Classification: | Class II |
| Panel: | 870 Cardiovascular |
| Classification Name: | Compressible Limb Sleeve |
| Common/usual name:<br>Device | Intermittent External Pneumatic Compression |
| Product Code: | JOW |
Proprietary Name: Model #'s:
Recovery+ and DVTherapy PS511 and PS511a
#### Identification of Predicate Devices [21CFR 807.92 a(3)]:
- DVT Therapy - Doctor's Orders, Inc. DVTcare CA5 K061125
- . DVT Therapy - BioCompression Systems, Bioarterial Plus K072666
- . Thermal Therapy - ThermoTek, Inc. NanoTherm/VascuTherm K061866
#### Name Clarification
The device as a whole will be marketed as the Recovery+, Model PS511. However, the DVT therapy portion of the device is designed to be portable and used as a stand-alone device. Therefore, the therapy portion of the device is named DVTherapy and assigned Model PS511a. Throughout the submission, when the discussion involves the full device it will be referred to as the Recovery+. When referring to the thermal only portion of the device, it will be referred to as the Recovery+ thermal compression unit. When the
{2}------------------------------------------------
discussion involves the portable DVT therapy portion of the device, it will be referred to as DVTherapy.
## Device Description [21 CFR 807.92 a(4)]
#### Intended Use
The Pulsar Scientific device is not a life-supporting or life-sustaining device. It is a new, prescription use only [21 CFR 801 Subpart D] device for use with restricted mobility and non-ambulatory patients, or other patients as deemed appropriate by the physician. It is not intended for over-the-counter use [21 CFR 801 Subpart C]. The device is intended to function as an intermittent, external pneumatic compression device. The intended therapy of the device is to reduce the risk of the formation of deep venous thrombosis (DVT) by aiding in blood flow back to the heart, to increase blood flow to various body areas, as determined by the physician, to aid in the reduction and control of edema and venous stasis ulcers.
#### Physical Description
The Recovery+ device is comprised of an electronically controlled, compartmentalized reusable pump used with a variety of single-patient use inflatable anatomical cuffs/wraps. The unit is light-weight and portable. The Recovery+ thermal component with standard pressure component and the DVT component (DVTherapy) of the pump are housed in separate compartments so that they can be used together, separately, or removed for portable usage of the DV.Therapy component. For purposes of this application we will refer to the components as the Recovery+ thermal compression unit and the DVTherapy unit.
#### Therapy Modalities Used
The Recovery+ device is designed to provide compression therapy and/or thermal therapy (cold or hot). The therapies may be used together or separately, as ordered by the physician. The thermal therapy provides chilled or heated fluid to the cuff affixed on an extremity therapy site and compresses the treatment site for optimum performance and fluid transfer. The compression therapy is accomplished by air pumped through the cuffs, which are affixed on an extremity therapy site, to provide intermittent, sequential pressure. This facilitates venous flow towards the center of the body and simulates the muscle contraction and blood flow of ambulation.
#### Safety Features
The device utilizes microprocessor control with multiple sensors to ensure patient safety and system functionality, and to provide consistent and repeatable therapy modalities. Indicators are both visual on the unit display and audible. Settings are in place to detect
{3}------------------------------------------------
any potentially unsafe situation and to terminate therapy to protect the patient and the system. The unit is light weight and portable. The thermal compression runs only on 115 VAC while the DVT unit can operate on batteries, connected to 115 VAC mains, or while docked in the thermal compression unit. The thermal compression unit has internal power supply while the DVT unit has an external power supply. The unit is protectively housed in a plastic shell, except for the outer membrane switch, the plastic quick-connect locks for tub connection, and an external power supply input jack.
# Statement of Intended Use [21 CFR 807.92 a(5)]
The Recovery+ device is used on patient populations for which the Recovery+ device is applicable, based on prescription by a physician.
## Indications
Including, but not limited to:
- Decrease the risk of deep venous thrombosis (DVT); .
- . Aid the blood flow back to the heart;
- Treat and assist healing of cutaneous ulceration (wounds), reduce wound healing . time, enhance arterial circulation (blood flow), reduce compartmental pressures, reduce edema (swelling), reduce the need for anticoagulant (blood thinning) medications;
- . Treatment of pain and swelling of acute periarticular processes, treatment of pain and swelling following mobilization of shoulder stiffness under anesthesia. treatment of pain and swelling postoperatively for bones, joints and soft tissue, treatment of pain and swelling caused by musculoskeletal contusions and athletic injury:
- Treatment of disorders associated with vascular or lymphatic insufficiency such . as Chronic Venous Insufficiency (CVI), venous stasis ulcers, post-mastectomy edema and chronic edema:
- Reduction of edema associated with soft tissue injuries such as ligament sprains, . postoperative edema, and burns.
### Contraindications
As with all predicate devices, the device should not be used in the following instances:
- . Presumptive evidence of congestive heart failure (CHF)
- 트 Suspected/observed pre-existing deep vein thrombosis (DVT) or pulmonary embolism
- Suspected/observed deep acute venal thrombosis (phlebothrombosis)
- l Suspected/observed inflammatory phlebitis process
- l Suspected/observed pulmonary edema
- Suspected/observed acute inflammations of the veins (thrombophlebitis)
- 트 Suspected/observed decompensated cardiac insufficiency
- Suspected/observed arterial dysregulation
- l Suspected/observed erysipelas
{4}------------------------------------------------
- 1 Suspected/observed carcinoma and carcinoma metastasis in the affected extremity
- 다 Suspected/observed decompensated hypertonia
- 포 Suspected/observed acute inflammatory skin diseases or infection
- I Suspected/observed venous or arterial occlusive disease
- I Determine venous and lymphatic return is undesirable
- Suspected/observed patient has Raynaud's Disease
- 미 Suspected/observed poor peripheral circulation
- I Suspected/observed hypersensitivity to cold
- l Patient therapy contact on extremity containing a fracture
- 트 Extremities that are not sensitive to pain
It is also not recommended that the Recovery+ device be used on a limb where the cuffs would interfere with open wounds, vein ligation, gangrene, dermatitis, recent skin graft, massive edema, or extreme deformity of the leg. They should not be used where increased venous or lymphatic return is undesirable.
い、
#### Differences in Indications
The indications for the Pulsar Scientific Recovery+ device are generally the same as those published for the predicate devices.
## Technological Characteristics [21 CFR 807.92a(6)1]
The technological characteristics of this Recovery+ device are the same in overall design function, materials, energy source, and mode of operation as the predicate units. The material in contact with the patient's skin is medical grade, non-latex, non-woven nylon and polyester, designed for patient comfort and convenience. The inflation deflation tubes are polyurethane.
The Pulsar Scientific Recovery+ device has the same performance characteristics as the predicate devices. The pneumatic control circuitry is a microprocessor-controlled system. Multiple safety redundancies are built into the system including: high and low temperature indicators, indicators for unit malfunctions, and system malfunction overpressure safety via a pressure vent valve. Power is supplied via 115 VAC line current, or battery. The surface contact temperature range is microprocessor controlled in cooling from 43ºF to 49ºF and to 105ºF in heating mode.
#### Testing Conducted
Non-clinical validation, including electromagnetic compatibility (emissions, immunity and safety), mechanical integrity, environmental and life cycle testing, have shown that the Recovery+ has performance characteristics substantially equivalent to or superseding the listed predicate devices. The Recovery+ device has been validated by Pulsar Scientific at a design validation level based on requirement of UL 60601-1, CSA C22.2 NO 601.1 and CENELEC EN 60601-1 and confirmed by an accredited test lab.
{5}------------------------------------------------
Additional bench testing has verified equivalent pressure delivery, cuff fill time, cycle time, temperature management and system operation as the predicate devices listed.
Clinical testing is not appropriate to this device and was therefore not done.
Per the requirements of 21 CR 807, non-clinical validation testing and the information provided in this submission, Pulsar Scientific concludes that the Recovery+ system is safe, effective, and performs in a manner that is substantially equivalent to the predicate devices listed in this submission.
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle or bird symbol, with three overlapping wing-like shapes, representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Pulsar Scientific c/o Ms. Joy Long 8 Stony Brook Lane Ludlow, MA 01056
# NOV 2 3 2011
Re: K111748
Trade/Device Name: Recovery+/DVTherapy (Models PS511 and PS511a) Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible limb sleeve Regulatory Class: II Product Code: JOW Dated: November 20, 2011 Received: November 23, 2011
Dear Ms. Long:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{7}------------------------------------------------
Page 2 - Ms. Joy Long
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
M.A. Hielm
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use
Pulsar Scientific - 510(k) Number: K111748
Device Name: Recovery+
Indications For Use:
Compression therapy - Including, but not limited to:
- Decrease the risk of deep venous thrombosis (DVT); .
- Aid the blood flow back to the heart: .
- Treat and assist healing of cutaneous ulceration (wounds). reduce wound healing time, . enhance arterial circulation (blood flow), reduce compartmental pressures, reduce edema (swelling), reduce the need for anticoagulant (blood thinning) medications;
- Treatment of pain and swelling of acute periarticular processes, treatment of pain and . swelling following mobilization of shoulder stiffness under anesthesia, treatment of pain · and swelling postoperatively for bones, joints and soft tissue, treatment of pain and
- swelling caused by musculoskeletal contusions and athletic injury;
- Treatment of disorders associate with vascular or lymphatic insufficiency such as . Chronic Venous Insufficiency (CVI), venous stasis ulcers, post-mastectomy edema and chronic edema;
- Reduction of edema associated with soft tissue injuries such as ligament sprains, . postoperative edema, and burns.
Thermal therapy - Including but not limited to:
- Reduction of general edema (swelling); ●
- Reduction of edema associated with soft tissue injuries such as ligament sprains ● postoperative edema, and burns;
- Treatment of pain and swelling of acute periarticular processes, treatment of pain and . swelling following mobilization of shoulder stiffness under anesthesia, treatment of pain and swelling postoperatively for bones, joints and soft tissue, treatment of pain and swelling caused by musculoskeletal contusions and athletic injury.
Prescription Use × AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K111748
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.