← Product Code [DWJ](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/DWJ) · K232638

# FilteredFlo Warming Blankets (K232638)

_Iob Medical, Inc. · DWJ · Oct 24, 2023 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/DWJ/K232638

## Device Facts

- **Applicant:** Iob Medical, Inc.
- **Product Code:** [DWJ](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/DWJ.md)
- **Decision Date:** Oct 24, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5900
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic, Pediatric, 3rd-Party Reviewed

## Indications for Use

The FilteredFlo Warming Blanket is intended to be used with the IOB-507 patient warming system. The IOB-507 Temperature Management system is indicated for hypothermic patients or normothermic patients for whom induced hyperthermia or localized increase in temperature is clinically indicated. These blankets are intended for adult and pediatric use.

## Device Story

FilteredFlo Warming Blankets are single-use, disposable accessories for the IOB-507 Temperature Management System. The system draws ambient air through a 0.2-micron particulate filter, warms it to a user-selected temperature, and delivers it via hose to the blanket. Blankets consist of nonwoven polypropylene fabric coated with polyethylene, forming internal air distribution channels. Perforations on the patient-facing side disperse warmed air over or around the patient. Used in clinical settings to manage patient body temperature; operated by healthcare providers. Output is convective heat transfer to the patient, aiding in the treatment of hypothermia or facilitating induced hyperthermia. Benefits include maintenance of normothermia or therapeutic temperature regulation.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including temperature uniformity, simulated transport testing (ASTM D4169), and bubble leak testing (ASTM F2096). Biocompatibility data was leveraged from the predicate device (K231596).

## Technological Characteristics

Materials: Two layers of nonwoven polypropylene fabric bonded to a polyethylene layer. Principle: Convective air warming via internal distribution channels and perforations. Dimensions: Various sizes (e.g., 235x125 cm to 100x80 cm). Connectivity: Connects to IOB-507 system via hose. Sterilization: Non-sterile.

## Regulatory Identification

A thermal regulating system is an external system consisting of a device that is placed in contact with the patient and a temperature controller for the device. The system is used to regulate patient temperature.

## Predicate Devices

- IOB Temperature Management System Model IOB-507 ([K231596](/device/K231596.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 24, 2023

IOB Medical Inc % Dave Yungvirt CEO Third Party Review Group, LLC 25 Independence Blvd Warren, New Jersey 07059

Re: K232638

Trade/Device Name: FilteredFlo Warming Blankets Regulation Number: 21 CFR 870.5900 Regulation Name: Thermal regulating system Regulatory Class: Class II Product Code: DWJ Dated: October 10, 2023 Received: October 11, 2023

Dear Dave Yungvirt:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Nicole M. Gillette -S

Nicole Gillette Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K232638

Device Name FilteredFlo Warming Blankets

Indications for Use (Describe)

The FilteredFlo Warming Blanket is intended to be used with the IOB-507 patient warming system. The IOB-507 Temperature Management system is indicated for hypothermic patients for whom induced hyperthermia or localized increase in temperature is clinically indicated. These blankets are intended for adult and pediatric use.

| Type of Use (Select one or both, as applicable)            |
|------------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) |
| <span></span> Over-The-Counter Use (21 CFR 801 Subpart C)  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(k) Summary

## K232638

#### I. SUBMITTER

IOB Medical Inc. 12011 Mosteller Rd. Cincinnati, OH 45241 IOB Medical is a part of Gentherm Medical.

Tel: 1(301)250-0831 Fax: 1(301)916-6213

Contact Person: Joe Shia Email: shiajl@yahoo.com

Date: October 21, 2023

#### II. DEVICE

Name of Device: FilteredFlo Warming Blankets Common or Usual Name: Warming Blanket Classification Name: Thermal Regulating System (21 CFR 870.5900) Regulatory Class: II Product Code: DWJ

#### III. PREDICATE DEVICE

K231596 IOB Temperature Management System Model IOB-507

#### IV. DEVICE DESCRIPTION

The FilteredFlo warming blankets are used with the IOB Temperature Management System (previously cleared k231596) that draws ambient- temperature air through a 0.2 micron particulate air filter. The filtered air is warmed to a selected temperature. The warmed air enters the FilteredFlo Warming Blanket through the hose and is distributed through delivery channels. Perforations on the patient side of the air delivery channels in the warming blanket gently disperse the warmed air over and around the patient.

The FilteredFlo warming blankets in the submission are the following:

{4}------------------------------------------------

FF-243 ADULT/PACU BLANKET FF-244 PEDIATRIC BLANKET FF-246 INFANT UNDERBODY BLANKET FF-247 LARGE PEDIATRIC UNDERBODY BLANKET FF-248 ADULT UNDERBODY BLANKET FF-443 UPPER BODY BLANKET FF-442 LOWER BODY BLANKET FF-344 TORSO BLANKET FF-145 WARMING TUBE BLANKET

These blankets are single-use and disposable. Each blanket consists of two layers of nonwoven polypropylene fabric coated with a layer of polyethylene. The layers are bonded together to form a distribution network of air delivery channels. The warm air is distributed around the blanket through the delivery channels and exits the blanket through a specially designed series of perforations in the patient side of the blanket.

#### V. INTENDED FOR USE

The FilteredFlo Warming Blanket is intended to be used with the IOB-507 patient warming system. The IOB-507 Temperature Management system is indicated for hypothermic patients or normothermic patients for whom induced hyperthermia or localized increase in temperature is clinically indicated. These blankets are intended for adult and pediatric use.

## VI. SUBSTANTIAL EQUIVALENCE INFORMATION

A summary comparison of features of the FilteredFlo Warming Blankets and the predicate device is provided in following Table 1.

| Parameters             | Predicate Device K231596                                                                                                                   | Proposed Device                                                                                                                                       |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
|                        | IOB Warming Blankets (models<br>IOB-004, IOB-007, IOB-028, IOB-<br>009, IOB-006, IOB-003, IOB-002,<br>IOB-001, IOB-015)                    | FilteredFlo Warming Blankets (models<br>FF-243, FF-244, FF-246, FF-247, FF-<br>248, FF-443, FF-442, FF-344, FF-145)                                   |
| Indications For<br>Use | IOB Temperature Management<br>system is indicated for hypothermic<br>patients or normothermic patients<br>for whom induced hyperthermia or | The FilteredFlo Warming Blanket is<br>intended to be used with the IOB-507<br>patient warming system. The IOB-507<br>Temperature Management system is |

Table 1: Comparison between the FilteredFlo Warming Blankets and the predicate device

{5}------------------------------------------------

|                       | localized increase in temperature is<br>clinically indicated.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |              | indicated for hypothermic patients or<br>normothermic patients for whom induced<br>hyperthermia or localized increase in<br>temperature is clinically indicated.<br>These blankets are intended for adult and<br>pediatric use. |              |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Material Design       | Consists of two layers of non-<br>woven polypropylene fabric<br>bonded to a fusion layer of<br>polyethylene.<br>The layers are bonded together to<br>form a distribution network of air<br>delivery channels.<br>The warm air is distributed around<br>the patient's body through the<br>delivery channels and exits the<br>blanket through a specially<br>designed series of perforations in<br>the patient side layer of the blanket.<br>The distribution of air is designed to<br>minimize temperature differences of<br>delivered air at different blanket<br>locations. |              | SAME                                                                                                                                                                                                                            |              |
| Sterility             | Sterile and Non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |              | Non-sterile                                                                                                                                                                                                                     |              |
|                       | IOB-004                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 210 x 120 cm | FF-243                                                                                                                                                                                                                          | 235 x 125 cm |
|                       | IOB-007                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 170 x 100 cm | FF-244                                                                                                                                                                                                                          | 142 x 100 cm |
|                       | IOB-028                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 100 x 100 cm | FF-246                                                                                                                                                                                                                          | 100 x 80 cm  |
|                       | IOB-009                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 160 x 80 cm  | FF-247                                                                                                                                                                                                                          | 160 x 80 cm  |
| Blanket<br>Dimensions | IOB-006                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 215 x 100 cm | FF-248                                                                                                                                                                                                                          | 215 x 100 cm |
|                       | IOB-003                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 202 x 64 cm  | FF-443                                                                                                                                                                                                                          | 220 x 80 cm  |
|                       | IOB-002                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 142 x 120 cm | FF-442                                                                                                                                                                                                                          | 135 x 100 cm |
|                       | IOB-001                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 142 x 120 cm | FF-344                                                                                                                                                                                                                          | 110 x 100 cm |
|                       | IOB-015                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 180 x 74 cm  | FF-145                                                                                                                                                                                                                          | 180 x 74 cm  |

# VII. SAFETY AND PERFORMANCE CHARACTERISTICS

#### 1. Nonclinical Tests

- a. Temperature uniformity tests were performed by measuring five testing points on blanket surface at different IOB Warmer settings. All test results show temperature uniformity equivalence between the FilteredFlo Warming Blankets and the predicate.

{6}------------------------------------------------

- b. Simulated transport testing was performed according to ASTM D4169. No package damage was observed. All product hold integrity after the transport testing.
- d. Bubble testing was carried out according to the ASTM F2096. No leakage was found.
- e. Biocompatibility tests were reported in the previously cleared k231596.
- 2. Clinical Studies

Not applicable.

## VIII. CONCLUSION

Based on the information presented in this 510K premarket notification including nonclinical tests of temperature uniformity tests, transport testing, bubble testing and biocompatibility testing, the FilteredFlo Warming Blankets are substantially equivalent to the predicates.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/DWJ/K232638](https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/DWJ/K232638)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
