← Product Code [DWJ](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/DWJ) · K162679

# IOB Temperature Management System (K162679)

_Iob Medical, Inc. · DWJ · Apr 28, 2017 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/DWJ/K162679

## Device Facts

- **Applicant:** Iob Medical, Inc.
- **Product Code:** [DWJ](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/DWJ.md)
- **Decision Date:** Apr 28, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5900
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The IOB Temperature Management system is indicated for hypothermic patients or normothermic patients for whom induced hyperthermia or localized increase in temperature is clinically indicated.

## Device Story

IOB Temperature Management System consists of warming unit and disposable blankets. Unit draws ambient air through 0.2-micron filter, warms air to selected temperature (32°C, 38°C, 43°C, or ambient), and delivers it via hose to blanket. Blankets feature distribution channels and perforations to disperse air over patient. Servo-controlled by thermistor at hose-blanket connection; control thermistor in unit adjusts heater power. Used in clinical settings to manage patient temperature. Healthcare providers select temperature settings on control panel. System maintains temperature despite ambient variations. Benefits include thermal regulation for patients undergoing surgery or requiring temperature therapy.

## Clinical Evidence

No clinical data. Bench testing only, including temperature uniformity, accelerated stability (3-year shelf life), electrical safety/EMC, software validation, and biocompatibility (ISO 10993).

## Technological Characteristics

Materials: non-woven polypropylene fabric coated with polyethylene. Sensing: thermistor-based servo-control. Energy: 110-120V/230V AC electric heater. Connectivity: standalone. Sterilization: non-sterile (select models sterile). Software: embedded control logic for temperature regulation and safety monitoring.

## Regulatory Identification

A thermal regulating system is an external system consisting of a device that is placed in contact with the patient and a temperature controller for the device. The system is used to regulate patient temperature.

## Predicate Devices

- Bair Hugger Temperature Management System ([K021473](/device/K021473.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of three human profiles facing right, with flowing lines suggesting movement or connection.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 28, 2017

IOB Medical, Inc. % Joe Shia Regulatory Consultant LSI International, Inc. 504 East Diamond Ave., Suite F Gaithersburg, Maryland 20877

Re: K162679

Trade/Device Name: IOB Temperature Management System Regulation Number: 21 CFR 870.5900 Regulation Name: Thermal Regulating System Regulatory Class: Class II Product Code: DWJ Dated: March 10, 2017 Received: March 20, 2017

Dear Joe Shia:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Fernando
Aguel-S

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K162679

#### Device Name

IOB Temperature Management System

Indications for Use (Describe)

The IOB Temperature Management system is indicated for hypothermic patients or normothermic for whom induced hyperthermia or localized increase in temperature is clinically indicated.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) SUMMARY

| 1. Date:           | April 28, 2017                                                                                                                                                            |
|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. Submitter:      | IOB Medical Inc<br>504E Diamond Ave., Suite I<br>Gaithersburg, MD 20877                                                                                                   |
| 3. Contact person: | Joe Shia<br>LSI International Inc.<br>504 East Diamond Ave., Suite F<br>Gaithersburg, MD 20877<br>Telephone: 240-505-7880<br>Fax: 301-916-6213<br>Email: shiajl@yahoo.com |

- 4. Device Name: IOB Temperature Management System
- 5. Classification: Class:

Class II

| Product Code | CFR #    | Product Name              |
|--------------|----------|---------------------------|
| DWJ          | 870.5900 | Thermal Regulating System |

- 6. Predicate Devices: K021473 Bair Hugger Temperature Management System
- 7. Intended Use

The IOB Temperature Management system is indicated for hypothermic patients or normothermic patients for whom induced hyperthermia or localized increase in temperature is clinically indicated.

- 8. Device Description
The IOB Temperature Management System consists of the IOB warming unit and the IOB warming The IOB warming unit draws ambient- temperature air through a 0.2 micron air filter. The blankets. filtered air is warmed to a selected temperature. The warmed air enters the IOB Warming Blanket through the hose and is distributed through delivery channels. Perforations on the patient side of the air delivery channels in the warming blanket gently disperse the warmed air over and around the patient. The warming unit has three temperature settings of 32°C, 38°C, and 43°C. These temperature settings are servo-controlled by a thermistor placed at the end of hose where the hose connects to the blanket. The unit can also delivers ambient-temperature air. The temperature indicated on the control panel is the temperature of air being delivered to the blanket at the end of the hose. A control thermistor in the warming unit adjusts the power applied to the heater to maintain the selected temperature. This enables the system to maintain the selected temperature under variations in ambient temperature. Besides, the warming unit has high and low air flow for choose.

The IOB Warming Blankets in this submission are the following:

{4}------------------------------------------------

Torso IOB-001 Lower Body IOB-002 Upper Body IOB-003 Full Body IOB-004 Pediatric Underbody IOB-005 Adult Underbody IOB-006 Pediatric Full Body IOB-007 Full Body Surgical IOB-008 Large Pediatric Underbody IOB-009 Spinal Underbody IOB-010 Lithotomy Underbody IOB-011 Pediatric Lower Body IOB-012 Pediatric Long IOB-014 Cath Lab IOB-015 Surgical Access IOB-016 Chest Access IOB-017 Multi-Access IOB-018 Dual Port Torso IOB-019 Cardiac Access IOB-020 XL Upper Body IOB-021 Outpatient Care IOB-022 Cardiac IOB-023

These blankets are single-use and disposable. Each blanket consists of two lavers of non-woven polypropylene fabric coated with a layer of polyethylene. The lavers are bonded together to form a distribution network of air delivery channels. The warm air is distributed around the blanket through the delivery channels and exits the blanket through a specially designed series of perforations in the patient side of the blanket.

- 9. Substantial Equivalence Information
A summary comparison of features of the IOB Temperature Management system and the predicate devices is provided in following Tables.

| Parameters | Predicate devices K021473                                                                                                        | Proposed devices                                                                                                                                                                                                       |
|------------|----------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|            | Bair Hugger blankets (models<br>540,525,522,300,555,635,310,610,550,575,<br>585,537,530,560,570,305,315,542,645,523,<br>110,630) | IOB Warming Blankets<br>(IOB-001,IOB-002,IOB-003,IOB-004,IOB-005,<br>IOB-006,IOB-007,IOB-008,IOB-009,IOB-010<br>IOB-011, IOB-012,IOB-014,IOB-15,IOB-16<br>IOB-17, IOB-18, IOB-019,IOB-020,IOB-021,<br>IOB-022,IOB-023) |

|  |  | Table 1: Comparison between IOB Warming Blankets and predicate devices. |  |  |
|--|--|-------------------------------------------------------------------------|--|--|
|  |  |                                                                         |  |  |

{5}------------------------------------------------

| INDICATIONS<br>FOR USE                 | The Bair Hugger temperature management<br>system is indicated for hyper- or hypothermic<br>patients or normothermic patients for whom<br>induced hyper- or hypothermia or localized<br>temperature therapy is clinically indicated. In<br>addition, the Bair Hugger temperature<br>management system can be used to provide<br>patient thermal comfort when conditions exist<br>that may cause patients to become too warm or<br>too cold. The Bair Hugger temperature<br>management system can be used with adult and<br>pediatric patients.                   | The IOB Temperature Management system is<br>indicated for hypothermic patients or<br>normothermic patients for whom induced<br>hyperthermia or localized increase in<br>temperature is clinically indicated.                                                                                                                                                                                                                                                       |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| MATERIAL<br>DESIGN                     | Consists of two layers of non- woven<br>polypropylene fabric bonded to a fusion layer of<br>polyethylene.<br>The layers are bonded together to form a<br>distribution network of air delivery channels<br>The warm air is distributed around the patient's<br>body through the delivery channels and exits the<br>blanket through a specially designed series of<br>perforations in the patient side layer of the<br>blanket.<br>The distribution of air is designed to minimize<br>temperature differences of delivered air at<br>different blanket locations. | SAME                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Shelf Life                             | 3 years                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | SAME                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Sterility                              | Non-sterile except 645 and 630                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Non-sterile except IOB-14, IOB-16, IOB-020 and<br>IOB-023                                                                                                                                                                                                                                                                                                                                                                                                          |
| Blanket<br>Dimensions<br>(approximate) | 540 107×91cm<br>525 152×91cm<br>522 188×61cm<br>300 213×91cm<br>635 221×91cm<br>610 183×91cm<br>575 188×91cm<br>585 188×91cm<br>560 23×335cm<br>570 213×91cm<br>305 183×91cm<br>315 213×91cm<br>542 107×91cm<br>523 213×91cm<br>110 213×91cm<br>310 152×91cm<br>555 91×84cm<br>550 152×81cm<br>537 89×61cm<br>530 188×21cm<br>645 23x274cm<br>630 152x91cm                                                                                                                                                                                                      | IOB-001 142×120cm<br>IOB-002 142×120cm<br>IOB-003 195×80cm<br>IOB-004 210×120cm<br>IOB-006 215×100cm<br>IOB-008 210×120cm<br>IOB-010 215×100cm<br>IOB-011 170×100cm<br>IOB-015 17x180cm<br>IOB-016 210×120cm<br>IOB-017 180×120cm<br>IOB-018 210×120cm<br>IOB-019 142×120cm<br>IOB-021 215×80cm<br>IOB-022 210×120cm<br>IOB-007 170×100cm<br>IOB-005 91×80cm<br>IOB-009 160×80cm<br>IOB-012 142×100cm<br>IOB-014 110×17cm<br>IOB-020 17x150cm<br>IOB-023 142x120cm |
| Air Velocity                           | 28-30cfm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Air<br>Temperature                     | Hose End Temperature<br>43°C +/- 3.0°C<br>38°C +/- 3.0°C<br>32°C +/- 3.0°C<br>Ambient                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Hose End Temperature<br>43°C +/- 3.0°C<br>38°C +/- 3.0°C<br>32°C +/- 3.0°C<br>Ambient                                                                                                                                                                                                                                                                                                                                                                              |
| System<br>Power                        | 110-120VAC,60Hz,10A<br>220-240VAC,50Hz,4.5A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 110-120V, 60 Hz, 12A<br>230V, 50/60 Hz, 8A                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Heater<br>Power                        | 850 Watts                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 1000 W                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Dimensions                             | 33X 25 X 28cm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 28 X 22X 22cm                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Weight                                 | 5.2kg                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 4.5kg                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| EMI/EMC<br>Compliant                   | IEC60601-1, IEC 60601-1-2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | IEC60601-1, IEC 60601-1-2                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Forced air<br>Over<br>Temperature      | Auto-shuts heater off at 47°C +/- 2°C                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Auto-shuts heater off at 47°C +/- 2°C.                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Hose with<br>Secure<br>Locking         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Air Filter                             | Replaceable 0.2 micron                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Replaceable 0.2 micron                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Temperature<br>Display                 | Front panel LCD display                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Front panel digital display                                                                                                                                                                                                                                                                                                                                                                                                                                        |

{6}------------------------------------------------

{7}------------------------------------------------

10. Safety and Performance Characteristics

- Temperature uniformity tests show equivalence between the IOB warming blankets and the 1. predicate.
- 2. Accelerated stability tests show three years shelf-life of the IOB warming blankets.
- 3. Electrical safety/EMC testing, and software validation show that the device meets its design specifications, performs as intended.
- 4. Biocompatibility testing (cytotoxicity, irritation and sensitivity) according to ISO 10993 for a limited contact device was demonstrated to be suitable for the intended use of the product.
- 5. Clinical Studies Not applicable

## 11.Conclusion

Based on the information presented in this 510K premarket notification, the IOB Temperature Management System is substantially equivalent to the predicate.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/DWJ/K162679](https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/DWJ/K162679)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
