← Product Code [DWJ](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/DWJ) · K123083

# COVIDIEN WARMTOUCH CONVECTIVE WARMING UNIT (K123083)

_Covidien · DWJ · Apr 17, 2013 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/DWJ/K123083

## Device Facts

- **Applicant:** Covidien
- **Product Code:** [DWJ](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/DWJ.md)
- **Decision Date:** Apr 17, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5900
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The WarmTouch™ Convective Warming Unit is a portable device that is used in conjunction with a WarmTouch™ Blanket as a system that is intended for prescription use. The warming unit requlated the management of normothermia from pre-operative to post-operative, which aligns with the Intended Use of the currently cleared (predicate) device. There are no clinical implications, no new diagnostic or therapeutic information that is not normally associated with other general uses of the device, nor are there any additional outcomes that could influence patient management.

## Device Story

Electro-mechanical convective warming unit; delivers heated air through flexible hose to specialized blankets draped over patient. Used in clinical settings (pre-operative to post-operative) to prevent/treat hypothermia and maintain normothermia. Operated by healthcare professionals. System includes blower unit and blankets; heated air distributed via blanket perforations. Modifications from predicate include updated GUI, blower design for IEC 80601-2-35:2009 compliance, and RoHS-compliant PCBA. Output is controlled thermal air flow; assists in patient temperature management; improves clinical outcomes by maintaining normothermia.

## Clinical Evidence

Bench testing only. Design verification and validation confirmed mechanical and software features met requirements. Usability evaluated via formative and summative studies in simulated operating environments to ensure safety and performance.

## Technological Characteristics

Electro-mechanical convective warming unit. Features include GUI, blower, and PCBA. Complies with IEC 80601-2-35:2009 and RoHS Directive 2011/65/EU. Employs temperature sensors for thermal regulation. Standalone device.

## Regulatory Identification

A thermal regulating system is an external system consisting of a device that is placed in contact with the patient and a temperature controller for the device. The system is used to regulate patient temperature.

## Predicate Devices

- WarmTouch™ Model 5300A Patient Warming System ([K020604](/device/K020604.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K123083

Image /page/0/Picture/1 description: The image contains the word COVIDIEN in bold, sans-serif font. To the left of the word is a black and white symbol that looks like a plus sign. The symbol is made up of a white plus sign on a black background.

# APR 1 7 2013

# 510(k) Summary

Date Summary Prepared: 28-September-2012 Date Summary Revised: 20-November-2012

### 510(k) Submitter / Holder

Covidien IIc 6135 Gunbarrel Avenue Boulder, CO 80301

### Contact

Name: Stacev A. Strand Title: Senior Regulatory Affairs Specialist Office Tel: 303.305.2631 720.539.9617 Mobile Tel: 303.305.2212 Fax: Email: Stacey. Strand@Covidien.com

# Name of Device

WarmTouch™ Convective Warming Unit Trade Name: Cataloq Number: 5016000 Common Name: Thermal Requiating System Cardiovascular (21 CFR § 870.5900, Class II, DWJ) Classification Name:

# Purpose of Submission

The purpose of this submission is to introduce the Covidien WarmTouch™ Convective Warming Unit which effectively provides a means for treating or preventing hypothermia in adult and pediatric patients in clinical settings. The design and intended use of the WarmTouch™ Convective Warming Unit Warming Unit is fundamentally similar to the predicate device. The new WarmTouch™ Convective Warming Unit includes modifications of the: hardware interface to a Graphic User Interface (GUI), blower's design for compliance with the IEC 80601-2-35:2009 (200 edition) standard, Printed Circuit Board Assembly (PCBA) to comply with the European Restriction of Hazardous Substances (RoHS) Directive 2011/65/EU.

{1}------------------------------------------------

This submission followed the "Guidance for Industry and FDA Staff: Format for Traditional and Abbreviated 510(k)s, 2005," the "Guidance for Industry on General / Specific Intended Use, 1998, " and the "Draft Guidance for Industry and FDA Staff. Factors to Consider when Making Benefit-Risk Determinations in Medical Device Premarket Review, 2011."

### Predicate Device

The WarmTouch™ Convective Warming Unit Warming Unit was compared and found to be substantially equivalent to the following product of comparable type in commercial distribution:

| Trade Name:         | WarmTouch™ Model 5300A Patient Warming System |
|---------------------|-----------------------------------------------|
| Device Common Name: | Thermal Regulating System                     |
| 510(k) Number:      | K020604 (cleared April 2002)                  |
| Manufacturer:       | Covidien, formerly Nellcor Puritan Bennett,   |
|                     | a division of Tyco Healthcare                 |

This WarmTouch™ Convective Warming Unit Warming Unit 510(k) submission is the only initial premarket notification for this new blower unit. There have not been any prior 510(k) premarket notification submissions, regarding this new blower unit, withdrawn or deleted.

#### Device Description

The WarmTouch™ Convective Warming Unit is an electro-mechanical blower that delivers heated air. The WarmTouch™ Convective Warming Unit is part of the WarmTouch™ Convective Warming System that consists of the blower warming unit and Covidien's WarmTouch™ Blankets. The blower actively delivers heated air from the blower's flexible hose to a lightweight blanket. During this process, the blanket has already been draped over the patient's body, before the blanket begins distributing the heated air through numerous small perforations in the blanket, which allows the air to reach the targeted areas of the patient's body.

# Indications for Use

.

The WarmTouch Convective Warming System (warming unit and blanket) is intended for prevention and treatment of hypothermia, and for the management of appropriate normothermia.

{2}------------------------------------------------

# Intended Use

The WarmTouch™ Convective Warming Unit is a portable device that is used in conjunction with a WarmTouch™ Blanket as a system that is intended for prescription use. The warming unit requlated the management of normothermia from pre-operative to post-operative, which aligns with the Intended Use of the currently cleared (predicate) device. There are no clinical implications, no new diagnostic or therapeutic information that is not normally associated with other general uses of the device, nor are there any additional outcomes that could influence patient management.

# Technological and Performance Characteristics

The design technology of the proposed and predicate blower units have the same method of operation, power control, temperature sensor, materials, manufacturing method, and accessories. The main design technology differences between the proposed and predicate blower units are the control mechanism, motor, and user interface.

#### Performance Data

The design of the WarmTouch™ Convective Warming Unit features the same ambient blower operating range and air flow rate as the predicate device. The proposed blower unit alarms have improved accuracy, along with the improved performance of air temperature exiting the blower.

Design verification and validation testing was performed to confirm the blower unit's mechanical and software features met specified requirements. All verification and validation activities met product requirements.

Animal testing was not required to demonstrate that the proposed device met its design requirements.

#### Usability / Human Factors

Usability was evaluated with users in simulated operating environments. These studies consisted of formative and summative studies, which demonstrate the device provides adequate assurance of safety and performance (in regards to human factors/usability aspects) for the patient and operator.

#### Substantial Equivalence

In establishing substantial equivalence of the WarmTouch™ Convective Warming Unit with the predicate device, Covidien evaluated the intended use, indications for use, and technological characteristics. The use of the WarmTouch™ Convective Warming Unit in patient monitoring environments does not raise any new types of questions of safety and effectiveness compared with the predicate device that is currently in use.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an eagle or bird in flight, composed of three curved lines.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 17, 2013

Covidien c/o Ms. Stacey Strand 6135 Gunbarrel Avenue Boulder, CO 80301

Re: K123083

Trade/Device Name: WarmTouch Convective Warming Unit Regulation Number: 21 CFR 870.5900 Regulation Name: Thermal Regulating System Regulatory Class: Class II Product Code: DWJ Dated: February 15, 2013 Received: February 19, 2013

Dear Ms. Strand:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to '

http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Matthew G. Hillebrenner

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use Statement

K123083 510(k) Number (if known):

# Device Name:

Covidien WarmTouch™ Convective Warming System

# Indications for Use:

The WarmTouch Convective Warming System (warming unit and blanket) is intended for prevention and treatment of hypothermia, and for the management of appropriate normothermia.

# Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)

# CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)

# Matthew G퀘咖llebrenner

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/DWJ/K123083](https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/DWJ/K123083)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
