← Product Code [DWJ](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/DWJ) · K093244

# UNIQUE TEMP WARMING BLANKETS, MODEL 8001140XXX (K093244)

_Geratherm Medical AG · DWJ · Oct 19, 2010 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/DWJ/K093244

## Device Facts

- **Applicant:** Geratherm Medical AG
- **Product Code:** [DWJ](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/DWJ.md)
- **Decision Date:** Oct 19, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5900
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

Geratherm UniqueTemp® is a thermal regulating system for adult patients in order to prevent hypothermia in the operating theater. It is suitable for pre-, intra- and post-operative use. The device is not life-supporting. It is intended for use by appropriately trained healthcare professionals in clinical environments.

## Device Story

Geratherm UniqueTemp® is a thermal regulating system designed to prevent hypothermia in adult patients within clinical operating theaters. The system comprises a control unit with a 3.5" TFT touch screen, connecting cables, and various warming blankets (arm/shoulder, torso, leg) with reusable, washable, steam-permeable covers. The device operates by providing active warming to the patient; healthcare professionals select temperatures between 37°C and 42°C in 0.1°C increments via the control unit. The blankets are flexible to accommodate complex patient positioning and can be covered with surgical drapes. The system is not life-supporting and is intended for use by trained medical professionals. By maintaining patient normothermia during lengthy operations, the device aims to improve patient outcomes. The control unit is constructed from anodized aluminum with a glass screen, while blankets utilize PU-coated materials and silicone cabling.

## Clinical Evidence

Bench testing only. The device underwent testing for biocompatibility, electromagnetic compatibility (IEC 60601-1-2), electrical safety (IEC 60601-1), and performance requirements for heating devices (IEC 80601-2-35). No clinical trial data was presented.

## Technological Characteristics

Thermal regulating system; anodized aluminum control unit with 3.5" TFT touch screen; PU-coated blankets; silicone cabling. Temperature range: 37°C to 42°C (0.1°C increments). Standards met: IEC 60601-1, IEC 60601-1-2, IEC 80601-2-35. Reusable covers are steam-permeable and washable up to 95°C.

## Regulatory Identification

A thermal regulating system is an external system consisting of a device that is placed in contact with the patient and a temperature controller for the device. The system is used to regulate patient temperature.

## Predicate Devices

- Hot Dog Patient Warming System ([K052392](/device/K052392.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

**Geratherm**
Medical Diagnostic Systems

K093244

## 510(k) Summary as required by section 807.92(c)

10/19/2010

OCT 1 9 2010

.

Submission Applicant: Geratherm Medical AG Thomas Robst Fahrenheitstraße 1 98716 Geschwenda / Germany Establishment Registration Number: 3003591690

Phone: +49 36205 98-0 Fax: +49 36205 98-116 E-mail: t.robst@geratherm.com

Application Correspondent/Contact: think! Andrea Pecsi Schwarzwaldstraße 5 78532 Tuttlingen Germany

Phone: +49-7462-924 051 Fax: +49-7462-924 128 E-mail: andrea@thinkworks.biz

Trade Name: Geratherm UniqueTemp®

Common Name: Warming systems / blankets

Classification Name: Thermal regulating systems (21 CFR 870.5900, Product code DWJ)

Substantial Equivalence Warming System: 510(k)-Number: K052392 - Hot Dog Patient Warming System Firm: AUGUSTINE BIOMEDICAL & DESIGN, LLC.

# Description of the Device:

Geratherm UniqueTemp® has been developed for the prevention of hypothermia. The Geratherm warming blankets can be used for covering the patient in various ways without restricting the

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for "Geratherm Medical Diagnostic Systems". The word "Geratherm" is in a bold, sans-serif font, with a registered trademark symbol to the right of the word. Below "Geratherm" is the text "Medical Diagnostic Systems" in a smaller, sans-serif font.

surgical area and provide active warming to the patient. All the blankets can be freely combined. The flexibility of the blankets makes it possible to warm the patient, even if complicated positioning on the operating table is required. After being placed over the patient, Geratherm UniqueTemp® can itself be covered with surgical drapes.

Geratherm UniqueTemp® is suitable for use with adult patients of all sizes. The blankets should not be used for children.

Geratherm UniqueTemp® is intended to be used during lengthy operations as well as by hypothermic patients. The device is not life-supporting. Geratherm UniqueTemp® should be operated only by qualified medical professionals.

The Geratherm UniqueTemp9 consists of the following components:

- . Control unit
	- · Holding claw
	- Mains cable
	- · Arm/shoulder blanket
	- · Reusable cover for arm/shoulder blanket
	- · Torso blanket
	- · Reusable cover for torso blanket
	- · Leg blankets
	- · Reusable cover for leg blankets
	- · Connecting cable
	- · Body temperature sensor

The Controller is made of anodized aluminum (colorless or black); furthermore, the control unit has a glass screen. All blankets and covers have a PU coated material. The connecting cable is made of silicone.

#### Indications for Use:

Geratherm UniqueTemp® is a thermal regulating system for adult patients in order to prevent hypothermia in the operating theater. It is suitable for pre-, intra-, and post-operative use. The device is not life-supporting. It is intended for use by appropriately trained healthcare professionals in clinical environments.

{2}------------------------------------------------

### Comparison with Predicate Device:

The Geratherm UniqueTemp® is identical to the predicate device in intended use, indications for use, where it is used, method of operation, safety characteristics, standards which are met (whereas the Geratherm UniqueTemp9 complies with a wider range of standards than the predicate device), used material and sterilization instructions. However, unlike the predicate device, the UniqueTemps has reusable covers which are steam permeable and washable up to 95°C. Therefore, the new device can be cleaned much easier.

The main technical characteristics are similar between the new and predicate device, but the UniqueTemp® has some new components. For instance, the new device has a color (TFT 3.5") temperature display with a touch screen and individual blanket control. Furthermore, the UniqueTemp® is equipped with a temperature pre-selection from 37°C up to 42°C in increments of 0.1°C (the predicate device has 3 possible preset temperatures of 38°C, 40°C and 43°C).

The UniqueTemp® is different from the predicate device in terms of the target population. The new device should be used for adult patients of all sizes, but not for children, whereas the Hot Dog Patient Warming System is intended for all patients.

Both devices are similar regarding the performance characteristics and reach a similar maximum temperature of 42°C and 43°C respectively. The only difference is the output power of the control unit, which does not raise questions regarding safety and effectiveness. Thus, the UniqueTemp® Patient Warming System is substantially equivalent to the predicate device.

### Non-Clinical Performance Data:

Geratherm UniqueTemp9 meets the requirements of Council Directive 93/42 EEC of 14.06.93 on Medical Devices and of the standard IEC 60601-1 (Medical electrical equipment, Part 1: General requirements for safety), IEC 60601-1-2 (Medical electrical equipment Part 1-2: General requirements for safety - Collateral standard: Electromagnetic compatibility - Requirements and Tests), and IEC 80601-2-35 (EMC, Medical electrical equipment - Part 2-35: Particular requirements for the basic safety and essential performance of heating devices using blankets, pads and mattresses and intended for heating in medical use). An EN ISO 13485 certified quality management system ensures that these requirements are met and entitles the manufacturer to use the CE 0118 label. Geratherm certifies compliance with relevant ISO/IEC/EN and other device-related standards.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for "Geratherm Medical Diagnostic Systems". The word "Geratherm" is in a bold, sans-serif font, with a registered trademark symbol to the right of the word. Below the word "Geratherm" are the words "Medical Diagnostic Systems" in a smaller, sans-serif font. The logo is black and white.

### Summary:

The substantial equivalence of the Geratherm UniqueTemp® and the referenced and approved earlier warming system, Hot Dog Patient Warming System (K052392), is based upon substantially similar methods for warming patients in the operating theater.

The presented data which was conducted using the Geratherm UniqueTemp®, in comparison to the predicate device, shows that the product is safe and effective for its intended use. The product components which are covered by this 510(k) premarket notification have been successfully tested for biocompatibility, electromagnetic compatibility, functionality and safety according to international standards.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three wing-like shapes, positioned above the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The text is arranged in a circular fashion around the bird-like figure.

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

Geratherm Medical AG c/o Ms. Andrea Pecsi Think! Schwarzwaldstrasse 5 78532 Tuttlingen Germany

OCT 1 9 2010

Re: K093244

Trade Name: Geratherm UniqueTemp® warming blankets Dated: September 15, 2010 Received: October 7, 2010

Dear Ms. Pecsi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{5}------------------------------------------------

Page 2 - Ms. Andrea Pecsi

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Bram D. Zuckerman, M.D.

am D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

## Indication for Use

510(k) Number (if known): K093244

Device Name: Geratherm UniqueTemp® Patient Warming System

OCT 1 9 2010

Indications for Use:

Geratherm UniqueTemp® is a thermal regulating system for adult patients in order to prevent hypothermia in the operating theater. It is suitable for pre-, intra- and post-operative use. The device is not life-supporting. It is intended for use by appropriately trained healthcare professionals in clinical environments.

Prescription Use _X AND/OR (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K93244

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/DWJ/K093244](https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/DWJ/K093244)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
