← Product Code [DWJ](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/DWJ) · K051336

# GERATHERM SOS-RESCUE BAG (K051336)

_R.G. Enterprises, Inc. · DWJ · Jun 29, 2006 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/DWJ/K051336

## Device Facts

- **Applicant:** R.G. Enterprises, Inc.
- **Product Code:** [DWJ](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/DWJ.md)
- **Decision Date:** Jun 29, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5900
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Geratherm SOS-Rescue Bag is a temperature controlled electric whole body warming system intended for use in preventing or treating the effects of hypothermia in adults while at the rescue scene or during transport by air or ground to a medical facility.

## Device Story

The Geratherm SOS-Rescue Bag is a portable, electrically heated thermal regulation system designed for pre-hospital and transport use. It consists of a sleeping-bag-style enclosure with an actively warmed upper panel containing resistive carbon fiber elements and an insulated bottom panel. The device is operated by rescue personnel in field or transport settings (ambulance, aircraft). A control module, powered by an auxiliary battery or vehicle/aircraft DC power, regulates temperature via a feedback circuit using two thermistor sensors. Operators select between 37°C or 42°C settings; the module provides visual LED status and audible alerts. The system includes a removable, washable protective liner for hygiene. By providing controlled conductive heat, the device prevents or counteracts hypothermia, reduces shock susceptibility, and decreases patient pain perception during transport.

## Clinical Evidence

Clinical evaluation performed at the Mayo Clinic (Kober et al., 2001) compared resistive carbon fiber active warming to passive warming in patients with hypothermia associated with minor trauma. Results indicated that patients treated with the active warming system experienced significantly lower degrees of hypothermia compared to those using passive warming techniques.

## Technological Characteristics

Thermal regulation system using resistive carbon fiber heating elements. Materials: polyurethane/polyester cloth laminate, Dacron® polyester hollow fiber fleece. Power: 12V/24-28V DC (battery, automotive, or aircraft). Control: Feedback circuit with dual thermistor sensors; high-impact polycarbonate housing. Connectivity: Standalone. Safety: Multiple fault sensors, circuit breaker, fused controller. Standards: EN 60601-1-2, IEC 60601-1, IEC 60601-2-35, EN 60601-2-32.

## Regulatory Identification

A thermal regulating system is an external system consisting of a device that is placed in contact with the patient and a temperature controller for the device. The system is used to regulate patient temperature.

## Predicate Devices

- ChillBuster™ Portable Electric Blanket (Thermogear, Inc. [K991684](/device/K991684.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a logo with the letters 'RG' stacked on top of each other. Below the letters, there is a zig-zag line. Underneath the zig-zag line, the words 'medical DIAGNOSTICS' are written in a smaller font. The logo appears to be for a medical diagnostics company.

## 051336 RG Medical Diagnostics

JUN 2 9 2006

## 510(k) Premarket Notification Summary

# Geratherm® SOS-Rescue Bag

Common Name - Warming Bag Classification name- Thermal Regulating System (870.5900) Predicate Device - ChillBuster™ Portable Electric Blanket (Thermogear, Inc. K991684)

The following material presents the 510 (k) Summary for the Geratherm® SOS-Rescue Bag

21130 Bridge Street • Southfield, MI 48034 Phone: 888-596-9498 ● Fax: 248-750-0187 ● Website: www.rgmd.com

Section 5 nage 1 of 7

{1}------------------------------------------------

#### Premarket (510K) Summary

#### Submitter Information

RG Enterprises, Inc. (dba) RG Medical Diagnostics 21130 Bridge Street Southfield. MI 48034 888-596-9498 Date Prepared 04/01/05

#### Device Name:

Proprietary name: Geratherm SOS-Rescue Bag (Thermamed "Smartcare") Common name: Hypothermic Therapy System CDRH Product Regulation: Thermal Regulation System (21 CFR, 870.5900)

#### Establishment Registration Number: 1835242

Classification: II (DWJ)

#### Predicate Device:

RG Medical Diagnostics, 21130 Bridge Street, Southfield, Michigan, 48034 has designated The Chillbuster™ Portable Electric Blanket System, Model 8001, manufactured by Thermogear, Inc. 18005 Lower Boones Ferry Rd., Togart, OR 977224 as the Predicate Device for the Geratherm SOS-Rescue Bag, manufactured by Geratherm AG, Fahrenheitstrasse 1, D-98716 Geschwenda/Germany.

#### Statement of Substantial Equivalence:

Geratherm SOS-Rescue Bag is substantially equivalent to:

1. ChillBuster™ Portable Electric Blanket (Thermogear, Inc K991684)

#### Description of Device:

The Geratherm SOS-Rescue Bag is made up of five major components:

1. Bag: The main functional element is 78.75 inches (200cm) by 31.5 inches (80 cm) comprised of two primary elements:

{2}------------------------------------------------

(a) The bag is constructed like a sleeping bag with an actively warmed upper panel and insulated bottom panel. The upper warming cover uses electrically resistive woven carbon fiber elements to generate warmth. The warming elements are sandwiched between two layers of Dacron® polyester hollow fiber insulating fleece. Lastly, the polvester fleece/carbon fiber layers are encapsulated with a waterproof polyurethane / polyester cloth laminate outside layer. A permanently attached power is cable located near the upper corner of the active warming cover. The free end is plugged into the power control module via color-coded quick-connect power connectors. The top warming cover is permanently sewn along one entire side to the insulated "bottom" panel. This arrangement allows top panel to be folded over the patient and secured once the patient has been placed in a prone position on the bottom panel. The bottom panel has 9 (nine) handholds around its perimeter that lets rescue personnel carry or move a supine patient without the use of a stretcher. (b) A removable waterproof polyurethane / polyester cloth laminate protective liner is attached to the bag's inner surface. The protective cover is held in place to the bag's inner surfaces with Velcro® attachments. This attachment system permits easy removal for machine washing and disinfecting.

2. Control Module: The control module is a impact plastic housing that contains the electronic circuitry. The module regulates bag temperature through a feedback circuit employing two thermistor temperature sensing systems. Operator control is achieved through the use of four buttons located on the face of the module: an on button, off button, a 37° C (98.6°F) select button, and a 42° C (107.6°F) select button. The control module uses LED arrays to display both bag temperature and battery power status. Normal functions are monitored both through visual displays and audible signal. Power sources include the following: the auxiliary battery, an automotive type power supply outlet, or an aircraft power supply outlet. Either 12 volt DC automotive or 24-28 volt DC aircraft power sources may be used to power the unit.

3. Battery: The Geratherm SOS-Rescue Bag uses a sealed lead-acid gel rechargeable battery of 10-ampere hour capacity.

4. Cabling: The Geratherm SOS-Rescue Bag has a power cable permanently attached to the upper corner. The cable attaches directly to the Control Module, The cable supplying power to the power module has a color-coded quick connect on the module end and a connecter that fits the auxiliary battery or automotive/aircraft DC power outlets on the other.

5. AC Charger: The Geratherm SOS-Rescue Bag employs a 12 volt, DC, 2ampere hour output battery charger to charge the battery or maintain full battery status. The charger may be used with either 110 or 240 Volt, 50 or 60 Hertz, AC input electric current.

{3}------------------------------------------------

#### Intended Use

The Geratherm® SOS-Rescue Bag is designed to prevent and counteract accidental hypothermia in accident victims and patients who are being transported with or without a stretcher, by ambulance, rotary or fixed wing aircraft. Using the Geratherm® SOS Rescue Bag System lowers susceptibility to shock caused by the cold. In addition, patients experience a subjective reduced perception of pain.

#### Non-Clinical Performance Data

The studies reported in this subsection were conducted by CE-LAB GmbH, iMG gGmbH, TREO Electrooptik GmbH, and TÜV GmbH

- The Geratherm SOS-Rescue Bag has been tested to the following Standards: standards:
General Medical Safety:

EMC Specification EN 60601-1-2:2001. NATO Standard DCS01, DCE01. Generic Specification of SPAME ( Special Purpose Aeromedical Equipment) Issue 3, page 16, No. 62 (vibration). IEC 60601-1:1990 + A1:1993 + A2:1995 IEC 60601-2-35:1996 EN 60601-2-32:1996 CE 0118 ISO 13485:2003 UL AWM Style 1007 & 1589 UL VW-1 FT

#### Clinical Performance Data

A formal clinical evaluation was performed at the Mayo Clinic comparing passive warming and active warming using Geratherm resistive carbon fiber electrical warming technology.

Alexander Kober, MD; Thomas Scheck, BS; Béla Fülesdi, MD; Frank Lieba, BS: Wolfgang Vlach, BS: Alexander Friedman, MD: & Daniel I. Sessler, MD . "Effectiveness of Resistive Heating Compared With Passive Warming in Treating Hypothermia Associated With Minor Trauma: A Randomized Trial," Mayo Clinic Proc. 2001:76:369-375

Results: Patients who were transported with Geratherm (Thermamed) electric / carbon fiber active warming systems experienced significantly lower degrees of hypothermia than those who were transported using passive warming techniques.

21130 Bridge Street · Southfield, MI 48034 Phone: 888-596-9498 • Fax: 248-750-0187 • Website: www.rgmd.com Section 5, page 4 of 7

{4}------------------------------------------------

#### Verification of Operation Under Non-Laboratory Conditions

The Geratherm SOS-Rescue Bag was tested according to the Operating Manual.and was found to prevent or reduce hypothermia by the following organizations:

CE 0118

SAMU 38 - Helicopter Mountain Rescue, Grenoble, France

ADAC Helicopter Helicopter Rescueservice, Munich, Germany

Royal Dutch Army, Helicopter Catastrophe Rescueteam, Netherlands

REGA - Mountain rescue service, REGA - Centre Zürich Airport, Zürich, Switzerland

#### Conclusions Drawn from Clinical and Non-Clinical Testing

The Geratherm SOS-Rescue Bag performed according to specification and compared favorably to the Predicate Device. No new issues of safety or effectiveness were found to arise in the Geratherm SOS-Rescue Bag when compared to the Predicate Device.

{5}------------------------------------------------

| Features                       | Geratherm® SOS-Rescue Bag                                                                           | ChillBuster™ Model 8001                |
|--------------------------------|-----------------------------------------------------------------------------------------------------|----------------------------------------|
| Indications for Use            | preventing / counteracting hypothermia                                                              | preventing / counteracting hypothermia |
| Function                       | low-level distributed heat to patient                                                               | low-level distributed heat to patient  |
| Heat Delivery Mechanism        | conduction                                                                                          | conduction                             |
| Heat Source                    | resistive carbon fiber cloth in insulated cover                                                     | thermal wire in blanket                |
| Electrical Requirements        | 12 VDC & 24-28 VDC @ 7 Amp                                                                          | 12 VDC @ 3.2 Amp                       |
| Heating Element Power          | 120 W (max)                                                                                         | 40 W (approx)                          |
| Heat Presented to Patient      | 37° C (98.6° F) or, 42° C (107.6° F)                                                                | 105° F                                 |
| User Heat Output Control       | Regulated 37° C or, Regulated 42° C                                                                 | Un-calibrated continuous 20W-40W       |
| Thermal Temperature Cutoff     | 37° C (98.6° F) or, 42° C (107.6° F)                                                                | 105° F                                 |
| Alarms                         | Audible and Visual                                                                                  | none disclosed                         |
| Circuit Protection             | multiple fault sensors, circuit breaker and fused controller                                        | Fused positive battery lead            |
| Internal Diagnostics           | Continuous automatic self testing                                                                   | none disclosed                         |
| Safety Agency Approvals        | TUV Munich                                                                                          | TUV Rheinland                          |
| EMC Compatibility Testing      | IEC 601-1-2                                                                                         | IEC 601-1-2                            |
| Cross-Contamination Protection | Disposable single use liner                                                                         | Disposable single use liner            |
| Cytotoxicity                   | L 929-Proliferation; EN 30993-5,-12; OSI 10993-5,-12; LM SOP 4-06-01                                |                                        |
| Controller Diagnostic Display  | cover temperature, battery power level, fault detection, fault identification                       |                                        |
| Bag Material                   | polyurethane film and Dacron® Polyester cloth laminate with Dacron® hollow filament fleece interior | Oxford nylon and nylon acrylic         |
| Control Unit Construction      | High Impact Polycarbonate                                                                           | Flame retardant polycarbonate          |
| Bag Cleaning                   | Machine wash - protective cover. Hand wash bag.                                                     | Machine wash and dry                   |
| System Weight                  | 9.6 lbs (4.35 kg) /w control unit                                                                   | 8 lbs                                  |
| Disinfection                   | Suitable for spray or sponge applied water based disinfectants                                      | No information                         |

### Comparison of Technological Characteristics ( SOS-Rescue Bag vs Predicate device)

21130 Bridge Street

Section 5, page 6 of 7

{6}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 2 9 2006

R.G. Enterprises, Inc. c/o Mr. Ronald G. Letoutneau President 21130 Bridge Street Southfield, MI 48034

Re: K051336

Geratherm SOS-Rescue Bag Regulation Number: 21 CFR 870.5900 Regulation Name: Thermal Regulating System Regulatory Class: Class II (Two) Product Code: DWJ Dated: June 21, 2006 Received: June 23, 2006

Dear Mr. Letoutneau:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may. therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices. good manufacturing practice. labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning vour device in the Federal Register.

{7}------------------------------------------------

Page 2 -- Mr. Ronald G. Letoutneau

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

. . . . . . .

Sincerely yours.

Duna R. bohner

ABram D. Zuckerman, M.D. -Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

### Indications for Use

510(k) Number (if known): K051336

Device Name: Geratherm SOS-Rescue Bag

Indications For Use:

The Geratherm SOS-Rescue Bag is a temperature controlled electric whole body warming system intended for use in preventing or treating the effects of hypothermia in adults while at the rescue scene or during transport by air or ground to a medical facility.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Duna R. Lochner

Division Sign-Off) Division Sign-Olf)
Division of Cardiovascular Devices

510(k) Number Koss (336

Page 1 of

Section 12. page 2 of 2

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/DWJ/K051336](https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/DWJ/K051336)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
