← Product Code [DRN](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/DRN) · K173483

# Pure Flow External Counter-Pulsation Device (K173483)

_Xtreem Pulse, LLC · DRN · May 30, 2018 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/DRN/K173483

## Device Facts

- **Applicant:** Xtreem Pulse, LLC
- **Product Code:** [DRN](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/DRN.md)
- **Decision Date:** May 30, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5225
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Pure Flow External Counter-Pulsation device is intended for the treatment of chronic stable angina that is refractory to optimal anti-anginal medical therapy and without options for revascularization. In addition, it is intended for use in healthy patients to provide improvement in vasodilation, increased blood flow. It is intended for use under the oversight of a healthcare professional.

## Device Story

Mechanical external counter-pulsation (ECP) device; inputs include real-time EKG signals and SpO2 data; device inflates/deflates series of fabric cuffs (calf, thigh, buttocks) synchronized with cardiac cycle; inflation during early diastole increases aortic pressure to augment myocardial perfusion; deflation during early systole reduces ventricular workload/afterload; operated by healthcare professionals in clinical settings; output provides hemodynamic support; benefits include restored blood flow to oxygen-deprived heart muscle; treatment typically 1 hour/day for 35 sessions.

## Clinical Evidence

No clinical data provided. Substantial equivalence supported by bench testing, including pressure control timing, pressure accuracy, and electrical safety/EMC testing.

## Technological Characteristics

Pneumatic compression system; nylon/cotton cuffs; Windows-based microprocessor; EKG/SpO2 sensors; 100-240V power; dimensions 1250mm x 700mm x 500mm; weight 120kg; software-controlled inflation/deflation; IEC 60601-1 and 60601-1-2 compliant.

## Regulatory Identification

An external counter-pulsating device is a noninvasive, prescription device used to assist the heart by applying positive or negative pressure to one or more of the body's limbs in synchrony with the heart cycle.

## Special Controls

*Classification.* (1) Class II (special controls) when the device is intended for the treatment of chronic stable angina that is refractory to optimal anti-anginal medical therapy and without options for revascularization. The special controls for this device are:(i) Nonclinical performance evaluation of the device must demonstrate a reasonable assurance of safety and effectiveness for applied pressure, synchronization of therapy with the appropriate phase of the cardiac cycle, and functionality of alarms during a device malfunction or an abnormal patient condition;
(ii) Reliabilities of the mechanical and electrical systems must be established through bench testing under simulated use conditions and matched by appropriate maintenance schedules;
(iii) Software design and verification and validation must be appropriately documented;
(iv) The skin-contacting components of the device must be demonstrated to be biocompatible;
(v) Appropriate analysis and testing must be conducted to verify electrical safety and electromagnetic compatibility of the device; and
(vi) Labeling must include a detailed summary of the device-related and procedure-related complications pertinent to use of the device.
(2) Class III (premarket approval) for the following intended uses: Unstable angina pectoris; acute myocardial infarction; cardiogenic shock; congestive heart failure; postoperative treatment of patients who have undergone coronary artery bypass surgery; peripheral arterial disease associated with ischemic ulcers rest pain or claudication, threatened gangrene, insufficient blood supply at an amputation site, persisting ischemia after embolectomy or bypass surgery, and/or pre- and post-arterial reconstruction to improve runoff; diabetes complicated by peripheral arterial disease or other conditions possibly related to arterial insufficiency including nocturnal leg cramps and/or necrobiosis diabeticorum; venous diseases, including prophylaxis of deep vein thrombophlebitis, edema (e.g., chronic lymphedema) and/or induration (e.g., stasis dermatitis) associated with chronic venous stasis, venous stasis ulcers, and/or thrombophlebitis; athletic injuries, including Charley horses, pulled muscles and/or edematous muscles; necrotizing cellulitis.
(c)
*Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with FDA on or before March 31, 2014, for any external counter-pulsating device, with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976, or that has, on or before March 31, 2014, been found to be substantially equivalent to any external counter-pulsating device, with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976. Any other external counter-pulsating device with an intended use described in paragraph (b)(2) of this section shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- Renew™ NCP-5 External Counterpulsation System ([K152115](/device/K152115.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 30, 2018

Xtreem Pulse, LLC % Rhonda Alexander Sr. Director, Regulatory & Leadership Development IUVO Consulting, LLC 1820 Perla Drive Virginia Beach, Virginia 23456

Re: K173483

Trade/Device Name: Pure Flow External Counter-Pulsation Device Regulation Number: 21 CFR 870.5225 Regulation Name: External Counter-Pulsating Device Regulatory Class: Class II Product Code: DRN Dated: April 16, 2018 Received: April 20, 2018

Dear Rhonda Alexander:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Nicole G. Ibrahim -S

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K173483

Device Name

PureFlow External Counter-Pulsation Device

#### Indications for Use (Describe)

The Pure Flow External Counter-Pulsation device is intended for the treatment of chronic stable angina that is refractory to optimal anti-anginal medical therapy and without options for revascularization. In addition, it is intended for use in healthy patients to provide improvement in vasodilation, increased blood flow. It is intended for use under the oversight of a healthcare professional.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary K173483

#### 1. SUBMITTER

XTREEM PULSE LLC Andrew Barile, CEO 353 W. 29 ST. SUITE 3 NEW YORK, NY 10001 UNITED STATES

Contact Person: Rhonda Alexander, MS, MPA Sr. Director, Regulatory & Leadership Development IUVO Consulting, LLC (757) 582-4337 ralexander@iuvoconsulting.com

Date Prepared: 22 May 2018

### II. DEVICE

| Name of Subject Device: | Pure Flow External Counter-Pulsation Device |
|-------------------------|---------------------------------------------|
| Classification Name:    | External Counter-Pulsation Device           |
| Regulatory Class:       | Class 2                                     |
| Product Code:           | DRN                                         |

### III. PREDICATE DEVICE

Renew™ NCP-5 External Counterpulsation System (K152115) Renew Group Ptc Ltd.

No reference devices were used in this submission.

### IV. DEVICE DESCRIPTION

The Pure Flow External Counter-Pulsation device is a mechanical device used to increase aortic pressure during early native heart diastole to augment myocardial perfusion, and to decrease aortic pressure during early native heart systole to reduce ventricular workload and vascular afterload.

The PureFlow consists of a main unit, and applied parts (computer monitor, keyboard, flexible tubes which connect to fabric cuffs, inside each cuff is an inflatable plastic balloon), EKG wires, and SpO2 sensors.

PureFlow works on the basis of external counter pulsation. The pressure is created by inflating a series of cuffs (similar to blood-pressure cuffs) that squeeze the extremities (calf's, thighs, buttocks) forcing blood back up to the heart on each heart cycle. This process ultimately restores (by force) sufficient

{4}------------------------------------------------

blood flow to parts of the heart that is lacking oxygen. The treatment process generally spans for a period of 1 hour each day for a period of 35 total treatments. ECP Treatment is proven to develop existing pathways of oxygenated blood back to the heart muscle. In other words, External Counterpulsation Therapy restores pathways for adequate & necessary blood flow back to heart through vessels that already exist.

The device also utilizes the following cleared components: Cable/Lead wire (ECG/EKG, SpO2, and Invasive Blood Pressure): K120010 LNCS/M-LNCS Oximetry Sensor: K101896

## V. INDICATIONS FOR USE

The Pure Flow External Counter-Pulsation device is intended for the treatment of chronic stable angina that is refractory to optimal anti-anginal medical therapy and without options for revascularization. In addition, it is intended for use in healthy patients to provide improvement in vasodilation, increased VO2, and increased blood flow. It is intended for use under the oversight of a healthcare professional.

## VI. SUMMARY OF COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

At a high level, the subject and predicate devices are based on the following same technological elements:

| CHARACTERISTICS                                  | PREDICATE DEVICE                                                                                                                                                                                                                                                                                                                                                                              | SUBJECT DEVICE                                                                                                                                                                                                                                                                                                                                                                                      | CONCLUSION |
|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
|                                                  | RENEW NCP-5 (K152115)                                                                                                                                                                                                                                                                                                                                                                         | PureFlow                                                                                                                                                                                                                                                                                                                                                                                            |            |
| Intended Use                                     | Stable and Unstable Angina, Congestive<br>Heart Failure, cardiogenic shock, and<br>acute myocardial infarction, increase in<br>circulation, increase VO2. In addition, it<br>is intended for use in healthy patients<br>to provide improvement in<br>vasodilation, increased VO2, and<br>increased blood flow. It is intended for<br>use under the oversight of a healthcare<br>professional. | Stable and Unstable Angina,<br>Congestive Heart Failure,<br>cardiogenic shock, and acute<br>myocardial infarction, increase in<br>circulation, increase VO2. In<br>addition, it is intended for use in<br>healthy patients to provide<br>improvement in vasodilation,<br>increased VO2, and increased<br>blood flow. It is intended for use<br>under the oversight of a healthcare<br>professional. | SAME       |
| Classification Name<br>Mechanism of<br>operation | Counter-Pulsating, External                                                                                                                                                                                                                                                                                                                                                                   | Counter-Pulsating, External                                                                                                                                                                                                                                                                                                                                                                         | SAME       |
| Triggering Mechanism                             | T-Wave trigger for inflation and P-<br>Wave deflation timing                                                                                                                                                                                                                                                                                                                                  | T-Wave trigger for inflation and P-<br>Wave deflation timing                                                                                                                                                                                                                                                                                                                                        | SAME       |
| Microprocessor                                   | Windows Based                                                                                                                                                                                                                                                                                                                                                                                 | Windows Based                                                                                                                                                                                                                                                                                                                                                                                       | SAME       |

{5}------------------------------------------------

| Emergency System<br>power-down                              | Red-Emergency Stop Button                                                                                                              | Red-Emergency Stop Button                                                                                                            | SAME      |
|-------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Maximum pressure<br>used for<br>treatment/Timer<br>Settings | 6.0 PSI<br>Timer- 1-60 minutes<br>(10 minute increments)                                                                               | 4.0 PSI<br>Timer: 1-60 minutes<br>(Single minute increments)                                                                         | DIFFERENT |
| Cuff system                                                 | Nylon fabric cuffs, three parts, calf,<br>thigh and buttocks.                                                                          | Nylon fabric cuffs, three parts, calf,<br>thigh and buttocks.                                                                        | SAME      |
| Major components                                            | Base Unit, Air-Tubes, EKC, SPO2, EKG<br>and three Cuffs                                                                                | Base Unit, Air-Tubes, EKC, SPO2,<br>EKG and three Cuffs                                                                              | SAME      |
| Voltage /HZ/ Wattage                                        | 100-240V/ 50/60HZ                                                                                                                      | 100-240 V 50/60HZ                                                                                                                    | SAME      |
| Dimension/Weight                                            | 1930mm (L) x 718mm (W) x 927mm (H)<br>102KG                                                                                            | 1250mmX700mmX 500mm<br>120KG                                                                                                         | SAME      |
| Major components                                            | Base Unit, Air-Tubes, EKC, SPO2, EKG<br>and three cuffs                                                                                | Base Unit, Air-Tubes, EKC, SPO2,<br>EKG and three nylon and cotton<br>cuffs                                                          | SAME      |
| Operating<br>Environment                                    | Temperature: 16° to 30° ( 60.8° F to 86<br>° F)<br>Humidity: 30% to 70% RH<br>(non-condensing)<br>Barometric Pressure: 80kPA to 100kPA | Temperature: 10° to 30° ( 50 °F- 86<br>°F)<br>Humidity: 30% to 75% RH<br>(non-condensing)<br>Barometric Pressure: 80kPA to<br>100kPA | SAME      |
| Electrical Safety and<br>EMC Safety Testing                 | Passed IEC 60601                                                                                                                       | Passed IEC 60601                                                                                                                     | SAME      |
| Biocompatibility Testing                                    | ISO 10993                                                                                                                              | NONE                                                                                                                                 | DIFFERENT |

There are no differences between the subject device and the predicate device with respect to indications and intended use. Electrical safety and EMC testing was performed per IEC 60601 on the subject device. The device passed the testing, just as the predicate device did. Thus, in electrical safety and EMC testing, the devices are equivalent.

The difference in the length of the device is only a design preference. Renew has decided to incorporate a built in treatment bed. The core dimension of the machine are the same. The additional length of the unit is due to the extension of the bed components, which have no therapeutic or functional effect on the treatment. Our device can be wheeled bedside to the patient who can lie comfortably on specialized beds if required.

{6}------------------------------------------------

The increased weight of PureFlow, as compared to the predicate, is due to the materials used for the shell of the housing of the unit. It weighs approximately 18kg more but this has no direct effect on the clinical application.

No biocompatibility testing was conducted on PureFlow, as none of its components or parts are intended to touch the skin. The cuffs of the device are intended to be used on top of clothing. The device will only use 510(k) cleared ECG electrodes and pulse oximeters.

### VII. NON-CLINICAL TESTING

The following performance data were provided in support of the substantial equivalence determination.

## Performance Testing

### Shelf-Life Testing

Shelf-life was calculated to be three years, based on information provided by the manufacturer estimated use conditions.

## Pressure Control Timing Measurement Test

The device was evaluated for synchronization of therapy with the appropriate phase of cardiac cycle or functionality of alarms per the special control in 21 CFR 870.5225. The reference standard used was JIS T0601-2-204: Particular requirements for the basic safety and essential performance of pneumatic compression equipment. Timer accuracy should be within +/- 5% of the set value. Three models were tested, and each met the pass criteria.

### Pressure Accuracy Test

Pressure was set for each device, and the actual pressure was measured to check if accuracy is verified. Referenced standard used was JIS T 0601-2-204 Particular requirements for the basic safety and essential performance of pneumatic compression equipment 201.12.1. Pressure must be +/- 20% from the set pressure value. Three models were tested, and each met the pass criteria.

### Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the subject device, and was found to comply with IEC 60601-1 and 60601-1-2.

## Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was determined to be of "moderate" level of concern.

#### VIII. CLINICAL TESTING

Clinical data were not required in this submission to support a finding of substantial equivalence.

### IX. CONCLUSIONS

{7}------------------------------------------------

We conclude that the information provided in this submission are sufficient to demonstrate that the Pure Flow External Counter-Pulsation Device is substantially equivalent to the predicate device which is currently marketed for the same intended use.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/DRN/K173483](https://fda.innolitics.com/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/DRN/K173483)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
