AECP-A, AUTOMATIC EXTERNAL COUNTERPULSATION DEVICE
Device Facts
| Record ID | K021340 |
|---|---|
| Device Name | AECP-A, AUTOMATIC EXTERNAL COUNTERPULSATION DEVICE |
| Applicant | External Counterpulsation Lab |
| Product Code | DRN · Cardiovascular |
| Decision Date | Jan 30, 2003 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.5225 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The AECP-A is a non-invasive external counterpulsation device for treatment of patients suffering from stable or unstable angina pectoris, acute myocardial infarction and cardiogenic shock
Device Story
AECP-A is a non-invasive external counterpulsation device; treats patients with stable/unstable angina, acute myocardial infarction, and cardiogenic shock. Device applies sequential pressure to lower extremities synchronized with cardiac cycle to enhance diastolic blood flow and reduce cardiac workload. Used in clinical settings under physician supervision. Benefits include improved coronary perfusion and hemodynamic support.
Clinical Evidence
No clinical data provided in the document; substantial equivalence determination based on regulatory review.
Technological Characteristics
Non-invasive external counterpulsation system; utilizes sequential pneumatic compression synchronized with the cardiac cycle. Class III device (21 CFR 870.5225).
Indications for Use
Indicated for patients suffering from stable or unstable angina pectoris, acute myocardial infarction, and cardiogenic shock.
Regulatory Classification
Identification
An external counter-pulsating device is a noninvasive, prescription device used to assist the heart by applying positive or negative pressure to one or more of the body's limbs in synchrony with the heart cycle.
Special Controls
*Classification.* (1) Class II (special controls) when the device is intended for the treatment of chronic stable angina that is refractory to optimal anti-anginal medical therapy and without options for revascularization. The special controls for this device are:(i) Nonclinical performance evaluation of the device must demonstrate a reasonable assurance of safety and effectiveness for applied pressure, synchronization of therapy with the appropriate phase of the cardiac cycle, and functionality of alarms during a device malfunction or an abnormal patient condition; (ii) Reliabilities of the mechanical and electrical systems must be established through bench testing under simulated use conditions and matched by appropriate maintenance schedules; (iii) Software design and verification and validation must be appropriately documented; (iv) The skin-contacting components of the device must be demonstrated to be biocompatible; (v) Appropriate analysis and testing must be conducted to verify electrical safety and electromagnetic compatibility of the device; and (vi) Labeling must include a detailed summary of the device-related and procedure-related complications pertinent to use of the device. (2) Class III (premarket approval) for the following intended uses: Unstable angina pectoris; acute myocardial infarction; cardiogenic shock; congestive heart failure; postoperative treatment of patients who have undergone coronary artery bypass surgery; peripheral arterial disease associated with ischemic ulcers rest pain or claudication, threatened gangrene, insufficient blood supply at an amputation site, persisting ischemia after embolectomy or bypass surgery, and/or pre- and post-arterial reconstruction to improve runoff; diabetes complicated by peripheral arterial disease or other conditions possibly related to arterial insufficiency including nocturnal leg cramps and/or necrobiosis diabeticorum; venous diseases, including prophylaxis of deep vein thrombophlebitis, edema (e.g., chronic lymphedema) and/or induration (e.g., stasis dermatitis) associated with chronic venous stasis, venous stasis ulcers, and/or thrombophlebitis; athletic injuries, including Charley horses, pulled muscles and/or edematous muscles; necrotizing cellulitis. (c) *Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with FDA on or before March 31, 2014, for any external counter-pulsating device, with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976, or that has, on or before March 31, 2014, been found to be substantially equivalent to any external counter-pulsating device, with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976. Any other external counter-pulsating device with an intended use described in paragraph (b)(2) of this section shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.
Related Devices
- K012141 — AECP-A, AUTOMATIC EXTERNAL COUNTERPULSATION DEVICE · External Counterpulsation Lab · Oct 25, 2001
- K022078 — SOLA, COUNTER-PULSATION, MODEL SECP-S · Sola Medical Devices, LLC · Jan 10, 2003
- K130439 — EXTERNAL COUNTERPULSATION DEVICE WITH SP02 MORNITORING · Chongqing Psk Sci-Tech Development Co., Ltd. · Dec 12, 2013
- K023427 — CARDIASSIST COUNTERPULSATION SYSTEM, MODEL MARK 3000 · Cardiomedics, Inc. · Jan 7, 2003
- K173483 — Pure Flow External Counter-Pulsation Device · Xtreem Pulse, LLC · May 30, 2018