MODIFICATION TO: AUTOPULSE RESUSCITATION SYSTEM MODEL 100
Applicant
Revivant Corp.
Product Code
DRM · Cardiovascular
Decision Date
Nov 12, 2003
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.5200
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The AutoPulse Resuscitation System Model 100 is intended to be used as an adjunct to manual CPR, on adult patients only, in cases of clinical death as defined by lack of spontaneous breathing and pulse.
Device Story
AutoPulse Resuscitation System Model 100 is a portable, battery-powered automated external cardiac compressor. Device consists of a reusable platform containing a control panel, drive mechanism, control system, and rechargeable battery, plus a single-use chest compression assembly (CCA) with a patient liner. Used as an adjunct to manual CPR to provide chest compressions to adult patients. Operator places patient on platform and secures CCA; device performs automated compressions. Intended to assist in resuscitation efforts by providing consistent chest compressions, potentially improving perfusion during cardiac arrest.
Clinical Evidence
Bench testing only; no clinical data provided. Testing conducted to evaluate conformance to product specifications and substantial equivalence to predicate devices.
Technological Characteristics
Portable, battery-powered automated chest compression system. Components: reusable platform (control panel, drive mechanism, control system, battery) and single-use chest compression assembly (CCA) with patient liner. Materials are biocompatible and previously cleared in other products.
Indications for Use
Indicated for adult patients in clinical death (lack of spontaneous breathing and pulse) as an adjunct to manual CPR.
Regulatory Classification
Identification
An external cardiac compressor is an externally applied prescription device that is electrically, pneumatically, or manually powered and is used to compress the chest periodically in the region of the heart to provide blood flow during cardiac arrest. External cardiac compressor devices are used as an adjunct to manual cardiopulmonary resuscitation (CPR) when effective manual CPR is not possible (e.g., during patient transport or extended CPR when fatigue may prohibit the delivery of effective/consistent compressions to the victim, or when insufficient EMS personnel are available to provide effective CPR).
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Nonclinical performance testing under simulated physiological conditions must demonstrate the reliability of the delivery of specific compression depth and rate over the intended duration of use.
(2) Labeling must include the following:
(i) The clinical training necessary for the safe use of this device;
(ii) Adjunctive use only indication prominently displayed on labels physically placed on the device and in any device manuals or other labeling;
(iii) Information on the patient population for which the device has been demonstrated to be effective (including patient size and/or age limitations,
*e.g.,* adult, pediatric and/or infant); and(iv) Information on the time necessary to deploy the device as demonstrated in the performance testing.
(3) For devices that incorporate electrical components, appropriate analysis and testing must demonstrate that the device is electrically safe and electromagnetically compatible in its intended use environment.
(4) Human factors testing and analysis must validate that the device design and labeling are sufficient for effective use by the intended user, including an evaluation for the time necessary to deploy the device.
(5) For devices containing software, software verification, validation, and hazard analysis must be performed.
(6) Components of the device that come into human contact must be demonstrated to be biocompatible.
AutoPulse™ Resuscitation System Model 100 (K011046)
AutoPulse™ Resuscitation System Model 100 (K022345)
Related Devices
K072572 — PRIMACONNEX RD AND WD ESTHETIC CONTOUR ZI ABUTMENTS · Lifecore Biomedical, Inc. · Oct 5, 2007
K063602 — AUTOPULSE RESUSCITATION SYSTEM MODEL100 · Zoll Circulation · Dec 21, 2006
K040453 — AUTOPULSE RESUSCITATION SYSTEM, MODEL 100 · Revivant Corp. · Mar 11, 2004
K022345 — AUTOPULSE RESUSCITATION SYSTEM · Revivant Corp. · Aug 15, 2002
K112998 — AUTOPULSE RESUSCITATION SYSTEM MODEL 1000 · Zoll Circulation · Mar 15, 2012
Submission Summary (Full Text)
{0}------------------------------------------------
K032852
NOV 1 2 2003
## 510(k) Summary
General Information
| Classification | Class III |
|----------------|-------------------------------------------|
| Trade Name | AutoPulse™ Resuscitation System Model 100 |
| Submitter | Revivant Corporation |
| | 775 Palomar Avenue |
| | Sunnyvale, CA 94085 |
| | 408-524-3500 |
| Contact | Bob H. Katz |
| | Vice President, Operations |
### Intended Use
The AutoPulse Resuscitation System Model 100 is intended to be used as an adjunct to manual CPR, on adult patients only, in cases of clinical death as defined by lack of spontaneous breathing and pulse.
#### Predicate Devices
| K851139 | Thumper™ Model 1005 |
|---------|-------------------------------------------|
| K011046 | AutoPulse™ Resuscitation System Model 100 |
| K022345 | AutoPulse™ Resuscitation System Model 100 |
#### Device Description
AutoPulse Resuscitation System Model 100 ("Device") is an automated, portable, battery powered device that compresses the chest of an adult human as an adjunct to manual CPR. The Device consists of a single use chest compression assembly (CCA) that includes a patient liner, and a reusable platform that contains a user control panel, a drive mechanism, a control system, and a power system (rechargeable battery).
### Materials
All materials used in the manufacture of the AutoPulse Resuscitation System Model 100 are suitable for this use and have been used in numerous previously cleared products.
{1}------------------------------------------------
## Testing
11.
Appropriate product testing was conducted to evaluate conformance to product specification and substantial equivalence to predicate devices.
## Summary of Substantial Equivalence
The AutoPulse Resuscitation System Model 100 is equivalent to the predicate products. The indications for use, basic overall function, and materials used are substantially equivalent.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines above them that resemble wings or feathers.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 1 2 2003
Mr. Bob H. Katz Vice President, Operations Revivant Corporation 775 Palomar Ave. Sunnyvale, CA 94085
Re: K032852
Trade/Device Name: Autopulse Resuscitation System, Model 100 Regulation Number: 870.5200 Regulation Name: External Cardiac Compressor Regulatory Class: III Product Code: DRM Dated: October 31, 2003 Received: November 3, 2003
Dear Mr. Katz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898, In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 -- Mr. Bob H. Katz
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known):
This application K032852
Device Name:
AutoPulse™ Resuscitation System Model 100
Indications for Use:
The AutoPulse Resuscitation System Model 100 is intended to be used as an adjunct to manual CPR, on adult patients only, in cases of clinical death as defined by lack of spontaneous breathing and pulse.
f. W. Aterlin
Infection
510(k) Number: K032852
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED '
### Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use V (Per 21 CFR 801.109)
OR
Over-The-Counter Use ________ (Optional Format 1-2-96)
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.