ITCLAMP 50

K140805 · Innovative Trauma Care, Inc. · DXC · Oct 9, 2014 · Cardiovascular

Device Facts

Record IDK140805
Device NameITCLAMP 50
ApplicantInnovative Trauma Care, Inc.
Product CodeDXC · Cardiovascular
Decision DateOct 9, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4450
Device ClassClass 2
AttributesTherapeutic

Intended Use

The iTClamp50 is a trauma clamp device for the temporary control of severe bleeding in the extremities, axilla, inguinal areas, scalp, and neck.

Device Story

iTClamp50 is a self-locking surgical clamp designed for temporary control of severe bleeding by sealing skin edges to create a contained hematoma. Device components include suture needles, plastic shell, locking mechanism, and lock release. Needles penetrate skin to evert edges between pressure bars, anchoring device to reduce slippage. Adjustable locking mechanism allows modulation of pressure to achieve fluid-tight seal. Used by first responders or medical personnel in trauma settings; provides temporary stabilization until surgical repair. Benefits include stable clot formation and hemorrhage control. Device is sterile and single-use.

Clinical Evidence

No clinical trials performed. Evidence includes cadaver studies comparing iTClamp50 to standard of care (manual pressure/balloon catheter) for neck wounds, confirming suitability. User assessment (n=15 first responders) using bleeding arm simulator (80-90 mmHg) across various environmental conditions (low light, wet/cold hands, etc.) demonstrated successful application/removal with <1% needle stick rate in 270 applications. Supportive data provided by unsolicited OUS and off-label case studies for neck wounds.

Technological Characteristics

Self-locking surgical clamp; plastic shell construction; suture needles for skin eversion; adjustable locking mechanism for pressure control. Sterile, single-use. No changes to physical composition, sterilization, or packaging from predicate.

Indications for Use

Indicated for temporary control of severe bleeding in extremities, axilla, inguinal areas, scalp, and neck in patients requiring trauma care.

Regulatory Classification

Identification

A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of three human profiles facing to the right, stacked on top of each other. The profiles are black against a white background. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 October 9, 2014 Innovative Trauma Care, Inc. Richard Waite Sr. Director of Quality Assurance/ Regulatory Affairs 3463 Magic Dr., Suite 120 San Antonio, Texas 78229 Re: K140805 Trade/Device Name: iTClamp50 Regulation Number: 21 CFR 870.4450 Regulation Name: Vascular Clamp Regulatory Class: Class II Product Code: DXC Dated: August 15, 2014 Received: August 29, 2014 Dear Mr. Waite: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Kenneth J. Cavanaugh -S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K140805 Device Name iTClamp50 Indications for Use (Describe) The iTClamp50 is a trauma clamp device for the temporary control of severe bleeding in the extremities, axilla, inguinal areas, scalp, and neck. X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### 510(k) Premarket Notification Innovative Trauma Care, Inc. iTClamp™50 ## 510(k) Summary In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the iTClamp™50 is provided below. | Device Common Name: | Vascular Clamp | |-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Proprietary Name: | iTClamp™50 | | Submitter: | Innovative Trauma Care, Inc.<br>3463 Magic Dr., Suite 120<br>San Antonio, TX 78229 | | Contact: | Richard Waite<br>Sr. Director of Quality Assurance/ Regulatory Affairs,<br>3463 Magic Dr., Suite 120<br>San Antonio, TX 78229<br>Office: 210-582-5850 ext. 1932<br>FAX: 210-582-5851<br>Cell: 214-662-9277<br>rwaite@innovativetraumacare.com | | Date Prepared: | February 17, 2014 | | Classification<br>Regulation: | 870.4450 | | Panel: | Cardiovascular | | Product Code: | DXC | | Predicate Device | K132651 - iTClamp™50 | #### Indication for Use The iTClamp™50 is a trauma clamp device for the temporary control of severe bleeding in the extremities, axilla, inguinal areas, scalp, and neck. The purpose of this 510(k) is to modify the already cleared indication for use to include use of the device on severe bleeding areas of the neck. #### Device Description The iTClamp™50 is a clamp device that quickly controls critical bleeding by closing the skin to create a temporary, contained hematoma until surgical repair. The iTClamp50 is a selflocking surgical clamp with suture needles that penetrate the skin to evert the skin edges (similar to sutures or staples) between pressure bars of the device and anchor it to the skin to {4}------------------------------------------------ #### 510(k) Premarket Notification Innovative Trauma Care, Inc. iTClamp™50 reduce slippage and leakage. Pressure is evenly distributed across the bars, which seal the skin over a wound. An adjustable locking mechanism increases or decreases pressure across the wound to achieve a fluid tight seal. The iTClamp controls bleeding by sealing the skin closed to apply direct pressure to the cut edges of the skin and create a temporary pool of blood (hematoma) under pressure. This permits formation of a stable clot until the patient can receive medical care and/or surgical repair. The device is provided sterile and is for single use. The iTClamp consists of the following components: - 1) Suture needles - 2) Plastic shell - 3) Locking mechanism - 4) Lock release mechanism #### Biocompatibility Testing Additional Biocompatibility Testing is not necessary to support the modified indication for use. Physical composition of the product, sterilization and packaging has not changed in anyway; therefore, biocompatibility testing provided in the previous 510(k) (K132651) is applicable to support this 510(k). #### Performance Testing - Bench Additional bench performance testing is not necessary. The physical composition of the product, packaging and sterilization has not changed in any way. Therefore, the bench testing provided in the previous 510(k) (K132651) is applicable to this 510(k). Innovative Trauma Care has performed cadaver testing to demonstrate substantial equivalence of the iTClamp50 for the new proposed indication for use on neck wounds. In addition, a second cadaver study was generated which compared the performance of the iTClamp50 and standard of care, i.e. tamponade by direct manual pressure and balloon catheter inflation in a wound. The results of the Cadaver Studies demonstrate that the device is suitable for use on the neck and is substantially equivalent to the previous 510(k) K132651. A user assessment was performed utilizing 15 trained first responders (fire fighters) to ensure the Directions for Use (DFU) were appropriate to support the clearance of K132651. The assessment utilized a bleeding arm simulator model. The simulator model uses a pulsing pump motor to delivery pulses of simulated blood (water) through a wound in the arm with pressure of 80-90 mmHg. The environmental conditions they were exposed to were: Normal conditions, cold hands, non-dominant hand, and dominant hands, low light, strobe lights, in a shower and underwater using a skin pad, fatigued hand exhausted), with fire gloves {5}------------------------------------------------ #### 510(k) Premarket Notification Innovative Trauma Care, Inc. iTClamp™50 on. These were completed using both wet hands (to simulate blood they were wetted with glycerol/water solution) and dry hands wearing vinyl exam gloves (except when wearing firefighting gloves and gear). The primary assessment was external leakage of water from the wound while the pump continued to run. Additionally, user needle sticks were evaluated. The outcome of the user assessment identified that the training based upon the DFU provided sufficient information for all participants to successfully apply and remove the device; the occurrence of needle stick was <1% in 270 device applications (Environmental use of the iTClamp™ 50 in a group of firefighters, Mckee, J., Mckee, I., Gao, H., Frost, G., Kirkpatrick, A. W.) The DFU content for the application and removal of the device has not changed and the technique remains the same within the scope of the expanded Indications for Use, therefore no additional user testing was conducted to support the temporary control of severe bleeding in neck wounds. #### Performance Testing - Animal Not applicable. Animal Performance Testing is not necessary to establish the substantial equivalence of the device. #### Performance Testing - Clinical Study Case studies were captured to support product performance for the cleared indications. For expansion of the indication to include temporary control of severe bleeding in the neck, data from unsolicited case studies of OUS and off-label use provided supportive data on the device performance for treatment of neck wounds. #### Substantial Equivalence The iTClamp50 is the same device that was previously cleared in K132651 to provide temporary bleeding control to wounds in the extremities, axilla, inguinal areas, and scalp until medical and/or surgical repair can be obtained. The iTClamp™ used for the new indication on the neck raises no new or different types of effectiveness questions because direct pressure is the standard of care for bleeding wounds to the neck. The available performance data support use of the iTClamp50 for neck wounds. The iTClamp50, therefore, is substantially equivalent to the already cleared iTClamp50 classified under 870.4450. | | Subject Device | Predicate Device | |--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) Number | TBD | K132651 | | Classification / Procode | 870.4450 / DXC | 870.4450 / DXC | | Device Name | iTClampTM50 | iTClampTM50 | | Manufacturer | Innovative Trauma Care, Inc. | Innovative Trauma Care, Inc. | | | Subject Device | Predicate Device | | Indication for Use | The iTClamp50 is a trauma<br>clamp device for the<br>temporary control of severe<br>bleeding in the extremities,<br>axilla, inguinal areas, scalp<br>and neck. | The iTClamp50 is a trauma<br>clamp device for the<br>temporary control of severe<br>bleeding in the extremities,<br>axilla, inguinal areas and<br>scalp. | | Device Design | iTClamp50 is a self-locking<br>surgical clamp with suture<br>needles that penetrate the<br>skin to evert the skin edges<br>between pressure bars of the<br>device and anchor it to the<br>skin to reduce slippage and<br>leakage. Pressure is evenly<br>distributed across the bars,<br>which seal the skin over a<br>wound. An adjustable<br>locking mechanism<br>increases or decreases<br>pressure across the wound<br>to achieve a fluid tight seal<br>or wound closure as desired. | iTClamp50 is a self-locking<br>surgical clamp with suture<br>needles that penetrate the<br>skin to evert the skin edges<br>between pressure bars of the<br>device and anchor it to the<br>skin to reduce slippage and<br>leakage. Pressure is evenly<br>distributed across the bars,<br>which seal the skin over a<br>wound. An adjustable<br>locking mechanism<br>increases or decreases<br>pressure across the wound<br>to achieve a fluid tight seal<br>or wound closure as desired. | | Device Operation | Application of pressure by<br>applying clamp to<br>temporarily seal wound site | Application of pressure by<br>applying clamp to<br>temporarily seal wound site | | Picture | Image: iTClamp50 | Image: iTClamp50 | {6}------------------------------------------------ ## 510(k) Premarket Notification Innovative Trauma Care, Inc. Innovative Trauma Care, Inc. iTClamp™ iTClamp™50
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...