BOOMERANG WIRE

K061075 · Cardiva Medical, Inc. · DXC · Sep 1, 2006 · Cardiovascular

Device Facts

Record IDK061075
Device NameBOOMERANG WIRE
ApplicantCardiva Medical, Inc.
Product CodeDXC · Cardiovascular
Decision DateSep 1, 2006
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.4450
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Cardiva Medical Boomerang™ Wire is intended to promote hemostasis at arteriotomy sites as an adjunct to manual compression. The Boomerang™ Wire is indicated for use in patients undergoing diagnostic femoral artery catheterization procedures, using 5 or 6 Fr introducer sheaths.

Device Story

Boomerang Wire system consists of sterile disposable wire and clip; used as adjunct to manual compression for femoral artery hemostasis. Device inserted through existing introducer sheath post-catheterization; distal tip deploys low-profile disc within femoral artery lumen. Disc pulled against arteriotomy site; upward tension applied to conform disc to vessel intima, sealing site. External clip at skin surface maintains tension; allows continued distal perfusion during hemostasis. Device facilitates natural vessel wall recoil and clotting; disc collapsed and wire removed post-procedure. No material left in body. Used in clinical setting by medical professionals.

Clinical Evidence

Performance supported by bench testing, biocompatibility testing, and animal testing. Post-market clinical data included to support safety and effectiveness claims.

Technological Characteristics

Sterile, disposable wire and clip system. Mechanical deployment of internal disc for vessel sealing. No electrical components or software. Materials evaluated for biocompatibility.

Indications for Use

Indicated for patients undergoing diagnostic femoral artery catheterization using 5 or 6 Fr introducer sheaths to promote hemostasis at arteriotomy sites as an adjunct to manual compression.

Regulatory Classification

Identification

A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K@61075 ps 1 of 2 Boomerang™ Wire 510(k) Notification ## 510(k) SUMMARY Cardiva Medical, Inc. Boomerang™ Wire 510(k) Notification SEP - 1 2006 ## GENERAL INFORMATION | Manufacturer: | Cardiva Medical, Inc. | |-----------------|-----------------------------------------------| | | 2585 Leghorn Street | | | Mountain View, CA 94043 | | | Phone: (650) 964-8900 | | | Facsimile: (650) 964-8911 | | | Establishment Registration Number: 3004182619 | | Contact Person: | Glenn Foy | | | President | | Date Prepared: | 4/14/2006 | ## DEVICE INFORMATION | Trade name: | Boomerang™ Wire | |-----------------------|-------------------------------------------------------------| | Classification Names: | Vascular Clamp (21 C.F.R. § 870.4450); | | | Catheter, Intravascular, Diagnostic (21 C.F.R. § 870.1200); | | | Surgical Vessel Dilator (21 C.F.R. § 870.4475); | | | Blood Access Device and Accessories (21 C.F.R. § 870.5540); | | Classification: | Class II | #### PREDICATE DEVICES Cardiva Medical VasoStasis Vascular Closure System (K041486) Radi Medical Systems AB, FemoStop™ System (K915280) CardioThoracic Systems, Inc., CTS FloCoil™ Shunt (K970638) ## INTENDED USE/INDICATIONS FOR USE The Cardiva Medical Boomerang™ Wire is intended to promote hemostasis at arteriotomy sites as an adjunct to manual compression. The Boomerang™ Wire is indicated for use in patients undergoing diagnostic femoral artery catheterization procedures, using 5 or 6 Fr introducer sheaths. {1}------------------------------------------------ Boomerang™ Wire 510(k) Notification #### DEVICE DESCRIPTION The Boomerang™ Wire consists of a sterile disposable Boomerang Wire and a sterile disposable Boomerang Clip (refer to the detailed Figures 1,2,3 in Section VIII). In conjunction with manual compression, the Boomerang Wire provides hemostasis at a femoral access site after femoral arterial catheterization while allowing continued distal perfusion. After completion of catheterization, the Boomerang Wire is inserted into the artery through the existing introducer sheath. After insertion, the distal tip of the Boomerang Wire is deployed, which opens the ffat, low-profile Boomerang Disc within the lumen of the femoral artery. The Boomerang Disc is then pulled back gently to the distal end of the introducer sheath. The introducer sheath is then removed from the vessel over the Boomerang Wire and the low-profile Boomerang Disc is positioned against the inside of the arteriotomy. Gentle upward tension is applied to the Boomerang Wire, which conforms the Boomerang Disc to the contours of the vessel and secures it against the intima, sealing the arteriotomy. The tension is then held in place by the external Boomerang Clip at the surface of the skin at the puncture site. The tension between the Boomerang Disc and the Boomerang Clip creates a site-specific compression of the arteriotomy and tract and establishes temporary hemostasis. This allows natural recoil of the smooth muscle of the vessel wall to occur at the arteriotomy site while the body's natural clotting process begins. Following the procedure, the Boomerang Disc is collapsed and the Boomerang Wire is completely removed from the artery. No part of the device is left behind nor is there any material introduced to alter the body's own natural clotting process. Final closure of the vessel occurs with manual compression to close the remaining needle puncture site left by removing the Boomerang Wire. #### SUBSTANTIAL EQUIVALENCE The Boomerang Wire is substantially equivalent to predicate devices currently being marketed. The marketed predicates are identified above. The Boomerang Wire is substantially equivalent to the predicate devices with regard to function, intended use, physical characteristics, materials and performance testing. All necessary testing was performed on the Boomerang Wire to ensure the product is substantially equivalent to the predicates and that any differences do not have a significant effect on safety and effectiveness. #### PERFORMANCE TESTING Various testing which included bench, biocompatibility, and animal testing was performed on the Boomerang Wire to ensure the product and the product materials were adequately tested and evaluated to demonstrate the product meets or exceeds the performance requirements and is safe and effective for its intended use. In addition, post market clinical data have been included to further support that the product meets or exceeds the performance requirements and is safe and effective for its intended use. #### CONCLUSION The Boomerang Wire was properly designed, tested and shown to be substantially equivalent to the identified predicate devices. {2}------------------------------------------------ Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP - 1 2006 Cardiva Medical, Inc. c/o Mr. Michael J. Billig Regulatory Consultant Experien Group, LLC 155 Moffett Park Drive, Suite A-101 Sunnyvale, California 94089 Re: K061075 Cardiva Medical Boomerang™ Wire System, Models 56 and 610 Regulation Number: 21 CFR 870.4450 Regulation Name: Vascular Clamp Regulatory Class: Class II (two) Product Code: DXC Dated: August 1, 2006 Received: August 3, 2006 Dear Mr. Billig: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Michael J. Billig Please be advised that FDA's issuance of a substantial equivalence determination does not mean r reaso de aar made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set OF R rate 077, lacems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. product railable. I vew you to begin marketing your device as described in your Section 510(k) I mo lower notification. The FDA finding of substantial equivalence of your device to a legally premated predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Elia Mallis Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Section IV Boomerang™ Wire 510(k) Notification # Indications for Use K061075 510(k) Number (if known): __ Device Name: Cardiva Medical Boomerang™ Wire Indications For Use: The Cardiva Medical Boomerang™ Wire is intended to promote hemostasis at The Cardiva Medical Doomerails arteriotomy sites as an adjunct to manual compression. The Boomerang™ Wire atterformy sites as an acjents undergoing diagnostic femoral artery 15 microation procedures, using 5 or 6 Fr introducer sheaths. Prescription Use _X (Part 21 CFR 801 Subpart D) OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) edmme (Division Sian Off) Division of Cardiovascular Devices KO61075 510(k) Number
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...