Geomed Vascular Dilators

K183438 · Geomed Medizin-Technik GmbH & Co. · DWP · Sep 3, 2019 · Cardiovascular

Device Facts

Record IDK183438
Device NameGeomed Vascular Dilators
ApplicantGeomed Medizin-Technik GmbH & Co.
Product CodeDWP · Cardiovascular
Decision DateSep 3, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4475
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Geomed Vascular Dilators are devices used to enlarge or calibrate vessels during coronary artery bypass or angioplasty procedures. They are designed to locate orifices, to trace the course of abnormal vessels, and to perform various maneuvers of dilation and measurement of annulus and lumen diameters.

Device Story

Geomed Vascular Dilators are reusable stainless-steel surgical instruments used by surgeons to enlarge or calibrate blood vessels during coronary artery bypass or angioplasty. The device consists of a tapered tip connected to a handle by a wire. Surgeons manually manipulate the dilator to locate orifices, trace abnormal vessels, and measure lumen or annulus diameters. The device is provided non-sterile and requires steam sterilization by the user before clinical application. It is intended for transient contact (<15 minutes) with blood vessels.

Clinical Evidence

Bench testing only. Testing included tensile strength of tip-to-wire and handle-to-wire bonds, dimensional verification, and validation of cleaning, drying, and steam sterilization processes per FDA guidance.

Technological Characteristics

Reusable stainless steel (ASTM F 899-12b) surgical instrument. Design features a tapered tip connected to a handle via wire. Available in Garrett (1.0-5.0 mm tip, 140/210 mm length) and DeBakey (1.0-10.0 mm tip, 190 mm length) configurations. Non-sterile; requires user-performed steam sterilization.

Indications for Use

Indicated for patients undergoing coronary artery bypass or angioplasty procedures requiring vessel enlargement, calibration, or measurement of annulus and lumen diameters.

Regulatory Classification

Identification

A surgical vessel dilator is a device used to enlarge or calibrate a vessel.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and below that, the word "ADMINISTRATION" in blue. September 18, 2019 Geomed Medizin-Technik GmbH & Co. % Angelika Scherp Regulatory Affairs Consultant Business Support International Amstel 320-1 1017 AP Amsterdam The Netherlands Re: K183438 Trade/Device Name: Geomed Vascular Dilators Regulation Number: 21 CFR 870.4475 Regulation Name: Surgical vessel dilator Regulatory Class: Class II Product Code: DWP Dated: July 26, 2019 Received: July 30, 2019 Dear Angelika Scherp: This letter corrects our substantially equivalent letter of September 3, 2019. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, ## Brian D. Pullin -S Brian Pullin, M.S. Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K183438 Device Name Geomed Vascular Dilators #### Indications for Use (Describe) Geomed Vascular Dilators are devices used to enlarge or calibrate vessels during coronary artery bypass or angioplasty procedures. They are designed to locate orifices, to trace the course of abnormal vessels, and to perform various maneuvers of dilation and measurement of annulus and lumen diameters. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| | <div> <span> <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | |--------------------------------------------------------------------------------------------------------------------| | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## K183438 # 5. 510(k) SUMMARY | Date: | September 1, 2019 | |-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter: | Name: Geomed Medizin-Technik GmbH & Co. KG<br>Address: Ludwigstaler Strasse 27<br>D-78532 Tuttlingen<br>Germany | | | Contact Person: Hanno Haug<br>Telephone: +49.7461.93550 | | Product: | Name of Device: Geomed Vascular Dilators<br>Common Name: Vascular Dilator<br>Classification Name: Surgical Vessel Dilator<br>CFR Section: 870.4475 | | Predicate Device: | K100518 – Instrumed Vessel Dilators<br>• This predicate has not been subject to a recall. | | Device Description: | Vascular Dilators are reusable stainless-steel surgical instruments that are used by qualified surgeons to enlarge or calibrate vessels. They are provided in two basic models and various sizes to accommodate individual needs, based on the anatomy of the site and the type of surgical procedure.<br>The device is provided non-sterile for steam sterilization by the user. | | Indications for Use: | Geomed Vascular Dilators are devices used to enlarge or calibrate vessels during coronary artery bypass or angioplasty procedures. They are designed to locate orifices, to trace the course of abnormal vessels, and to perform various maneuvers of dilation and measurement of annulus and lumen diameters. | | Technological Characteristics | The technological and performance characteristics of subject and predicate devices are identical, as shown by the following comparison table: | {4}------------------------------------------------ | Manufacturer | Geomed Medizin-Technik | Instrumed K100518 | |--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for Use | Geomed Vascular Dilators are<br>devices used to enlarge or calibrate<br>vessels during coronary artery<br>bypass or angioplasty procedures.<br>They are designed to locate<br>orifices, to trace the course of<br>abnormal vessels, and to perform<br>various maneuvers of dilation and<br>measurement of annulus and lumen<br>diameters. | INSTRUMED vessel dilators are<br>devices used to enlarge or calibrate<br>vessels during coronary artery<br>bypass or angioplasty procedures.<br>They are designed to locate orifices,<br>to trace the course of abnormal<br>vessels, and to perform various<br>maneuvers of dilation and<br>measurement of annulus and lumen<br>diameters. | | Design | Tapered tip connected to a handle<br>by wire. | Tapered tip connected to handle by<br>wire. | | Garrett Vascular Dilator | | | | Tip Diameter | 1.0-5.0 mm | 1.0-5.0 mm | | Length | 140 mm / 210 mm | 140 mm / 210 mm | | DeBakey Vascular Dilator | | | | Tip Diameter | 1.0-10.0 mm | 1.0-10.0 mm | | Length | 190 mm | 190 mm | | Material | ASTM F 899-12b Stainless Steel | ASTM F 899-07 Stainless Steel | | Patient Contact | Blood vessels, transient (<15 min.) | Blood vessels, transient (<15 min.) | | Sterility | Non-sterile | Non-sterile | | Reusable | Yes | Yes | The subject Geomed Vascular Dilators are very similar to the predicate device in terms of intended use and technology. Design verification and validation testing to support determination of substantial equivalence consisted of the following tests: - > Tensile strength testing of the tip-to-wire and handle-to-wire bonds, indicating that the bond strength is sufficient to withstand forces experienced during clinical use - 人 Verification of device dimensions Performance Testing: - Σ Validation testing of the recommended end user manual and automated cleaning procedures - Σ Validation testing of the recommended end user steam sterilization process - ∑ Full-cycle validation testing of drying time after sterilization Cleaning and sterilization validation testing was conducted in accordance with the recommendations outlined in FDA Guidance Document "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling" (03/17/2015). Acceptance criteria were met for all tests performed. The subject device Geomed Vascular Dilators has the same technological Conclusion: characteristics and indications for use as the predicate device. The information provided in this 510(k) submission, including results of nonclinical testing. indicates that Geomed Vascular Dilators are substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...