MEDTRONIC DLP ARTERIAL CANNULA WITH 3D TIP - 22 FR.

K013013 · Medtronic Cardiac Surgical Products · DWF · Dec 4, 2001 · Cardiovascular

Device Facts

Record IDK013013
Device NameMEDTRONIC DLP ARTERIAL CANNULA WITH 3D TIP - 22 FR.
ApplicantMedtronic Cardiac Surgical Products
Product CodeDWF · Cardiovascular
Decision DateDec 4, 2001
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4210
Device ClassClass 2
AttributesTherapeutic

Intended Use

These cannulae are intended for the perfusion of the ascending aorta during short-term (6 hours or less) cardiopulmonary bypass.

Device Story

Arterial cannula for cardiopulmonary bypass; 22 Fr size; features baffled tip design with side exit holes and scoop-shaped ports. Baffled tip promotes diffuse blood flow pattern compared to open-tip designs. Used in surgical settings by clinicians for short-term aortic perfusion. Device facilitates blood delivery during bypass; minimizes localized flow turbulence. Biocompatible materials; non-clinical testing confirms hemolytic properties comparable to predicate devices.

Clinical Evidence

Bench testing only. Biocompatibility testing conducted per ISO 10993-1 for externally communicating devices in contact with blood (<24 hours). In vitro dynamic hemolysis assessment performed using heparinized bovine blood in short circuits; results showed hemolysis rates comparable to comparator cannulae.

Technological Characteristics

Cardiovascular bypass vascular cannula; 22 Fr size. Features baffled tip with side exit holes and scoop-shaped ports. Biocompatible materials compliant with ISO 10993-1. Mechanical flow-diffusing design.

Indications for Use

Indicated for patients requiring perfusion of the ascending aorta during short-term (≤6 hours) cardiopulmonary bypass.

Regulatory Classification

Identification

A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # KU13013 DEC 0 4 2001 ## Summary of Safety and Effectiveness Information {1}------------------------------------------------ ## DEC 0 4 2001 # 510(k) Summary [As Required by 21 CFR 807.92] Submitter: Roger Brink Medtronic Cardiac Surgical Products 620 Watson SW Grand Rapids, MI 49504 | Telephone: | (616) 643-7337 | |------------|----------------| | Fax: | (616) 643-1044 | September 4, 2001 Date Summary Prepared: - Medtronic DLP Arterial Cannulae modified Trade Name of Device: with 3D Tip - 22 Fr. - Cardiovascular cannula Common Name of Device: - Classification Name of Device: "Cardiopulmonary bypass vascular cannula," Class II at 21 CFR 870.4210 Predicate Substantially Medtronic DLP Arterial Cannulae, Class II at Equivalent Devices: 21 CFR 870.4210, Curved Tip Models cleared under 510(k) Number K840002, and 24 Fr. 3D Tip Models cleared under 510(k) Number K000776. - The Medtronic DLP Arterial Cannulae with 3D Tip Description of Device: represent modified versions of existing Medtronic DLP Arterial Cannulae. The proposed change involves the incorporation of a baffled tip design into the 22 Fr. This baffled tip imparts a more diffuse cannula. pattern of blood flow exiting from the cannula tip than exists with the existing open tip cannula design. - Intended Use of Device: These cannulae are intended for the perfusion of the ascending aorta during short-term (6 hours or less) cardiopulmonary bypass. | Special 510(k) Notification | 22 Fr. Arterial Cannulae with 3D Tip | Confidential | |-------------------------------------|--------------------------------------|--------------| | Medtronic, Inc. | | | | Medtronic Cardiac Surgical Products | 9/4/01 | 33 | {2}------------------------------------------------ #### Comparison to Existing Predicate Devices The Medtronic DLP Arterial Cannulae with 3D Tip - 22 Fr. are substantially equivalent to existing Medtronic DLP Arterial Cannulae. The existing cannulae have been modified to include a baffled tip, which features side exit holes and scoop-shaped exit ports on the outside of the tip surface. The indications for use for both the existing and modified devices are identical, and the addition of a baffled tip feature does not represent a change to the fundamental scientific technology of the devices. #### Summary of Non-Clinical Performance Data Material biocompatibility testing was conducted in accordance with the ISO 10993-1 standard. Under this standard these cannulae are categorized as externally communicating devices in contact with circulating blood for a limited (<24 hour) contact duration. An in vitro assessment of the dynamic hemolytic properties of the cannulae was also performed. Short circuits containing 1000 ml of freshly collected, heparinized bovine blood were constructed. The blood was circulated through test circuits which included the 22 Fr. 3D Tip model and competitive comparator models. Rates of hemolysis that were comparable to those measured for the comparator cannulae were observed for the modified device as well. #### Conclusions of Non-Clinical Tests The results of the non-clinical tests summarized above support an assertion that the Medtronic DLP Arterial Cannula 22 Fr. modified with a 3D Tip are as safe and effective as the existing Medtronic DLP Arterial Cannulae. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image is a seal for the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. In the center of the seal is an abstract image of an eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC 0 4 2001 Mr. Roger W. Brink Director RA/QA Medtronic Cardiac Surgical Products 620 Watson SW Grand Rapids, MI 49504-6393 Re: K013013 Trade Name: Medtronic DLP Arterial Cannulae with 3D Tip - 22 Fr Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiovascular Bypass Vascular Catheter Regulatory Class: II (two) Product Code: DWF Dated: September 4, 2001 Received: September 7, 2001 Dear Mr. Brink: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Mr. Roger W. Brink Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4586. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, V. D. Teller Bu Bram Zuckerman, M.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Medtronic DLP Arterial Cannulae with 3D Tip – 22 Fr. Device Name: Indications for Use: These cannulae are intended for use in perfusion of the ascending These canndiao are internation for tess) cardiopulmonary bypass. This Indication for Use is identical to the predicate device. Medtronic DLP Arterial Cannulae with 3D Tip - 24 Fr. ### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE) Division of Cardiovascular & Respiratory Devices | 510(k) Number | K013015 | |---------------|---------| |---------------|---------| | Prescription Use (Per 21 CFR 801.109 | <div style="display:inline-block;">✓</div> | OR | Over-The-Counter-Use | |--------------------------------------|--------------------------------------------|----|----------------------| |--------------------------------------|--------------------------------------------|----|----------------------| | Special 510(k) Notification | 22 Fr. Arterial Cannulae with 3D Tip | Confidential | |-------------------------------------|--------------------------------------|--------------| | Medtronic, Inc. | | | | Medtronic Cardiac Surgical Products | 9/4/01 | 35 |
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...