K212027 · Medtronic, Inc. · MCW · Aug 24, 2021 · Cardiovascular
Device Facts
Record ID
K212027
Device Name
TurboHawk Plus Directional Atherectomy System
Applicant
Medtronic, Inc.
Product Code
MCW · Cardiovascular
Decision Date
Aug 24, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4875
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The TurboHawk Plus directional atherectomy system is intended for use in atherectomy of the peripheral vasculature. The TurboHawk Plus catheter is indicated for use in conjunction with the SpiderFX embolic protection device in the treatment of severely calcified lesions. The TurboHawk Plus catheter is not intended for use in the coronary, carotid, iliac, or renal vasculature.
Device Story
System consists of flexible catheter and battery-driven cutter driver; used for peripheral artery atherectomy. Catheter features distal cutting unit with rotating inner blade within tubular housing; tracks over 0.014-inch guidewire. Operator uses thumb switch on catheter to engage/disengage cutter while advancing across lesion; excised tissue captured/stored in device tip. Used in clinical setting by physicians. Device shaves occlusive material from artery; helps restore blood flow. Pairing with SpiderFX embolic protection device mitigates distal embolization risk during treatment of heavily calcified plaque.
Clinical Evidence
Bench testing only. Performance evaluation included device inspections, simulated use (trackability, rotational fatigue, cycling, cutting), kink resistance, heat generation, torsional/tensile strength, cut mass/pass efficiency, capture efficiency, embolization analysis, and coating integrity. Biocompatibility testing conducted per ISO 10993-1 for limited (<24 hour) circulating blood contact.
Technological Characteristics
Catheter-based atherectomy system; battery-powered cutter driver. Flexible shaft tracks over 0.014-inch guidewire. Distal cutting unit with rotating blade. Lubricious coating. Biocompatible materials per ISO 10993-1. Sterilization method consistent with predicate. Mechanical cutting principle.
Indications for Use
Indicated for atherectomy of peripheral vasculature to treat de novo and restenotic atherosclerotic calcified and non-calcified lesions in native peripheral arteries. Use with SpiderFX embolic protection device for severely calcified lesions. Contraindicated for coronary, carotid, iliac, or renal vasculature.
Regulatory Classification
Identification
An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)
Predicate Devices
TurboHawk Peripheral Plaque Excision System (K170191)
K161361 — HawkOne Directional Atherectomy System · Medtronic Vascular, Inc. · Oct 14, 2016
K111723 — TURBOHAWK PERIPHERAL PLAQUE EXCISION SYSTEM · Ev3, Inc. · Oct 27, 2011
K141801 — HAWKONE DIRECTIONAL ATHERECTOMY SYSTEM · Covidien, LLC · Oct 16, 2014
K093301 — TURBOHAWK PERIPHERAL PLAQUE EXCISION SYSTEM · Ev3, Inc. · Nov 6, 2009
K170191 — SilverHawk Peripheral Plaque Excision System, TurboHawk Peripheral Plaque Excision System · Medtronic Vascular, Inc. · Jun 16, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 24, 2021
Medtronic, Inc Jenny Andersen Senior Director of Regulatory Affairs 4600 Nathan Lane North Plymouth, Minnesota 55442
Re: K212027
Trade/Device Name: TurboHawk Plus Directional Atherectomy System Regulation Number: 21 CFR 870.4875 Regulation Name: Intraluminal Artery Stripper Regulatory Class: Class II Product Code: MCW Dated: June 29, 2021 Received: June 30, 2021
Dear Jenny Andersen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212027
Device Name
TurboHawk Plus Directional Atherectomy System
Indications for Use (Describe)
The TurboHawk Plus directional atherectomy system is intended for use in atherectomy of the peripheral vasculature. The TurboHawk Plus catheter is indicated for use in conjunction with the SpiderFX embolic protection device in the treatment of severely calcified lesions. The TurboHawk Plus catheter is not intended for use in the coronary, carotid, iliac, or renal vasculature.
Type of Use (Select one or both, as applicable)
| <span style="font-size:12pt;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------------------------------|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "Medtronic" in blue font. The font is sans-serif and the letters are all capitalized except for the "d", "t", "r", "o", "n", and "i". The "i" has a circular dot above it.
K212027
TurboHawk Plus Directional Atherectomy System 510(k) Summary
#### 510(k) Summary TurboHawk Plus™ Directional Atherectomy System
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 C.F.R § 807.92.
### 1. Submitter Information
| Applicant | Medtronic, Inc<br>710 Medtronic Parkway<br>Minneapolis, MN 55432<br>Tel: 763-514-4000 |
|----------------|------------------------------------------------------------------------------------------|
| Contact Person | Angela Huber<br>Regulatory Affairs Manager<br>612-704-2418<br>angela.huber@medtronic.com |
| Date Prepared | June 24, 2021 |
#### 2. Subject Device
| Device Trade Name | TurboHawk Plus Directional Atherectomy System |
|----------------------|-------------------------------------------------------------------|
| Device Common Name | Catheter, Peripheral, Atherectomy |
| Classification Name | Intraluminal Artery Stripper<br>21 CFR 870.4875, Product Code MCW |
| Classification Panel | Cardiovascular |
#### 3. Predicate Device
| Device Trade Name | TurboHawk™ Peripheral Plaque Excision System |
|-------------------|----------------------------------------------|
| 510(k) Number | K1701911 |
| Clearance Date | June 16, 2017 |
#### 4. Reference Device
| Device Trade Name | HawkOne Directional Atherectomy System | |
|-----------------------|----------------------------------------|------------------|
| 510(k) Number | K161361 | K141801 |
| 510(k) Clearance Date | October 14, 2016 | October 16, 2014 |
<sup>1</sup> Medtronic considers the TurboHawk Plus Directional Atherectomy System to be substantially equivalent to the currently marketed TurboHawk Peripheral Plaque Excision Systems, K170191. Previous 510k clearances include; K111723, K103618, K093301.
{4}------------------------------------------------
K212027
TurboHawk Plus Directional Atherectomy System 510(k) Summary
# 5. Device Description
The TurboHawk Plus Directional Atherectomy System (TurboHawk Plus Catheter and cutter driver) is designed for the treatment of de novo and restenotic atherosclerotic calcified and noncalcified lesions located in native peripheral arteries. When treating complex, hard, calcified lesions, pairing the TurboHawk Plus catheter with the SpiderFX embolic protection device mitigates risk of distal embolization that may be generated when heavily calcified plaque breaks down.
The TurboHawk Plus catheter consists of a flexible shaft designed to track with a 0.36 mm (0.014 in) guidewire. The distal end of the TurboHawk catheter is comprised of a small cutting unit with an inner blade that rotates within a tubular housing. The proximal end of the TurboHawk Plus catheter contains a connector and cutter positioning lever (thumb switch) designed to fit into the cutter driver is a battery-driven, internally powered device, designed to power the TurboHawk Plus directional atherectomy catheter.
The TurboHawk Plus directional atherectomy system has two switches: 1) the main power switch on the cutter driver and 2) the cutter positioning lever (thumb switch) on the TurboHawk Plus catheter. The main power switch on the cutter driver supplies power to the device when turned on. When the thumb switch is pulled proximally to the On position, the TurboHawk Plus catheter activates the drive shaft and the cutter. With the cutter engaged, the TurboHawk Plus catheter is slowly advanced across the lesion, shaving occlusive material from the artery. The excised tissue is captured and stored in the tip of the device. The cutting process is completed by advancing the TurboHawk Plus catheter thumb switch distally, deactivating the drive shaft and disengaging the cutter. When the TurboHawk Plus catheter thumb switch is fully advanced distally to the Off position, excised tissue is packed into the tip. This cutting sequence is repeated as necessary to achieve the desired degree of plaque excision.
### 6. Intended Use
The TurboHawk Plus is intended for treatment of de novo and restenotic atherosclerotic calcified and non-calcified lesions located in native peripheral arteries.
### 7. Indications for Use
The TurboHawk Plus directional atherectomy system is intended for use in atherectomy of the peripheral vasculature. The TurboHawk Plus catheter is indicated for use in conjunction with the SpiderFX embolic protection device in the treatment of severely calcified lesions. The TurboHawk Plus catheter is not intended for use in the coronary, carotid, iliac, or renal vasculature.
# 8. Comparison of Technological Characteristics
The TurboHawk Plus is comprised of HawkOne catheter (reference) combined with the TurboHawk cutter driver (predicate). The TurboHawk Plus device uses similar design and
{5}------------------------------------------------
#### K212027 TurboHawk Plus Directional Atherectomy System 510(k) Summary
materials as the predicate devices. Modifications have been made to the TurboHawk Plus device design, so the proposed device catheter can function with the predicate cutter driver. Additionally, modifications are made to allow for improved tracking, crossing, torqueability, radiopacity, as well as improved procedural efficiency with the pre-loaded distal flush tool when compared to the predicate.
TurboHawk Plus shares the following similarities to the predicate device and reference devices.
- Same intended use ●
- Similar indications for use ●
- Same fundamental scientific technology ●
- Same principle of operation
- Similar catheter and cutter driver materials ●
- Similar device dimensions and diameter of vessels
- Same lubricious coating
- Same sterility assurance level and method of sterilization
- Same packaging material and configuration .
In addition, device materials, and manufacturing site and methods and labeling are similar between the proposed and legally marketed predicate device and reference devices.
#### 9. Performance Testing Summary
To demonstrate substantial equivalence of the proposed TurboHawk Plus to the predicate device. the following performance tests were performed using internal Risk Analysis procedures:
- Device Inspections ●
- Simulated use (trackability, rotational fatigue, cycling and cutting) ●
- Kink resistance ●
- Heat generation ●
- Torsional Strength
- Tensile Strength
- Catheter to cutter driver interaction
- Cut Mass/Pass (Plaque Removal Efficiency) ●
- Capture Efficiency (Debris removal and collection)
- Cycle & Life
- Cutting embolization analysis
- Consistency of Tissue Removal ●
- Coating integrity ●
- Particulate evaluation ●
The results from these tests demonstrate that TurboHawk Plus meets the product performance specifications. The technological characteristics and performance criteria of the TurboHawk Plus
{6}------------------------------------------------
#### TurboHawk Plus Directional Atherectomy System 510(k) Summary K212027
devices are comparable to the predicate device and the proposed TurboHawk Plus device performs in a manner equivalent to the predicate device currently on the market.
# 10. Biocompatibility Testing
The biocompatibility evaluation for the TurboHawk Plus was conducted in accordance with the principles of the ISO 10993:2018 Part 1, "Evaluation and Testing within a risk management process ".
The biocompatibility testing included the following tests:
- Cytotoxicity
- Sensitization ●
- Irritation
- Acute Systemic Toxicity ●
- Pyrogenicity
- Haemocompatibility -Hemolysis ●
- Haemocompatibility -Complement Activation
- Haemocompatibility -Partial Thromboplastin Time (PTT)
- Haemocompatibility Thromboresistance ●
- Hemocompatibility Platelet and Leukocyte Count ●
The TurboHawk Plus is considered biocompatible for its intended use under ISO 10993-1 category: externally communicating device, circulating blood contact with limited (<24 hour) exposure.
### 11.Conclusions
Based on the same intended use and indications for use, similar technological characteristics, and safety and performance testing included in this submission, Medtronic considers the TurboHawk Plus Directional Atherectomy System to be substantially equivalent to the currently marketed TurboHawk Peripheral Plaque Excision Systems, K170191.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.