K973011 · Medtronic Bio-Medicus, Inc. · KFM · Feb 25, 1998 · Cardiovascular
Device Facts
Record ID
K973011
Device Name
BPX80 BIO-PUMP (BPX80)
Applicant
Medtronic Bio-Medicus, Inc.
Product Code
KFM · Cardiovascular
Decision Date
Feb 25, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4360
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Medtronic Bio-Medicus Bio-Pump® centrifugal blood pump is indicated for use only with the Medtronic Bio-Medicus Bio-Console to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to six hours). It is also indicated for use in extracorporeal circulatory support systems (for periods up to six hours) not requiring complete cardiopulmonary bypass (e.g., valvuloplasty, circulatory support during mitral valve reoperation, surgery of the vena cava or aorta, liver transplants etc).
Device Story
BPX-80 Bio-Pump is a single-use, disposable centrifugal blood pump for extracorporeal circulatory support. Device consists of polycarbonate housing, three smooth rotating cones, magnet, and bearing/shaft assembly. Operates by centrifugal force; venous blood gravity drains into inlet; rotating cones accelerate blood toward outlet. Pump speed (RPM) controlled by Medtronic Bio-Console. Used in clinical settings (e.g., OR) by trained medical professionals. Provides circulatory support for up to six hours. Benefits include controlled blood flow and pressure during bypass or surgical procedures. Available with optional heparin-bonded Carmeda Bioactive Surface.
Clinical Evidence
No clinical data. Evidence based on in-vitro bench testing and biocompatibility evaluation. Bench testing included hydraulic performance, priming volume, chemical resistance, environmental testing, Carmeda coating integrity, complement activation, and HIMA hemolysis testing. Biocompatibility testing performed per ISO 10993-1 and USP XXIII. Sterilization validated per ANSI/AAMI/ISO 11137 (gamma) and 11135 (EtO).
Technological Characteristics
Polycarbonate housing; three smooth rotating cones; magnet/bearing assembly. UV-cured adhesive bonding. 3/8" inlet/outlet ports. Optional Carmeda® Bioactive Surface. Sterilization: Gamma radiation (ANSI/AAMI/ISO 11137) or EtO gas (ANSI/AAMI/ISO 11135). Driven by external Bio-Console. Mechanical centrifugal pump principle.
Indications for Use
Indicated for patients requiring extracorporeal circulatory support during cardiopulmonary bypass or other procedures (e.g., valvuloplasty, mitral valve reoperation, vena cava/aorta surgery, liver transplants) for up to six hours.
Regulatory Classification
Identification
A nonroller-type cardiopulmonary and circulatory bypass blood pump is a prescription device that uses a method other than revolving rollers to pump the blood through an extracorporeal circuit for periods lasting less than 6 hours for the purpose of providing either:(i) Full or partial cardiopulmonary bypass (i.e., circuit includes an oxygenator) during open surgical procedures on the heart or great vessels; or(ii) Temporary circulatory bypass for diversion of flow around a planned disruption of the circulatory pathway necessary for open surgical procedures on the aorta or vena cava.
Special Controls
*Classification* —Class II (special controls). The special controls for this device are:(i) Non-clinical performance testing must perform as intended over the intended duration of use and demonstrate the following: Operating parameters, dynamic blood damage, heat generation, air entrapment, mechanical integrity, and durability/reliability;
(ii) The patient-contacting components of the device must be demonstrated to be biocompatible;
(iii) Sterility and shelf life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components; and
(iv) Labeling must include information regarding the duration of use, and a detailed summary of the device- and procedure-related complications pertinent to use of the device.
(b)
*Nonroller-type temporary ventricular support blood pump* —(1)*Identification.* A nonroller-type temporary ventricular support blood pump is a prescription device that uses any method resulting in blood propulsion to provide the temporary ventricular assistance required for support of the systemic and/or pulmonary circulations during periods when there is ongoing or anticipated hemodynamic instability due to immediately reversible alterations in ventricular myocardial function resulting from mechanical or physiologic causes. Duration of use would be less than 6 hours.(2)
*Classification.* Class III (premarket approval).(c)
*Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with FDA on or before September 8, 2015, for any nonroller-type temporary ventricular support blood pump that was in commercial distribution before May 28, 1976, or that has, on or before September 8, 2015, been found to be substantially equivalent to any nonroller-type temporary ventricular support blood pump that was in commercial distribution before May 28, 1976. Any other nonroller-type temporary ventricular support blood pump shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
Medtronic Bio-Medicus, Inc. - BP-80 Bio-Pump® (K852698)
{0}------------------------------------------------
# SUMMARY OF SAFETY AND EFFECTIVENESS
#### FEB 2 5 1998
## 510(k) SUMMARY
## COMPANY NAME AND CONTACT PERSON
August 12, 1997
Medtronic Bio-Medicus Inc. 9600 West 76th Street Eden Prairie, MN 55344 tel. (612)944-7784 fax (612)944-7557
Thomas K. Johnsen Product Regulations Manager
## DEVICE NAME
BPX-80 Bio-Pump®
#### COMMON NAME
Centrifugal Blood Pump
## CLASSIFICATION NAME
Non-roller type cardiopulmonary bypass blood pump (21 CFR - 870.4360)
## PREDICATE DEVICE OR LEGALLY MARKETED DEVICE
Medtronic Bio-Medicus, Inc. - BP-80 Bio-Pump® (K852698)
Medtronic Cardiopulmonary - CBBP-80 Bio-Pump® (K852698, K891687 and K896978)
{1}------------------------------------------------
## DEVICE DESCRIPTION
The BPX-80 Bio-Pump centrifugal blood pump is a single use, disposable, non-pyrogenic device designed to move blood through the extracorporeal circuit by centrifugal force created by smooth rotating cones. Venous blood gravity drains into the inlet port of the Bio-Pump". The smooth cones rotate in the polycarbonate housing. As the cones rotate, energy in the form of pressure and velocity is transferred from the cones to the blood is gently accelerated toward the outlet of the pump. The Bio-Pump moves the blood through the circuit at a desired pressure and flow rate by increasing the speed of the rotating cones. This is accomplished by adjusting the Bio-Console's RPMs (revolutions per minute). Arterial blood is returned to the patient from the extracorporeal circuit.
The BPX-80 Bio-Pump consists of a housing and backplate, three smooth surface cones, a magnet, and a bearing/shaft assembly. The plastic components of the BPX-80 Bio-Pump are molded from polycarbonate. The joints are bonded with an ultraviolet light cured adhesive. The inlet port (3/8") is located at the top center of the Bio-Pump", and the outlet port (3/8") is located tangentially on the outer diameter. Notches on the backplate allow the BPX-80 Bio-Pump® to be securely locked onto the pump receptacle of the Bio-Console . The BPX-80 Bio-Pump is designed to be used only with Medtronic Bio-Medicus Bio-Consoles®
The BPX-80 Bio-Pump® centrifugal blood pump can be sterilized by either gamma radiation or ethylene oxide (EtO) gas. The Bio-Pump is also available with a heparin bonded (nonleaching) Carmeda® Bioactive Surface.
## INTENDED USE
The Medtronic Bio-Medicus Bio-Pump® centrifugal blood pump is indicated for use only with the Medtronic Bio-Medicus Bio-Console to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to six hours). It is also indicated for use in extracorporeal circulatory support systems (for periods up to six hours) not requiring complete cardiopulmonary bypass (e.g., valvuloplasty, circulatory support during mitral valve reoperation, surgery of the vena cava or aorta, liver transplants etc). . 我在
Page 10
*
{2}------------------------------------------------
## TECHNOLOGICAL CHARACTERISTICS
The technological characteristics are identical to the BP-80 Bio-Pump®, with the exception of the following modifications; material change (acrylic to polycarbogate), dimensional change (internal dimensions of the port), and UV adhesive in place of solvent bond. The packaging material, sterilization methods/cycles, optional heparin bonding (Carmeda® Coating), drive magnet, bearing assembly, internal seal and component specifications are identical to the currently marketed BP-80 Bio-Pump®.
## SUMMARY OF PERFORMANCE DATA
#### In-vitro Bench Testing:
In-vitro bench testing demonstrated that when compared to the predicate devices (BP-80 Bio-Pump and CBBP-80 Bio-Pump"), the BPX-80 Bio-Pump and CBBPX-80 Bio-Pump do not significantly affect safety and effectiveness and are substantially equivalent to other commercially distributed centrifugal blood pumps. The in-vitro bench testing included analysis of:
- Hydraulic Performance and Priming Volume Tests .
- Chemical Resistance Tests .
- Environmental Tests .
- Carmeda® Coating Tests .
- Complement Activation .
- HIMA Hemolysis Testing �
#### Biocompatibility:
Biocompatibility testing of the BPX-80 Bio-Pump was performed in accordance with the FDA Blue Book Memorandum - #G95-1 and Biological Evaluation of Medical Devices Guidance - International Standard ISO 10993-1, and in accordance with United States Pharmacopoeia - XXIII.
Based on the results of the biocompatibility testing performed, the BPX-80 Bio-Pump was determined to be biocompatible and nontoxic and, therefore, safe for its intended use. Biocompatibility on the CBBPX-80 Bio-Pump has been substantiated by over 8 years of the Carmeda Bioactive Surface being applied to Cardiopulmonary bypass products, many of which have polycarbonate housing materials.
નદ
{3}------------------------------------------------
#### Sterilization:
Sterilization of the BPX-80 Bio-Pump® has been validated to assure a sterility assurance level (SAL) of 10°.
Gamma radiated Bio-Pumps® are sterilized in accordance with ANSI/AAMI/ISO 11137 - Method 1, Sterilization of Health Care Product Requirements for Validation and Routine Control - Radiation Sterilization.
EtO sterilized Bio-Pumps are sterilized in accordance with the American National Standards Institute, Inc. (ANSI) standard ANSI/AAMI/ISO 11135-1994 (Medical Devices - Validation and Routine Control of Ethylene Oxide Sterilization).
#### EtO Residuals:
EtO dissipation curves are used for routine product release to assure EtO sterilized Bio-Pumps meet the limits for residual concentrations of ethylene oxide (<25ppm), ethylene chlorohydrin (<25ppm), and ethylene glycol (<250ppm) as published in ANSI Standard Number ANSI/AAMI/ISO 10993-7 (Biological Evaluation of Medical Devices - Part 7: Ethylene Oxide Sterilization Residuals).
#### Pyrogens:
Routine Pyrogen Testing is performed using the Limulus Amebocyte Lysate (LAL) method. Product testing and release criteria (less than .5 EU/ml) is in accordance to the December 1987 Guideline issued by the Food and Drug Administration, Office of Compliance ("Guideline on Validation of the Limulus Amebocyte Lysate Test as an End-Product Endotoxin Test for Human and Animal Parenteral Drugs, Biological Products, and Medical Devices").
#### Conclusion
BPX-80 Bio-Pump® is substantially equivalent to the BP-80 Bio-Pump®. Performance, functional, and biocompatibility testing demonstrated that there are no new safety and effectiveness questions raised by the modifications made to the currently marketed BP-80 Bio-Pump®.
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows a logo for the U.S. Department of Health and Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized, curved shapes that resemble abstract human figures or waves, stacked on top of each other.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20856
FEB 2 5 1998
Mr. Thomas K. Johnsen Product Regulation Manager Medtronic Cardiac Surgery Medtronic Bio-Medicus, Inc. 9600 West 76th Street Eden Prairie, MN 55344
Re : K973011 BPX-80 Bio-Pump Centrifugal Blood Pump Regulatory Class: III Product Code: KFM Dated: November 25, 1997 November 28, 1997 Received:
Dear Mr. Johnsen:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the The general controls provisions of the Act include Act. requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
Page 2 - Thomas K. Johnsen
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html."
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
# INDICATIONS FOR USE
CONFIDENTIAL
K 973011 510(k) Number: __
BPX-80 Bio-Pump® Centrifugal Blood Pump Device Name:
#### Indications for use:
The Medtronic Bio-Medicus Bio-Pump centrifugal blood pump is indicated for use only with the Medtronic Bio-Medicus Bio-Console to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to six hours). It is also indicated for use in extracorporeal circulatory support systems (for periods up to six hours) not requiring complete cardiopulmonary bypass (e.g., valvuloplasty, circulatory support during mitral valve reoperation, surgery of the vena cava or aorta, liver transplants etc).
#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | K973011 |
|--------------------------------------|------------------------------------------|
| | (Division Sign-Off) |
| | Division of Cardiovascular, Respiratory, |
| | and Nemalogical Devices |
| Prescription Use (Per 21 CFR 801.109 | X |
| OR | Over-The-Counter-Use ________ |
(Optional Format 1-2-96)
| Medtronic Bio-Medicus, Inc. | 510(k) Notification - BPX-80 Bio-Pump® | Confidential |
|-----------------------------|----------------------------------------|--------------|
|-----------------------------|----------------------------------------|--------------|

August 12, 1997