K080824 · Medtronic Perfusion Systems · DWA · May 7, 2008 · Cardiovascular
Device Facts
Record ID
K080824
Device Name
BIO-CONSOLE 560
Applicant
Medtronic Perfusion Systems
Product Code
DWA · Cardiovascular
Decision Date
May 7, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4380
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Medtronic centrifugal blood pumping system is intended to pump blood through the extracorporeal bypass circuit for extracorporeal support for periods appropriate to cardiopulmonary bypass procedures (up to 6 hours).
Device Story
Bio-Console 560 is an extracorporeal blood pumping console; comprises base unit with display, touch screen user interface, and integrated flow control knob. System integrates with external drive motor, disposable centrifugal pump, flow transducer with disposable insert, and emergency handcrank. Used in clinical settings (e.g., OR) by perfusionists or clinicians to manage extracorporeal bypass circuits. Device transforms user inputs into pump speed control; monitors sensors and alarms; manages coast speed; ensures pump speed exceeds coast speed to maintain clamp opening. Output provides real-time flow control and system status monitoring to assist clinicians in maintaining hemodynamic support during bypass. Benefits include precise control of blood flow during cardiac procedures.
Clinical Evidence
Bench testing only. Software verification and validation testing confirmed that the device and its software-controlled functional characteristics are substantially equivalent to the predicate device. All test data satisfied documented product and performance specifications.
Technological Characteristics
Extracorporeal blood pumping console; base unit with touch screen interface and flow control knob. Compatible with external drive motor, disposable centrifugal pump, flow transducer, and emergency handcrank. Software-controlled sensor and alarm management. Class II device (21 CFR 870.4380).
Indications for Use
Indicated for patients requiring extracorporeal blood pumping support during cardiopulmonary bypass procedures for durations up to 6 hours.
Regulatory Classification
Identification
A cardiopulmonary bypass pump speed control is a device used that incorporates an electrical system or a mechanical system, or both, and is used to control the speed of blood pumps used in cardiopulmonary bypass surgery.
{0}------------------------------------------------
K080824
## 510(k) SUMMARY of Safety and Effectiveness
# MAY - 7 2008
#### l. Applicant Information:
| Date Prepared: | March 21, 2008 |
|-----------------------------------|------------------------------------------------------------|
| Submitter: | Medtronic, Inc. |
| Address: | 7611 Northland Drive<br>Brooklyn Park, MN 55428 |
| Establishment<br>Registration No. | 2184009 |
| Contact Person: | Preeti Jain<br>Director of Regulatory and Clinical Affairs |
| Telephone Number: | (763) 391-9533 |
| Fax Number: | (763) 391-9100 |
### II. Device Information:
.
| Trade Name: | Bio-Console® 560 |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Cardiopulmonary bypass pump speed control |
| Classification Name: | Control, Pump Speed, Cardiopulmonary Bypass |
| Classification: | Class II, 21 CFR 870.4380 |
| Product Code: | DWA |
| Predicate Device: | Medtronic Bio-Console® 560<br>510(k) No. K070286, Reg. No. 870.4380; Product Code: DWA |
| | Device Intended Use: The Medtronic centrifugal blood<br>pumping system is intended to pump blood through the<br>extracorporeal bypass circuit for extracorporeal support for<br>periods appropriate to cardiopulmonary bypass procedures (up<br>to 6 hours). |
{1}------------------------------------------------
- The Bio-Console® 560 is an extracorporeal blood pumping Device Description: console consisting of a Base Unit with a display and a User Interface (touch screen) with integrated flow control knob. In order to perform its intended function, the Bio-Console 560 is compatible with the Medtronic External Drive Motor, disposable Centrifugal Pump, Flow Transducer with disposable Insert, and Emergency Handcrank.
- The Medtronic centrifugal blood pumping system is intended to Intended Use: pump blood through the extracorporeal bypass circuit for extracorporeal support for periods appropriate to cardiopulmonary bypass procedures (up to 6 hours).
- Comparison to The Bio-Console 560 has the same intended use, hardware Predicate Device: design and basic software as the previously cleared Bio-Console 560. The only change to the device is the incorporation of some minor software modifications, which modify the way the Base Unit and User Interface control sensors and alarms, control the Coast Speed and ensure that the Pump Speed is greater than the Coast Speed to open the clamp.
- Software verification and validation testing confirms that the Test Data: function of the Bio-Console 560 and its software-controlled functional characteristics are substantially equivalent to the predicate device. All test data obtained satisfied the documented product and performance specifications.
- Based upon the technical information, intended use, and in Summary: vitro verification and validation information provided in previous pre-market notifications, the Bio-Console 560 addressed in this submission has been shown to be substantially equivalent to the currently marketed predicate device.
{2}------------------------------------------------
Public Health Service
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the three branches of government. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
ood and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
#### MAY - 7 2008
Medtronic, Inc. c/o Ms. Preeti Jain Director of Regulatory and Clinical Affairs 7601 Northland Drive Minneapolis, MN 55428
Re: K080824
Bio-Console® 560 Regulation Number: 21 CFR 870.4380 Regulation Name: Cardiopulmonary bypass pump speed control Regulatory Class: Class II (two) Product Code: DWA Dated: March 21, 2008 Received: March 24, 2008
Dear Ms. Jain:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Preeti Jain
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as sct forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Ouna R. Kirchner
/ Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): Ko80824
Bio-Console® 560 Device Name:
Indications for Use:
The Medtronic centrifugal blood pumping system is intended to pump blood through the extracorporeal bypass circuit for extracorporeal support for periods appropriate to cardiopulmonary bypass procedures (up to 6 hours).
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-the-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Dana E. Lechner
(Division Sign-Off) (Division Sign-On)
Division of Cardiovascular Devices
510(k) Number_Ko&ox24