MEDTRONIC RESTING HEART SYSTEM, MODEL AAR1000 AND 711

K031700 · Medtronic Perfusion Systems · DTN · Aug 29, 2003 · Cardiovascular

Device Facts

Record IDK031700
Device NameMEDTRONIC RESTING HEART SYSTEM, MODEL AAR1000 AND 711
ApplicantMedtronic Perfusion Systems
Product CodeDTN · Cardiovascular
Decision DateAug 29, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Medtronic Resting Heart System is intended for use in surgical procedures requiring extracorporeal gas exchange, circulatory support, and thermal regulation. This device is indicated for use in procedures requiring blood flow rate of 1 to 6 liters/min and lasting up to six hours. The system is indicated for use only with the Bio-Console.

Device Story

Medtronic Resting Heart System is a low-prime volume cardiopulmonary bypass (CPB) circuit for patient support during surgery (up to 6 hours). System includes a controller for active air removal and a disposable module. Venous return connects to a Venous Air Removal Device (VARD), which utilizes an arterial filter (38 micron) modified for the venous side, equipped with ultrasonic crystals to sense liquid levels. Automated air removal occurs via a vent port. System replaces traditional cardiotomy venous reservoirs or bags. Operated by perfusionists in surgical settings; requires Bio-Console. Output includes real-time air removal and flow management, assisting clinicians in maintaining extracorporeal circulation and reducing air-related risks during bypass.

Clinical Evidence

Bench testing only. Performance data provided for air handling capabilities, blood trauma, and pressure drops to demonstrate equivalence to standard extracorporeal circuits. Comprehensive testing completed on the Active Air Removal Controller, including performance and software verification and validation.

Technological Characteristics

Extracorporeal circuit components including oxygenator, heat exchanger, centrifugal blood pump, arterial filter, flow probe, tri-optic measurement cell, tubing, and connectors. VARD utilizes 38-micron arterial filter with ultrasonic crystals for liquid level sensing. Active air removal controller manages vent port. System is designed for 1-6 L/min flow rates. Sterilization and material standards not specified.

Indications for Use

Indicated for surgical patients requiring extracorporeal gas exchange, circulatory support, and thermal regulation during procedures with blood flow rates of 1-6 L/min for up to 6 hours. Must be used with Bio-Console.

Regulatory Classification

Identification

A cardiopulmonary bypass blood reservoir is a device used in conjunction with short-term extracorporeal circulation devices to hold a reserve supply of blood in the bypass circulation.

Special Controls

*Classification.* Class II (special controls), except that a reservoir that contains a defoamer or filter is classified into the same class as the defoamer or filter. The device, when it is a cardiopulmonary bypass blood reservoir that does not contain defoamers or blood filters, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ AUG 2 9 2003 Medtronic Resting Heart System 510(k) Submi # APPENDIX II 510(k) Summary Medtronic Resting Heart System (As required by 21CFR 807.92) ### Submitter Information A. | Submitter's Name: | Medtronic Perfusion Systems | |---------------------------|------------------------------------------| | Address: | 7611 Northland Drive N | | | Minneapolis, Minnesota 55428-1088 U.S.A. | | Telephone Number: | 763.391.9000 | | Contact Person: | Preeti Jain | | Date Submission Prepared: | May 30, 2003 | ### B. Device Information | Device Trade Name: | Medtronic Resting Heart System | |-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Common or usual Name: | Cardiopulmonary Bypass Surgery Circuit with Active Air<br>Removal, Oxygenator, Centrifugal blood Pump, Arterial<br>Filter, flow probe, tri-optic measurement cell, tubings and<br>connectors, holding bag. | | Classification Name: | For the new/changed components:<br>Cardiopulmonary bypass blood reservoir,<br>Cardiopulmonary bypass adaptors, fittings, manifold, | | | For the unchanged components: | | | Cardiopulmonary Bypass Oxygenator, | | | Cardiopulmonary bypass heat exchanger, | | | Cardiopulmonary Bypass blood pump (Non Roller type), | | | Cardiopulmonary bypass arterial line blood filter, | | | Cardiopulmonary bypass tubing, | | | Cardiopulmonary bypass inline gas sensor, | | | Cardiopulmonary bypass pump speed control | {1}------------------------------------------------ Predicate Devices(For new/changed components): | | 1. AFFINITY® Venous Reservoir Bag (K935717) | |----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | 2. Carmeda™ Coated Tubings and Connectors (K891687) | | | 3. Medtronic Overpressure/Vacuum Relief Valve<br>(K953564) | | | 4. CardioVentions CORx System (K012325) | | Device Description: | The Medtronic Resting Heart System is a low prime<br>volume alternative to existing Cardiopulmonary Bypass<br>(CPB) perfusion circuit. The system is used for patient<br>support during cardiopulmonary procedures lasting up to 6<br>hours. The System consists of the Medtronic Resting Heart<br>System Controller for Active Air Removal and the<br>Medtronic Resting Heart Disposable Module | | | The venous return portion of the CPB circuit is connected<br>to a Venous Air Removal Device (VARD) that is equipped<br>with an automated air removal system connected to the<br>vent port on the device. The VARD is an AFFINITY<br>Arterial Filter (38 micron) moved to the venous side of the<br>circuit and equipped with ultrasonic crystals that sense the<br>liquid level in the filter. The VARD replaces the Affinity™<br>Cardiotomy Venous Reservoir (CVR) or the Affinity™<br>Venous Reservoir Bag (VRB) that is normally used to<br>remove air from the venous blood entering the CPB circuit. | | Indications for Use: | The Medtronic Resting Heart System is intended for use<br>in surgical procedures requiring extracorporeal gas<br>exchange, circulatory support, and thermal regulation. This<br>device is indicated for use in procedures requiring blood<br>flow rate of 1 to 6 liters/min and lasting up to six hours.<br>The system is indicated for use only with the Bio-Console | ## C. Comparison of Required Technological Characteristics Medtronic Resting Heart System is a system comprised of standard components of an extracorporeal circuit for use during cardiopulmonary bypass. The system has the same technological characteristics as a traditional circuit and the single components have the same technological characteristics as the predicate. {2}------------------------------------------------ #### D. Performance Data Performance data, such as, Air Handling capabilities, blood trauma, pressure drops have been provided in the 510(k) submission to show equivalence of the Medtronic Resting Heart System to a standard extracorporeal circuit. In addition comprehensive testing has been completed on the Active Air Removal Controller including performance and Software Verification and Validation. ## Conclusion Medtronic Resting Heart System is substantially equivalent to the noted predicate device based on the similarities of technological characteristics, indications for use and the results of performance comparative testing. {3}------------------------------------------------ Public Health Service Image /page/3/Picture/2 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. In the center of the seal is an abstract image of an eagle with its wings spread. AUG 2 9 2003 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Medtronic Perfusion Systems c/o Ms. Preeti Jain 7611 Northland Drive Minneapolis, MN 55428 Re: K031700 Medtronic Resting Heart™ System Regulation Number: 21 CFR 870.4290, 870.4400 Regulation Name: Cardiopulmonary Bypass Blood Reservoir and Bypass Adaptor, Stopcock, Manifold, or Fitting Regulatory Class: Class II (two) Product Code: DTN, DTL Dated: May 30, 2003 Received: June 2, 2003 Dear Ms. Jain: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Ms. Preeti Jain Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Jhsa DAz fr Bram D. Zuckerma Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Medtronic Resting Heart System 510(k) Submission # APPENDIX IV # Indications for Use Statement | Page | 1 of 1 | |---------------------------|--------------------------------| | 510(k) Number (if known): | K031700 | | Device Name: | Medtronic Resting Heart System | Indications for Use: ﻜ Prescription Use (Per 21 CFR 801.109) The Medtronic Resting Heart System is intended for use in surgical procedures requiring extracorporeal gas exchange, circulatory support, and thermal regulation. This device is indicated for use in procedures requiring blood flow rate of 1 to 6 liters/min and lasting up to six hours. The system is indicated for use only with the Bio-Console. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) ______________________________________________________________________________________________________________________________________________________________________________ ______________________________________________________________________________________________________________________________________________________________________________ Concurrence of CDRH, Office of Device Evaluation (ODE) | (Division Sign-Off) | <i>[Signature]</i> | |------------------------------------|--------------------| | | <i>[Signature]</i> | | Division of Cardiovascular Devices | | | 510(k) Number | K031700 | | Prescription Use<br>(Per 21 CFR 801.109) | X | OR | Over-The-Counter Use<br>(Optional Format 1-2-96) | _ | |------------------------------------------|---|----|--------------------------------------------------|---| |------------------------------------------|---|----|--------------------------------------------------|---| N
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...