K210293 · AtriCure, Inc. · PZX · Mar 3, 2021 · General, Plastic Surgery
Device Facts
Record ID
K210293
Device Name
AtriClip LAA Exclusion System
Applicant
AtriCure, Inc.
Product Code
PZX · General, Plastic Surgery
Decision Date
Mar 3, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The AtriClip LAA Exclusion System is indicated for the exclusion of the heart's left atrial appendage, performed under direct visualization and in conjunction with other cardiac surgical procedures. Direct visualization, in this context, requires that the surgeon is able to see the heart directly, with or without assistance from a camera, endoscope, etc., or other appropriate viewing technologies.
Device Story
AtriClip LAA Exclusion System is a single-use, sterile, implantable clip preloaded on a disposable applier. Used by cardiac surgeons during open or minimally invasive cardiac procedures to exclude the left atrial appendage (LAA). The clip applies uniform pressure to the LAA, facilitating mechanical closure and electrical isolation. The system includes a handle, shaft, suture anchors, and articulation controls. The device is provided sterile via gamma irradiation. It is intended to reduce the risk of thrombus formation in the LAA. No software or electronic components are involved.
Clinical Evidence
No new clinical data was provided. The submission relies on previously cleared bench and animal data (Kamohara 2005, 2006; Starck 2012) which remain valid for the modified device.
Technological Characteristics
Implantable clip and applier; titanium clip components; polyethylene terephthalate suture material. Dimensions: 35mm, 40mm, 45mm, 50mm lengths. Sterilization: Gamma irradiation (10^-6 SAL). Biocompatibility assessed per ISO 10993. Non-pyrogenic. No software or electronic components.
Indications for Use
Indicated for patients undergoing cardiac surgical procedures requiring exclusion of the left atrial appendage (LAA) under direct visualization.
Regulatory Classification
Identification
An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 3, 2021
AtriCure, Inc. Mary Galeano Senior Regulatory Affairs Specialist 7555 Innovation Way Mason, Ohio 45040
Re: K210293
Trade/Device Name: AtriClip LAA Exclusion System Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable Clip Regulatory Class: Class II Product Code: PZX Dated: January 29, 2021 Received: February 2, 2021
Dear Mary Galeano:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Rachel Neubrander Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210293
### Device Name
AtriClip LAA Exclusion System with preloaded Gillinov-Cosgrove Clip, AtriClip LAA Exclusion System with preloaded V-Clip
### Indications for Use (Describe)
The AtriClip LAA Exclusion System is indicated for the exclusion of the heart's left atrial appendage, performed under direct visualization and in conjunction with other cardiac surgical procedures.
Direct visualization, in this context, requires that the surgeon is able to see the heart directly, with or without assistance from a camera, endoscope, etc., or other appropriate viewing technologies.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### l. Applicant Information
| Manufacturer: | AtriCure, Inc.<br>7555 Innovation Way<br>Mason, Ohio 45040 |
|--------------------|-----------------------------------------------------------------------------|
| Contact Person: | Mary Galeano, RAC<br>Sr. Regulatory Affairs Specialist<br>P: 513-257-6279 |
| Alternate Contact: | Jonathan McElwee, RAC<br>Sr. Manager, Regulatory Affairs<br>P: 513-795-9127 |
| Date Prepared: | 03/03/2021 |
#### ll. Device Information
| Proprietary Name: | AtriClip™ LAA Exclusion System with preloaded Gillinov-<br>Cosgrove Clip<br>(LAA0, ACH1, ACH2, PRO1, and PRO2)<br>AtriClip™ LAA Exclusion System with preloaded V-Clip<br>(PROV and ACHV) |
|-------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Implantable Clip and Clip Applier |
| Classification: | Implantable Clip and Clip Applier<br>Regulatory Class: Class II; per 21 CFR 878.4300<br>Product Code: PZX<br>Classification Panel: Cardiovascular |
| Predicate Device: | The device proposed for modification in this submission is the<br>AtriClip LAA Exclusion System cleared via K191413 on August 21<br>2019. |
#### III. Device Description
The AtriClip LAA Exclusion System consists of a single use, sterile, self-closing, implantable Clip preloaded on a Single Use Clip Applier along with a selection guide. When closed, the Clip applies uniform pressure over the length of the Clip to ensure consistent, reproducible, and secure exclusion of the left atrial appendage (LAA). Preclinical animal studies (Kamohara 2005,2006) demonstrate that complete exclusion with the Clip also results in acute and chronic electrical isolation of the LAA. A human clinical study (Starck 2012) has demonstrated acute electrical isolation. Chronic electrical isolation has not been evaluated in human clinical studies.
{4}------------------------------------------------
The Clip is available in the following lengths to accommodate difference sizes of LAA: 35 mm, 40 mm, 45 mm, and 50 mm. The Clip Appliers are disposable device with a handle, shaft, suture anchors, articulation controls and deployment loop which contains the Clip.
## Intended Use/ Indications for Use
The AtriClip LAA Exclusion System is indicated for the exclusion of the heart's left atrial appendage, performed under direct visualization and in conjunction with other cardiac surgical procedures.
Direct visualization, in this context, requires that the surgeon is able to see the heart directly, with or without assistance from a camera, endoscope, etc., or other appropriate viewing technologies.
### IV. Comparison of Technological Characteristics with the Predicate Device (AtriClip LAA Exclusion System - K191413)
| # | Feature | Previously Cleared AtriClip™ LAA<br>Exclusion System | Modified AtriClip™ LAA Exclusion<br>System | Equivalence<br>Comparison |
|------|---------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| 1 | Proprietary Name | AOD1: AtriClip LAA Exclusion System with<br>Preloaded Gillinov-Cosgrove Clip<br><br>AOD2: AtriClip LAA Exclusion System with<br>Preloaded PROV Clip | AOD1: AtriClip LAA Exclusion System with<br>Preloaded Gillinov-Cosgrove Clip<br><br>AOD2: AtriClip LAA Exclusion System with<br>PROV Clip | Same |
| | | 2 | Clip Applier<br>Product Code(s) | |
| 3 | Device<br>Classification | 21 CFR 878.4300,<br>product code: PZX | 21 CFR 878.4300,<br>product code: PZX | Same |
| 4 | Regulatory Class | Class II | Class II | Same |
| 5 | FDA Division | Cardiovascular | Cardiovascular | Same |
| 6 | Performance<br>Standards | No performance standards applicable to<br>this product have been promulgated by<br>FDA | No performance standards applicable to<br>this product have been promulgated by<br>FDA | Same |
| 7 | Intended Use | The AtriClip LAA Exclusion System is<br>indicated for the exclusion of the heart's left<br>atrial appendage, performed under direct<br>visualization and in conjunction with other<br>cardiac surgical procedures.<br>Direct visualization, in this context, requires<br>that the surgeon is able to see the heart<br>directly, with or without assistance from a<br>camera, endoscope, etc., or other<br>appropriate viewing technologies. | The AtriClip LAA Exclusion System is<br>indicated for the exclusion of the heart's left<br>atrial appendage, performed under direct<br>visualization and in conjunction with other<br>cardiac surgical procedures.<br>Direct visualization, in this context, requires<br>that the surgeon is able to see the heart<br>directly, with or without assistance from a<br>camera, endoscope, etc., or other<br>appropriate viewing technologies. | Same |
| # | Feature | Previously Cleared AtriClip™ LAA<br>Exclusion System | Modified AtriClip™ LAA Exclusion<br>System | Equivalence<br>Comparison |
| 8 | Device<br>Description in the<br>Instructions for<br>Use | The AtriClip LAA Exclusion System<br>contains the Gillinov-Cosgrove LAA Clip<br>(Clip) for exclusion of the heart's left atrial<br>appendage (LAA). Preclinical animal<br>studies (Kamohara 2005, 2006)<br>demonstrate that complete exclusion with<br>the Clip also results in acute and chronic<br>electrical isolation of the LAA. A human<br>clinical study (Starck 2012) has<br>demonstrated acute electrical isolation.<br>Chronic electrical isolation has not been<br>evaluated in human clinical studies.<br>The Clip is pre-loaded on a disposable<br>Clip applier. The AtriClip LAA Exclusion<br>System with preloaded Gillinov-Cosgrove<br>Clip does not contain natural rubber latex<br>components. | The AtriClip LAA Exclusion System<br>contains the Gillinov-Cosgrove LAA Clip<br>(Clip) for exclusion of the heart's left atrial<br>appendage (LAA). Preclinical animal<br>studies (Kamohara 2005, 2006)<br>demonstrate that complete exclusion with<br>the Clip also results in acute and chronic<br>electrical isolation of the LAA. A human<br>clinical study (Starck 2012) has<br>demonstrated acute electrical isolation.<br>Chronic electrical isolation has not been<br>evaluated in human clinical studies.<br>The Clip is pre-loaded on a disposable<br>Clip applier. The AtriClip LAA Exclusion<br>System with preloaded Gillinov-Cosgrove<br>Clip does not contain natural rubber latex<br>components. | Same |
| 9 | Software | This product does not contain software. | This product does not contain software. | Same |
| 10 | Shelf Life | 3 years. | 3 years. | Same |
| 11 | Biocompatibility | Passed. Materials have been assessed<br>based on ISO 10993 and are commonly<br>employed in tissue contacting devices. | Passed. Materials have been assessed<br>based on ISO 10993 and are commonly<br>employed in tissue contacting devices. | Same |
| 12 | Sterilization | Gamma Irradiation. | Gamma Irradiation. | Same |
| 13 | Sterility<br>Assurance Level | 10-6 SAL | 10-6 SAL | Same |
| 14 | Pyrogen | Non-Pyrogenic. | Non-Pyrogenic. | Same |
| Clip | | | | |
| 15 | Suture Material | Polyethylene Terephthalate | Polyethylene Terephthalate | Same |
| 16 | Preloaded Clip | AOD1: Preloaded Gillinov-Cosgrove Clip<br>AOD2: Preloaded V-Clip | AOD1: Preloaded Gillinov-Cosgrove Clip<br>AOD2: Preloaded V-Clip | Same |
| 17 | Clip Opening | AOD1: Opens 12.5mm ± 1.5mm as<br>measured between the beams of the<br>AOD1 clip.<br>AOD2: Opens AOD2 Clip to a minimum of<br>12mm at the distal end and 4mm at the<br>proximal end of the clip. | AOD1: Opens 12.5mm ± 1.5mm as<br>measured between the beams of the<br>AOD1 clip.<br>AOD2: Opens AOD2 Clip to a minimum of<br>12mm at the distal end and 4mm at the<br>proximal end of the clip. | Same |
| 19 | Biocompatibility | Passed. Materials have been assessed<br>based on ISO 10993 and are commonly<br>employed in tissue contacting devices. | Passed. Materials have been assessed<br>based on ISO 10993 and are commonly<br>employed in tissue contacting devices. | Same |
| 20 | Sterilization | Gamma Irradiation. | Gamma Irradiation. | Same |
AtriCure added an alternate raw material supplier of the titanium for the AOD1 clip.
{5}------------------------------------------------
- The devices have the same intended use, and;
- No changes were made in operating principle, or specifications of performance.
- · The contraindications, warnings, and precautions remain the same
- The results of the verification and validation testing:
- o Demonstrated equivalency in performance
- Device biocompatibility remains unchanged o
- Did not raise any new issues of safety o
{6}------------------------------------------------
#### V. Performance Data
A review of the risk analysis concluded there is no overall change to the risk profile of the proposed AtriClip LAA Exclusion System versus the previously cleared AtriClip LAA Exclusion System as the modifications to the proposed AtriClip LAA Exclusion System do not add or remove any features of the device or change the clinical application.
The proposed changes do not include any chanqe to design or performance specifications of the AtriClip LAA Exclusion System. Additionally, the proposed changes did not modify the intended use, therefore the previously submitted bench and animal data remain valid for the AtriClip LAA Exclusion System.
#### VI. Conclusions
AtriCure has demonstrated that the modifications made to the AtriClip LAA Exclusion System are substantially equivalent in fundamental design, technology, function, device materials, packaging, sterilization, operating principle, and intended use / indication for use as the previously cleared AtriClip LAA Exclusion System) per K191413.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.