MC3 QuickFlow Dual Lumen Catheter

K171610 · Mc3 Incorporated · PZS · Nov 16, 2017 · Cardiovascular

Device Facts

Record IDK171610
Device NameMC3 QuickFlow Dual Lumen Catheter
ApplicantMc3 Incorporated
Product CodePZS · Cardiovascular
Decision DateNov 16, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4100
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The MC3 26 Fr. QuickFlow™ Dual Lumen Catheter is a single use dual lumen catheter, which provides both venous drainage and reinfusion of blood via the femoral vein, that is indicated for use for up to 72 hours in patients with acute respiratory failure requiring partial Veno-Venous Extracorporeal Membrane Oxygenation (full hemodynamic support is not required and predicted flows needed to maintain adequate oxygenation do not exceed 31/min) where other available treatment options have failed and continued clinical deterioration is expected or the risk of death is imminent. The MC3 28 Fr. QuickFlow™ Dual Lumen Catheter is a single use dual lumen catheter, which provides both venous drainage and reinfusion of blood via the femoral vein, that is indicated for use for up to 72 hours in patients with acute respiratory failure requiring partial Veno-Venous Extracorporeal Membrane Oxygenation (full hemodynamic support is not required and predicted flows needed to maintain adequate oxygenation do not exceed 3.5l/min) where other available treatment options have failed and continued clinical deterioration is expected or the risk of death is imminent.

Device Story

MC3 QuickFlow Dual Lumen Catheter is a single-use, wire-reinforced, dual-lumen cannula for femoral vein access; provides simultaneous venous drainage and blood reinfusion during partial Veno-Venous Extracorporeal Membrane Oxygenation (VV ECMO). Device includes an introducer for wire-guided placement; features radiopaque markers for positioning. Used in clinical settings by physicians for patients with acute respiratory failure. Catheter design allows connection to extracorporeal circuits via barbed connectors. By facilitating extracorporeal gas exchange, the device supports patients failing conventional therapies, potentially preventing imminent death. Physician selects cannula size based on patient femoral vein size and cardiac output requirements.

Clinical Evidence

No clinical studies involving human patients were conducted. Substantial equivalence was demonstrated through bench testing and in vivo evaluations. Bench testing included simulated use durability, pressure/flow characteristics, kink resistance, tensile strength, leak/burst testing, and introducer testing. Hemolysis and recirculation testing were performed. In vivo thrombogenicity evaluation was conducted over a 72-hour duration.

Technological Characteristics

Dual-lumen, co-axial design; silicone and polyurethane construction; spring wire reinforcement. Radiopaque markers at tip and drainage holes. Includes introducer for 0.035"/0.038" guide wires. Barbed connectors for circuit integration. Biocompatible per ISO 10993-1:2009. Sterile, single-use. No software or electronic components.

Indications for Use

Indicated for patients with acute respiratory failure requiring partial Veno-Venous Extracorporeal Membrane Oxygenation (VV ECMO) for up to 72 hours, where other treatments failed and clinical deterioration or death is imminent. Requires partial support; full hemodynamic support not required. Flow limits: 3 L/min (26 Fr) or 3.5 L/min (28 Fr).

Regulatory Classification

Identification

An extracorporeal circuit and accessories for long-term respiratory/cardiopulmonary support (>6 hours) is a system of devices and accessories that provides assisted extracorporeal circulation and physiologic gas exchange of the patient's blood in patients with acute respiratory failure or acute cardiopulmonary failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent. The main devices and accessories of the system include, but are not limited to, the console (hardware), software, and disposables, including, but not limited to, an oxygenator, blood pump, heat exchanger, cannulae, tubing, filters, and other accessories (e.g., monitors, detectors, sensors, connectors).

Special Controls

*Classification* —Class II (special controls). The special controls for this device are:(1) The technological characteristics of the device must ensure that the geometry and design parameters are consistent with the intended use, and that the devices and accessories in the circuit are compatible; (2) The devices and accessories in the circuit must be demonstrated to be biocompatible; (3) Sterility and shelf-life testing must demonstrate the sterility of any patient-contacting devices and accessories in the circuit and the shelf life of these devices and accessories; (4) Non-clinical performance evaluation of the devices and accessories in the circuit must demonstrate substantial equivalence of the performance characteristics on the bench, mechanical integrity, electromagnetic compatibility (where applicable), software, durability, and reliability; (5) In vivo evaluation of the devices and accessories in the circuit must demonstrate their performance over the intended duration of use, including a detailed summary of the clinical evaluation pertinent to the use of the devices and accessories to demonstrate their effectiveness if a specific indication (patient population and/or condition) is identified; and (6) Labeling must include a detailed summary of the non-clinical and in vivo evaluations pertinent to use of the devices and accessories in the circuit and adequate instructions with respect to anticoagulation, circuit setup, performance characteristics with respect to compatibility among different devices and accessories in the circuit, and maintenance during a procedure.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. November 16, 2017 MC3 Incorporated Martha Rumford Director RA/QA 2555 Bishop Circle West Dexter, Michigan 48130 Re: K171610 Trade/Device Name: MC3 QuickFlow Dual Lumen Catheter Regulation Number: 21 CFR 870.4100 Regulation Name: Extracorporeal Circuit And Accessories For Long-Term Respiratory/Cardiopulmonary Failure Regulatory Class: Class II Product Code: PZS Dated: October 13, 2017 Received: October 16, 2017 Dear Martha Rumford: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); {1}------------------------------------------------ and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Nicole G. Ibrahim -S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Image /page/2/Picture/0 description: The image contains the logo for MC3 Cardiopulmonary. The logo consists of a red abstract symbol on the left, followed by the text "MC3" in a dark blue, bold font. Below "MC3" is the word "Cardiopulmonary" in a smaller, regular font, also in dark blue. The overall design is clean and professional. #### Section 4 - Indication for Use MC3 QuickFlow™ Dual Lumen Catheter | 510(k) Number: | K171610 | |----------------|------------------------------------| | Device Name: | MC3 QuickFlow™ Dual Lumen Catheter | Indications for Use: The MC3 26 Fr. QuickFlow™ Dual Lumen Catheter is a single use dual lumen catheter, which provides both venous drainage and reinfusion of blood via the femoral vein, that is indicated for use for up to 72 hours in patients with acute respiratory failure requiring partial Veno-Venous Extracorporeal Membrane Oxygenation (full hemodynamic support is not required and predicted flows needed to maintain adequate oxygenation do not exceed 31/min) where other available treatment options have failed and continued clinical deterioration is expected or the risk of death is imminent. The MC3 28 Fr. QuickFlow™ Dual Lumen Catheter is a single use dual lumen catheter, which provides both venous drainage and reinfusion of blood via the femoral vein, that is indicated for use for up to 72 hours in patients with acute respiratory failure requiring partial Veno-Venous Extracorporeal Membrane Oxygenation (full hemodynamic support is not required and predicted flows needed to maintain adequate oxygenation do not exceed 3.5l/min) where other available treatment options have failed and continued clinical deterioration is expected or the risk of death is imminent. | Prescription Use | X | |---------------------------|---| | (21 CFR 801 Subsection D) | | or | Over-The Counter Use | | |---------------------------|--| | (21 CFR 801 Subsection C) | | (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for MC3 Cardiopulmonary. The logo consists of a red abstract symbol on the left, followed by the text "MC3" in large, bold, dark blue letters. Below "MC3" is the word "Cardiopulmonary" in a smaller, lighter font, also in dark blue. # 510(k) Summary MC3 QuickFlow Dual Lumen Catheter Date Prepared: November 16, 2017 #### Sponsor Information: Owner/Applic ant/Submitter: MC3 Incorporated 2555 Bishop Circle West Dexter, MI 48130 1-734-995-9089 Registration number: 3011468686 Contact Person: Martha Rumford Director RA/QA 2555 Bishop Circle West Dexter, MI 48130 734.995.9089, Extension 232 # Device Names/Classification: | Device Trade Name: | MC3 QuickFlow Dual Lumen Catheter | |----------------------|---------------------------------------------------------------------------------------------| | Device Common Name: | Dual lumen ECMO cannula | | Classification Name: | Extracorporeal circuit and accessories for long-term<br>respiratory/cardiopulmonary failure | | Regulation Number: | 21 CFR 870.4100 | | Product Code: | PZS | | Predicate Device: | 81 FR 7451, Feb. 12, 2016 | Secondary Predicate Device: Avalon Elite Bi-Caval Dual Lumen Catheter (K081820) # Indications for Use: The MC3 26 Fr. QuickFlow™ Dual Lumen Catheter is a single use dual lumen catheter, which provides both venous drainage and reinfusion of blood via the femoral vein, that is indicated for use for up to 72 hours in patients with acute respiratory failure requiring partial Veno-Venous Extracorporeal Membrane Oxygenation (full hemodynamic support is not required and predicted flows needed to maintain adequate oxygenation do not exceed 31/min) where other available treatment options have failed and continued clinical deterioration is expected or the risk of death is imminent. The MC3 28 Fr. QuickFlow™ Dual Lumen Catheter is a single use dual lumen catheter, which provides both venous drainage and reinfusion of blood via the femoral vein, that is indicated for use for up to 72 hours in patients with acute respiratory failure requiring partial Veno-Venous Extracorporeal {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for MC3 Cardiopulmonary. The logo consists of a red abstract symbol on the left, followed by the text "MC3" in large, bold, dark blue letters. Below "MC3" is the word "Cardiopulmonary" in a smaller, lighter font, also in dark blue. The red symbol appears to be three curved shapes arranged in a circular pattern. Membrane Oxygenation (full hemodynamic support is not required and predicted flows needed to maintain adequate oxygenation do not exceed 3.5Vmin) where other available treatment options have failed and continued clinical deterioration is expected or the risk of death is imminent. # Device Description: The MC3 QuickFlow Dual Lumen Catheters are dual lumen catheters supplied with an introducer to facilitate wire guided placement into the vasculature by normal access techniques. The introducer is designed to follow a pre-positioned standard 0.038" (0.97 mm) or 0.035" (0.89 mm) guide wire (not included). The catheter is wire-reinforced for flexibility and kink-resistance and includes depth marks. Both the introducer and catheter are made of radiopaque materials and the catheter has radiopaque markers at the tip and most proximal set of drainage holes. The devices are comprised of two separate thin-walled catheters made of silicone and polyurethane polymer and reinforced with a spring wire. The dual lumen design involves two catheters of different diameters joined together in a co-axial configuration. An arrangement of side holes in the outer (drainage) catheter allows for egress of blood. An open tip on the inner (return) catheter allow s for blood ingress. Unreinforced tube extensions with barbed connectors, allowing clamping and connection to the extracorporeal circuit, are joined to the wye junction. A tapered tip introducer is included to facilitate placement of the catheter into the blood vessel over a standard guide wire (not included). Securement of the device may be accomplished via a groove designed in the Y-connector to accept a suture and a suture around a 3/8" barb connector. The cannula size and type is chosen by the physician based on the patient's femoral vein size and clinical needs, the following table may be used as an aid to choosing the cannula: | BSA (m²) | Expected cardiac<br>output (C0) l/min<br>(based on 2.5l/min/m²) | Minimum predicted<br>flow requirements for<br>adequate oxygenation<br>(l/min) (60% of<br>expected CO¹) | Recommended MC3<br>cannula | |----------|-----------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|----------------------------| | ≤ 1.6 | 4 | 2.4 | 26 or 28 Fr. | | 1.8 | 4.5 | 2.7 | 26 or 28 Fr. | | 2.0 | 5 | 3 | 26 or 28 Fr. | | 2.2 | 5.5 | 3.3 | 28 Fr. | - 1. Schmidt Met al. Intensive Care Med (2013) 39:838-846; DOI 10.1007/s00134-012-2785-8 In patients on VV ECMO who are hyperdynamic, single access cannulae may not be sufficient to achieve satisfactory gas exchange. #### Performance Evaluations: Clinical studies involving patients are not necessary to demonstrate the substantial equivalence of the subject devices. Performance assessments for substantial equivalence were accomplished through bench and in vivo studies that included the following evaluations: - Simulated Use Durability Testing - Pressure/Flow Characteristics - Kink Resistance Testing {5}------------------------------------------------ Image /page/5/Picture/0 description: The image contains the logo for MC3 Cardiopulmonary. The logo consists of a red abstract symbol on the left, resembling three curved shapes arranged in a circular pattern. To the right of the symbol is the text "MC3" in a bold, dark blue font. Below "MC3" is the word "Cardiopulmonary" in a smaller, lighter font, also in dark blue. - Tensile Strength Testing ● - Leak and Burst Testing - . Introducer Testing - Hemolysis Testing - . Recirculation Testing - In-vivo Testing ### Substantial Equivalence Comparison: Substantial equivalence analysis includes comparison to the special controls of FDA's final order, 81 FR 7451, Feb. 12, 2016, as well as comparison to the Secondary Predicate Device. This is due to the fact that there has been no device yet cleared under regulation 21 CFR 870.4100 to date. #### Predicate (Special Controls): The MC3 QuickFlow Dual Lumen Catheter meets all special controls required by 21 CFR 870.4100. Special Controls met are: - . Technological Characteristics: Geometry and design parameters are consistent with the devices intended use in extracorporeal life support procedures. The subject device is designed to be compatible with other extracorporeal circuit devices and accessories. - Biocompatibility: The subject device is demonstrated to be biocompatible up to 72 hours device in accordance with ISO 10993-1:2009 in accordance with GLP (21 CFR 58). - Sterility and Shelf-life: Testing demonstrates the sterility of the subject device as provided and that it maintains its sterility, integrity, durability, and reliability over the stated Shelf-life of the device. - Non-clinical Performance: Substantial equivalence is demonstrated by performance characteristics ● on the bench, mechanical integrity, durability, and reliability. - In vivo Evaluation: In vivo thrombogenicity evaluation demonstrates the subject device's ● performance over a 72-hour duration of use. - Labeling: The instructions for use includes a detailed summary of the non-clinical and in vivo evaluations pertinent to the device's use in an extracorporeal circuit. Adequate instructions are also included with respect to anticoagulation, circuit setup, maintenance during a procedure, and performance characteristics relevant to compatibility among different devices and accessories in the circuit. A 72-hour duration of use is specified in the Indication for Use. #### Secondary Predicate: A comparison of the MC3 QuickFlow Dual Lumen Catheter to the Secondary Predicate Device shows the following similarities: - Comparison of Intended Use: The intended use of both the subject and predicate devices includes veno-venous perfusion during extracorporeal life support procedures. {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for MC3 Cardiopulmonary. The logo consists of a red abstract symbol on the left, followed by the text "MC3" in large, bold, dark blue letters. Below "MC3" is the word "Cardiopulmonary" in a smaller, lighter font, also in dark blue. The red symbol appears to be three curved shapes arranged in a circular pattern. - Comparison of Labeling: The labeling that will be used for the MC3 QuickFlow Dual Lumen Catheter is similar to the labeling used with the Secondary Predicate Device. Labeling is designed to provide clarity for the user and satisfy current regulatory requirements. The instructions accurately present the directions necessary for the end-user to employ the device in extracorporeal life support procedures. - . Comparison of Operation and Technology: The MC3 OuickFlow Dual Lumen Catheters and the Secondary Predicate Device utilized the same general technologies and principles of operation. - Comparison of Design: With respect to the design of the MC3 QuickFlow Dual Lumen Catheters, there are no significant design differences from the Secondary Predicate Device. Each of the nonsignificant differences has been determined to have no adverse impact on performance. - . Comparison of Materials: The biocompatible materials of construction used in the MC3 QuickFlow Dual Lumen Catheters are the same generic materials that are used in the Secondary Predicate Device. - . Comparison of Performance: MC3 Incorporated has conducted performance studies with the MC3 QuickFlow Dual Lumen Catheters to ensure that these devices satisfy appropriate device performance specifications and satisfy user needs. There are no appreciable differences between the subject devices and the Secondary Predicate Device. # Conclusion: The information and data included in this 510(k) Notification demonstrate that the MC3 QuickFlow Dual Lumen Catheter is substantially equivalent to the "Predicate Device" for both venous drainage and reinfusion of blood via the femoral vein during extracorporeal life support procedures of up to 72 hours for acute respiratory or acute cardiopulmonary failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...