← Product Code [MJN](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/MJN) · K960715

# PRUITT-INAHARA CAROTID SHUNT 400-40-8F/INAHARA-PRUITT CAROTID SHUNT 500-50-8F (K960715)

_Ideas For Medicine, Inc. · MJN · Sep 9, 1996 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/MJN/K960715

## Device Facts

- **Applicant:** Ideas For Medicine, Inc.
- **Product Code:** [MJN](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/MJN.md)
- **Decision Date:** Sep 9, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.4450
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The intended use of these devices is to provide a temporary bypass during carotid endarterectomy.

## Device Story

Dual lumen balloon shunts; channel/divert blood in carotid artery during endarterectomy. Features: T-port for patency confirmation/back-bleeding; external safety balloon on inflation port. Safety balloon inflates at 17 psi to prevent overinflation of internal carotid balloon; reduces vessel damage risk. Used in clinical setting by surgeons. Provides temporary bypass; allows procedure in smaller arteries.

## Clinical Evidence

No clinical data; bench testing only. Materials certified to meet USP Class VI biocompatibility requirements.

## Technological Characteristics

Dual lumen balloon catheter; 8 French diameter. Materials identical to predicate models; USP Class VI biocompatible. Mechanical safety balloon inflation mechanism (17 psi limit). Supplied sterile and non-pyrogenic.

## Regulatory Identification

A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.

## Predicate Devices

- Pruitt-Inahara Shunts (9 French)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

Ideas for Medicine, Inc.
Pruitt-Inahara Carotid Shunt (Model 400-40-8F)
Inahara-Pruitt Carotid Shunt (Model 500-50-8F)
510(k) Notification

# SAFETY and EFFECTIVENESS SUMMARY

K960715

## PROPRIETARY NAME:

Pruitt-Inahara Carotid Shunt and
Inahara-Pruitt Carotid Shunt

SEP - 9 1996

## COMMON USUAL NAME:

Vascular Balloon Catheter

## CLASSIFICATION:

Class II

## CLASSIFICATION NUMBER:

Unknown

## DEVICE DESCRIPTION:

The Pruitt-Inahara Carotid Shunt (Model 400-40-8F) and the Inahara-Pruitt Carotid Shunt (Model 500-50-8F) are dual lumen balloon shunts which channel and divert blood in the carotid artery during endarterectomy. The 8 French shunts are for use in arteries too small to accommodate a 9 French shunt. Both the 400-40-8F and 500-50-8F have a T Port to confirm patency and allow back-bleeding after insertion. There is an external safety balloon on the inflation port for the balloon which lies in the internal carotid artery. The safety balloon is designed to inflate when the internal carotid balloon reaches 17 psi; this reduces the possibility of balloon overinflation and resultant vessel damage. All the materials are identical to those used in other models of Ideas for Medicine’s shunts. These materials have been used for years and were certified to meet the requirements of the USP Class VI testing for biocompatibility. Every lot is supplied sterile and non-pyrogenic.

## SUBSTANTIAL EQUIVALENT:

The intended use of these devices is to provide a temporary bypass during carotid endarterectomy. The 8 French Pruitt Shunts are identical in both materials and intended use to the 9 French Pruitt-Inahara Shunts. The smaller OD of the shaft will allow it to be used in smaller arteries.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/MJN/K960715](https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/MJN/K960715)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
