← Product Code [MJN](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/MJN) · K160067

# RenovoCath (K160067)

_Renovorx, Inc. · MJN · Mar 7, 2017 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/MJN/K160067

## Device Facts

- **Applicant:** Renovorx, Inc.
- **Product Code:** [MJN](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/MJN.md)
- **Decision Date:** Mar 7, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.4450
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The RenovoCath™ RC120 Catheter is intended for the isolation of blood flow and delivery of fluids, including diagnostic and/or therapeutic agents, to selected sites in the peripheral vascular system. The RenovoCath™ RC120 is also indicated for temporary vessel occlusions including arteriography, preoperative occlusion, and chemotherapeutic drug infusion. The RenovoCath™ RC120 is intended for general intravascular use in the peripheral vasculature in arteries 3mm and larger. The RenovoCath™ RC120 is intended for use in arteries from 3mm in diameter for yessel entry and to occlude vessels ranging between 4mm to 11mm in diameter. The diagnostic and/or therapeutic agents are to be used in accordance with specifications outlined by the respective agent manufacturer. The RenovoCath™ RC120 Catheter is not intended for use in coronary and intracranial arteries. The RenovoCath™ RC120 Catheter is not intended for embolic protection or as an aspiration catheter.

## Device Story

RenovoCath RC120 is a multi-lumen, dual-balloon catheter with adjustable distance between balloons. Used in peripheral vasculature to isolate blood flow and deliver fluids/agents. Operated by clinicians under fluoroscopic guidance; includes two 3cc syringes for precise balloon inflation. Device enables targeted delivery of therapeutic agents (e.g., chemotherapy) or diagnostic contrast media to specific vascular sites. Benefits include localized treatment delivery and temporary flow control. Not for coronary/intracranial use.

## Clinical Evidence

Bench testing only. Testing included biocompatibility, sterilization validation, shipping/packaging, accelerated aging, and design verification. Comparative performance testing against the predicate included balloon inflation time and infusion flow rate. Additional characterization included extractables/leachables profiling and SEM analysis of material changes.

## Technological Characteristics

Multi-lumen, dual-balloon catheter with adjustable balloon spacing. Mechanical inflation via 3cc syringes. Designed for peripheral vasculature (4-11mm vessel diameter).

## Regulatory Identification

A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.

## Predicate Devices

- Boston Scientific Equalizer™ Occlusion Balloon Catheter ([K140273](/device/K140273.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 7, 2017

RenovoRx, Inc. % Ronald S. Warren Sr. Director, Regulatory Affairs Experien Group, LLC 755 N. Mathilda Avenue, Suite 100 Sunnyvale, CA 94085

Re: K160067

Trade/Device Name: RenovoCath™ RC120 Catheter Regulation Number: 21 CFR 870.4450 Regulation Name: Vascular Clamp Regulatory Class: Class II Product Code: MJN Dated: February 23, 2017 Received: February 24, 2017

Dear Ronald S. Warren:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Fernando
Aguel-S
Fernando Agnel

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K160067

Device Name RenovoCath™ RC120 Catheter

#### Indications for Use (Describe)

The RenovoCath™ RC120 Catheter is intended for the isolation of blood flow and delivery of fluids, including diagnostic and/or therapeutic agents, to selected sites in the peripheral vascular system. The RenovoCath™ RC120 is also indicated for temporary vessel occlusions including arteriography, preoperative occlusion, and chemotherapeutic drug infusion. The RenovoCath™ RC120 is intended for general intravascular use in the peripheral vasculature in arteries 3mm and larger. The RenovoCath™ RC120 is intended for use in arteries from 3mm in diameter for yessel entry and to occlude vessels ranging between 4mm to 11mm in diameter.

The diagnostic and/or therapeutic agents are to be used in accordance with specifications outlined by the respective agent manufacturer.

The RenovoCath™ RC120 Catheter is not intended for use in coronary and intracranial arteries.

The RenovoCath™ RC120 Catheter is not intended for embolic protection or as an aspiration catheter.

| Type of Use (Select one or both, as applicable)                               |                                                                              |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span>    Prescription Use (Part 21 CFR 801 Subpart D)   </div> | <div> <span> </span>    Over-The-Counter Use (21 CFR 801 Subpart C)   </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(k) Notification K160067

#### GENERAL INFORMATION

#### Applicant:

RenovoRx, Inc. 4546 El Camino Real, Suite 282 Los Altos, CA 94022 USA Phone: 1-650-284-4433

## Contact Person:

Ronald S. Warren Regulatory Consultant for RenovoRx, Inc. Experien Group, LLC. 224 Airport Parkway, Suite 250 San Jose. CA 95110 U.S.A. Phone: 1-408-505-3926 FAX: 1-408-400-0865 Date Prepared: March 7, 2017

DEVICE INFORMATION

Trade/Proprietary Name: RenovoCath™ RC120 Catheter Generic/Common Name: Catheter, Intravascular Occluding, Temporary Classification: 21 CFR§870.4450, Vascular clamp Product Code: MJN, Catheter, Intravascular Occluding, Temporary

PREDICATE DEVICE(S)

Boston Scientific Equalizer™ Occlusion Balloon Catheter (K140273)

This predicate has not been subject to a design-related recall.

#### INDICATIONS FOR USE

The RenovoCath™ RC120 Catheter is intended for the isolation of blood flow and delivery of fluids, including diagnostic and/or therapeutic agents, to selected sites in the peripheral vascular system. The RenovoCath™ RC120 is also indicated for temporary vessel occlusion in applications including arteriography, preoperative occlusion, and chemotherapeutic drug infusion. The RenovoCath™ RC120 is intended for general intravascular use in the peripheral vasculature in arteries 3mm and larger. The RenovoCath™ RC120 is intended for use in arteries 

{4}------------------------------------------------

from 3mm in diameter for vessel entry and to occlude vessels ranging between 4mm to 11mm in diameter.

The diagnostic and/or therapeutic agents are to be used in accordance with specifications outlined by the respective agent manufacturer.

The RenovoCath™ RC120 Catheter is not intended for use in coronary and intracranial arteries.

The RenovoCath™ RC120 Catheter is not intended for embolic protection or as an aspiration catheter.

## PRODUCT DESCRIPTION

The RenovoCath™ RC120 Catheter ("RenovoCath") is a multi-lumen, dual-balloon catheter. The distance between the proximal and distal balloons is adjustable. The RenovoCath is provided with two off-the-shelf 3cc syringes that facilitate precise inflation of the balloons under fluoroscopic guidance.

## Summary of Intended Use and Technological Characteristics Comparison

The indications for use for the RenovoCath are substantially equivalent to the indications for use for the predicate device. Both devices have the same intended use and similar technological characteristics. In addition, the bench testing on the predicate device confirms that functionally the Equalizer Catheter and RenovoCath have similar design features to facilitate the delivery of fluids to the peripheral vasculature. Any differences in the technological characteristics between the devices do not raise any new issues of safety or effectiveness. Thus, the RenovoCath is substantially equivalent to the predicate device.

# TESTING IN SUPPORT OF SUBSTANTIAL EQUIVALENCE DETERMINATION

There are no design changes of the RenovoCath requiring new testing. Reference is made to previous bench testing conducted on the RenovoCath which supported the original 510(k) clearance of K141175. That testing included biocompatibility, sterilization validation, shipping and packaging, accelerated aging, and design verification testing. For purposes of establishing substantial equivalence to the Equalizer Occlusion Balloon Catheter, the predicate device was tested for the following using the same methods previously performed on RenovoCath:

- Balloon Inflation Time ●
- . Infusion Flow Rate

RenovoCath underwent additional performance and characterization testing including:

- . Determination of extractables/leachables profile after exposure to simulated use conditions (pre-conditioning to simulate expected clinical and infusate properties)
- Evaluation of material changes through functional testing and SEM analysis

The collective results of the testing demonstrate that the RenovoCath meets its specifications and performs as intended. In addition, the bench testing on the predicate device confirms that functionally the Equalizer Catheter and RenovoCath have similar design features to facilitate the delivery of fluids to the peripheral vasculature.

{5}------------------------------------------------

## CONCLUSION

The RenovoCath has the same intended use and similar technological characteristics as the Equalizer Occlusion Balloon Catheter predicate device. Functionally, the Equalizer Catheter and RenovoCath have similar design features to facilitate the delivery of fluids to the peripheral vasculature. On this basis, the RenovoCath is substantially equivalent to the predicate device and the indications for use have been updated to reflect the indications for use of the predicate device.

## SUMMARY

The RenovoCath is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/MJN/K160067](https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/MJN/K160067)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
