K160067 · Renovorx, Inc. · MJN · Mar 7, 2017 · Cardiovascular
Device Facts
Record ID
K160067
Device Name
RenovoCath
Applicant
Renovorx, Inc.
Product Code
MJN · Cardiovascular
Decision Date
Mar 7, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4450
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The RenovoCath™ RC120 Catheter is intended for the isolation of blood flow and delivery of fluids, including diagnostic and/or therapeutic agents, to selected sites in the peripheral vascular system. The RenovoCath™ RC120 is also indicated for temporary vessel occlusions including arteriography, preoperative occlusion, and chemotherapeutic drug infusion. The RenovoCath™ RC120 is intended for general intravascular use in the peripheral vasculature in arteries 3mm and larger. The RenovoCath™ RC120 is intended for use in arteries from 3mm in diameter for yessel entry and to occlude vessels ranging between 4mm to 11mm in diameter. The diagnostic and/or therapeutic agents are to be used in accordance with specifications outlined by the respective agent manufacturer. The RenovoCath™ RC120 Catheter is not intended for use in coronary and intracranial arteries. The RenovoCath™ RC120 Catheter is not intended for embolic protection or as an aspiration catheter.
Device Story
RenovoCath RC120 is a multi-lumen, dual-balloon catheter with adjustable distance between balloons. Used in peripheral vasculature to isolate blood flow and deliver fluids/agents. Operated by clinicians under fluoroscopic guidance; includes two 3cc syringes for precise balloon inflation. Device enables targeted delivery of therapeutic agents (e.g., chemotherapy) or diagnostic contrast media to specific vascular sites. Benefits include localized treatment delivery and temporary flow control. Not for coronary/intracranial use.
Clinical Evidence
Bench testing only. Testing included biocompatibility, sterilization validation, shipping/packaging, accelerated aging, and design verification. Comparative performance testing against the predicate included balloon inflation time and infusion flow rate. Additional characterization included extractables/leachables profiling and SEM analysis of material changes.
Technological Characteristics
Multi-lumen, dual-balloon catheter with adjustable balloon spacing. Mechanical inflation via 3cc syringes. Designed for peripheral vasculature (4-11mm vessel diameter).
Indications for Use
Indicated for isolation of blood flow and delivery of diagnostic/therapeutic fluids in peripheral vasculature (arteries ≥3mm). Used for temporary vessel occlusion (4mm-11mm diameter) for arteriography, preoperative occlusion, and chemotherapeutic infusion. Contraindicated for coronary and intracranial arteries; not for embolic protection or aspiration.
Regulatory Classification
Identification
A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.
Predicate Devices
Boston Scientific Equalizer™ Occlusion Balloon Catheter (K140273)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 7, 2017
RenovoRx, Inc. % Ronald S. Warren Sr. Director, Regulatory Affairs Experien Group, LLC 755 N. Mathilda Avenue, Suite 100 Sunnyvale, CA 94085
Re: K160067
Trade/Device Name: RenovoCath™ RC120 Catheter Regulation Number: 21 CFR 870.4450 Regulation Name: Vascular Clamp Regulatory Class: Class II Product Code: MJN Dated: February 23, 2017 Received: February 24, 2017
Dear Ronald S. Warren:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Fernando
Aguel-S
Fernando Agnel
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K160067
Device Name RenovoCath™ RC120 Catheter
#### Indications for Use (Describe)
The RenovoCath™ RC120 Catheter is intended for the isolation of blood flow and delivery of fluids, including diagnostic and/or therapeutic agents, to selected sites in the peripheral vascular system. The RenovoCath™ RC120 is also indicated for temporary vessel occlusions including arteriography, preoperative occlusion, and chemotherapeutic drug infusion. The RenovoCath™ RC120 is intended for general intravascular use in the peripheral vasculature in arteries 3mm and larger. The RenovoCath™ RC120 is intended for use in arteries from 3mm in diameter for yessel entry and to occlude vessels ranging between 4mm to 11mm in diameter.
The diagnostic and/or therapeutic agents are to be used in accordance with specifications outlined by the respective agent manufacturer.
The RenovoCath™ RC120 Catheter is not intended for use in coronary and intracranial arteries.
The RenovoCath™ RC120 Catheter is not intended for embolic protection or as an aspiration catheter.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Notification K160067
#### GENERAL INFORMATION
#### Applicant:
RenovoRx, Inc. 4546 El Camino Real, Suite 282 Los Altos, CA 94022 USA Phone: 1-650-284-4433
## Contact Person:
Ronald S. Warren Regulatory Consultant for RenovoRx, Inc. Experien Group, LLC. 224 Airport Parkway, Suite 250 San Jose. CA 95110 U.S.A. Phone: 1-408-505-3926 FAX: 1-408-400-0865 Date Prepared: March 7, 2017
DEVICE INFORMATION
Trade/Proprietary Name: RenovoCath™ RC120 Catheter Generic/Common Name: Catheter, Intravascular Occluding, Temporary Classification: 21 CFR§870.4450, Vascular clamp Product Code: MJN, Catheter, Intravascular Occluding, Temporary
PREDICATE DEVICE(S)
Boston Scientific Equalizer™ Occlusion Balloon Catheter (K140273)
This predicate has not been subject to a design-related recall.
#### INDICATIONS FOR USE
The RenovoCath™ RC120 Catheter is intended for the isolation of blood flow and delivery of fluids, including diagnostic and/or therapeutic agents, to selected sites in the peripheral vascular system. The RenovoCath™ RC120 is also indicated for temporary vessel occlusion in applications including arteriography, preoperative occlusion, and chemotherapeutic drug infusion. The RenovoCath™ RC120 is intended for general intravascular use in the peripheral vasculature in arteries 3mm and larger. The RenovoCath™ RC120 is intended for use in arteries
{4}------------------------------------------------
from 3mm in diameter for vessel entry and to occlude vessels ranging between 4mm to 11mm in diameter.
The diagnostic and/or therapeutic agents are to be used in accordance with specifications outlined by the respective agent manufacturer.
The RenovoCath™ RC120 Catheter is not intended for use in coronary and intracranial arteries.
The RenovoCath™ RC120 Catheter is not intended for embolic protection or as an aspiration catheter.
## PRODUCT DESCRIPTION
The RenovoCath™ RC120 Catheter ("RenovoCath") is a multi-lumen, dual-balloon catheter. The distance between the proximal and distal balloons is adjustable. The RenovoCath is provided with two off-the-shelf 3cc syringes that facilitate precise inflation of the balloons under fluoroscopic guidance.
## Summary of Intended Use and Technological Characteristics Comparison
The indications for use for the RenovoCath are substantially equivalent to the indications for use for the predicate device. Both devices have the same intended use and similar technological characteristics. In addition, the bench testing on the predicate device confirms that functionally the Equalizer Catheter and RenovoCath have similar design features to facilitate the delivery of fluids to the peripheral vasculature. Any differences in the technological characteristics between the devices do not raise any new issues of safety or effectiveness. Thus, the RenovoCath is substantially equivalent to the predicate device.
# TESTING IN SUPPORT OF SUBSTANTIAL EQUIVALENCE DETERMINATION
There are no design changes of the RenovoCath requiring new testing. Reference is made to previous bench testing conducted on the RenovoCath which supported the original 510(k) clearance of K141175. That testing included biocompatibility, sterilization validation, shipping and packaging, accelerated aging, and design verification testing. For purposes of establishing substantial equivalence to the Equalizer Occlusion Balloon Catheter, the predicate device was tested for the following using the same methods previously performed on RenovoCath:
- Balloon Inflation Time ●
- . Infusion Flow Rate
RenovoCath underwent additional performance and characterization testing including:
- . Determination of extractables/leachables profile after exposure to simulated use conditions (pre-conditioning to simulate expected clinical and infusate properties)
- Evaluation of material changes through functional testing and SEM analysis
The collective results of the testing demonstrate that the RenovoCath meets its specifications and performs as intended. In addition, the bench testing on the predicate device confirms that functionally the Equalizer Catheter and RenovoCath have similar design features to facilitate the delivery of fluids to the peripheral vasculature.
{5}------------------------------------------------
## CONCLUSION
The RenovoCath has the same intended use and similar technological characteristics as the Equalizer Occlusion Balloon Catheter predicate device. Functionally, the Equalizer Catheter and RenovoCath have similar design features to facilitate the delivery of fluids to the peripheral vasculature. On this basis, the RenovoCath is substantially equivalent to the predicate device and the indications for use have been updated to reflect the indications for use of the predicate device.
## SUMMARY
The RenovoCath is substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.