← Product Code [MJN](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/MJN) · K132990

# OCCLUSION BALLOON CATHETER (K132990)

_Boston Scientific Corp · MJN · Oct 22, 2013 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/MJN/K132990

## Device Facts

- **Applicant:** Boston Scientific Corp
- **Product Code:** [MJN](/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/MJN.md)
- **Decision Date:** Oct 22, 2013
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.4450
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Occlusion Balloon Catheters are indicated for temporary vessel occlusion in applications including arteriography, preoperative occlusion, emergency control of hemorrhage, chemotherapeutic drug infusion and renal opacification procedures. The Occlusion Balloon Catheter product line consists of two specific designs - Standard Occlusions Balloon Catheters and Berenstein™ Occlusion Balloon Catheter. Only the Berenstein Occlusion Balloon Catheter has been designed for coaxial delivery of small catheters or embolic agents.

## Device Story

Occlusion balloon catheter; compliant latex balloon mounted on distal tip of dual-lumen, radiopaque catheter shaft; proximal luer fittings. Used for temporary vessel occlusion during arteriography, preoperative procedures, hemorrhage control, drug infusion, or renal opacification. Physician-operated in clinical settings. Balloon inflation lumen allows expansion; central lumen facilitates guidewire passage, contrast medium infusion, or (Berenstein model) coaxial delivery of small catheters/embolic agents. Output is physical vessel occlusion; enables targeted therapy or diagnostic visualization. Benefits include controlled blood flow management and localized delivery of therapeutic agents.

## Clinical Evidence

Bench testing only. Testing included biocompatibility (cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, pyrogenicity, hemocompatibility, USP physicochemical tests) and in-vitro performance (deflated profile, tensile strength, inflated O.D., deflation time, multiple inflation, burst pressure, sheath compatibility).

## Technological Characteristics

Compliant latex balloon; dual-lumen radiopaque catheter shaft; proximal luer fittings. Mechanical occlusion principle. Sterilized device. No software or electronic components.

## Regulatory Identification

A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.

## Predicate Devices

- Occlusion Balloon Catheter ([K062202](/device/K062202.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K132990

Special 510(k) Submission
Occlusion Balloon Catheter

ﻘﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟ

. ......

## 510(k) Summary

### per 21 CFR §807.92

| Submitter's<br>Name and<br>Address                | Boston Scientific Corporation<br>One Scimed Place<br>Maple Grove, MN 55311                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|---------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Name<br>and<br>Information                | Diane Nelson<br>Regulatory Affairs Specialist<br>OCT 2 2 2013<br>Phone: 763-255-0813<br>763-494-2222<br>Fax:<br>e-mail: diane.nelson@bsci.com                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Date Prepared                                     | 21 October 2013                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Proprietary<br>Name                               | Occlusion Balloon Catheter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Common Name                                       | Vascular Clamp                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Product Code                                      | MJN - Catheter, Intravascular Occluding, Temporary                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Classification                                    | Class II, 21 CFR Part 870.4450 - Vascular Clamp                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Predicate<br>Device(s)                            | Occlusion Balloon Catheter<br>October 20, 2006<br>K062202                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Device<br>Description                             | The Occlusion Balloon Catheters are a compliant latex balloon mounted on<br>the distal tip of a dual lumen, radiopaque catheter shaft to which two luer<br>fittings are attached proximally. In addition to the balloon inflation lumen,<br>the central lumen is used to pass the catheter over the guidewire as well as<br>infusion of contrast medium, and in the case of the Berenstein™ Occlusion<br>Balloon Catheter, coaxial delivery of small catheters or embolic agents.                                                                                     |
| Intended Use/<br>Indications for<br>Use of Device | The Occlusion Balloon Catheters are indicated for temporary vessel<br>occlusion in applications including arteriography, preoperative occlusion,<br>emergency control of hemorrhage, chemotherapeutic drug infusion and<br>renal opacification procedures.<br>The Occlusion Balloon Catheter product line consists of two specific<br>designs - Standard Occlusions Balloon Catheters and Berenstein<br>Occlusion Balloon Catheter.<br>Only the Berenstein Occlusion Balloon Catheter has been designed for<br>coaxial delivery of small catheters or embolic agents. |
| Comparison of<br>Technological<br>Characteristics | The Occlusion Balloon Catheter will incorporate a substantially equivalent<br>design, packaging, fundamental technology, manufacturing, sterilization<br>and intended use as those featured in the predicate Occlusion Balloon<br>Catheter.                                                                                                                                                                                                                                                                                                                           |
| Performance<br>Data                               | Bench testing and biocompatibility testing were performed to support a<br>determination of substantial equivalence. The results of these tests provide<br>reasonable assurance that the proposed device has been designed and<br>tested to assure conformance to the requirements for its intended use. No<br>new safety or performance issues were raised during the testing and.<br>therefore, these devices may be considered substantially equivalent to the<br>predicate devices.                                                                                |

.

{1}------------------------------------------------

Special 510(k) Submission Occlusion Balloon Catheter

> The following biocompatibility and bench testing were completed on the Occlusion Balloon Catheter:

#### Biocompatibility

Cytotoxicity

Sensitization Intracutaneous Reactivity

Acute Systemic Toxicity

Materials Mediated Pyrogenicity

Hemocompatibility: Direct Hemolysis, Partial Thromboplastin Time (PTT),

Complement Activation, In Vitro Hemocompatibility

USP Physicochemical Tests for Plastics

The following in-vitro performance tests were completed on the Occlusion Balloon Catheter:

Bench

Deflated Balloon Profile Proximal Bond Tensile Inflated Balloon O.D. Balloon Deflation Time Multiple Inflation, Challenge Balloon Burst, Challenge Sheath Compatibility

Conclusion

Based on the Indications for Use, technological characteristics, safety and performance testing, the Occlusion Balloon Catheter has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the Occlusion Balloon Catheter (K062202 cleared October 20, 2006).

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.

October 22, 2013

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Boston Scientific Corporation C/O Diane Nelson, Regulatory Affairs Specialist One Scimed Place Maple Grove, MN 55311

Re: K132990

Trade/Device Name: Occlusion Balloon Catheter Regulation Number: 21 CFR 870.4450 Regulation Name: Catheter, Intravascular Occluding, Temporary Regulatory Class: Class II Product Code: MJN Dated: September 23, 2013 Received: September 24, 2013

Dear Ms. Neison:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Ms. Diane Nelson, Regulatory Affairs Specialist

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

# Bram D. 400 arman -S

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510(k) Number (if known): __K132990

Device Name: Occlusion Balloon Catheter

Indications for Use:

Occlusion Balloon Catheters are indicated for temporary vessel occlusion in applications including arteriography, preoperative occlusion, emergency control of hemorrhage, chemotherapeutic drug infusion and renal opacification procedures.

The Occlusion Balloon Catheter product line consists of two specific designs - Standard Occlusions Balloon Catheters and Berenstein™ Occlusion Balloon Catheter.

Only the Berenstein Occlusion Balloon Catheter has been designed for coaxial delivery of small catheters or embolic agents.

× Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Image /page/4/Picture/11 description: The image shows the text "Bram D. Zuckerman -S 2013.10.22 16:34:24 -04'00'". The text appears to be a timestamp or a record of some kind. The text is black and the background is white.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/MJN/K132990](https://fda.innolitics.com/submissions/CV/subpart-e%E2%80%94cardiovascular-surgical-devices/MJN/K132990)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
