ROTABLATOR ROTATIONAL ANGIOPLASTY SYSTEM WITH THE ROTALINK EXCHANGEABLE CATHETER
K993648 · Boston Scientific Corporation Northwest Technology · MCW · Nov 24, 1999 · Cardiovascular
Device Facts
Record ID
K993648
Device Name
ROTABLATOR ROTATIONAL ANGIOPLASTY SYSTEM WITH THE ROTALINK EXCHANGEABLE CATHETER
Applicant
Boston Scientific Corporation Northwest Technology
Product Code
MCW · Cardiovascular
Decision Date
Nov 24, 1999
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.4875
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The peripheral Rotablator system is intended for percutaneous use in peripheral vessels in patients with occlusive atherosclerotic disease who are acceptable candidates for bypass graft surgery or percutaneous transluminal angioplasty.
Device Story
Rotablator system performs rotational atherectomy to restore luminal patency in peripheral vessels. System components include advancer, catheter with diamond-coated burr, console, foot pedal, and guide wire. During operation, burr rotates at 140,000-190,000 RPM to ablate occlusive atherosclerotic material into fine particles; particles are carried distally and cleared via reticuloendothelial system. Used by physicians in clinical settings. Device provides mechanical intervention to treat peripheral arterial disease, offering an alternative to bypass surgery or standard angioplasty.
Clinical Evidence
Bench testing only.
Technological Characteristics
System comprises advancer, catheter with diamond-coated burr, console, foot pedal, and guide wire. Operates via high-speed rotational ablation (140,000-190,000 RPM). Materials and construction are similar to the predicate device.
Indications for Use
Indicated for percutaneous use in peripheral vessels in patients with occlusive atherosclerotic disease who are candidates for bypass graft surgery or percutaneous transluminal angioplasty.
Regulatory Classification
Identification
An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)
Predicate Devices
Peripheral Rotablator® Rotational Angioplasty System with the RotaLink® Exchangeable Catheter (K970296)
Related Devices
K121774 — ROTABLATOR ROTATIONAL ATHERECTOMY SYSTEM WITH PERIPHERAL ROTALINK PLUS · Boston Scientific Corp · Sep 13, 2012
K970296 — ROTABLATOR ROTATIONAL ANGIOPLASTY SYSTEM WITH THE ROTALINK EXCHANGEABLE CATHETER · Boston Scientific Corp · Aug 29, 1997
K220962 — ROTAPRO Rotational Atherectomy System · Boston Scientific Corporation · Sep 6, 2022
K212351 — Revolution Peripheral Atherectomy System · Rex Medical, L.P. · Dec 15, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Special 510(k): Device Modificatio Rotablator® Angioplasty System with RotaLink® Exchangeable Catheter for Peripheral Use
| 14 m . r . 1 . 1 . 1 . 1 . 1 | 11 - 1 - 1 - 1 - 1 - - - - f -- Darimbanal II -- |
|------------------------------|--------------------------------------------------|
| Section III - 510(k) Summary | NOV 24 1999 |
|------------------------------|-------------|
|------------------------------|-------------|
### General Information
| Submitter | Boston Scientific Corporation Northwest Technology<br>Center, Inc.<br>17425 N.E. Union Hill Road<br>Redmond, WA 98052 |
|----------------------|------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Cyndy Adams<br>425-556-1570 (phone)<br>425-558-1400 (fax) |
| Classification Name | Catheter, Peripheral, Atherectomy<br>(per 21 CFR 870.4875) |
| Common or Usual Name | Rotational Angioplasty System<br>Rotablator® Rotational Angioplasty System with the<br>RotaLink® Exchangeable Catheter |
| Classification Panel | Cardiovascular |
# Name of Predicate Device
Predicate Device 510(k) Reference No. K970296 Peripheral Rotablator® Rotational Angioplasty System with the RotaLink® Exchangeable Catheter
{1}------------------------------------------------
### Device Description
The Rotablator® Rotational Angioplasty system uses a high speed, rotating, diamondcoated burr to ablate occlusive material and restore luminal patency. The burr spins at 140.000-190.000 RPM and ablates material into very fine particles that are carried distally and removed via the reticuloendothelial system. The Rotablator System with Exchangeable Catheter comprises four main components: advancer, catheter with diamond-coated burr, console and foot pedal, and guide wire.
### Intended Use
The peripheral Rotablator system is intended for percutaneous use in peripheral vessels in patients with occlusive atherosclerotic disease who are acceptable candidates for bypass graft surgery or percutaneous transluminal angioplasty.
### Summary of Technological Characteristics
The proposed Rotablator system is similar in construction and materials to the currently marketed Rotablator system.
### Test Summary
The proposed Rotablator system is considered to be substantially equivalent to the previously marketed Rotablator system based on a comparison of the intended uses and designs and results of the testing and evaluations performed.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an emblem of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 2 4 1999
Ms. Cyndy Adams Sr. Regulatory Affairs Specialist Boston Scientific Corporation 17425 N.E. Union Hill Road Redmond, WA 98052-3376
Re: K993648
> Peripheral Rotablator® Rotational Angioplasty System Trade Name: With the RotaLink™ Exchangeable Catheter Regulatory Class: II Product Code: MCW Dated: October 28, 1999 Received: October 29, 1999
Dear Ms. Adams:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to A substantially equivalent determination assumes compliance with 895. the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements concerning action. your device in the Federal Register. Please note: this response to
{3}------------------------------------------------
Page 2 - Ms. Cyndy Adams
your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) Additionally, for questions on the promotion and advertising 594-4586. of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Ohto M. Mth for
Celia M. Witten, Ph.D., M.D. Acting Director Division of Cardiovascular, Respiratory and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### Section I - Indications for Use
510(k) Number (if known)
| Device Name | Peripheral Rotablator® Rotational Angioplasty System with the RotaLink™ Exchangeable Catheter |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The peripheral Rotablator system is intended for percutaneous use in peripheral vessels in patients with occlusive atherosclerotic disease who are acceptable candidates for bypass graft surgery or percutaneous transluminal angioplasty. |
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109) OR Over The Counter Use
(Division Sign-Off)
Division of Cardiovascular, Respiratory,
and Neurological Devices
510(k) Number K993648
(Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.