THROMBOLYTIC BRUSH CATHETER

K963925 · Micro Therapeutics, Inc. · MCW · Aug 8, 1997 · Cardiovascular

Device Facts

Record IDK963925
Device NameTHROMBOLYTIC BRUSH CATHETER
ApplicantMicro Therapeutics, Inc.
Product CodeMCW · Cardiovascular
Decision DateAug 8, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4875
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Micro Therapeutics Thrombolytic Brush Catheter is intended for percutaneous dissolution of acute thrombus (i.e., less than two weeks old) located in artificial arteriovenous (A-V) grafts. The Thrombolytic Brush Catheter is designed to augment the area of interface between clot and pharmacologic agent by simultaneous thrombolysis and clot maceration. Clinical studies demonstrate effective dissolution of thrombus in A-V grafts when this product is used in conjunction with urokinase. The Thrombolytic Brush Catheter is not intended for use in native vessels. The device should not be used on patients with a history of significant pulmonary disease or pulmonary hypertension.

Device Story

System consists of catheter with proximal Y-connector, soft nylon brush on flexible stainless steel drive cable, and hand-held battery-powered motor drive. Used in clinical setting by physicians for percutaneous treatment of thrombosed artificial A-V grafts. Device inserted into graft; motor rotates brush to macerate clot while simultaneously delivering pharmacologic agent (urokinase) through catheter. Increased surface area contact between clot and drug accelerates thrombolysis. Output is visual confirmation of thrombus dissolution under fluoroscopy. Benefits include reduced procedure time and lower drug dosage compared to pulse-spray infusion, facilitating rapid restoration of graft patency.

Clinical Evidence

Prospective randomized trial of 81 patients (45 treatment, 36 control) with thrombosed hemodialysis grafts. Primary endpoints: lytic procedure duration, residual thrombus, drug dosage, and graft patency. Results: Treatment group showed significantly shorter lytic procedure duration (17 vs 28 min, p=.0001), lower residual thrombus (6.5% vs 25.5%, p=.0001), and reduced urokinase/heparin requirements (p=.0001). Acute success (graft patent within 30 min) was 98% vs 43% (p=.0001). No significant difference in 4.5-month primary patency. Animal studies (GLP) confirmed safety and efficacy.

Technological Characteristics

Catheter features medical-grade plastic body, stainless steel drive cable, nylon brush tip, and platinum alloy radiopaque marker band. Battery-powered motor drive handle. System is percutaneous, fluoroscopically guided. Biocompatibility per ISO 10993-1. Sterilized product.

Indications for Use

Indicated for percutaneous dissolution of acute thrombus (<2 weeks old) in artificial A-V grafts. Contraindicated in patients with significant pulmonary disease or pulmonary hypertension. Not for use in native vessels.

Regulatory Classification

Identification

An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} Jul-22-97 13:42 MTI/MICRO THERAPEUTICS 1 714 361 0210 P.10 K963925 # 510(k) SUMMARY Micro Therapeutics, Inc. 1062-F Calle Negocio San Clemente, CA 92673 (714) 361-0616 AUG - 8 1997 ## Contact Person John L. Gehrich, Ph.D. Vice President of Regulatory &amp; Clinical Affairs ## Trade Name Thrombolytic Brush Catheter and Motor Drive ## Classification Name Graft Thrombectomy Catheter Infusion Catheter ## Substantially Equivalent Devices Fogarty® Graft Thrombectomy Catheter (Baxter Corp.) Mewissen Infusion Catheter (MediTech) Amplatz Thrombectomy Device (Microvena) ## Description The Micro Therapeutics Thrombolytic Brush Catheter is intended for the percutaneous dissolution of thrombus located in artificial arteriovenous (A-V) grafts. The Micro Therapeutics Thrombolytic Brush Catheter is designed to augment the area of interface between clot and pharmacologic agent by simultaneous thrombolysis and clot maceration. The integral system utilizes a catheter with proximal Y-connector, a soft nylon brush attached to a stainless steel flexible drive cable, and a hand-held battery-powered motor drive. ## Intended Use The Micro Therapeutics Thrombolytic Brush Catheter is intended for percutaneous dissolution of acute thrombus (i.e., less than two weeks old) located in artificial arteriovenous (A-V) grafts. The Thrombolytic Brush Catheter is designed to augment the area of interface between clot and pharmacologic agent by simultaneous thrombolysis and clot maceration. Clinical studies demonstrate effective dissolution of thrombus in A-V grafts when this product is used in conjunction with urokinase. The Thrombolytic Brush Catheter is not intended for use in native vessels. The device should not be used on patients with a history of significant pulmonary disease or pulmonary hypertension. ## Technological Characteristics This product is equivalent in *intended use* as well as *dimensional characteristics, composition* and *function* to the legally marketed Baxter Fogarty® Graft Thrombectomy Catheter, manufactured by Baxter Healthcare Corporation, as well as the MediTech Mewissen Infusion Catheter. This product is equivalent in *intended use* as well as *composition* and *function* to the legally marketed Amplatz Thrombectomy Device, (K954205) manufactured by Microvena, Corporation. {1} Jul-22-97 13:43 MTI/MICRO THERAPEUTICS 1 714 361 0210 P.11 # Summary of Studies ## Performance Data ### In Vitro Tests Sample devices were subjected to extensive physical bench testing. In vitro tests were conducted which included complete dimensional measurements, bristle and wire cable strength characterization, motor drive integrity testing, catheter flow rates, bond strengths, burst pressure and performance under simulated conditions. Additionally, electromagnetic and patient safety tests were conducted by an independent laboratory to evaluate the electromagnetic and leakage current potential of the battery operated motor drive handle. Based on the results from these tests, it was concluded that the design offered a considerable safety margin in critical areas and is suitable for its intended use. ## Radiopacity The radiopacity of the Thrombolytic Brush Catheter is comparable to other vascular catheters. The materials used in the manufacture of the Thrombolytic Brush Catheter are the same as the predicate devices listed in the 510(k), utilizing medical grade plastic, stainless steel and radiopaque positioning markers. The stainless steel cable drive and brush tip of the Thrombolytic Brush Catheter are adequately visualized under fluoroscopy and the catheter body has a platinum alloy marker band at the distal tip. Adequate radiopacity was demonstrated in both the animal trial and in the clinical study by the ability of all investigators to visualize the device under fluoroscopy. ## Biocompatibility Tests Tests for biocompatibility of materials for the Micro Therapeutics Thrombolytic Brush Catheter were performed to establish that the materials used in the device met the qualifications for short-term use in the vascular system in accordance with ISO 10993-1. Biocompatibility testing was performed on sterile product. In determining biocompatibility test design, testing was selected as deemed appropriate for the type of tissue/device interface and the duration of patient exposure. The results of these tests demonstrate the toxicological safety of the Thrombolytic Brush Catheter for its intended use. ### In Vivo Tests In vivo animal tests were performed to assess ease of use, suitable flexibility/stiffness required, safety and efficacy. Histologic examinations were performed both in the arterial and venous portion of the anastomosis as well as macro- and microscopic examination of sliced 3mm sections of lung. Animal studies were performed at the University of Illinois College of Medicine at Peoria in Peoria, Illinois. The study was approved by the Animal Research Committee of the above-mentioned institution and all studies were performed under GLP guidelines. Animal studies demonstrated the Micro Therapeutics Thrombolytic Brush Catheter to be a safe and efficacious device for percutaneous administration of pharmacologicals for dissolution of thrombus located in artificial A-V grafts. {2} Jul-22-97 13:43 MTI/MICRO THERAPEUTICS 1 714 361 0210 P.12 # Results of Clinical Trials Eighty-one patients with thrombosed artificial looped hemodialysis grafts were randomized at five institutions in the United States. Forty-five subjects were randomized to treatment with the Thrombolytic Brush Catheter and thirty-six were randomized to treatment with pulse-spray infusion. Initial success was defined as &lt;20% residual thrombus in the graft. Following the lysis procedure (Thrombolytic Brush Catheter or pulse-spray infusion) patients underwent thrombectomy to remove residual clot and balloon angioplasty to treat stenotic lesions. Patients were followed for 4½ months post procedure to determine primary graft patency (the ability to dialyze with no further interventions). The results of this study are detailed in the table below. | Parameter | Thrombolytic Brush Treatment Group | Pulse-Spray Infusion Control Group | Statistical Comparison | | --- | --- | --- | --- | | Number of patients enrolled | 45 | 36 | n/a | | Number of grafts treated | 43 | 35 | n/a | | Duration of lytic procedure | 17 minutes | 28 minutes | Significant (p=.0001) | | Residual thrombus after lytic procedure | 6.5% | 25.5% | Significant (p=.0001) | | Urokinase dose used | 215,435 Units | 455,882 Units | Significant (p=.0001) | | Heparin dose used | 2,570 Units | 4926 Units | Significant (p=.0001) | | Acute success of lytic procedure Graft patent within 30 minutes | 42/43 (98%) | 15/35 (43%) | Significant (p=.0001) | | Stenotic lesion visualized | 41/41 (100%) (No Secondary Intervention 2 patients) | 33/33 (100%) (No Secondary Intervention 2 patients) | Not Significant (p=1.000) | | Thrombectomy performed to remove residual thrombus | 14/41 (34%) | 14/33 (42%) | Not Significant (p=.481) | | PTA performed to treat stenotic lesion | 41/41 (100%) | 32/32 (100%) 1 Missing | Not Significant (p=1.000) | | Residual thrombus after thrombectomy and/or PTA | 2.2% | 1.6% | Not Significant (p=.6607) | | Duration of entire procedure | 70 minutes | 84 minutes | Not Significant (p=.1735) | | Procedural success after all interventions¹ Graft patent, no major complications | 39/43 (91%) | 34/35 (97%) | Not Significant (p=.621) | | Successful dialysis after all interventions | 34/39 (87%) | 30/34 (88%) | Lifetable analysis: Not significant¹ | | Primary Patency at 3 months² | 16/38 (42%) (No follow-up 1 patient) | 15/31 (48%) (No follow-up 3 patients) | Lifetable analysis: Not significant¹ | | Failures within 3 months (Rethrombosed/New Graft/Died/Transplant) | 22/38 (58%) (No follow-up 1 patient) | 16/31 (52%) (No follow-up 3 patients) | Lifetable analysis: Not significant¹ | | Primary Patency at 4½ months² | 14/38 (37%) (No follow-up 1 patient) | 14/30 (47%) (No follow-up 4 patients) | Lifetable analysis: Not significant¹ | | Failures within 4½ months (Rethrombosed/New Graft/Died/Transplant) | 24/38 (63%) (No follow-up 1 patient) | 16/30 (53%) (No follow-up 4 patients) | Lifetable analysis: Not significant¹ | ¹Procedure Failures (4 treatment group, 1 control group) were not included in the Lifetable analyses for primary patency calculations. ²Patients who rethrombosed, had new grafts placed, died or received transplants were considered failures for Primary Patency at the time of rethrombosis, death or intervention and at the subsequent follow-up interval(s). ³Lifetable analyses demonstrate no statistically significant difference between treatment and control for primary patency during the 4½ month follow-up period. **Conclusion:** Clinical data demonstrate the Thrombolytic Brush Catheter is a safe, effective and rapid alternative to pulse-spray thrombolysis for patients with thrombosed hemodialysis access grafts. {3} DEPARTMENT OF HEALTH &amp; HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20856 John L. Gehrich, Ph.D. Vice President, Regulatory and Clinical Affairs Micro Therapeutics, Inc. 1062-F Calle Negocio San Clemente, California 92673 AUG - 8 1997 Re: K963925 Thrombolytic Brush Catheter and Motor Drive Regulatory Class: II (two) Product Code: MCW Dated: May 20, 1997 Received: May 21, 1997 Dear Dr. Gehrich: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, {4} or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5} Jul-22-97 13:41 MTI/MICRO THERAPEUTICS 1 714 361 0210 P.09 510(k) Number (if known): Device Name: Micro Therapeutics Thrombolytic Brush Catheter and Motor Drive Indications for Use: The Micro Therapeutics Thrombolytic Brush Catheter is intended for percutaneous dissolution of acute thrombus (i.e., less than two weeks old) located in artificial arteriovenous (A-V) grafts. The Thrombolytic Brush Catheter is designed to augment the area of interface between clot and pharmacologic agent by simultaneous thrombolysis and clot maceration. Clinical studies demonstrate effective dissolution of thrombus in A-V grafts when this product is used in conjunction with urokinase. The Thrombolytic Brush Catheter is not intended for use in native vessels. The device should not be used on patients with a history of significant pulmonary disease or pulmonary hypertension. (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number K963925 (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of DCRH, Office of Device Evaluation (ODE) Prescription Use ☑ OR Over the Counter Use
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...