Diamondback 360 Peripheral Orbital Atherectomy System
K152694 · Cardiovascular Systems, Inc. · MCW · Oct 21, 2015 · Cardiovascular
Device Facts
Record ID
K152694
Device Name
Diamondback 360 Peripheral Orbital Atherectomy System
Applicant
Cardiovascular Systems, Inc.
Product Code
MCW · Cardiovascular
Decision Date
Oct 21, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.4875
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Diamondback 360 Peripheral Orbital Atherectomy System is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy. The OAS supports removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). The OAS is a percutaneous orbital atherectomy system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty.
Device Story
Percutaneous orbital atherectomy system for removal of stenotic material from peripheral arteries and AV dialysis fistulae. Device utilizes eccentrically rotating diamond-coated crown to ablate stenotic material into particles small enough for bodily absorption. System components include Orbital Atherectomy Device (OAD), atherectomy guide wire, saline infusion pump, and lubricant. Operated by clinicians in interventional settings. Modifications include reduced shaft diameter, updated crown design, and revised saline sheath/glue plug components. Output is physical removal of vessel wall obstructions, facilitating improved blood flow. Benefits include minimally invasive treatment of occlusive disease.
Clinical Evidence
Bench testing only. No clinical data presented. Testing included system life, stall, introducer compatibility, temperature, tensile, flexibility, torque, orbit, guide wire brake, saline flow, track, packaging/distribution, and biocompatibility (ISO 10993-5, ISO 10993-4, ISO 10993-18).
Technological Characteristics
Orbital atherectomy system; diamond-coated eccentrically rotating crown; saline-infused. Components: OAD, guide wire, saline infusion pump, lubricant. Biocompatibility per ISO 10993-5 (cytotoxicity), ISO 10993-4 (hemolysis), and ISO 10993-18 (chemical characterization).
Indications for Use
Indicated for patients with occlusive atherosclerotic disease in peripheral arteries or occluded hemodialysis grafts (AV shunts) who are candidates for percutaneous transluminal atherectomy or angioplasty.
Regulatory Classification
Identification
An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)
Diamondback 360 Peripheral Orbital Atherectomy System (K150732, K151260)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized human figure. The FDA logo is on the right and features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue and "ADMINISTRATION" in a smaller font size below.
October 01, 2021
Cardiovascular Systems Inc. Kim Wallner Senior Regulatory Affairs Specialist 1225 Old Highway 8 NW Saint Paul, Minnesota 55112
Re: K152694
Trade/Device Name: Diamondback 360 Peripheral Orbital Atherectomy System Regulation Number: 21 CFR 870.4875 Regulation Name: Intraluminal artery stripper Regulatory Class: Class II Product Code: MCW
Dear Kim Wallner:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated October 21, 2015. Specifically, FDA is updating this SE Letter because of a typo in the 510(k) Summary header, as an administrative correction.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Gregory O'Connell, OHT2: Office of Cardiovascular Devices, (301) 796-6075, gregory.oconnell@fda.hhs.gov.
Sincerely,
For
Sara M. Digitally signed by
Sara M. Royce -S
Date: 2021.10.01
Royce -S 13:43:30 -04'00'
11
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 21, 2015
Cardiovascular Systems, Inc. Ms. Kim Wallner Senior Regulatory Affairs Specialist 1225 Old Highway 8 NW Saint Paul, MN 55112
Re: K152694
Trade/Device Name: Diamondback 360 Peripheral Orbital Atherectomy System Regulation Number: 21 CFR 870.4875 Regulation Name: Intraluminal Artery Stripper Regulatory Class: Class II Product Code: MCW Dated: September 18, 2015 Received: September 21, 2015
Dear Ms. Wallner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K152694
Device Name
Diamondback 360 Peripheral Orbital Atherectomy Sytstem
Indications for Use (Describe)
The Diamondback 360 Peripheral Orbital Atherectomy System is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy.
The OAS supports removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). The OAS is a percutaneous orbital atherectomy system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-----------------------------------------------------------------------------------------|
| <span style="font-family: Arial;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
| Company Name: | Cardiovascular Systems, Inc.<br>1225 Old Highway 8 NW<br>Saint Paul, MN 55112 | | |
|----------------------------|-------------------------------------------------------------------------------|--|---------------------------------------------------------------------------------------------------------------|
| Contact: | Kim Wallner | | |
| Phone: | (651) 202-4917 | | |
| Fax: | (651) 305-7734 | | |
| Summary Date: | September 18, 2015 | | |
| Trade Name: | Diamondback 360® Peripheral Orbital Atherectomy System | | |
| Product Code: | MCW—Catheter, Peripheral, Atherectomy | | |
| Classification Regulation: | 21 CFR 870.4875—Intraluminal Artery Stripper | | |
| Classification: | II | | |
| Primary Predicate Device: | | | |
| | 510(k) Number:<br>Manufacturer:<br>Trade Name: | | K110389<br>Cardiovascular Systems, Inc.<br>Stealth 360® Orbital PAD System |
| Reference Devices: | | | |
| | 510(k) Numbers:<br>Manufacturer:<br>Trade Name: | | K122987, K131092, K151260<br>Cardiovascular Systems, Inc.<br>Stealth 360® Orbital PAD System |
| | 510(k) Number:<br>Manufacturer:<br>Trade Name: | | K150732, K151260<br>Cardiovascular Systems, Inc.<br>Diamondback 360® Peripheral Orbital<br>Atherectomy System |
{5}------------------------------------------------
## Description of Device
The Diamondback 360 Peripheral Orbital Atherectomy System (OAS) is intended for use in the treatment of peripheral arteries and A-V graft (shunt) stenosis.
The OAS provides a method of removing stenotic material from peripheral arteries and A-V grafts. The Diamondback 360 uses an eccentrically rotating sarface (crown) to remove stenotic material on the vessel wall. The stenotic particles that are removed are small enough to be absorbed by the body.
The Diamondback 360 Peripheral Orbital Atherectomy System consists of the following components:
- 1) Orbital Atherectomy Device (OAD)
- 2) Atherectomy Guide Wire (e.g., ViperWire Advance)
- 3) Saline Infusion Pump (SIP)
- 4) Atherectomy Lubricant (e.g., ViperSlide)
## Description of Change
The 1.50 Solid Peripheral Orbital Atherectomy Device was modified to reduce the diameter of the shaft (drive shaft, drive shaft tip bushing, and saline sheath) including modification of the crown for compatibility with the smaller diameter drive shaft. In addition, the saline sheath material and design for the 1.50 Solid, 2.00 Solid, 1.50 Classic and 2.00 Classic OAD configurations were changed and an updated glue plug component was added to accomodate the new saline sheath design.
#### Intended Use
The Diamondback 360 Peripheral Orbital Atherectomy System is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy.
The OAS supports removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). The OAS is a percutaneous orbital atherectomy system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty.
#### Technology
The Diamondback 360 Peripheral Orbital Atherectomy System provides a method of removing or reducing occlusive atherosclerotic or stenotic material. The OAS applies a diamond coated, eccentrically rotating surface to remove stenotic material on the vessel wall. The stenotic particles that are removed are small enough to be absorbed by the body. This is the same technology that was cleared to market for use in the peripheral arteries for Stealth 360
{6}------------------------------------------------
Orbital PAD System in 510(k) K110389, K122987, and K131092 and for Diamondback 360 Peripheral OAS in 510(k) K150732.
## Performance Data
The Diamondback 360 Peripheral Orbital Atherectomy System was evaluated using the following performance bench testing to confirm the performance characteristics as compared to the predicate device.
- System Life Testing including guide wire ●
- Stall Testing
- Introducer Compatibility Testing ●
- Temperature Testing ●
- Tensile Testing ●
- Flexibility Testing
- Delivered Torque Testing ●
- Orbit Testing
- Guide Wire Brake Testing
- Saline Flow Testing
- o Track Testing
- Packaging/Simulated Distribution Testing
- o Biocompatibility
- Cytotoxicity per ISO 10993-5 o
- ASTM Hemolysis per ISO 10993-4 O
- Chemical Characterization per ISO 10993-18 O
All test results demonstrate that the materials chosen, the manufacturing process, and the design utilized for the Diamondback 360 Peripheral Orbital Atherectomy System met the established specifications necessary for consistent performance during its intended use.
#### Conclusions
The Diamondback 360 Peripheral Orbital Atherectomy System met all predetermined acceptance criteria of design verification and validation testing as specified by applicable standards, test protocols, and/or customer inputs. Testing results demonstrate that the Diamondback 360 Peripheral Orbital Atherectomy System is substantially equivalent to the legally marketed predicate devices and does not raise any new safety or effectiveness questions.