2.4mm X 127 cm Phoenix Atherectomy Deflecting System
K151145 · Volcano Atheromed, Inc. · MCW · May 26, 2015 · Cardiovascular
Device Facts
Record ID
K151145
Device Name
2.4mm X 127 cm Phoenix Atherectomy Deflecting System
Applicant
Volcano Atheromed, Inc.
Product Code
MCW · Cardiovascular
Decision Date
May 26, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.4875
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Phoenix® Atherectomy System is in atherectomy of the peripheral vasculature. It is not intended for use in coronary, carotid, iliac or renal vasculature.
Device Story
Phoenix Atherectomy System is a sterile, single-use, over-the-wire peripheral atherectomy device. System comprises a flexible front-cutting catheter and a battery-powered handle. Device is used in peripheral vasculature to debulk plaque. During operation, the handle motor rotates the distal cutter; excised plaque is continuously captured and conveyed proximally through the catheter into an external collection reservoir. The 2.4mm model features tip deflection capabilities to assist in navigation. The device is operated by a clinician in a clinical setting. The physician uses the device to remove obstructive plaque, potentially improving blood flow in the peripheral arteries. The system includes a wire support clip for guidewire management.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included dimensional/visual inspection, catheter sweep/deflection verification, simulated use, comparative predicate testing in simulated lesions, cutter/shaft torque testing, tensile testing, drive train stress testing, stall testing, temperature rise, kink bend radius, guidewire/sheath compatibility, flow rate, corrosion resistance, large vessel debulking diameter, and packaging/shelf-life validation.
Technological Characteristics
Sterile, single-use, battery-powered atherectomy system. Components: Phoenix Catheter (multi-strand stainless steel torque shaft, Teflon sheath) and Phoenix Handle. 2.4mm model features tip deflection via slider-actuated pre-shaped Nitinol section. Rotational speed: 10,000-12,000 RPM. Guidewire compatibility: 0.014". Sterilization: Ethylene Oxide. Mechanical debulking and conveyance principle.
Indications for Use
Indicated for atherectomy of peripheral vasculature. Not for use in coronary, carotid, iliac, or renal vasculature.
Regulatory Classification
Identification
An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 26, 2015
Volcano AtheroMed, Inc. c/o Ms. Jean Chang Vice President, Operations 1455 Adams Drive, Suite 1120 Menlo Park, CA 94025
Re: K151145
Trade/Device Name: Phoenix Atherectomy System Regulation Number: 21 CFR 870.4875 Regulation Name: Intraluminal Artery Stripper Regulatory Class: Class II Product Code: MCW Dated: April 28, 2015 Received: April 29, 2015
Dear Ms. Chang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Kenneth J. Cavanaugh -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017 See PRA Statement below.
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
# K151145
Device Name
Phoenix® Atherectomy System
#### Indications for Use (Describe)
The Phoenix® Atherectomy System is in atherectomy of the peripheral vasculature. It is not intended for use in coronary, carotid, iliac or renal vasculature.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# Submitter Information:
| Date of 510(k) Summary Preparation: | April 28, 2015 |
|-------------------------------------|------------------------------------------------------------------------------------------|
| Name and Address of Manufacturer: | Volcano AtheroMed, Inc.<br>1455 Adams Dr.<br>Menlo Park, CA 94025 |
| Contact Person: | Jean Chang<br>Vice President, Operations<br>Phone: (650) 473-6846<br>Fax: (650) 473-9927 |
| Subject Device: | |
| Device Trade Name: | Phoenix® Atherectomy System |
| Common Name: | Peripheral Atherectomy Catheter |
| Regulation Description: | Intraluminal Artery Stripper |
| Regulation Number: | 21 CFR 870.4875 |
| Product Code: | MCW |
| Device Class: | Class II |
| Classification Panel: | Cardiovascular |
# Predicate Device:
Trade Name: 510(k) Number: Manufacturer:
Phoenix Atherectomy System K132682 & K140944 Volcano AtheroMed, Inc.
{4}------------------------------------------------
# Device Description:
The Phoenix Atherectomy System is a sterile, single-use device designed for atherectomy of the peripheral vasculature. The Phoenix Atherectomy System has two main components: the Phoenix Catheter and the Phoenix Handle.
The Phoenix Catheter is a flexible, over-the-wire (OTW), front-cutting Catheter that continuously captures and clears debulked plaque proximally through the Catheter and Handle into a collection reservoir that resides outside the patient. For use, the Phoenix Catheter is inserted into the Phoenix Handle incorporates a self-contained battery-powered motor designed to drive and rotate the cutter of the Phoenix Atherectomy Catheter at its specified rotational speed. The device is activated by an ON/OFF slider switch on the top of the Handle. An accessory, the Phoenix Wire Support Clip, is also supplied and can be used to clip a guidewire torque device in a fixed position relative to the Phoenix Handle. The Catheter, Handle, and Wire Support Clip are each packaged separately as sterile, single-use components of the Phoenix Atherectomy System.
There are multiple models of the Phoenix Catheter. The smaller 1.8mm and 2.2mm Phoenix Catheter models are compatible with 5F and 6F sheaths, respectively, and track directly over the guidewire with no tip deflection capability. The larger 2.4mm Phoenix Catheter is 7F with the tip having deflection capability. All Phoenix Catheter models use the same Phoenix Handle and are compatible with specific commercially available 0.014" exchange length (260 cm or greater) guidewires listed in the product's instructions for use.
This 510(k) includes modifications to the 2.4mm tip diameter Phoenix Deflecting Catheter distal cutter assembly, as well as the Catheter and proximal chassis design to optimize the ease of use, tip deflection capabilities and manufacturability of the device. Table 9-1 summarizes the subject modifications relative to the predicate devices.
# Indications for Use:
The Phoenix® Atherectomy System is intended for use in atherectomy of the peripheral vasculature. It is not intended for use in coronary, carotid, iliac or renal vasculature.
# Testing Summary:
To demonstrate the substantial equivalence of the modified 2.4mm Deflecting Phoenix Atherectomy System to the predicate 2.4mm Deflecting Phoenix Atherectomy System, the performance and technological characteristics were evaluated by completion of the following testing:
- Dimensional and Visual Inspection
- . Catheter Sweep & Deflection Span Verification
- Simulated Use
{5}------------------------------------------------
- Comparative Predicate Testing in Simulated Lesion .
- . Cutter Torque Chain Torque-to-Failure Test
- . Functional Outer Shaft Torque Test
- Outer Shaft Torque Test, Deflecting Catheter
- . Thrust Bearing Tensile Test
- Outer Shaft to Slider Tensile Test
- Catheter Drive Train Stress Test
- . Cutter Stall Test
- Temperature Rise of Catheter During Simulated Use
- Kink Bend Radius Test
- Guidewire Compatibility
- . Sheath Compatibility
- . Sheath Flow Rate
- . Corrosion Resistance
- . Large Vessel Debulking Diameter for Deflecting Catheter
- Packaging and Shelf Life
The results from this testing demonstrate that the performance and technological characteristics of the modified Phoenix 2.4mm Deflecting Atherectomy System meet defined design requirements and that the modified device performs in a manner equivalent to the predicate Phoenix 2.4mm Deflecting Atherectomy System with the identical intended use.
{6}------------------------------------------------
| Table 9-1: Summary of Technological Characteristics for the Modified Phoenix | | |
|------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Atherectomy System | | |
| Technological | Predicate Phoenix Atherectomy<br>System (K132682), | Modified Phoenix Atherectomy<br>System, (Subject Device), |
| Characteristic | 7F Phoenix Catheters (FG1179) | 7F Phoenix Catheters (FG1728) |
| Rotational Speed | 10,000-12,000 RPM | Identical |
| Guidewire Exchange | Over-the-wire | Identical |
| Guidewire Compatibility | 0.014" | Identical |
| Sheath Compatibility | 7F | Identical |
| Catheter Working Length | 130 cm | 127 cm Straight<br>117 cm Deflected |
| Catheter Torque Shaft | Multi-Strand Stainless Steel (SS) | Multi-Strand Stainless Steel (SS) |
| Catheter Inner Shaft | Stainless Steel Outer Shaft and<br>Teflon Sheath | Stainless Steel Outer Shaft and Thicker<br>Teflon sheath |
| Catheter Outer Shaft | Stainless Steel Outer Shaft and<br>Teflon sheath | Identical |
| Catheter Shaft Diameter | 2.2mm | Identical |
| Distal Cutter Flute<br>Maximum Diameter | 2.3mm | 2.4mm |
| Tip Diameter and<br>Crossing Profile | 2.4mm | Identical |
| Cutting Tip Port | None | Single exit port conveys excised debris<br>from the inner guidewire lumen into the<br>Distal Cutting Flute channel |
| Second Stage<br>Maceration within<br>Housing | Yes | Identical |
| Cutter Housing | Coating | No Coating |
| Distal Tip Assembly<br>Coating | Yes | Yes |
| Minimum Vessel Size for<br>Device Use | 3.0mm | Identical |
| Deflection span<br>(Side-to-side) | ≥ 5.5mm | ≥ 6.0mm |
| Deflection Mechanism | Knob Advances/Retracts Inner Shaft<br>to straighten or buckle inner Nitinol<br>tube for deflection | Slider Advances/Retracts the Outer<br>Shaft with the pre-shaped Nitinol<br>section at the distal end to deflect or<br>straighten the tip of the Inner Shaft<br>housing the cutter assembly |
| Debris Collection &<br>Removal | Continuous collection and removal of<br>excised debris by mechanical<br>conveyance | Identical |
| Steering (Directional)<br>mechanism | Rotation of knob steers distal tip and<br>cutter by torqueing Catheter Inner<br>Shaft | Identical Mechanism: Rotation of knob<br>steers distal tip and cutter by torqueing<br>Catheter Outer Shaft |
| Catheter Coating | No | Identical |
| Sterilization | Ethylene Oxide | Identical |
| Single-use only | Yes | Identical |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.