JETSTREAM G2 SYSTEM, MODELS PV20300 (JETSTREAM G2 CATHETER AND CONTROL POD); PVCN100 (PV CONSOLE)
K090325 · Pathway Medical Technologies, Inc. · MCW · Mar 11, 2009 · Cardiovascular
Device Facts
Record ID
K090325
Device Name
JETSTREAM G2 SYSTEM, MODELS PV20300 (JETSTREAM G2 CATHETER AND CONTROL POD); PVCN100 (PV CONSOLE)
Applicant
Pathway Medical Technologies, Inc.
Product Code
MCW · Cardiovascular
Decision Date
Mar 11, 2009
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.4875
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Jetstream G2TM System is intended for use in atherectorny of the peripheral vasculature and to break apart and remove thrombus from upper and lower extremity peripheral arteries ≥3.0 mm in diameter. It is not intended for use in coronary, carotid, iliac or renal vasculature.
Device Story
Jetstream G2 System is an atherectomy catheter system for debulking and treating peripheral vascular disease. System comprises a sterile, single-use catheter with an electrically driven, differentially cutting tip and a reusable console. Catheter features separate lumens for continuous aspiration of excised tissue/thrombus and infusion of sterile saline. Console includes two peristaltic pumps, power supply, system controller, and user interface. Operated by physicians in clinical settings; console remains outside sterile field. Device provides mechanical debulking and thrombus removal; aspiration removes debris to an external collection bag. Benefits include effective treatment of peripheral arterial disease through simultaneous cutting, aspiration, and infusion.
Clinical Evidence
Bench testing only.
Technological Characteristics
System consists of an electrically driven catheter with a differentially cutting tip and a console with two peristaltic pumps. Features include continuous aspiration and infusion lumens. Dimensions/form factor: catheter with control pod and compact console mountable on I.V. stand. Sterilization: sterile, single-use catheter. Connectivity: keypad interface and LED indicators for operational status.
Indications for Use
Indicated for atherectomy and thrombus removal in upper and lower extremity peripheral arteries ≥3.0 mm in diameter. Contraindicated for use in coronary, carotid, iliac, or renal vasculature.
Regulatory Classification
Identification
An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)
Predicate Devices
Jetstream™ Pathway PVTM Atherectomy System (K083489)
K093456 — JETSTREAM G3 SYSTEM · Pathway Medical Technologies, Inc. · Dec 1, 2009
K083489 — JETSTREAM PATHWAY PC ATHERECTOMY SYSTEM, MODEL OV10300 · Pathway Medical Technologies, Inc. · Jan 23, 2009
K101221 — JETSTREAM G3 SYSTEM, MODEL PV31300, PV CONSOLE, MODEL PVCN100 · Pathway Medical Technologies, Inc. · Jun 2, 2010
K100462 — JETSTREAM G3 L SYSTEM, PV CONSOLE, MODELS PV31400, PV31300, PVCN100 · Pathway Medical Technologies, Inc. · May 19, 2010
K083837 — JETSTREAM G2 SYSTEM, MODELS PV20300 (JETSTREAM G2 CATHETER AND CONTROL POD); PVCN100 (PV CONSOLE) · Pathway Medical Technologies, Inc. · Jan 13, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
K090325 page 1 of 2
Image /page/0/Picture/1 description: The image shows the logo for Pathway Medical Technologies. The logo consists of a circle on the left, with a white shape inside that resembles a winding path. To the right of the circle is the company name, "PATHWAY" in large, bold letters, with "MEDICAL TECHNOLOGIES" in smaller letters underneath.
MAR 1 1 2009
## 510(k) SUMMARY
### General Information:
Date of Summary Preparation:
February 6, 2009
Name and Address of Manufacturer:
Pathway Medical Technologies, Inc. 10801 120th Ave NE Kirkland, Washington 98033
Contact Person:
Trade Name:
Common Name:
Regulation Number:
Regulation Name:
Regulatory Class:
Product Code:
Predicate Devices:
Classification Panel:
Brit Baird Regulatory Affairs Specialist Phone: 425-636-4137 Fax: 425-636-4001
Jetstream G2TM System
Peripheral Atherectomy Catheter
21 CFR 870.4875
Intraluminal Artery Stripper
A
Class II
Cardiovascular
MCW
Manufacturer: Pathway Medical Technologies, Inc. (1) Jetstream™ Pathway PVTM Atherectomy System (K083489) (2) Jetstream G2TM System (K083837)
Indications for Use: The Jetstream G2TM System is intended for use in atherectorny of the peripheral vasculature and to break apart and remove thrombus from upper and lower extremity peripheral arteries ≥3.0 mm in diameter. It is not intended for use in coronary, carotid, iliac or renal vasculature.
Pathway Medical Technologies, Inc.
Appendix 4
{1}------------------------------------------------
Device Description: The Jetstream G2TM System is an atherectomy catheter system designed with an expandable cutting tip for use in debulking and treating vascular disease in the peripheral vasculature. Separate lumens within the Catheter allow for continuous aspiration and infusion during device use. Excised tissue, thrombus, and fluid are aspirated from the peripheral treatment site through a port in the Catheter tip to a collection bag located on the Console. The distal portion of the Catheter also possesses infusion ports that provide continuous infusion of sterile saline during the atherectorny procedure.
The Jetstream G2TM System consists of two primary components: (1) a Catheter and Control Pod and (2) a Console, which are packaged separately. Each of these system components is described generally as follows:
- . Jetstream G2TM Catheter and Control Pod: A sterile, single-use unit consisting of an electrically driven Catheter and Control Pod. The Catheter utilizes a differentially cutting tip and includes both aspiration and infusion capabilities. The Control Pod provides a user interface with keypad controls. The unit, its electrical connectors, tubing, and aspirant collection bag are packaged in a double-pouched tray.
- . PV Console: A reusable compact Console, with two (2) peristaltic pumps for aspiration and infusion, power supply, system controller, keypad interface, and LED indicators for device operational status. The Console mounts on a standard I.V. stand and remains outside the sterile field during the procedure.
This 510(k) is for the incorporation of the identical design modifications recently cleared under 510(k) K083837 (for the Jetstream G2 System) into the predicate Jetstream System cleared under 510(k) K083489.
Substantial Equivalence: The Jetstream G2 System is substantially equivalent to the specified predicate devices. The device has the identical indications for use and the same technological characteristics. Bench testing was completed and provided to support the safety and effectiveness of the modifications that were the subject of this 510(k).
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 1 1 2009
Pathway Medical Technologies, Inc. c/o Mr. Brit Baird Regulatory Affairs Specialist 10801 120th Ave NE Kirkland, WA 98033
Re: K090325
> Jetstream G2 System Regulation Number: 21 CFR 870.4875 Regulation Name: Intraluminal Artery Stripper Regulatory Class: Class II (two) Product Code: MCW Dated: February 6, 2009 Received: February 9, 2009
#### Dear Mr. Baird:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2 - Mr. Brit Baird
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Qimel
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# INDICATIONS FOR USE
510(k) Number (if known): `K090325
Device Name: Jetstream G2TM System
Indications for Use: The Jetstream G2TM System is intended for use in atherectorny of the peripheral vasculature and to break apart and remove thrombus from upper and lower extremity peripheral arteries ≥3.0 mm in diameter. It is not intended for use in coronary, carotid, iliac or renal vasculature.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Dminer R.v. Ames
(Division Sign-Off) Division of Cardiovascular Devices
510(k) Number________________________________________________________________________________________________________________________________________________________________
Pathway Medical Technologies, Inc. 510(k)
CONFIDENTIAL
Appendix 1 000016
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.