K033793 · Rex Medical · MCW · Dec 16, 2003 · Cardiovascular
Device Facts
Record ID
K033793
Device Name
CLEANER II ROTATIONAL THROMBECTOMY SYSTEM
Applicant
Rex Medical
Product Code
MCW · Cardiovascular
Decision Date
Dec 16, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4875
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Rex Medical Cleaner II Rotational Thrombectomy System permits mechanical declotting and removal of thrombi and emboli in synthetic dialysis grafts. The Embolectory balloon component, located on the Cleaner II Device, may be used for subsequent Embolectomy procedures to remove emboli and thrombi in the arterial system communicating with the synthetic dialysis graft
Device Story
Battery-operated, hand-held, wall-contacting rotational thrombectomy device; used for mechanical declotting of synthetic dialysis grafts. Features rotational wire with integrated soft distal tip to macerate clot into non-harmful particulate size; includes integrated embolectomy balloon for removal of macerated and wall-adherent thrombi. Operated by clinicians in a clinical setting. Device restores patency to occluded grafts; balloon component allows for additional thrombi/emboli removal in communicating arterial systems. Benefits include atraumatic mechanical removal of obstructions.
Clinical Evidence
Bench testing only. Performance testing conducted in accordance with ISO 10555-1, ISO 10555-4, and FDA guidance for intravascular catheters. Tests included stall torque, tensile strength, valve compression, contrast infusion, wire fatigue, clot maceration efficacy, and balloon integrity. Statistical sampling based on ISO 2859-1.
Technological Characteristics
Battery-operated, hand-held rotational thrombectomy system. Features rotational wire with soft distal tip and integrated embolectomy balloon. Materials and design comply with ISO 10555-1 and ISO 10555-4 standards. Mechanical operation for clot maceration and balloon-based thrombus removal.
Indications for Use
Indicated for mechanical declotting and removal of thrombi and emboli in synthetic dialysis grafts, and for subsequent embolectomy procedures to remove emboli and thrombi in the arterial system communicating with the synthetic dialysis graft.
Regulatory Classification
Identification
An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)
K040252 — CLEANER ROTATIONAL THROMBECTOMY SYSTEM KIT · Rex Medical · Feb 26, 2004
K031610 — REX MEDICAL CLEANER ROTATIONAL THROMBECTOMY SYSTEM · Rex Medical · May 27, 2003
K060904 — CLEANER ROTATIONAL THROMBECTOMY SYSTEM · Rex Medical, L.P. · Apr 28, 2006
K030504 — AKONYA ELIMINATOR MECHANICAL THROMBECTOMY DEVICE, MODEL EL 10060060 · Idev Technologies, Inc. · Sep 16, 2003
K041374 — AKONYA ELIMINATOR MECHANICAL THROMBECTOMY DEVICE, MODEL EL10060060 · Idev Technologies, Inc. · Mar 2, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
DEC 1 6 2003
Attachment 1
# 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
K033793 The assigned 510(k) number is:
#### 1. Submitter's Identification:
Rex Medical 555 North Lane Suite 6101 Conshohochen, PA 19428 (610) 940-0665
Contact: Mr. Michael Paris
## Date Summary Prepared:
October 29, 2003
#### 2. Name of the Device:
Rex Medical Cleaner II Rotational Thrombectomy System".
#### 3. Common or Usual Name:
Intraluminal Artery Stripper (Peripheral Atherectomy Catheter)
#### 4. Predicate Device Information:
- 1) K031610: Cleaner™ Rotational Thrombectomy System
- 2) K902327: Embolectorny Catheter
- 3) K032393: Stopcock
- 4) K022170: Inner-Lock™ Sheath Introducer
- 5) K032569: Short Sheath Introducer
{1}------------------------------------------------
#### 5. Device Description:
The Rex Medical Cleaner II Rotational Thrombectorny System is a battery operated, hand held, wall contacting, rotational thrombectorny device which provides an effective means to restore patency to occluded synthetic dialysis grafts. The rotational wire, with integrated soft distal tip, provides an atraumatic approach to mechanical thrombectorny. The Cleaner II macerates clot into particulate size that is not harmful to the patient. The integrated Embolectorny balloon provides a means to remove macerated thrombi, as well as wall adherent thrombi, from the dialysis graft. If required, the Embolectomy balloon located on the Cleaner II device may be used to remove any additional thrombi or emboli from the patient after the initial thrombectorny procedure has been performed,
#### 6. Intended Use:
The Rex Medical Cleaner II Rotational Thrombectomy System permits mechanical declotting and removal of thrombi and emboli in synthetic dialysis grafts. The Embolectory balloon component, located on the Cleaner II Device, may be used for subsequent Embolectomy procedures to remove emboli and thrombi in the arterial system communicating with the synthetic dialysis graft
#### 7. Comparison to Predicate Devices:
The Cleaner II device has the same technological features as the predicate devices.
The following testing was performed to prove substantial equivalence of the Rex Medical Cleaner II Rotational Thrombectomy Device to the predicate devices:
| Stall Torque Test |
|-------------------------------------------------|
| Tensile Test |
| Valve Compression Test |
| Contrast Infusion Test |
| Wire Fatigue Test |
| Clot Maceration in Graft |
| Dimensional Analysis of Balloon |
| Catheter |
| Balloon Leakage and Damage<br>(Per ISO 10555-4) |
{2}------------------------------------------------
Testing proved that the Rex Medical Cleaner II device is as safe and effective as the predicate devices.
### 8. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
Performance testing, which includes testing protocols, testing results and conclusions, were based on ISO 10555-1, ISO 10555-4, and FDA's "Guidance on Pre-market Notification (510(k)) Submission for Short-Term and Long-Term Intravascular Catheters", was submitted with this submission.
Statistical sampling rationale for choosing the number of devices that were tested was based on ISO 2859-1 sampling plans in accordance with our projected lots (batch) size.
Testing results revealed that the Rex Medical Cleaner II Rotational Thrombectorny System device is substantially equivalent to the predicate devices.
#### ். Conclusions:
The subject device, Rex Medical Cleaner II Rotational Thrombectorny System, has the same intended use as the predicate devices. Moreover, bench testing contained in our submission and the non-clinical testing supplied demonstrates that there are no differences in their technological characteristics, thereby not raising any new questions of safety or effectiveness. Thus, the Rex Medical Cleaner II Rotational Thrombectomy System is substantially equivalent to the predicate devices.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure with three stylized wing segments.
DEC 1 6 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Rex Medical Inc. c/o Mr. Robert Mosenkis President Citech 5200 Butler Pike Plymouth Meeting, PA 19462-1298
Re: K033793 Cleaner II Rotational Thrombectomy System Regulation Number: 21 CFR 870.4875 Regulation Name: Peripheral Atherectomy Catheter Regulatory Class: Class II Product Code: MCW Dated: December 4, 2003 Received: December 5, 2003
Dear Mr. Mosenkis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Robert Mosenkis
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA S Issualice of a succession complies with other requirements of the Act
that FDA has made a determination that your device complies - You must that FDA has made a decertimation that your are res by other Federal agencies. You must a or any Federal statutes and regulations daminding, but not limited to: registration and listing (21 lot comply with all the Act s requirements, me; come, cannfacturing practice requirements as set a CFR Part 807); labeling (21 CFR 1 at 001); good if applicable, the electronic
forth in the quality systems (QS) regulation (21 CFR Part 820); and 11 app 1000, 1050 forth in the quality systems (QD) regallations (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control provisions (Declons of Pevice as described in your Section 5 10(k)
This letter will allow you to begin marketing your device as described in your de I his letter will anow you to ocgin marketing of substantial equivalence of your device to a legally
premarket notification. The FDA finding of substantial equivaliance and t premarket nonification. The I Dry Imaning of basification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice ion your do rice at (301) 594-4692. Also, please note the regulation entitled, Contact the Office of Compulation at (301) of the Part 807.97). You may obtain "Misoraliding by reference to premainterialities under the Act from the Division of Small other general information on your responsion.com at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address
(501) 445 059 059 050 000 / cdrh / dsma / dsmamain. html
Sincerely yours,
Ashley B. Banam
Image /page/4/Picture/6 description: The image shows a signature in black ink on a white background. The signature appears to be cursive and somewhat stylized, with loops and curves forming the letters. It is difficult to discern the exact letters or name represented by the signature due to its abstract nature.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Exhibit B
Page _ 1 of of 1
K033793 510(k) Number (if known):
Device Name; Rex Medical Cleaner II Rotational Thrombectomy System
Indications For Use:
The Rex Medical Cleaner II Rotational Thrombectomy System permits mechanical declotting and removal of thrombi and emboli in synthetic dialysis grafts. The Embolectorny balloon component, located on the Cleaner II Device, may be used for subsequent Embolectorny procedures to remove embol and thrombi in the arterial system communicating with the synthetic dialynisor afte
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Ashley B. Baum
(Division Sian-off) Division of Cardidnasgular Devices
510(K) N...
*K033793* (SM. K)
Over-The-Counter Use (Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.